Skip to content

Natus OBM00002Recalls & reported problems

Xltek · OBM Data Acquisition Box (DAB) · Class 2

All models of Natus (13)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00382830010825), received between May 4, 2023 and Jan 11, 2024 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Device Markings/Labelling Problem” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. May 7, 2025Open, Classified

    Electrode placement labels on the Digital Acquisition Box (DAB) were applied backwards. This may ultimately lead to misdiagnosis of seizure location in neonates monitored with this device.

    Z-1925-2025 · Root cause: Employee error · Recall details · FDA record ↗

Most reported problems

Device problems

  • Device Markings/Labelling Problem1
  • Failure of Device to Self-Test1
  • No Display/Image1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 11, 2024Device Markings/Labelling Problem; Device Markings/Labelling Problem

    PART OBM00002 OBM DAB (DIGITAL ACQUISITION BOX) MISLABELED. THE CUSTOMER REPORTED EXPERIENCING ISSUES WITH THEIR OBM (OLYMPIC BRAINS MONITOR) HAVING THE ELECTRODE PLACEMENT STICKER ON BACKWARDS, RESULTING IN THE ELECTRODES BEING APPLIED INCORRECTLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18490690 ↗
  • MalfunctionJun 12, 2023Failure of Device to Self-Test; Failure of Device to Self-Test

    PART OBM00002 OBM DAB (DIGITAL ACQUISITION BOX) - OBM DAB (DIGITAL ACQUISITION BOX) - THE CUSTOMER REPORTED A SIGNAL, "FAILED SELF CHECK." PER QUESTIONNAIRE, THE MEDICAL TEAM WERE WORKING WITH A BABY WHEN THE OBM SIGNALED "FAILED SELF CHECK" FOR 20 -30 MINUTES. THE REGISTERED NURSE TRIED TO TROUBLESHOOT, INCLUDING CHANGING TO A DIFFERENT USB. THE OBM UNIT FINALLY WORKED AFTER 7-8 ATTEMPTS BUT THEY WERE UNCLEAR WHAT MADE THE DIFFERENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17112782 ↗
  • MalfunctionMay 4, 2023No Display/Image; No Display/Image

    PART OBM00002 OBM DAB (DIGITAL ACQUISITION BOX): NATUS OLYMPIC BRAINZ MONITOR WAS CONNECTED TO A PATIENT AND FUNCTIONING. SUDDENLY THE MONITOR STOPPED WORKING, STARTED BEEPING, AND THE MONITORING STOPPED. UNABLE TO RESUME MONITORING AFTER ALL CONNECTIONS WERE CHECKED, MONITOR TURNED OFF AND BACK ON. NO PATIENT OR USER HARMED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16873181 ↗