Skip to content

Alinity 07P9030Recalls & reported problems

Abbott Ireland Diagnostics · Alinity i AFP Reagent Kit 1200 Tests · Class 3

Adverse event reports
5
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00380740130824), received between May 31, 2023 and Dec 9, 2025 — classified as 5 malfunctions.

The most frequently reported specific device problem is “High Test Results” (5 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • High Test Results5

Patient problems

  • No Clinical Signs, Symptoms or Conditions5

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 9, 2025High Test Results; High Test Results

    THE CUSTOMER REPORTED FALSELY ELEVATED ALINITY I AFP RESULTS GENERATED BY THE ALINITY I PROCESSING MODULE FOR THREE PATIENTS. NO INFORMATION WAS AVAILABLE REGARDING THE PATIENTS¿ DIAGNOSES. THE SAMPLES WERE REPEATED, AND LOWER RESULTS WERE OBTAINED. THE FOLLOWING DATA WAS PROVIDED: CUSTOMER¿S REFERENCE RANGE: =8.1 G/L. SID (B)(6) (59-YEAR-OLD MALE): (B)(6) 2025 ALINITY I AFP INITIAL RESULT = 42.4 G/L, REPEAT RESULT = 2.9 G/L. SID (B)(6) (83-YEAR-OLD FEMALE): (B)(6) 2025 ALINITY I AFP INITIAL RESULT = 23.2 G/L, REPEAT RESULT = 1.5 G/L SID (B)(6) (64- YEAR-OLD MALE): (B)(6) 2025 ALINITY I AFP INITIAL RESULT = 20.1 G/L, REPEAT RESULT = 12.6 G/L NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23757037 ↗
  • MalfunctionOct 29, 2025High Test Results; High Test Results

    THE CUSTOMER OBSERVED FALSELY ELEVATED ALINITY I AFP FOR ONE PATIENT. THE FOLLOWING RESULTS WERE PROVIDED (REFERENCE RANGE <=8.8 UG/L): ON (B)(6) 2025, SID (B)(6), INITIAL AFP RESULT (B)(6) = 559.12 UG/L; (B)(6) 2025, ANOTHER SAMPLE WAS COLLECTED, SID (B)(6), INITIAL RESULT (B)(6) = 6.37 UG/L; (B)(6) 2025, REPEAT RESULT= 5.78 UG/L, (B)(6) 2025, REPEAT RESULT ON ANOTHER ANALYZER (B)(6 )= 5.69 UG/L. ADDITIONAL RESULTS PROVIDED: CA125 RESULT= 3.6 U/ML; CEA RESULT= 1.00 UG/L THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23421164 ↗
  • MalfunctionAug 4, 2025High Test Results; High Test Results

    THE CUSTOMER OBSERVED A FALSELY ELEVATED ALINITY I AFP RESULT FOR ONE PATIENT. THE FOLLOWING DATA WAS PROVIDED (CUSTOMER¿S REFERENCE RANGE IS </=8.8 UG/L): SAMPLE ID (B)(6), INITIAL RESULT, ON (B)(6) 2025, WAS 62.65, REPEAT RESULT WAS 4.26, REPEAT RESULT, ON ANOTHER ANALYZER, WAS 3.90 UG/L. ADDITIONAL LABORATORY RESULTS WERE PROVIDED: CEA WAS 3.8 UG/L, TPSA WAS <0.025 UG/L. THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22708768 ↗
  • MalfunctionApr 10, 2025High Test Results; High Test Results

    THE CUSTOMER REPORTED FALSE ELEVATED ALINITY I AFP AND PROVIDED THE FOLLOWING DATA: ON (B)(6) 2025, SAMPLE (B)(6) INITIAL RESULT WAS 83.7 UG/LAND REPEAT RESULT ON ANOTHER ANALYZER WAS <2 UG/L. NO FURTHER PATIENT DATA WAS PROVIDED. PER THE CUSTOMER DISCREPANT RESULTS WERE NOT REPORTED OUT TO THE MEDICAL PROVIDER. THERE WAS NO REPORTED IMPACT TO PATIENT MANAGEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21819378 ↗
  • MalfunctionMay 31, 2023High Test Results; High Test Results

    THE CUSTOMER REPORTED FALSELY ELEVATED ALINITY I AFP RESULTS FOR ONE PATIENT SAMPLE. THE CUSTOMER WAS UNABLE TO PROVIDE THE PATIENT¿S MEDICAL HISTORY AND INFORMATION. THE FOLLOWING DATA WAS PROVIDED: ON (B)(6) 2023 SID (B)(6). INITIAL RESULT = 20.15 NG/ML RESULT ON DIFFERENT MODULE (METHOD UNKNOWN) = 2.1 REPEAT RESULT ON THE SAME ANALYZER = 2.63 NG/ML THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17035169 ↗

Other KIT, TEST, ALPHA-FETOPROTEIN FOR NEURAL TUBE DEFECTS devices

All KIT, TEST, ALPHA-FETOPROTEIN FOR NEURAL TUBE DEFECTS devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.