Alinity 07P9920Recalls & reported problems
Abbott GmbH · Alinity c Creatinine Reagent Kit 3000 Tests · Class 2
All models of Alinity (37)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This Abbott Alinity creatinine reagent kit is a lab test for measuring creatinine. The 24 reports mostly describe falsely high results that test within normal range on repeat testing. The reports mention no deaths and 1 injury. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 24
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 24 adverse event reports that list this exact device (DI 00380740135102), received between Dec 26, 2017 and Feb 13, 2024 — classified as 1 injury, 23 malfunctions.
The most frequently reported specific device problem is “High Test Results” (24 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- High Test Results24
Patient problems
- No Known Impact Or Consequence To Patient12
- No Clinical Signs, Symptoms or Conditions11
- Insufficient Information1
Recent adverse event reports
The 24 most recent of 24 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 13, 2024High Test Results; High Test Results
THE CUSTOMER OBSERVED A FALSELY ELEVATED ALINITY C CREATININE RESULT GENERATED FOR A 21 YEAR OLD MALE PATIENT SAMPLE. THE FOLLOWING DATA WAS PROVIDED (CUSTOMER¿S REFERENCE RANGE 0.70-1.20 MG/DL): (B)(6)2024 INITIAL RESULT = 5.32 MG/DL, (B)(6)2024 REPEAT RESULT = 1.13 MG/DL (B)(6)2024 SAME PATIENT NEW SAMPLE; RESULT ON ANOTHER ANALYZER = 1.01 MG/DL NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18698123 ↗MalfunctionFeb 6, 2024High Test Results; High Test Results
THE CUSTOMER REPORTED FALSELY ELEVATED ALINITY C CREATININE AND PROVIDED THE FOLLOWING DATA: CUSTOMER REFERENCE RANGE IS 50-98 UMOL/L. SAMPLE ID (B)(6)INITIAL RESULT WAS 1323.10 UMOL/L. REPEAT RESULT WAS 59.69 UMOL/L. REPEATED RESULT ON ANOTHER ANALYZER ((B)(6)) WAS 58 UMOL/L. THE CUSTOMER REPORTED THAT ALL CONTROL LEVELS FOR CREATININE ARE IN RANGE. PER THE CUSTOMER THE DISCREPANT RESULTS WERE NOT REPORTED OUT TO THE PATIENT¿S MEDICAL PROVIDER. THERE WAS NO REPORTED IMPACT TO PATIENT MANAGEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18655317 ↗MalfunctionJan 29, 2024High Test Results; High Test Results
THE CUSTOMER REPORTED A FALSELY ELEVATED CREATININE RESULT GENERATED ON THE ALINITY C PROCESSING MODULE ON A PATIENT. THE FOLLOWING RESULTS WERE PROVIDED: SID (B)(6) TWO PRIMARY SAMPLE TUBES WERE DRAWN OFF-SITE, THE CUSTOMER PREPARED AN ALIQUOT OF EACH TUBE. ALIQUOT 1 PROCESSED ON (B)(6) 2024 CREATININE = 189 UMOL/L (NORMAL RANGE: 64-110 UMOL/L) NEW SAMPLE TESTED ON (B)(6) 2024 WAS NORMAL. THE CUSTOMER THEN PROCESSED THE ORIGINAL PRIMARY TUBES AND THE 2 ALIQUOTS: PRIMARY TUBE 1 = 90.4 UMOL/L, PRIMARY TUBE 2 = 84.3 UMOL/L, ALIQUOT 1 = 132.4 / 127.2 UMOL/L, ALIQUOT 2 = 77 / 76.4 UMOL/L . NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18597080 ↗MalfunctionAug 28, 2023High Test Results; High Test Results
THE CUSTOMER STATED THAT FALSELY ELEVATED ALINITY C CREATININE RESULTS WERE GENERATED. THE FOLLOWING DATA WAS PROVIDED. SID (B)(6): INITIAL RESULT OF 347.0 UMOL/L RETESTED AT 73.2 UMOL/L AFTER RECENTRIFUGATION SID (B)(6): INITIAL RESULT OF 212.2 UMOL/L RETESTED AT 55.7 UMOL/L AFTER RECENTRIFUGATION NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17633802 ↗InjuryNov 22, 2022High Test Results; High Test Results
THE CUSTOMER OBSERVED A FALSE ELEVATED ALINITY C CREATININE RESULT FOR A FEMALE PATIENT. THE CUSTOMER REPORTED THE PATIENT WAS TREATED ACUTELY WITH AN UNSPECIFIED TREATMENT. THE FOLLOWING EXAMPLE RESULTS PROVIDED: (B)(6) 2022 SID (B)(6)= 576 UMOL/L. (B)(6) 2022 SID (B)(6)= REPEAT (1:10 DILUTION) 300 UMOL/L, REPEAT ON SECOND ALINITY IN THE LAB 300 UMOL/L EVEN THOUGH IT WAS REQUESTED, NO ADDITIONAL INFORMATION REGARDING THE ACUTE TREATMENT OR PATIENT HARM WAS PROVIDED AT THIS TIME.
Patient problems: Insufficient Information
MAUDE report 15843702 ↗MalfunctionJun 24, 2022High Test Results; High Test Results
THE CUSTOMER OBSERVED FALSELY ELEVATED ALINITY C CREATININE RESULTS FOR ONE PATIENT. THE FOLLOWING DATA WAS PROVIDED: (B)(6) 2022 SID (B)(6) INITIAL RESULT = 2.88 MG/DL, (B)(6) 2022 SID (B)(6) INITIAL RESULT = 1.94 MG/DL, REFERENCE LABORATORY RESULT = 1.9 MG/DL . NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14801368 ↗MalfunctionJun 17, 2022High Test Results; High Test Results
THE CUSTOMER OBSERVED FALSELY ELEVATED ALINITY C CREATININE RESULTS FOR ONE PATIENT WHO WAS REFERRED TO A SPECIALIST BASED ON THE ELEVATED RESULTS. THE FOLLOWING DATA WAS PROVIDED: SID (B)(6). MEASUREMENT 1: 2.88 MG/DL (B)(6) 2022. MEASUREMENT 2: 1.94 MG/DL (B)(6) 2022 NO ADVERSE IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14735409 ↗MalfunctionMay 6, 2022High Test Results; High Test Results
THE CUSTOMER OBSERVED A FALSELY ELEVATED ALINITY C CREATININE RESULT FOR ONE SAMPLE THAT WAS REPORTED OUT OF THE LABORATORY. THE FOLLOWING DATA WAS PROVIDED: (B)(6) 2022, SID (B)(6) INITIAL RESULT = 586.8 UMOL/L, REPEATS ON A 2ND ALINITY = 120.8 UMOL/L, 120.1 UMOL/L, AND 113.1 UMOL/L, REPEATS ON A 3RD ALINITY = 119.7 UMOL/L THE PATIENT WAS REFERRED TO A NEPHROLOGIST DUE TO THE DISCREPANT RESULT. NO ADDITIONAL IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14304343 ↗MalfunctionApr 14, 2022High Test Results; High Test Results
THE CUSTOMER STATED THAT A (B)(6) ALINITY C CREATININE RESULT OF (B)(6) WAS GENERATED FOR A PATIENT SAMPLE ((B)(6)) ON (B)(6), 2022 THAT RETESTED AT (B)(6). AN ADDITIONAL OCCURRENCE OCCURRED ON (B)(6), 2022. SID (B)(6) INITIAL RESULT WAS (B)(6) AND REPEAT RESULT WAS (B)(6). NO ADVERSE IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14114113 ↗MalfunctionJan 19, 2022High Test Results; High Test Results
THE CUSTOMER OBSERVED FALSELY ELEVATED ALINITY C CREATININE RESULTS FOR TWO PATIENTS. THE FOLLOWING DATA WAS PROVIDED: SID IDM (B)(6) INITIAL RESULT WAS 1106.94 UMOL/L, REPEATED 98.30 UMOL/L, REPEATED ON ANOTHER ANALYZER 93 UMOL/L, SID IDM (B)(6) INITIAL RESULT 1222.26 UMOL/L, REPEATED 105 UMOL/L, (REFERENCE RANGES FOR SERUM CREATININE: MEN 63.6-110.5 UMOL/L; WOMEN 50.4-98.1 UMOL/L). NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13291579 ↗MalfunctionAug 20, 2021High Test Results; High Test Results
THE CUSTOMER IDENTIFIED FALSELY ELEVATED ALINITY C CREATININE RESULTS FOR ONE PATIENT. THE FOLLOWING INFORMATION WAS PROVIDED: (B)(6) 2021 INITIAL RESULT 8 MG / DL, REPEATED 0.52 MG EACH / DL. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12348328 ↗MalfunctionJul 30, 2021High Test Results; High Test Results
THE CUSTOMER REPORTED FALSELY ELEVATED CREATININE RESULT ON ONE PATIENT GENERATED ON THE ALINITY C PROCESSING MODULE. THE RESULTS PROVIDED WERE: ON (B)(6) 2021 SID (B)(6) INITIAL = 375.6UMOL/L (NORMAL RANGE FOR FEMALE 50-98UMOLL) / RETESTED = 64UMOL/L / THE SAMPLE WAS THEN TESTED ON ALL 3 ANALYZERS WITH RESULTS 66, 67, 71UMOL/L. THERE WAS NO REPORTED IMPACT TO PATIENT MANAGEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12253982 ↗MalfunctionDec 23, 2020High Test Results; High Test Results
THE CUSTOMER REPORTED A (B)(6) ALINITY C CREATININE RESULT ON ONE MALE PATIENT. THE RESULTS PROVIDED WERE ON (B)(6) 2020 INITIAL = 400UMOL/L (NORMAL IS 110) / RETEST THE NEXT DAY ON A NEW SAMPLE, SAME PATIENT = 73UMOL/L. THERE WAS NO REPORTED IMPACT TO PATIENT MANAGEMENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 11065926 ↗MalfunctionOct 9, 2020High Test Results; High Test Results
THE ACCOUNT GENERATED FALSE ELEVATED ALINITY C CREATININE OF 2.7 MG/DL ON SID (B)(6) THAT REPEATED 0.66 MG/DL. THE SAMPLE WAS REPEATED ON ANOTHER ALINITY ANALYZER WITH NORMAL RESULTS. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10659282 ↗MalfunctionOct 8, 2020High Test Results; High Test Results
THE CUSTOMER REPORTED FALSE ELEVATED ALINITY C CREATININE RESULTS FOR A PATIENT. THE FOLLOWING DATA WAS PROVIDED: . ON (B)(6) 2020, SID: (B)(6) = CREATININE = 2.39 MG/DL, REPEAT = 2.43 MG/DL. ON (B)(6) 2020, CUSTOMER RETESTED SAME PATIENT SAMPLE, AND GENERATED A CREATININE RESULT OF 1.39 MG/DL. ON (B)(6) 2020, SID (B)(6) = CREATININE = 1.42 MG/DL (ON A DIFFERENT ALINITY C ANALYZER), REPEAT CREATININE = 1.71 MG/DL (AFTER RE-CALIBRATING ON ORIGINAL ALINITY C ANALYZER). THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10650093 ↗MalfunctionSep 10, 2020High Test Results; High Test Results
THE CUSTOMER REPORTED FALSELY ELEVATED ALINITY C CREATININE RESULTS ON TWO PATIENTS. THE RESULTS PROVIDED WERE: ON (B)(6) 2020 (B)(6) INITIAL = 6.62MG/DL / RETEST = 6.78 / AFTER PHYSICIAN QUESTIONED RETEST = 2.63 / NEW SAMPLE, SAME PATIENT = 1.0 / PLASMA SAMPLE FROM (B)(6) 2020 = 0.91MG/DL; (B)(6) INITIAL = 5.98 / RETEST = 4.42MG/DL / AFTER PHYSICIAN QUESTIONED = 4.01, 3.99MG/DL / NEW SAMPLE, SAME PATIENT =0.82MG/DL(NORMAL RANGE MALE= 0.7 - 1.2, FEMALE= 0.5 - 0.9MG/DL). THERE WAS NO REPORTED IMPACT TO PATIENT MANAGEMENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10514124 ↗MalfunctionJun 26, 2020High Test Results; High Test Results
THE CUSTOMER OBSERVED FALSELY ELEVATED CREATININE ON ALINITY C PROCESSING MODULE FOR MULTIPLE PATIENTS. THE RESULTS WERE PROVIDED: ON (B)(6) 2020 SID (B)(6) CREATININE=177 UMOL/L ON (B)(6) 2020 REPEATED ON ANOTHER ALINITY =98 UMOL/L(% DIFFERENCE=81%) (LABORATORY REFERENCE RANGE FOR MEN=63.6 TO 110.5 UMOL/L AND WOMEN=50.4 TO 98.1 UMOL/L).THERE WAS NO REPORTED IMPACT TO PATIENT MANAGEMENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10203144 ↗MalfunctionFeb 21, 2020High Test Results; High Test Results
THE CUSTOMER REPORTED A FALSE ELEVATED CREATININE RESULT WHEN PROCESSING ON THE ALINITY C. THE INITIAL RESULT WAS 17.75 MG/DL AND RETEST WAS 0.77 MG/DL. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9739782 ↗MalfunctionJan 28, 2020High Test Results; High Test Results
THE CUSTOMER OBSERVED FALSELY ELEVATED RESULTS WHILE USING ALINITY C CREATININE REAGENTS. THE FOLLOWING DATA WAS PROVIDED FOR THE SAME PATIENT. NO SPECIFIC PATIENT INFORMATION WAS PROVIDED. SID (B)(6) INITIAL 2.67 MG/DL. REPEAT (NO SID PROVIDED) 0.75, 0.71 MG/DL. REPEAT USING A NEW SAMPLE. (SID (B)(6)) 0.73 MG/DL. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9639133 ↗MalfunctionDec 26, 2019High Test Results; High Test Results
THE CUSTOMER REPORTED A FALSELY ELEVATED CREATININE RESULT GENERATED ON THE ALINITY C ANALYZER ON ONE PATIENT. RESULTS PROVIDED: SID 500010630 = 2.934 / 1.99 MG/DL. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9520943 ↗MalfunctionDec 3, 2019High Test Results; High Test Results
THE CUSTOMER REPORTED A FALSELY INCREASED CREATININE RESULT GENERATED ON THE ALINITY C ANALYZER ON ONE PATIENT. RESULTS PROVIDED: 10.86 / 0.86 MG/DL. CREATININE HISTORICALLY = 0.81 / 0.83 MG/DL. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9404450 ↗MalfunctionJun 12, 2019High Test Results; High Test Results
THE CUSTOMER REPORTED FALSELY ELEVATED CREATININE RESULTS ON ONE PATIENT GENERATED ON THE ALINITY C ANALYZER. THE RESULTS PROVIDED WERE: ON (B)(6) 2019:SID (B)(6) INITIAL = 1.04MG/DL / REPEATED ON SID (B)(6) ON (B)(6) 2019 = 2.00 / REPEATED ON (B)(6) 2019 = 1.35MG/DL. THERE WAS NO REPORTED IMPACT TO PATIENT MANAGEMENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8691838 ↗MalfunctionDec 3, 2018High Test Results; High Test Results
THE CUSTOMER OBSERVED FALSELY ELEVATED RESULTS WHILE USING ALINITY C CREATININE REAGENTS. THE FOLLOWING DATA WAS PROVIDED FOR THE SAME PATIENT. SID (B)(6) INITIAL 2.67 MG/DL. REPEAT (NO SID PROVIDED) 0.75, 0.71 MG/DL. REPEAT USING A NEW SAMPLE ((B)(6)) 0.73 MG/DL. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8126514 ↗MalfunctionDec 26, 2017High Test Results; High Test Results
THE CUSTOMER REPORTS THAT ONE PATIENT SAMPLE (SID (B)(6)) GENERATED AN INITIAL ALINITY CREATININE ASSAY RESULT OF 3.37 MG/DL ON AN ALINITY C PROCESSING MODULE. THE SAMPLE RETESTED AT 0.69 MG/DL. NO SUSPECT RESULTS WERE REPORTED FROM THE LAB. THERE IS NO IMPACT TO PATIENT MANAGEMENT REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7146078 ↗
FDA clearance and classification
The FDA regulates product code CGX (alkaline picrate, colorimetry, creatinine) under 21 CFR 862.1225 as a Class II device. 21 CFR 862.1225 ↗
Devices of this type normally need 510(k) clearance before sale.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the Alinity 07P9920
Has the Alinity 07P9920 been recalled?
No FDA recall record in our data names this exact device identifier (DI 00380740135102), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the Alinity 07P9920?
The FDA's MAUDE database lists 24 adverse event reports for this device, received between Dec 26, 2017 and Feb 13, 2024. In those reports, the most common specific device problem is “High Test Results” (24). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Alinity 07P9920?
None of the reports are classified as deaths, 1 as an injury and 23 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Who makes the Alinity 07P9920, and is it still sold?
The FDA's device database lists Abbott GmbH as the company responsible for it (brand Alinity, model 07P9920). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other ALKALINE PICRATE, COLORIMETRY, CREATININE devices
- Dimension® Flex® reagent cartridge CRE2 10872079Siemens Healthcare Diagnostics Inc. · ALKALINE PICRATE, COLORIMETRY, CREATININE1 recall9 reports
- ABX PENTRA CREATININE 120 CP 1220001933Horiba Abx Sas · Alkaline Picrate, Colorimetry, Creatinine1 recall0 reports
- ARCHITECT 3L81-22Abbott Laboratories · ALKALINE PICRATE, COLORIMETRY, CREATININE31 reports
- Atellica CH Crea_2 11097596Siemens Healthcare Diagnostics Inc. · ALKALINE PICRATE, COLORIMETRY, CREATININE19 reports
- Creatinine Jaffe Gen.2 08057532190Roche Diagnostics GmbH · ALKALINE PICRATE, COLORIMETRY, CREATININE16 reports
- Alinity 04T9120Abbott Ireland · ALKALINE PICRATE, COLORIMETRY, CREATININE15 reports
- ARCHITECT 3L81-32Abbott Laboratories · ALKALINE PICRATE, COLORIMETRY, CREATININE15 reports
- Alinity 07P9930Abbott GmbH · ALKALINE PICRATE, COLORIMETRY, CREATININE15 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page