Natus 046621Recalls & reported problems
Xltek · BrainWatch Amplifier · Class 2
All models of Natus (13)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00382830045742), received between Jul 17, 2025 and Jul 17, 2025 — classified as 1 malfunction.
The most frequently reported specific device problem is “Inaccurate Information” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Inaccurate Information1
- Incorrect, Inadequate or Imprecise Result or Readings1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 17, 2025Incorrect, Inadequate or Imprecise Result or Readings; Inaccurate Information; Incorrect, Inadequate or Imprecise Result or Readings; Inaccurate Information
AMPLIFIER, BRAINWATCH AND PERSYST ANALYZER -FAULTY SEIZURE DETECTIONS. THE CUSTOMER INDICATES FALSE SEIZURE BURDEN DETECTIONS ON STUDY (FALSE POSITIVE). FEEDBACK FROM THE CUSTOMER: IN ONE STUDY, THE BRAIWATCH SYSTEM CONCLUDED THAT THE PATIENT WAS HAVING A SEIZURE. THE HEATHCARE TEAM TREATED THE PATIENT FOR SEIZURE. LATER ON, WHILE NEUROLOGIST REVIEWED EEG STUDY, HE DETERMINED THAT IT'S ALL ARTIFACT AND NOT SEIZURE. IT WAS CONFIRMED BY RERUNNING THE STUDY IN CONVENTIONAL EEG SYSTEM (MODEL UNKNOWN) THAT PATIENT HAD NOT HAD A SEIZURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22530125 ↗
Other Reduced- Montage Standard Electroencephalograph devices
- NA B125 VSP3.0 Patient MonitorGE Medical Systems Information Technologies, Inc. · Reduced- montage standard electroencephalograph2 recalls0 reports
- BIS PMB4000DOCCovidien LP · Reduced- montage standard electroencephalograph4 reports
- Sedline 9829Masimo Corporation · Reduced- montage standard electroencephalograph2 reports
- Natus OBM00401Xltek · Reduced- Montage Standard Electroencephalograph1 report
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page