Skip to content

ARCHITECT 9D96-21Recalls & reported problems

Abbott Laboratories · CLIN CHEM C3 REAGENT · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00380740019730), received between Jun 5, 2019 and Jun 5, 2019 — classified as 1 malfunction.

The most frequently reported specific device problem is “Low Test Results” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Most reported problems

Device problems

  • Low Test Results1

Patient problems

  • No Patient Involvement1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 5, 2019Low Test Results; Low Test Results

    ABBOTT IDENTIFIED THAT AN UPDATE TO THE REAGENT AND SPECIMEN COLLECTION AND HANDLING/PREPARATION FOR ANALYSIS SECTIONS OF THE INSTRUCTIONS FOR USE (IFU) FOR THE INCLUDED ARCHITECT PRODUCTS WAS REQUIRED FOR THE FOLLOWING ISSUES: BASED ON RECENT FIBRINOGEN INTERFERENCE TESTING, IT HAS BEEN DETERMINED THAT EDTA PLASMA SAMPLES ARE NO LONGER ACCEPTABLE FOR USE WITH THE FOLLOWING ASSAYS: APOLIPOPROTEIN A1 (APO A1), COMPLEMENT C3 (C3), COMPLEMENT C4 (C4), HAPTOGLOBIN, IMMUNOGLOBULIN G (IGG), IMMUNOGLOBULIN A (IGA), AND IMMUNOGLOBULIN M (IGM). THERE IS A POTENTIAL FOR FALSELY DEPRESSED RESULTS DUE TO FIBRINOGEN INTERFERENCE WITH EDTA SPECIMENS FOR THESE PRODUCTS. ADDITIONALLY, FOR THE COMPLEMENT C4 ASSAY, THERE IS A POTENTIAL FOR FALSELY DEPRESSED RESULTS DUE TO FIBRINOGEN INTERFERENCE WITH HEPARIN TUBES AT THE CONCENTRATIONS > 1512 MG/DL IN LITHIUM HEPARIN TUBES AND >859 MG/DL IN SODIUM HEPARIN TUBES. ADDITIONALLY, BASED ON RECENT TESTING, THE SPECIMEN STORAGE INFORMATION IS BEING REVISED FOR THE FOLLOWING ASSAYS: IGG, IGA, IGM, AND C3. THE MAXIMUM STORAGE TIME FOR SAMPLES STORED AT 2-8 DEGREES C AND 20-25 DEGREES C WERE INACCURATELY STATED. THESE ISSUES WERE IDENTIFIED THROUGH INTERNAL T

    Patient problems: No Patient Involvement

    MAUDE report 8671652 ↗

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page