ARCHITECT 7D70-31Recalls & reported problems
Abbott Laboratories · Magnesium · Class 1
- Adverse event reports
- 16
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 16 adverse event reports that list this exact device (DI 00380740012939), received between Aug 1, 2017 and Dec 22, 2020 — classified as 16 malfunctions.
The most frequently reported specific device problem is “High Test Results” (12 reports).
1 FDA recall record name this device's identifier.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Recalls naming this device
- Jan 2, 2019Terminated
Abbott internal testing has identified that the Magnesium urine application demonstrates depressed urine result recovery. All samples >0.5mEq/L fail the linearity acceptance criteria for the bias specification and demonstrate depressed recovery of up to 37%. The specific cause of the negative bias/depressed results is currently under investigation.
Z-0971-2019 · Root cause: Under Investigation by firm · FDA record ↗
Most reported problems
Device problems
- High Test Results12
- Low Test Results3
- Incorrect Or Inadequate Test Results1
Patient problems
- No Known Impact Or Consequence To Patient12
- No Consequences Or Impact To Patient4
Recent adverse event reports
The 16 most recent of 16 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 22, 2020Low Test Results; Low Test Results
THE CUSTOMER OBSERVED A FALSELY DECREASED MAGNESIUM RESULT FOR ONE PATIENT ON AN ARCHITECT C16000 ANALYZER. THE FOLLOWING DATA WAS PROVIDED (CUSTOMERS REFERENCE RANGE IS 0.75-1.02 MMOL/L): SAMPLE (B)(6) INITIAL RESULT WAS 0.24, REPEATED, UNDER SAMPLE ID (B)(6), WAS 0.77 MMOL/L. THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 11058369 ↗MalfunctionOct 21, 2019High Test Results; High Test Results
THE CUSTOMER REPORTED A FALSELY ELEVATED MAGNESIUM (MG) RESULT GENERATED ON THE ARCHITECT C8000 ANALYZER. RESULTS PROVIDED: SID (B)(6) ON (B)(6) 2019 = 2.3 / 0.9 MEQ/L (NORMAL RANGE: 1.3-2.1 MEQ/L). THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9214936 ↗MalfunctionJul 29, 2019High Test Results; High Test Results
THE CUSTOMER REPORTED A FALSELY ELEVATED MAGNESIUM (MG) RESULT GENERATED ON THE ARCHITECT C8000 ANALYZER. RESULTS PROVIDED: SID (B)(6), (B)(6) 2019 = 2.3 / 0.9 MEQ/L (NORMAL RANGE: 1.3-2.1 MEQ/L). THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8837657 ↗MalfunctionMay 21, 2019High Test Results; High Test Results
THE CUSTOMER REPORTED A FALSE ELEVATED MAGNESIUM RESULT ON THE ARCHITECT C8000 ANALYZER. THE INITIAL RESULT WAS GREATER THAN 8.0 MG/DL, REPEATED AT 2.0 AND 2.1 MG/DL NO REPORTED IMPACT TO PATIENT MANAGEMENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8627875 ↗MalfunctionDec 22, 2018High Test Results; High Test Results
THE ACCOUNT GENERATED A FALSE ELEVATED ARCHITECT MAGNESIUM (7.33 MG/DL) THAT REPEATED ON ANOTHER ANALYZER (2.34, 2.39 MG/DL). NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED. NO SPECIFIC PATIENT INFORMATION WAS PROVIDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8193074 ↗MalfunctionNov 12, 2018High Test Results; High Test Results
THE CUSTOMER GENERATED A FALSELY ELEVATED MAGNESIUM RESULT WHILE USING THE CLINICAL CHEMISTRY MAGNESIUM ASSAY ON THE ARCHITECT C16000 ANALYZER. THE CUSTOMER PROVIDED THE FOLLOWING RESULT DATA: SID (B)(6): INITIAL 8.6 MG/DL, RETEST 1.7 MG/DL. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED. NO ADDITIONAL PATIENT INFORMATION WAS PROVIDED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8059823 ↗MalfunctionSep 21, 2018High Test Results; High Test Results
THE CUSTOMER OBSERVED A FALSELY ELEVATED MAGNESIUM RESULT ON THE ARCHITECT C8000 ANALYZER. THE FOLLOWING DATA WAS PROVIDED FOR A MALE PATIENT: (B)(6) INITIAL 6.3, REPEAT 1.8 AND 1.9 MG/DL. THE SAMPLE WAS CONFIRMED AT 1.7 ON ANOTHER METHOD. THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7899908 ↗MalfunctionAug 30, 2018High Test Results; High Test Results
THE ACCOUNT OBSERVED FALSELY ELEVATED ARCHITECT MAGNESIUM ON 2 PATIENT SAMPLES THAT REPEATED LOWER. (B)(6) WAS 3.03 BUT REPEATED 1.3 MG/DL. (B)(6) WAS 2.79 BUT REPEATED 1.3 MG/DL. THE ACCOUNT USES A NORMAL RANGE OF 1.6 TO 2.6 MG/DL. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED. NO SPECIFIC PATIENT INFORMATION WAS PROVIDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7836030 ↗MalfunctionAug 23, 2018High Test Results; High Test Results
THE CUSTOMER OBSERVED FALSELY ELEVATED MAGNESIUM RESULTS ON THE ARCHITECT C8000 ANALYZER. THE FOLLOWING DATA WAS PROVIDED (MG/DL): SID W10289 INITIAL 3.3, REPEATS 3.2, 1.2, 1.2; SID W11215 INITIAL 2.9, REPEATS 2.9, 1.3, 1.3; SID 5085854 INITIAL 3.3, REPEATS 1.2, 1.0, 1.9; SID 257642 INITIAL 2.9, REPEAT 1.3. THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7810395 ↗MalfunctionJun 28, 2018Low Test Results; Low Test Results
THE CUSTOMER OBSERVED MAGNESIUM IMPRECISION ON THE ARCHITECT C16000 SYSTEM. PATIENT 1: INITIAL RESULT OF <0.7 MG/DL RETESTED AT 2.7 MG/DL. PATIENT 2: INITIAL RESULT OF 0.7 MG/DL RETESTED AT 2.3 MG/DL ON A DIFFERENT ANALYZER. NO ADVERSE IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7647917 ↗MalfunctionJun 28, 2018Low Test Results; Low Test Results
THE CUSTOMER REPORTED A FALSELY DEPRESSED ARCHITECT MAGNESIUM RESULT ON THE ARCHITECT C16000 ANALYZER. SAMPLE 1 GENERATED AN INITIAL MAGNESIUM RESULT OF 1.0 MG/DL AND RETEST ON A SECOND ANALYZER GENERATED 2.0 MG/DL. NO ADVERSE IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7647853 ↗MalfunctionJun 27, 2018High Test Results; High Test Results
THE CUSTOMER REPORTED A FALSELY ELEVATED ARCHITECT MAGNESIUM RESULT FOR AN (B)(6) MALE. SAMPLE ID (B)(6): INITIAL 8.9 MG/DL, RETEST 2.9 MG/DL, RETEST ALTERNATE ARCHITECT 2.0 MG/DL. A NEW SAMPLE WAS COLLECTED (SAMPLE ID (B)(6)) GENERATED 2.3 MG/DL. NO ADVERSE IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7642498 ↗MalfunctionJun 26, 2018High Test Results; High Test Results
THE CUSTOMER OBSERVED FALSELY ELEVATED MAGNESIUM RESULTS ON THE ARCHITECT C16000 ANALYZER. THE FOLLOWING DATA WAS PROVIDED (MG/DL): PATIENT 1 INITIAL 7.5, REPEAT 2.1; PATIENT 2 INITIAL 6.0, REPEAT 2.3. THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7639579 ↗MalfunctionNov 16, 2017High Test Results; High Test Results
THE CUSTOMER REPORTED AN INITIAL FALSE ELEVATED ARCHITECT MAGNESIUM RESULT. SAMPLE (B)(6)GENERATED AN INITIAL RESULT OF 6.97 MG/DL AND REPEAT GENERATED A RESULT OF 2.1 MG/DL. THE CUSTOMER USES A NORMAL RANGE OF 1.6 TO 2.6 MG/DL. NO SPECIFIC PATIENT INFORMATION WAS PROVIDED AND NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7037177 ↗MalfunctionSep 15, 2017High Test Results; High Test Results
THE CUSTOMER REPORTED FALSE ELEVATED ARCHITECT MAGNESIUM RESULTS FOR TWO PATIENTS. THE CUSTOMER USES A NORMAL RANGE OF 1.6 TO 2.6 MG/DL. THE FOLLOWING DISCREPANT RESULTS WERE PROVIDED: PATIENT 3: INITIAL 7.3 AND REPEAT 1.5 MG/DL. PATIENT 5: INITIAL 6.9 AND REPEAT 1.7 MG/DL . NO SPECIFIC PATIENT INFORMATION WAS PROVIDED. NO ADVERSE IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6870692 ↗MalfunctionAug 1, 2017Incorrect Or Inadequate Test Results; Incorrect Or Inadequate Test Results
THE CUSTOMER OBSERVED MULTIPLE FALSELY ELEVATED MAGNESIUM RESULTS WHILE USING THE CLINICAL CHEMISTRY MAGNESIUM REAGENTS. THE CUSTOMER USES REFERENCE RANGE LESS THAN 2 MG/DL. THE CUSTOMER STATED THAT THE INITIAL RESULTS WERE GREATER THAN 8 AND THE REPEAT RESULTS WERE LESS THAN 2, HOWEVER SPECIFIC RESULTS PER PATIENT WERE NOT PROVIDED. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6758705 ↗
Other PHOTOMETRIC METHOD, MAGNESIUM devices
- ARCHITECT 7D70-21Abbott Laboratories · PHOTOMETRIC METHOD, MAGNESIUM1 recall37 reports
- Dimension Vista® Flex® reagent cartridge MG 10445158Siemens Healthcare Diagnostics Inc. · PHOTOMETRIC METHOD, MAGNESIUM1 recall5 reports
- Dimension® Flex® reagent cartridge MG 10444963Siemens Healthcare Diagnostics Inc. · PHOTOMETRIC METHOD, MAGNESIUM1 recall0 reports
- MAGNESIUM (Mg) MG8326Randox Laboratories Limited · Photometric Method, Magnesium1 recall0 reports
- MAGNESIUM (Mg) MG3880Randox Laboratories Limited · Photometric Method, Magnesium1 recall0 reports
- ARCHITECT 03P6822Abbott GmbH · PHOTOMETRIC METHOD, MAGNESIUM43 reports
- Alinity 08P1934Abbott Ireland · PHOTOMETRIC METHOD, MAGNESIUM30 reports
- ARCHITECT 03P6824Abbott Ireland · PHOTOMETRIC METHOD, MAGNESIUM22 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.