ARCHITECT 8K41-27Recalls & reported problems
Abbott Laboratories · ARCHITECT Insulin Reagent Kit · Class 1
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00380740017569), received between May 3, 2018 and Jul 15, 2020 — classified as 2 malfunctions.
The most frequently reported specific device problem is “High Test Results” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- High Test Results2
Patient problems
- No Consequences Or Impact To Patient1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 15, 2020High Test Results; High Test Results
THE ACCOUNT GENERATED FALSE ELEVATED ARCHITECT INSULIN OF 358 AND 376 PMOL/L ON A NON-DIABETIC PATIENT ((B)(6) YEAR OLD EUROPEAN, GENDER NOT PROVIDED). TWO ADDITIONAL SAMPLES FROM THE PATIENT GENERATED ARCHITECT INSULIN OF 120 AND 75 PMOL/L. THE 3 PATIENT SAMPLES WERE PROCESSED ON ANOTHER ARCHITECT ANALYZER WITH VALUES OF 371.7, 92, 71 PMOL/L. THE ACCOUNT USES A NORMAL INSULIN RANGE OF 17.8 TO 173 PMOL/L. THE PATIENT HISTORY WAS (B)(6) SINCE 1991 WITH TRIPLE THERAPY TREATMENT AND HAS NORMAL TPSA AND GLYCEMIA RESULTS. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10281037 ↗MalfunctionMay 3, 2018High Test Results; High Test Results
THE CUSTOMER STATED THAT AN INITIAL ARCHITECT INSULIN RESULT OF 399.7 PMOL/L WAS GENERATED ON A (B)(6) PATIENT SAMPLE. THE SAMPLE WAS NOT REPEATED AT THE TIME. THE PATIENT QUESTIONED THE RESULT AS SHE HAD HER BLOOD DRAWN AT ANOTHER HOSPITAL ON A DIFFERENT DAY AND THE RESULTS WERE 20 PMOL/L. THE METHOD WAS UNKNOWN. THE ORIGINAL SAMPLE HAD BEEN DISPOSED AND IS NOT AVAILABLE. THERE WAS NO REPORT OF IMPACT TO PATIENT MANAGEMENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7485180 ↗
Other RADIOIMMUNOASSAY, IMMUNOREACTIVE INSULIN devices
- Insulin HYE-5361DRG Instruments Gesellschaft mit beschränkter Haftung · Radioimmunoassay, Immunoreactive Insulin1 recall0 reports
- Elecsys Insulin 07027559190Roche Diagnostics GmbH · RADIOIMMUNOASSAY, IMMUNOREACTIVE INSULIN18 reports
- Elecsys Insulin 12017547122Roche Diagnostics GmbH · RADIOIMMUNOASSAY, IMMUNOREACTIVE INSULIN12 reports
- Alinity 04T7520Abbott GmbH · RADIOIMMUNOASSAY, IMMUNOREACTIVE INSULIN11 reports
- ARCHITECT 08K4128Abbott GmbH · RADIOIMMUNOASSAY, IMMUNOREACTIVE INSULIN8 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.