Skip to content

ARCHITECT 1L82-27Recalls & reported problems

Abbott Ireland Diagnostics · ARCHITECT AUSAB Reagent · Class 2

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00380740115364), received between Aug 17, 2018 and May 14, 2024 — classified as 4 malfunctions.

The most frequently reported specific device problem is “High Test Results” (4 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Apr 24, 2025Open, Classified

    transit delays of certain cold chain products due to severe thunderstorms which may caused the products to be stored outside the recommended labeled storage conditions. This failure could increase the probability of an incorrect test result or a delayed test result.

    Z-1842-2025 · Root cause: Storage · FDA record ↗

Most reported problems

Device problems

  • High Test Results4
  • False Positive Result1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • No Known Impact Or Consequence To Patient2

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 14, 2024High Test Results; High Test Results

    THE CUSTOMER OBSERVED A FALSE POSITIVE ARCHITECT AUSAB RESULT GENERATED ON THE ARCHITECT I2000SR PROCESSING MODULE FOR ONE SAMPLE. THE FOLLOWING RESULTS WERE PROVIDED: SID (B)(6) INITIAL AUSAB RESULT = 19.64 MIU/ML, REPEAT RESULT = 0.48 MIU/ML, REPEAT RESULT ON ((B)(6)) 0.12 MIU/ML NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19309515 ↗
  • MalfunctionMar 1, 2022False Positive Result; High Test Results; False Positive Result; High Test Results

    THE CUSTOMER OBSERVED FALSE POSITIVE AUSAB RESULTS GENERATED ON AN ARCHITECT I1000SR ANALYZER FOR MULTIPLE PATIENTS, WHICH REPEATED NEGATIVE. THE FOLLOWING SIDS WERE PROVIDED: (B)(6) INITIAL RESULT = 32.4, (B)(6) INITIAL RESULT = 19.56, (B)(6) INITIAL RESULT = 35.76. THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13632172 ↗
  • MalfunctionFeb 15, 2019High Test Results; High Test Results

    THE CUSTOMER REPORTED (B)(6) ARCHITECT (B)(6) RESULTS ON FIVE PATIENTS. THE RESULTS PROVIDED WERE: ON (B)(6) 2019 (B)(6); ON (B)(6); ON (B)(6). THERE WAS NO REPORTED IMPACT TO PATIENT MANAGEMENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8343928 ↗
  • MalfunctionAug 17, 2018High Test Results; High Test Results

    THE CUSTOMER OBSERVED (B)(6) RESULTS WHILE USING THE ARCHITECT AUSAB ASSAY. THE CUSTOMER INDICATED THAT ON (B)(6) 2018, 8 PATIENT SPECIMENS GENERATED RESULTS BETWEEN 20-30 MIU/ML. NONE OF THE SPECIMENS HAVE BEEN RETESTED. THE CUSTOMER BELIEVES THE RESULTS MAY HAVE BEEN HIGH SINCE THEY WERE EXPECTING (B)(6) RESULTS. THE SPECIMENS WERE TESTED IMMEDIATELY AFTER CAP SURVEY SPECIMENS THAT ALSO GENERATED HIGH RESULTS, WHICH WERE EXPECTED TO BE LOWER AS WELL. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED. NO SPECIFIC PATIENT INFORMATION WAS PROVIDED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7793011 ↗

Other Test, hepatitis b (b core, be antigen, be antibody, b core igm) devices

All Test, hepatitis b (b core, be antigen, be antibody, b core igm) devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.