ARCHITECT 7K61-27Recalls & reported problems
Abbott Ireland · ARCHITECT B12 Reagent Kit (100T) · Class 2
All models of ARCHITECT (38)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00380740014179), received between May 9, 2017 and Sep 27, 2022 — classified as 3 malfunctions.
The most frequently reported specific device problem is “High Test Results” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- High Test Results3
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 27, 2022High Test Results; High Test Results
THE CUSTOMER OBSERVED FALSELY ELEVATED ARCHITECT B12 RESULT FOR ONE PATIENT. THE SAMPLE WAS AUTOMATICALLY REPEATED BY THE INSTRUMENT WHICH GENERATED A LOWER RESULT. THE FOLLOWING DATA WAS PROVIDED: ON (B)(6) 2022 INITIAL RESULT = > 2000 PG/ML, REPEAT RESULT = 107 PG/ML THE CUSTOMER REPORTED THE 107 PG/ML. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15492108 ↗MalfunctionApr 1, 2022High Test Results; High Test Results
THE CUSTOMER OBSERVED FALSELY ELEVATED ARCHITECT B12 RESULTS FOR MULTIPLE SAMPLES WHEN COMPARED TO A REFERENCE LABORATORY. ADDITIONALLY, THE CUSTOMER NOTED THERE WAS A SHIFT IN QC RESULTS STARTING ON (B)(6) 2022. THE FOLLOWING DATA WAS PROVIDED (CUSTOMER¿S REFERENCE RANGE: 211 TO 1074 PG/ML, TEST DATE WAS (B)(6) 2022, UNITS OF MEASURE IS PG/ML): SID (B)(6) ARCHITECT RESULT = 315, ARUP RESULT = 200. SAMPLE 5: ARCHITECT RESULT = 360, ARUP RESULT = 255. SAMPLE 6: ARCHITECT RESULT = 383, ARUP RESULT = 273. SAMPLE 10: ARCHITECT RESULT = 405, ARUP RESULT = 254. SAMPLE 14: ARCHITECT RESULT = 371, ARUP RESULT = 235. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13978736 ↗MalfunctionMay 9, 2017High Test Results; High Test Results
THE CUSTOMER OBSERVED AN ELEVATED RESULT WHILE USING THE ARCHITECT B12 ASSAY. THE CUSTOMER PROVIDED THE FOLLOWING DATA (PG/ML): SID (B)(4). INITIAL (B)(6) 2017: 522 RETESTS: (B)(6) 2017: 303 AND 4/15/17: 281.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6553872 ↗
Other RADIOASSAY, VITAMIN B12 devices
- Access Vitamin B12 33000Beckman Coulter, Inc. · Radioassay, Vitamin B121 recall16 reports
- ADVIA Centaur® CKMB MCM 10309782Siemens Healthcare Diagnostics Inc. · RADIOASSAY, VITAMIN B121 recall0 reports
- Atellica IM VB12 10995715Siemens Healthcare Diagnostics Inc. · RADIOASSAY, VITAMIN B1224 reports
- Elecsys Vitamin B12 II 07028121190Roche Diagnostics GmbH · RADIOASSAY, VITAMIN B1221 reports
- Atellica IM VB12 10995714Siemens Healthcare Diagnostics Inc. · RADIOASSAY, VITAMIN B1213 reports
- ADVIA Centaur® VB12 10309970Siemens Healthcare Diagnostics Inc. · RADIOASSAY, VITAMIN B128 reports
- ARCHITECT 7K61-37Abbott Ireland · RADIOASSAY, VITAMIN B124 reports
- ADVIA Centaur® Active B12 10995088Siemens Healthcare Diagnostics Inc. · RADIOASSAY, VITAMIN B121 report
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page