Skip to content

Alinity 08P3221Recalls & reported problems

Abbott GmbH · Alinity i CA 19-9XR Reagent Kit 200 Tests · Class 2

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00380740155919), received between Mar 23, 2020 and Feb 13, 2025 — classified as 4 malfunctions.

The most frequently reported specific device problem is “High Test Results” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • High Test Results4

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 13, 2025High Test Results; High Test Results

    THE CUSTOMER REPORTED A FALSELY ELEVATED ALINITY CA 19-9XR RESULT GENERATED ON THE ALINITY I AM PROCESSING MODULE FOR ONE PATIENT WITH A HISTORY OF PANCREATIC CANCER. THE FOLLOWING DATA WAS PROVIDED: ON (B)(6) 2025, SID (B)(6): INITIAL CA 19-9 RESULT = > 4000 U/ML. REPEAT CA 19-9 RESULT ON ANOTHER ALINITY I ((B)(6)) = 4438.19 U/ML (REPORTED TO THE DOCTOR). DUE TO THE ELEVATED CA 19-9 RESULT, THE PATIENT¿S SURGERY WAS CANCELLED. IT IS NOT KNOWN WHAT TYPE OF SURGERY WAS CANCELLED AND IF THE SURGERY WAS RESCHEDULED. THE CUSTOMER STATED THAT THE PATIENT WAS TESTED AT A NEIGHBORING FACILITY ((B)(6)) ON (B)(6) 2025 AND OBTAINED A CA 19-9 RESULT OF 600. THERE WAS NO FURTHER IMPACT TO PATIENT MANAGEMENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21380671 ↗
  • MalfunctionJul 2, 2023High Test Results; High Test Results

    THE CUSTOMER OBSERVED FALSELY ELEVATED ALINITY I CA 19-9XR RESULT FOR AN 83-YEAR-OLD FEMALE PATIENT WHO UNDERGOES REGULAR TESTING EVERY SIX MONTHS OR SO. DUE TO THE ELEVATED ALINITY I CA 19-9XR RESULT, THE FOLLOW-UP TESTING PERIOD WAS REDUCED, AND THE PATIENT WAS TESTED AGAIN ON (B)(6) 2023 WHICH SHOWED A LOWER RESULT. IT IS UNKNOWN IF THE PATIENT WAS DIAGNOSED WITH PANCREATIC CANCER. THE FOLLOWING DATA WAS PROVIDED: CA 19-9XR RESULT (2 MONTHS PRIOR) = 90 U/ML. FOLLOW-UP RESULT = (B)(6) 2023 CA 19-9XR RESULT = 17 U/ML. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17244384 ↗
  • MalfunctionMar 14, 2023High Test Results; High Test Results

    UPDATE: RECEIVED ADDITIONAL INFORMATION FROM CUSTOMER ON 27 MAR 2023. CUSTOMER STATED THEY PERFORMED A HETEROPHILE BLOCKING TUBE TEST FOR THE SAMPLE AND GENERATED THE SAME RESULT OF > 96,000 U/ML.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16538843 ↗
  • MalfunctionMar 23, 2020High Test Results; High Test Results

    THE CUSTOMER OBSERVED A FALSELY ELEVATED CA19-9 RESULT ON THE ALINITY ANALYZER. THE FOLLOWING DATA WAS PROVIDED: CASE 2 (B)(6): MALE DOB (B)(6) 1969 DIAGNOSED WITH AN UNKNOWN TYPE OF CANCER: INITIAL 812.92, REPEAT 1717.82 U/ML (111% SHIFT). THERE WAS NO NEGATIVE IMPACT TO PATIENT MANAGEMENT REPORTED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9865054 ↗

Other SYSTEM, TEST, CARBOHYDRATE ANTIGEN (CA19-9), FOR MONITORING AND MANAGEMENT OF PANCREATIC CANCER devices

All SYSTEM, TEST, CARBOHYDRATE ANTIGEN (CA19-9), FOR MONITORING AND MANAGEMENT OF PANCREATIC CANCER devices →