Skip to content

NATUS EDS3 COLLECT BAG KIT 5 BAGS NT821733Recalls & reported problems

Natus Medical Incorporated · NATUS EDS3 COLLECT BAG KIT 5 BAGS · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00382830050562), received between May 24, 2025 and May 24, 2025 — classified as 1 malfunction.

The most frequently reported specific device problem is “Break” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

3 other recalls cover devices cleared under the same FDA submission (K061568); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Nov 28, 2023Class IIOpen, Classified

    Sterility assurance cannot be guaranteed for external drainage systems due to incomplete bioburden testing.

    Z-0634-2024 · Root cause: Process change control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Break1
  • Crack1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 24, 2025Break; Crack; Break; Crack

    NT821732C- PATIENT HAD EVD IN PLACE FROM OR. 2 RNS ATTEMPTED TO CHANGE DRAIN BAG AND EXPERIENCED DIFFICULTY IN DISCONNECTING BAG FROM TUBING. ALCOHOL WAS USED TO TRY TO LOOSEN CONNECTION. DRAIN BAG WITH CONTENTS WAS REMOVED FROM HOLDER TO TROUBLESHOOT CONNECTION. AFTER SEVERAL ATTEMPTS, BAG WAS REMOVED FROM TUBING. TUBING WAS ASSESSED AND NO ISSUES NOTED. NEW BAG ATTACHED. RN WENT TO CLEAN CONNECTION WITH CHG AFTER MANIPULATION/CHANGING OF BAG AS EXTRA PRECAUTION DUE TO MULTIPLE ATTEMPTS TO REPLACE BAG. WHEN THE CHG SWAB TOUCHED THE TUBING, THE TUBING IMMEDIATELY BROKE OFF THE BAG. THE MALE END OF THE TUBING BROKE OFF INTO THE FEMALE END OF THE BAG. REQUIRED TUBING CHANGE AT DRAIN SITE BY NEUROSURGERY TEAM, NEW EVD BAG, AND PATIENT STARTED ON ANTIBIOTICS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22082042 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K061568. One submission often covers many models, so check the product and lot details in each record.

  • Z-0091-2015 · Sep 15, 2014 · Terminated
    Codman Lumbar Drainage Catheter Kit II with EDS 3 (External Drainage System) Catalog Number: 82-1738 Indicated for draining cerebrospinal fluid (CFS) and other fluids of similar p
  • Z-0090-2015 · Sep 15, 2014 · Terminated
    Codman EDS 3 Drainage System (no Ventricular Catheter) (External Drainage System) Catalog Number: 82-1731 Indicated for draining cerebrospinal fluid (CFS) and other fluids of simil
  • Z-0089-2015 · Sep 15, 2014 · Terminated
    Codman EDS 3 Drainage System with Ventricular Catheter (External Drainage System) Catalog Number: 82-1730 Indicated for draining cerebrospinal fluid (CFS) and other fluids of simil

Other Shunt, Central Nervous System And Components devices

All Shunt, Central Nervous System And Components devices →