ARCHITECT 02R9825Recalls & reported problems
Abbott Ireland · ARCHITECT STAT High Sensitivity Troponin-I Reagent Kit 100 Tests · Class 2
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00380740175498), received between Mar 7, 2023 and Nov 16, 2023 — classified as 2 malfunctions.
The most frequently reported specific device problem is “High Test Results” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- High Test Results2
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 16, 2023High Test Results; High Test Results
THE CUSTOMER OBSERVED A FALSELY ELEVATED ARCHITECT STAT HIGH SENSITIVITY TROPONIN-I RESULT FOR ONE MALE PATIENT. THE FOLLOWING DATA WAS PROVIDED (NORMAL PATIENT RANGE MALE 0 - 0.033 NG/ML): SID (B)(6) INITIAL RESULT 0.316 NG/ML, REPEATED 0.022 / 0.033 / 0 / 0 / 0 / 0.008 NG/ML. REPEATED ON DIFFERENT ARCHITECT WITH A RESULT OF 0 NG/ML. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18152908 ↗MalfunctionMar 7, 2023High Test Results; High Test Results
THE CUSTOMER OBSERVED A FALSELY ELEVATED ARCHITECT STAT TROPONIN-I RESULT FOR ONE PATIENT SAMPLE ON THE ARCHITECT I2000SR, SERIAL NUMBER (B)(4). THE SAMPLE WAS REPEATED AND THE RESULT WAS IN THE CUSTOMERS NORMAL RANGE. THE FOLLOWING DATA WAS PROVIDED: INITIAL RESULT = 21 NG/L, REPEAT RESULTS ON ISR64746 <4, <4, AND <4 RESULT FROM ANOTHER ANALYZER = <4 NG/L CUSTOMER¿S REFERENCE RANGE IS <4 = NORMAL NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16495511 ↗
Other IMMUNOASSAY METHOD, TROPONIN SUBUNIT devices
- i-STAT 03P90-25Abbott Point Of Care Inc. · IMMUNOASSAY METHOD, TROPONIN SUBUNIT474 reports
- Atellica IM TNIH 10997841Siemens Healthcare Diagnostics Inc. · IMMUNOASSAY METHOD, TROPONIN SUBUNIT463 reports
- Access hsTnI B52669Beckman Coulter, Inc. · Immunoassay Method, Troponin Subunit2 recalls170 reports
- i-STAT 06P23-25Abbott Point Of Care Inc. · IMMUNOASSAY METHOD, TROPONIN SUBUNIT195 reports
- VITROS 6802301Ortho Clinical Diagnostics · IMMUNOASSAY METHOD, TROPONIN SUBUNIT1 recall139 reports
- Atellica IM TNIH 10997840Siemens Healthcare Diagnostics Inc. · IMMUNOASSAY METHOD, TROPONIN SUBUNIT166 reports
- Dimension EXL LOCI Troponin I High Sensitive 10471068Siemens Healthcare Diagnostics Inc. · IMMUNOASSAY METHOD, TROPONIN SUBUNIT1 recall76 reports
- Access AccuTnI+3 A98143Beckman Coulter, Inc. · Immunoassay Method, Troponin Subunit1 death106 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.