Skip to content

neoBLUE CozyRecalls & reported problems

Natus Medical Incorporated · The neoBLUE cozy™ Phototherapy System is a mobile phototherapy light that delivers a narrow band of high-intensity blue light via blue light emitting diodes (LEDs) to provide treatment for neonatal hyperbilirubinemia. The neoBLUE cozy light was designed for use within a patient enclosure, positioned under the baby · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00382830053198), received between Aug 9, 2018 and Aug 9, 2018 — classified as 1 malfunction.

The most frequently reported specific device problem is “Electromagnetic Compatibility Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Electromagnetic Compatibility Problem1

Patient problems

  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 9, 2018Electromagnetic Compatibility Problem; Electromagnetic Compatibility Problem

    IT WAS REPORTED TO NATUS THAT A NEOBLUE COZY UNIT WAS GENERATING EXCESSIVE ARTIFACTS IN EKG EQUIPMENT WHILE EKG MONITORING. THE COMPLAINANT TRIED MULTIPLE CONFIGURATIONS OF THE EKG EQUIPMENT AND THE NEOBLUE COZY TO LIMIT THE ARTIFACT BUT WAS UNSUCCESSFUL. IT WAS REPORTED THAT THE ARTIFACT ONLY OCCURRED IN THE PRESENCE OF A NEOBLUE COZY AND NOT IN THE PRESENCE OF AN OVERHEAD NEOBLUE UNIT. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR ENVIRONMENTAL/SAFETY CONCERNS.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7769051 ↗

Other Unit, Neonatal Phototherapy devices

All Unit, Neonatal Phototherapy devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page