Abbott GmbH · Alinity s Anti-HCV II Reagent Kit

All models of Alinity (37)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00380740177706), received between Jun 5, 2024 and Jul 23, 2025 — classified as 5 malfunctions.

The most frequently reported specific device problem is “False Positive Result” (5 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 2 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 2 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • False Positive Result5

Patient problems

  • No Clinical Signs, Symptoms or Conditions5

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 23, 2025False Positive Result; False Positive Result

    THE CUSTOMER OBSERVED FALSE REACTIVE ALINITY S ANTI-HCV II RESULTS FOR NUMEROUS TISSUE AND ORGAN DONORS. THE CUSTOMER REPORTED THAT FOR TISSUE DONORS FROM (B)(6) 2025, 24 DONOR SAMPLES WERE REPEAT REACTIVE WITH THE ALINITY S ANTI-HCV ASSAY. THOSE 24 REACTIVE SAMPLES WERE TESTED WITH NAT AND 19 SAMPLES WERE NEGATIVE. THOSE 19 CONFIRMED NEGATIVE SAMPLES WERE THEN TESTED WITH AN ORTHO ANTI-HCV ASSAY AND 13 WERE NONREACTIVE AND 6 WERE REACTIVE. THE CUSTOMER WAS USING ORTHO ANTI-HCV AS THEIR PRIMARY MEANS OF TESTING TISSUE DONORS FROM (B)(6) 2024 TO (B)(6), 2025, AND THERE WERE 13 DONOR SAMPLES THAT WERE REPEAT REACTIVE. THOSE 13 SAMPLES WERE TESTED WITH NAT AND 8 WERE NEGATIVE. THE 8 CONFIRMED NEGATIVES SAMPLES WERE TESTED WITH THE ALINITY S ANTI-HCV ASSAY AND 3 WERE NONREACTIVE AND 5 WERE REPEAT REACTIVE. THE CUSTOMER REPORTED THAT FOR ORGAN DONORS FROM (B)(6) 2025, 20 DONOR SAMPLES WERE REPEAT REACTIVE WITH THE ALINITY S ANTI-HCV ASSAY. THOSE 20 REACTIVE SAMPLES WERE TESTED WITH NAT AND 8 SAMPLES WERE NEGATIVE. THOSE 8 CONFIRMED NEGATIVE SAMPLES WERE THEN TESTED WITH AN ORTHO ANTI-HCV ASSAY AND 3 WERE NONREACTIVE AND 5 WERE REACTIVE. THE CUSTOMER WAS USING ORTHO ANTI-HCV AS THEIR PRI

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22582816 ↗
  • MalfunctionJul 23, 2025False Positive Result; False Positive Result

    THE CUSTOMER OBSERVED FALSE REACTIVE ALINITY S ANTI-HCV II RESULTS FOR NUMEROUS TISSUE AND ORGAN DONORS. THE CUSTOMER REPORTED THAT FOR TISSUE DONORS FROM (B)(6) 2025, 24 DONOR SAMPLES WERE REPEAT REACTIVE WITH THE ALINITY S ANTI-HCV ASSAY. THOSE 24 REACTIVE SAMPLES WERE TESTED WITH NAT AND 19 SAMPLES WERE NEGATIVE. THOSE 19 CONFIRMED NEGATIVE SAMPLES WERE THEN TESTED WITH AN ORTHO ANTI-HCV ASSAY AND 13 WERE NONREACTIVE AND 6 WERE REACTIVE. THE CUSTOMER WAS USING ORTHO ANTI-HCV AS THEIR PRIMARY MEANS OF TESTING TISSUE DONORS FROM (B)(6) 2024 TO (B)(6) 2025, AND THERE WERE 13 DONOR SAMPLES THAT WERE REPEAT REACTIVE. THOSE 13 SAMPLES WERE TESTED WITH NAT AND 8 WERE NEGATIVE. THE 8 CONFIRMED NEGATIVES SAMPLES WERE TESTED WITH THE ALINITY S ANTI-HCV ASSAY AND 3 WERE NONREACTIVE AND 5 WERE REPEAT REACTIVE. THE CUSTOMER REPORTED THAT FOR ORGAN DONORS FROM (B)(6) 2025, 20 DONOR SAMPLES WERE REPEAT REACTIVE WITH THE ALINITY S ANTI-HCV ASSAY. THOSE 20 REACTIVE SAMPLES WERE TESTED WITH NAT AND 8 SAMPLES WERE NEGATIVE. THOSE 8 CONFIRMED NEGATIVE SAMPLES WERE THEN TESTED WITH AN ORTHO ANTI-HCV ASSAY AND 3 WERE NONREACTIVE AND 5 WERE REACTIVE. THE CUSTOMER WAS USING ORTHO ANTI-HCV AS THEIR PRIM

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22582724 ↗
  • MalfunctionMar 20, 2025False Positive Result; False Positive Result

    THE CUSTOMER OBSERVED FALSE REPEAT REACTIVE ALINITY S ANTI-HCV II RESULTS FOR 21 SPECIMENS, WHILE THE HCV RESULTS WITH GRIFOLS NAT AND/OR ORTHO ELISA WERE NEGATIVE. THERE WAS NO SPECIFIC ACTUAL PATIENT OR DONOR RESULTS PROVIDED BY THE CUSTOMER. THERE WAS NO REPORTED IMPACT ON PATIENT OR DONOR MANAGEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21652244 ↗
  • MalfunctionMar 20, 2025False Positive Result; False Positive Result

    THE CUSTOMER OBSERVED FALSE REPEAT REACTIVE ALINITY S ANTI-HCV II RESULTS FOR 21 DONORS, WHILE THE HCV RESULTS WITH GRIFOLS NAT AND/OR ORTHO ELISA WERE NEGATIVE. THERE WAS NO SPECIFIC ACTUAL PATIENT OR DONOR RESULTS PROVIDED BY THE CUSTOMER. THERE WAS NO REPORTED IMPACT TO PATIENT MANAGEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21652059 ↗
  • MalfunctionJun 5, 2024False Positive Result; False Positive Result

    THE CUSTOMER OBSERVED FALSE REACTIVE ALINITY S ANTI-HCV II RESULTS FOR LIVING ORGAN SAMPLES THAT WERE NOT CONSISTENT WITH NAT TESTING. THE FOLLOWING DATA WAS PROVIDED: (B)(6) 2024 SAMPLE ID (B)(6) INITIAL RESULT ON ANALYZER (B)(6) = 87.59 S/CO, REPEAT RESULT = 86.24 S/CO, 87.93 S/CO, 89.36 S/CO, 88.67 S/CO (B)(6) 2024 SAMPLE ID (B)(6) INITIAL RESULT ON ANALYZER (B)(6) = 82.81 S/CO, REPEAT RESULTS = 83.54 S/CO, 84.31 S/CO, 79.80 S/CO PATIENT MEDICAL HISTORY: BOTH ORGAN DONORS HAVE A KNOWN PAST MEDICAL HISTORY OF HCV AND/OR HBV. CUSTOMER STATED THAT DUE TO INCREASED REACTIVE RESULTS, THEY WANT ALL POSITIVES/REACTIVES INVESTIGATED. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19466482 ↗

Common questions about the Alinity 04W5660

Has the Alinity 04W5660 been recalled?

No FDA recall record in our data names this exact device identifier (DI 00380740177706), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the Alinity 04W5660?

The FDA's MAUDE database lists 5 adverse event reports for this device, received between Jun 5, 2024 and Jul 23, 2025. In those reports, the most common specific device problem is “False Positive Result” (5). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Alinity 04W5660?

None of the reports are classified as deaths, 0 as injuries and 5 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Who makes the Alinity 04W5660, and is it still sold?

The FDA's device database lists Abbott GmbH as the company responsible for it (brand Alinity, model 04W5660). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.