Skip to content

AlinIQ 03R8965Recalls & reported problems

Abbott Srl · AlinIQ AMS · Class 1

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00380740212575), received between Apr 2, 2025 and Mar 9, 2026 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Computer Software Problem” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 2 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Computer Software Problem4

Patient problems

  • No Clinical Signs, Symptoms or Conditions4

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 9, 2026Computer Software Problem

    THE CUSTOMER OBSERVED THE ALINIQ AMS REPORTED THE HPVOTHER, WHICH IS A COMBINED PLACEHOLDER FOR THE HPV OTHER A AND HPV OTHER B, AS NEGATIVE RESULT EVEN THOUGH THE HPV RESULT WAS OTHER HR HPV A, SHOULD HAVE RESULTED AS POSITIVE. HPV ¿OTHER¿ RESULTS ARE CONSOLIDATED INTO A SINGLE OUTPUT TEST (HPVOTHER), WHICH CONFLICTED WITH THE EXISTING VALIDATION RULE BY EVALUATING MULTIPLE OUTCOMES (OTHER A AND OTHER B) AGAINST ONE COMBINED RESULT. ALL PATIENT RESULTS WERE BLOCKED AND HELD IN THE LIS AND NO RESULTS WERE RELEASED OUT OF THE LABORATORY. A QPL ADJUSTMENT WAS IMPLEMENTED TO ENSURE THE RULE CORRECTLY INTERPRETS THE COMBINED HPV OTHER RESULTS AND PRODUCES A CONSISTENT OUTCOME. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24544175 ↗
  • MalfunctionJan 6, 2026Computer Software Problem

    THE CUSTOMER REPORTED THE ALINIQ AMS AUTO-VALIDATED A DISCREPANT ALINITY I AFP (ALPHA1 FETOPROTEIN) RESULT FOR ONE PATIENT SAMPLE. MULTIPLE SAMPLES WERE BEING RUN FOR THE ALINITY I AFP (ALPHA1 FETOPROTEIN) AND THE ANALYZER GENERATED ERROR CODE 1094 - UNABLE TO PROCESS TEST, NO REAGENT CARTRIDGE AVAILABLE. ALL SAMPLES WERE REPEATED WITH A DILUTION FACTOR UPDATED BY THE USER, AND GENERATED RESULTS OF <16.6. FOUR OUT OF FIVE OF THE SAMPLES WERE HELD IN AMS AND MANUALLY VALIDATED BY THE CUSTOMER, AND ONE SAMPLE AUTO-VALIDATED. THE FOLLOWING INFORMATION WAS PROVIDED: SID (B)(6) - RESULT VALIDATED BY USER = <16.6, CORRECT RESULT = 4.7. SID (B)(6) - VALIDATED BY USER = <16.6, CORRECT RESULT = <1.7. SID (B)(6) - VALIDATED BY USER = <16.6, CORRECT RESULT = 9.5. SID (B)(6) - AUTO-VALIDATED RESULT = <16.6, CORRECT RESULT = 4.8. SID (B)(6) - VALIDATED BY USER = <16.6, CORRECT RESULT = 5.7. THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23973549 ↗
  • MalfunctionApr 15, 2025Computer Software Problem; Computer Software Problem

    THE CUSTOMER REPORTED DISCREPANCIES IN HBSAG RESULTS FOR TWO SAMPLES (SID (B)(6) WHERE THE AMS SYSTEM REPORTED A VALUE OF 54, WHICH DIFFERS SIGNIFICANTLY FROM THE ORIGINAL VALUES (100.862 AND 5821.777, RESPECTIVELY). THE RESULTS WERE INCORRECT DUE TO MANUALLY OVERWRITING OR INSERTED WITH THE BARCODE SCANNER. ADDITIONALLY, THE BARCODE SCANNERS AT THE AMS WORKSTATIONS WERE DISABLED DUE TO CONTINUOUS SCANNING OF RANDOM BARCODES, WHICH COULD INADVERTENTLY ALTER SAMPLE RESULTS. MANUAL EDITS WERE FOUND IN THE EVENT HISTORY FOR VARIOUS SPECIMENS, ALL LINKED TO THE AMS USER ACCOUNT "A". THE EXAMPLES OF THIS WERE PROVIDED: SAMPLE: (B)(6), SPECIMEN: ANTI-HCV, RESULT: 1537.00. SPECIMEN: ALBBCG, RESULT: 15.00, REQUEST: 270325436. SPECIMEN: CHE, RESULT: 9.29, REQUEST: 030425113. SPECIMEN: ALKP, RESULT: 78.00, REQUEST: 270325019. SPECIMEN: ALT, RESULT: 78.00, REQUEST: 270325019. SPECIMEN: AST, RESULT: 78.00, REQUEST: 270325019. THE CUSTOMER CONFIRMED THAT THE INCORRECT RESULTS WERE NOT RELEASED OUT OF THE LAB. THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21844270 ↗
  • MalfunctionApr 2, 2025Computer Software Problem; Computer Software Problem

    THE CUSTOMER REPORTED THE ALINIQ AMS WILL CANCEL THE HIL INDEX (HEMOLYSIS, ICTERIC, LIPEMIA) ORDER ON SAMPLES THAT HAD A CORRESPONDING ASSAY THAT WAS CANCELED. THIS WAS IDENTIFIED WHEN THE CUSTOMER FOUND MULTIPLE SAMPLES THAT HAD NO HIL MEASUREMENTS COMPLETED ON THEM. AFTER FURTHER REVIEW, AN ALINIQ AMS RULE WAS CONFIGURED INCORRECTLY. THIS RULE IS RESPONSIBLE FOR ADDING HIL ORDERS TO CERTAIN DEFINED TESTS THAT THE LIS SENDS TO THE AMS. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21751073 ↗

FDA clearance and classification

The FDA regulates product code JQP (calculator/data processing module, for clinical use) under 21 CFR 862.2100 as a Class I device. 21 CFR 862.2100 ↗

Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the AlinIQ 03R8965

Has the AlinIQ 03R8965 been recalled?

No FDA recall record in our data names this exact device identifier (DI 00380740212575), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the AlinIQ 03R8965?

The FDA's MAUDE database lists 4 adverse event reports for this device, received between Apr 2, 2025 and Mar 9, 2026. In those reports, the most common specific device problem is “Computer Software Problem” (4). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the AlinIQ 03R8965?

None of the reports are classified as deaths, 0 as injuries and 4 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the AlinIQ 03R8965 FDA approved?

Devices of this type (product code JQP) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Abbott Srl.

Who makes the AlinIQ 03R8965, and is it still sold?

The FDA's device database lists Abbott Srl as the company responsible for it (brand AlinIQ, model 03R8965). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Calculator/data processing module, for clinical use devices

All Calculator/data processing module, for clinical use devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page