Skip to content

ARCHITECT 02R9835Recalls & reported problems

Abbott Ireland · ARCHITECT STAT High Sensitivity Troponin-I Reagent Kit 500 Tests · Class 2

Adverse event reports
5
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00380740175504), received between Nov 16, 2021 and Nov 17, 2025 — classified as 5 malfunctions.

The most frequently reported specific device problem is “High Test Results” (5 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • High Test Results5

Patient problems

  • No Clinical Signs, Symptoms or Conditions4
  • Cardiac Enzyme Elevation1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 17, 2025High Test Results; High Test Results

    THE CUSTOMER OBSERVED FALSELY ELEVATED ARCHITECT STAT HIGH SENSITIVITY TROPONIN-I AND PROVIDED THE FOLLOWING DATA: A 17 YEAR OLD MALE PRESENTED WITH CHEST PAIN RADIATING TO HIS ARMS. THE PATIENT HAD EKG'S, CARDIAC MRI, AND AN ECHO ALONG WITH BLOOD WORK. THERE WAS NO PATIENT HARM. THE PATIENT WAS DIAGNOSED WITH MYOCARDITIS. THE CUSTOMER REPORTED THAT THE CARDIAC MRI WAS DUE TO TROPONIN RESULT AND EKG AND THAT THE DIAGNOSTIC WORK-UP WAS DUE TO THE SYMPTOMS RELATED TO THE MYOCARDITIS DIAGNOSIS. SAMPLES WERE SENT FOR FURTHER TESTING AND THE RESULTS WERE LOWER AND THE CUSTOMER REPORTED THIS CORRELATED WITH PATIENT EKG, MRI, AND ECHO RESULTS. THE CUSTOMER PROVIDED THE FOLLOWING LABORATORY DATA: ON (B)(6) 2025, SAMPLE ID (B)(6), RESULT WAS >50,000 NG/L. TROPONIN T WAS 145 AND BECKMAN PLATFORM RESULT WAS 6,280 NG/L. ON (B)(6) 2025, SAMPLE ID (B)(6), RESULT WAS 820 NG/L. TROPONIN T WAS 18 AND BECKMAN PLATFORM RESULT WAS 26.1 NG/L. PATIENT NORMAL RANGES: MALE 4 - 34 NG/L. . THERE WAS NO FURTHER IMPACT TO PATIENT MANAGEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23570027 ↗
  • MalfunctionMay 30, 2025High Test Results; High Test Results

    THE CUSTOMER REPORTED FALSELY ELEVATED ARCHITECT STAT HIGH SENSITIVITY TROPONIN-I AND PROVIDED THE FOLLOWING DATA: PATIENT INFORMATION: 40-YEAR-OLD FEMALE. PATIENT PAST MEDICAL HISTORY INCLUDES TREATMENT FOR SYPHILIS. THE PATIENT ALSO HAS HAD A NON-PRODUCTIVE COUGH FOR THE PAST MONTH. ON (B)(6) 2025, THE PATIENT PRESENTED WITH CHEST DISCOMFORT, SHORTNESS OF BREATH, AND A FEELING OF NERVOUS/ANXIETY. EKG SHOWED NORMAL SINUS RHYTHM, NONSPECIFIC T WAVE ABNORMALITY, INVERTED T WAVE III AND IT WAS DETERMINED THAT THE EKG AS ABNORMAL. INITIAL TROPONIN WAS 17, 380 AND REPEAT ON 15 MAY WAS 12,000 (CUSTOMER NORMAL RANGE <18). THE PATIENT WAS BROUGHT TO THE CARDIAC CATHETERIZATION LAB, WHICH THE PATIENT REPORTEDLY HAD A NEGATIVE CATH RESULT. ADDITIONAL INFORMATION: PATIENT ALSO HAD ELEVATED D-DIMER FOR WHICH CTA OF THE THORAX WAS PERFORMED WHICH SHOWED: NO EVIDENCE OF PULMONARY EMBOLISM, MULTIFOCAL BILATERAL MILD GROUND GLASS OPACITIES WITH ASSOCIATED AREAS OF BRONCHIAL WALL THICKENING, SUGGESTIVE OF AN INFECTIOUS OR INFLAMMATORY PROCESS/ BRONCHIOLITIS. CLINICAL CORRELATION RECOMMENDED. THE DOCTOR CALLED QUESTIONING THE RESULT AFTER THE NEGATIVE CATH LAB. THE CUSTOMER THEN SENT THE SAMPLE TO

    Patient problems: Cardiac Enzyme Elevation

    MAUDE report 22114716 ↗
  • MalfunctionNov 8, 2023High Test Results; High Test Results

    THE CUSTOMER REPORTED A FALSELY ELEVATED ARCHITECT STAT HIGH SENSITIVE TROPONIN-I RESULT ON A 68-YR OLD FEMALE PATIENT WHO PRESENTED TO THE EMERGENCY ROOM WITH CHEST PAIN. THE FOLLOWING RESULTS WERE PROVIDED: SID (B)(6) ¿ (B)(6) 2023 13:35 = 388 NG/L, REPEATED SAME SAMPLE AT 19:26 = < 4 NG/L. SUBSEQUENT DRAWS: SID - DATE/TIME COLLECTED RESULT (B)(6) ¿ (B)(6) 2023 14:35, 4 NG/L. (B)(6) ¿ (B)(6) 2023 15:40 , 4 NG/L. NORMAL REFERENCE RANGE: FEMALE <18 NG/L THE PATIENT HAD AN ABNORMAL ECG T-WAVE. PATIENT WAS ADMINISTERED THE FOLLOWING: ASPIRIN, PLAVIX, HEPARIN DRIP FOR MANAGEMENT OF ACS. THE PATIENT WAS DISCHARGED ON (B)(6) 2023 WITH A DISCHARGE CONDITION OF: GOOD. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18090784 ↗
  • MalfunctionMar 25, 2022High Test Results; High Test Results

    THE CUSTOMER OBSERVED A FALSE POSITIVE ARCHITECT STAT HIGH SENSITIVE TROPONIN-I RESULT FOR A PATIENT SAMPLE WHILE RUNNING ON THE ARCHITECT I2000 ANALYZER. THE FOLLOWING DATA WAS PROVIDED: ON (B)(6) 2022 SID (B)(6): INITIAL TROPONIN-I RESULT: 116 NG/L (CUSTOMER¿S CUTOFF > 35NG/L; REPEAT ON ANOTHER ARCHITECT: 7.4 NG/L; RE-CENTRIFUGED SAMPLE AND REPEATED ON ORIGINAL ANALYZER: 7.8 NG/ML THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13906229 ↗
  • MalfunctionNov 16, 2021High Test Results; High Test Results

    THE CUSTOMER OBSERVED FALSELY ELEVATED ARCHITECT STAT HIGH SENSITIVITY TROPONIN-I RESULTS FOR SEVERAL PATIENTS. THE ELEVATED RESULTS WERE STILL ELEVATED AFTER REPEAT TESTING AND WERE QUESTIONED BY PHYSICIANS. THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12823682 ↗

Other IMMUNOASSAY METHOD, TROPONIN SUBUNIT devices

All IMMUNOASSAY METHOD, TROPONIN SUBUNIT devices →