BD BACTEC™ MGIT™ Mycobacteria Growth Indicator Tubes 245122 (UDI 00382902451228)Recalls & reported problems

Becton, Dickinson And Company · BD BBL™ MGIT™ Mycobacteria Growth Indicator Tubes, 7 mL, carton of 100 tubes · Class 1

All models of BD BACTEC MGIT Mycobacteria Growth Indicator Tubes (2)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This is a BD MGIT Mycobacteria Growth Indicator tube, used in culture testing. Most of the 83 reports describe incorrect or unreliable results, including false negatives and unexplained growth, and some describe contamination of the tubes. There are no deaths and no injuries. No recalls name this device. Reports rose sharply in the last 12 months, with 73 compared with 5 in the 12 months before.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
82
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 82 adverse event reports that list this exact device (DI 00382902451228), received between Jul 12, 2021 and Aug 13, 2026 — classified as 82 malfunctions.

The most frequently reported specific device problem is “Incorrect, Inadequate or Imprecise Result or Readings” (73 reports).

2 other recalls cover devices cleared under the same FDA submission (K974883); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 5 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 63 reportsApr 2026: 6 reportsMay 2026: 2 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reports20242025202663
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Incorrect, Inadequate or Imprecise Result or Readings73
  • Microbial Contamination of Device6
  • Device Ingredient or Reagent Problem1
  • Device Markings/Labelling Problem1
  • False Negative Result1

Patient problems

  • No Clinical Signs, Symptoms or Conditions82

Recent adverse event reports

The 25 most recent of 82 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 13, 2026Incorrect, Inadequate or Imprecise Result or Readings

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50240107 ↗
  • MalfunctionMay 22, 2026Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING BD BACTEC¿ MGIT¿ MYCOBACTERIA GROWTH INDICATOR TUBES, AN UNSPECIFIED NUMBER OF ATYPICAL PATIENT RESULTS WERE OBTAINED FROM A MGIT320 INSTRUMENT. IT WAS NOTED THE CUSTOMER RECEIVED "UNEXPLAINED GROWTH UNITS IN MGIT TUBES" AND NO ADDITIONAL INFORMATION IS AVAILABLE. THERE WERE NO HEALTH IMPACT OR CONSEQUENCES REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25257663 ↗
  • MalfunctionMay 22, 2026Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING BD BACTEC¿ MGIT¿ MYCOBACTERIA GROWTH INDICATOR TUBES, AN UNSPECIFIED NUMBER OF ATYPICAL PATIENT RESULTS WERE OBTAINED FROM A MGIT320 INSTRUMENT. IT WAS NOTED THE CUSTOMER RECEIVED "UNEXPLAINED GROWTH UNITS IN MGIT TUBES" AND NO ADDITIONAL INFORMATION IS AVAILABLE. THERE WERE NO HEALTH IMPACT OR CONSEQUENCES REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25257644 ↗
  • MalfunctionApr 21, 2026Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING BD BACTEC¿ MGIT¿ MYCOBACTERIA GROWTH INDICATOR TUBES, ONE (1) FALSE NEGATIVE PATIENT RESULT WAS OBTAINED FROM A MGIT INSTRUMENT. CONFIRMATORY TESTING WAS NOTED TO HAVE BEEN PERFORMED AND THE ERRONEOUS FALSE NEGATIVE RESULT WAS NOT REPORTED TO THE CLINICIAN. THERE WERE NO HEALTH IMPACT OR CONSEQUENCES REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24941053 ↗
  • MalfunctionApr 17, 2026Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING BD BACTEC¿ MGIT¿ MYCOBACTERIA GROWTH INDICATOR TUBES, AN UNSPECIFIED NUMBER OF FALSE NEGATIVE PATIENT RESULTS WERE OBTAINED FROM A MGIT INSTRUMENT. THE CUSTOMER ALSO DID NOT PROVIDE A BATCH NUMBER, BUT NOTED THIS ISSUE HAS OCCURRED WITH RANDOM BATCH NUMBERS. THE AFFECTED PATIENT SAMPLES ARE OBSERVED TO HAVE BACILLUS GROWTH AND ARE CONFIRMED POSITIVE FOR BACILLI BY BACILLOSCOPY. NO ERRONEOUS FALSE NEGATIVE RESULTS WERE RELEASED TO THE CLINICIAN. THERE WERE NO HEALTH IMPACT OR CONSEQUENCES REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24911580 ↗
  • MalfunctionApr 8, 2026Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING BD BACTEC¿ MGIT¿ MYCOBACTERIA GROWTH INDICATOR TUBES, ONE (1) FALSE NEGATIVE PATIENT RESULT WAS OBTAINED FROM A MGIT INSTRUMENT. THE ERRONEOUS RESULT WAS OBVIOUS TO THE CUSTOMER SINCE THERE WAS GROWTH ON SOLID MEDIA AND THE SAMPLE WAS FROM A KNOWN PATIENT. THERE WERE NO HEALTH IMPACT OR CONSEQUENCES REPORTED. THIS IS REPORT 2 OF 2.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24817160 ↗
  • MalfunctionApr 8, 2026Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING BD BACTEC¿ MGIT¿ MYCOBACTERIA GROWTH INDICATOR TUBES, ONE (1) FALSE NEGATIVE PATIENT RESULT WAS OBTAINED FROM A MGIT INSTRUMENT. THE ERRONEOUS RESULT WAS OBVIOUS TO THE CUSTOMER SINCE THERE WAS GROWTH ON SOLID MEDIA AND THE SAMPLE WAS FROM A KNOWN PATIENT. THERE WERE NO HEALTH IMPACT OR CONSEQUENCES REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24815092 ↗
  • MalfunctionApr 8, 2026Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING BD BACTEC¿ MGIT¿ MYCOBACTERIA GROWTH INDICATOR TUBES, ONE (1) FALSE NEGATIVE PATIENT RESULT WAS OBTAINED FROM A MGIT INSTRUMENT. THE ERRONEOUS RESULT WAS OBVIOUS TO THE CUSTOMER SINCE THERE WAS GROWTH ON SOLID MEDIA AND THE SAMPLE WAS FROM A KNOWN PATIENT. THERE WERE NO HEALTH IMPACT OR CONSEQUENCES REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24815091 ↗
  • MalfunctionApr 7, 2026Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING BD BACTEC¿ MGIT¿ MYCOBACTERIA GROWTH INDICATOR TUBES, ONE (1) NEGATIVE PATIENT RESULT WAS OBTAINED FROM A MGIT 960 INSTRUMENT. UPON PERFORMING PARALLEL TESTING WHICH INCLUDED CULTURING USING LOWENSTEIN¿JENSEN MEDIA AND TB-ICT, A POSITIVE RESULT WAS OBTAINED. THERE WERE NO HEALTH IMPACT OR CONSEQUENCES REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24801203 ↗
  • MalfunctionMar 26, 2026Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING THE BD BACTEC¿ MGIT¿ MYCOBACTERIA GROWTH INDICATOR TUBES, ONE (1) FALSE RESISTANT M. TUBERCULOSIS PATIENT RESULT WAS OBTAINED FROM A MGIT INSTRUMENT. THE CUSTOMER IS UNABLE TO PROVIDE THE NUMBER OF AFFECTED PATIENT SAMPLES OR TUBES USED WITH A SPECIFIC PRODUCT LOT. NO CONFIRMATORY TESTING WAS REPORTED; HOWEVER, THE CUSTOMER NOTED THAT NO ACTUAL GROWTH WAS OBSERVED. REPEAT TESTING OCCURRED AND NO ADVERSE HEALTH IMPACT WAS REPORTED. THIS IS REPORT 31 OF 61.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24700790 ↗
  • MalfunctionMar 26, 2026Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING THE BD BACTEC¿ MGIT¿ MYCOBACTERIA GROWTH INDICATOR TUBES, ONE (1) FALSE RESISTANT M. TUBERCULOSIS PATIENT RESULT WAS OBTAINED FROM A MGIT INSTRUMENT. THE CUSTOMER IS UNABLE TO PROVIDE THE NUMBER OF AFFECTED PATIENT SAMPLES OR TUBES USED WITH A SPECIFIC PRODUCT LOT. NO CONFIRMATORY TESTING WAS REPORTED; HOWEVER, THE CUSTOMER NOTED THAT NO ACTUAL GROWTH WAS OBSERVED. REPEAT TESTING OCCURRED AND NO ADVERSE HEALTH IMPACT WAS REPORTED. THIS IS REPORT 38 OF 61.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24700789 ↗
  • MalfunctionMar 26, 2026Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING THE BD BACTEC¿ MGIT¿ MYCOBACTERIA GROWTH INDICATOR TUBES, ONE (1) FALSE RESISTANT M. TUBERCULOSIS PATIENT RESULT WAS OBTAINED FROM A MGIT INSTRUMENT. THE CUSTOMER IS UNABLE TO PROVIDE THE NUMBER OF AFFECTED PATIENT SAMPLES OR TUBES USED WITH A SPECIFIC PRODUCT LOT. NO CONFIRMATORY TESTING WAS REPORTED; HOWEVER, THE CUSTOMER NOTED THAT NO ACTUAL GROWTH WAS OBSERVED. REPEAT TESTING OCCURRED AND NO ADVERSE HEALTH IMPACT WAS REPORTED. THIS IS REPORT 37 OF 61.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24700788 ↗
  • MalfunctionMar 26, 2026Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING THE BD BACTEC¿ MGIT¿ MYCOBACTERIA GROWTH INDICATOR TUBES, ONE (1) FALSE RESISTANT M. TUBERCULOSIS PATIENT RESULT WAS OBTAINED FROM A MGIT INSTRUMENT. THE CUSTOMER IS UNABLE TO PROVIDE THE NUMBER OF AFFECTED PATIENT SAMPLES OR TUBES USED WITH A SPECIFIC PRODUCT LOT. NO CONFIRMATORY TESTING WAS REPORTED; HOWEVER, THE CUSTOMER NOTED THAT NO ACTUAL GROWTH WAS OBSERVED. REPEAT TESTING OCCURRED AND NO ADVERSE HEALTH IMPACT WAS REPORTED. THIS IS REPORT 35 OF 61.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24700786 ↗
  • MalfunctionMar 26, 2026Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING THE BD BACTEC¿ MGIT¿ MYCOBACTERIA GROWTH INDICATOR TUBES, ONE (1) FALSE RESISTANT M. TUBERCULOSIS PATIENT RESULT WAS OBTAINED FROM A MGIT INSTRUMENT. THE CUSTOMER IS UNABLE TO PROVIDE THE NUMBER OF AFFECTED PATIENT SAMPLES OR TUBES USED WITH A SPECIFIC PRODUCT LOT. NO CONFIRMATORY TESTING WAS REPORTED; HOWEVER, THE CUSTOMER NOTED THAT NO ACTUAL GROWTH WAS OBSERVED. REPEAT TESTING OCCURRED AND NO ADVERSE HEALTH IMPACT WAS REPORTED. THIS IS REPORT 56 OF 61.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24700649 ↗
  • MalfunctionMar 26, 2026Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING THE BD BACTEC¿ MGIT¿ MYCOBACTERIA GROWTH INDICATOR TUBES, ONE (1) FALSE RESISTANT M. TUBERCULOSIS PATIENT RESULT WAS OBTAINED FROM A MGIT INSTRUMENT. THE CUSTOMER IS UNABLE TO PROVIDE THE NUMBER OF AFFECTED PATIENT SAMPLES OR TUBES USED WITH A SPECIFIC PRODUCT LOT. NO CONFIRMATORY TESTING WAS REPORTED; HOWEVER, THE CUSTOMER NOTED THAT NO ACTUAL GROWTH WAS OBSERVED. REPEAT TESTING OCCURRED AND NO ADVERSE HEALTH IMPACT WAS REPORTED. THIS IS REPORT 58 OF 61.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24700647 ↗
  • MalfunctionMar 26, 2026Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING THE BD BACTEC¿ MGIT¿ MYCOBACTERIA GROWTH INDICATOR TUBES, ONE (1) FALSE RESISTANT M. TUBERCULOSIS PATIENT RESULT WAS OBTAINED FROM A MGIT INSTRUMENT. THE CUSTOMER IS UNABLE TO PROVIDE THE NUMBER OF AFFECTED PATIENT SAMPLES OR TUBES USED WITH A SPECIFIC PRODUCT LOT. NO CONFIRMATORY TESTING WAS REPORTED; HOWEVER, THE CUSTOMER NOTED THAT NO ACTUAL GROWTH WAS OBSERVED. REPEAT TESTING OCCURRED AND NO ADVERSE HEALTH IMPACT WAS REPORTED. THIS IS REPORT 60 OF 61.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24700646 ↗
  • MalfunctionMar 26, 2026Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING THE BD BACTEC¿ MGIT¿ MYCOBACTERIA GROWTH INDICATOR TUBES, ONE (1) FALSE RESISTANT M. TUBERCULOSIS PATIENT RESULT WAS OBTAINED FROM A MGIT INSTRUMENT. THE CUSTOMER IS UNABLE TO PROVIDE THE NUMBER OF AFFECTED PATIENT SAMPLES OR TUBES USED WITH A SPECIFIC PRODUCT LOT. NO CONFIRMATORY TESTING WAS REPORTED; HOWEVER, THE CUSTOMER NOTED THAT NO ACTUAL GROWTH WAS OBSERVED. REPEAT TESTING OCCURRED AND NO ADVERSE HEALTH IMPACT WAS REPORTED. THIS IS REPORT 61 OF 61.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24700645 ↗
  • MalfunctionMar 26, 2026Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING THE BD BACTEC¿ MGIT¿ MYCOBACTERIA GROWTH INDICATOR TUBES, ONE (1) FALSE RESISTANT M. TUBERCULOSIS PATIENT RESULT WAS OBTAINED FROM A MGIT INSTRUMENT. THE CUSTOMER IS UNABLE TO PROVIDE THE NUMBER OF AFFECTED PATIENT SAMPLES OR TUBES USED WITH A SPECIFIC PRODUCT LOT. NO CONFIRMATORY TESTING WAS REPORTED; HOWEVER, THE CUSTOMER NOTED THAT NO ACTUAL GROWTH WAS OBSERVED. REPEAT TESTING OCCURRED AND NO ADVERSE HEALTH IMPACT WAS REPORTED. THIS IS REPORT 59 OF 61.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24700644 ↗
  • MalfunctionMar 26, 2026Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING THE BD BACTEC¿ MGIT¿ MYCOBACTERIA GROWTH INDICATOR TUBES, ONE (1) FALSE RESISTANT M. TUBERCULOSIS PATIENT RESULT WAS OBTAINED FROM A MGIT INSTRUMENT. THE CUSTOMER IS UNABLE TO PROVIDE THE NUMBER OF AFFECTED PATIENT SAMPLES OR TUBES USED WITH A SPECIFIC PRODUCT LOT. NO CONFIRMATORY TESTING WAS REPORTED; HOWEVER, THE CUSTOMER NOTED THAT NO ACTUAL GROWTH WAS OBSERVED. REPEAT TESTING OCCURRED AND NO ADVERSE HEALTH IMPACT WAS REPORTED. THIS IS REPORT 57 OF 61.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24700643 ↗
  • MalfunctionMar 26, 2026Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING THE BD BACTEC¿ MGIT¿ MYCOBACTERIA GROWTH INDICATOR TUBES, ONE (1) FALSE RESISTANT M. TUBERCULOSIS PATIENT RESULT WAS OBTAINED FROM A MGIT INSTRUMENT. THE CUSTOMER IS UNABLE TO PROVIDE THE NUMBER OF AFFECTED PATIENT SAMPLES OR TUBES USED WITH A SPECIFIC PRODUCT LOT. NO CONFIRMATORY TESTING WAS REPORTED; HOWEVER, THE CUSTOMER NOTED THAT NO ACTUAL GROWTH WAS OBSERVED. REPEAT TESTING OCCURRED AND NO ADVERSE HEALTH IMPACT WAS REPORTED. THIS IS REPORT 55 OF 61.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24700641 ↗
  • MalfunctionMar 26, 2026Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING THE BD BACTEC¿ MGIT¿ MYCOBACTERIA GROWTH INDICATOR TUBES, ONE (1) FALSE RESISTANT M. TUBERCULOSIS PATIENT RESULT WAS OBTAINED FROM A MGIT INSTRUMENT. THE CUSTOMER IS UNABLE TO PROVIDE THE NUMBER OF AFFECTED PATIENT SAMPLES OR TUBES USED WITH A SPECIFIC PRODUCT LOT. NO CONFIRMATORY TESTING WAS REPORTED; HOWEVER, THE CUSTOMER NOTED THAT NO ACTUAL GROWTH WAS OBSERVED. REPEAT TESTING OCCURRED AND NO ADVERSE HEALTH IMPACT WAS REPORTED. THIS IS REPORT 53 OF 61.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24700640 ↗
  • MalfunctionMar 26, 2026Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING THE BD BACTEC¿ MGIT¿ MYCOBACTERIA GROWTH INDICATOR TUBES, ONE (1) FALSE RESISTANT M. TUBERCULOSIS PATIENT RESULT WAS OBTAINED FROM A MGIT INSTRUMENT. THE CUSTOMER IS UNABLE TO PROVIDE THE NUMBER OF AFFECTED PATIENT SAMPLES OR TUBES USED WITH A SPECIFIC PRODUCT LOT. NO CONFIRMATORY TESTING WAS REPORTED; HOWEVER, THE CUSTOMER NOTED THAT NO ACTUAL GROWTH WAS OBSERVED. REPEAT TESTING OCCURRED AND NO ADVERSE HEALTH IMPACT WAS REPORTED. THIS IS REPORT 54 OF 61.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24700639 ↗
  • MalfunctionMar 26, 2026Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING THE BD BACTEC¿ MGIT¿ MYCOBACTERIA GROWTH INDICATOR TUBES, ONE (1) FALSE RESISTANT M. TUBERCULOSIS PATIENT RESULT WAS OBTAINED FROM A MGIT INSTRUMENT. THE CUSTOMER IS UNABLE TO PROVIDE THE NUMBER OF AFFECTED PATIENT SAMPLES OR TUBES USED WITH A SPECIFIC PRODUCT LOT. NO CONFIRMATORY TESTING WAS REPORTED; HOWEVER, THE CUSTOMER NOTED THAT NO ACTUAL GROWTH WAS OBSERVED. REPEAT TESTING OCCURRED AND NO ADVERSE HEALTH IMPACT WAS REPORTED. THIS IS REPORT 44 OF 61.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24700637 ↗
  • MalfunctionMar 26, 2026Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING THE BD BACTEC¿ MGIT¿ MYCOBACTERIA GROWTH INDICATOR TUBES, ONE (1) FALSE RESISTANT M. TUBERCULOSIS PATIENT RESULT WAS OBTAINED FROM A MGIT INSTRUMENT. THE CUSTOMER IS UNABLE TO PROVIDE THE NUMBER OF AFFECTED PATIENT SAMPLES OR TUBES USED WITH A SPECIFIC PRODUCT LOT. NO CONFIRMATORY TESTING WAS REPORTED; HOWEVER, THE CUSTOMER NOTED THAT NO ACTUAL GROWTH WAS OBSERVED. REPEAT TESTING OCCURRED AND NO ADVERSE HEALTH IMPACT WAS REPORTED. THIS IS REPORT 43 OF 61.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24700634 ↗
  • MalfunctionMar 26, 2026Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING THE BD BACTEC¿ MGIT¿ MYCOBACTERIA GROWTH INDICATOR TUBES, ONE (1) FALSE RESISTANT M. TUBERCULOSIS PATIENT RESULT WAS OBTAINED FROM A MGIT INSTRUMENT. THE CUSTOMER IS UNABLE TO PROVIDE THE NUMBER OF AFFECTED PATIENT SAMPLES OR TUBES USED WITH A SPECIFIC PRODUCT LOT. NO CONFIRMATORY TESTING WAS REPORTED; HOWEVER, THE CUSTOMER NOTED THAT NO ACTUAL GROWTH WAS OBSERVED. REPEAT TESTING OCCURRED AND NO ADVERSE HEALTH IMPACT WAS REPORTED. THIS IS REPORT 52 OF 61.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24700631 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K974883. One submission often covers many models, so check the product and lot details in each record.

  • Z-0028-2014 · Sep 9, 2013 · Terminated
    BBL (tm) MGIT (tm) Mycobacteria Growth Indicator Tube, part # 245122, packaged in 100 ct cartons and labeled in part Product Usage: The Mycobacteria Growth Indicator Tube suppl
  • Z-2890-2011 · Mar 28, 2011 · Terminated
    BD BACTEC(tm) MGI (tm) System Software, Catalog 445922, Lot V5.02A. Product Usage: The BACTEC(tm) MGIT (tm) System Software is used in the MGIT 960 and MGIT 320 instruments for

Other System, blood culturing devices