Alinity 08P3920Recalls & reported problems
Abbott Laboratories · Alinity c Iron Reagent Kit 2300 Tests · Class 1
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 0
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
No MAUDE adverse event reports in our data reference this device identifier directly.
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Recalls naming this device
- Sep 29, 2023Open, Classified
A high recovery of Iron assay has been observed at some customer sites. Internal studies confirmed the positive shift of Quality Control and patient samples results, up to 15% for Iron Products (Alinity c Iron Reagent REF. 08P3920) to be used on Abbott Alinity Instrument and to 30% for Iron Products (IRON REF. 6K95-41 and IRON REF. 6K95- 30) to be used on Abbott Architect Instrument. When using the affected lots, there is a potential for delay in reporting patient results due to QC failure, also a potential for falsely elevated results.
Z-0326-2024 · Root cause: Under Investigation by firm · FDA record ↗
Other Photometric method, iron (non-heme) devices
- Iron Gen.2 3183696122Roche Diagnostics GmbH · Photometric method, iron (non-heme)1 recall16 reports
- Atellica CH Iron_2 11097601Siemens Healthcare Diagnostics Inc. · Photometric method, iron (non-heme)1 recall1 report
- ADVIA® Chemistry IRON_2 Iron 2 Reagents 10377510Siemens Healthcare Diagnostics Inc. · Photometric method, iron (non-heme)1 recall1 report
- ARCHITECT 6K95-41Abbott Laboratories · Photometric method, iron (non-heme)1 recall0 reports
- ADVIA® Chemistry IRON_2 Reagents 10341118Siemens Healthcare Diagnostics Inc. · Photometric method, iron (non-heme)1 recall0 reports
- ARCHITECT 6K95-30Abbott Laboratories · Photometric method, iron (non-heme)1 recall0 reports
- Atellica CH Iron3 (Iron3) 11537211Siemens Healthcare Diagnostics Inc. · Photometric method, iron (non-heme)1 recall0 reports
- Iron Gen.2 08057931190Roche Diagnostics GmbH · Photometric method, iron (non-heme)5 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.