Skip to content

ARCHITECT 02R9825Recalls & reported problems

Abbott Ireland · ARCHITECT STAT High Sensitivity Troponin-I Reagent Kit 100 Tests · Class 2

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00380740175498), received between Mar 7, 2023 and Nov 16, 2023 — classified as 2 malfunctions.

The most frequently reported specific device problem is “High Test Results” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • High Test Results2

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 16, 2023High Test Results; High Test Results

    THE CUSTOMER OBSERVED A FALSELY ELEVATED ARCHITECT STAT HIGH SENSITIVITY TROPONIN-I RESULT FOR ONE MALE PATIENT. THE FOLLOWING DATA WAS PROVIDED (NORMAL PATIENT RANGE MALE 0 - 0.033 NG/ML): SID (B)(6) INITIAL RESULT 0.316 NG/ML, REPEATED 0.022 / 0.033 / 0 / 0 / 0 / 0.008 NG/ML. REPEATED ON DIFFERENT ARCHITECT WITH A RESULT OF 0 NG/ML. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18152908 ↗
  • MalfunctionMar 7, 2023High Test Results; High Test Results

    THE CUSTOMER OBSERVED A FALSELY ELEVATED ARCHITECT STAT TROPONIN-I RESULT FOR ONE PATIENT SAMPLE ON THE ARCHITECT I2000SR, SERIAL NUMBER (B)(4). THE SAMPLE WAS REPEATED AND THE RESULT WAS IN THE CUSTOMERS NORMAL RANGE. THE FOLLOWING DATA WAS PROVIDED: INITIAL RESULT = 21 NG/L, REPEAT RESULTS ON ISR64746 <4, <4, AND <4 RESULT FROM ANOTHER ANALYZER = <4 NG/L CUSTOMER¿S REFERENCE RANGE IS <4 = NORMAL NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16495511 ↗

Other IMMUNOASSAY METHOD, TROPONIN SUBUNIT devices

All IMMUNOASSAY METHOD, TROPONIN SUBUNIT devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.