BD BACTEC™ MGIT™ 960 SIRE Kit 245123Recalls & reported problems

Becton, Dickinson And Company · BD BACTEC™ MGIT™ 960 SIRE Kit, carton of 4 lyophilized drug vials, and 8 SIRE Supplements · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This BD BACTEC MGIT 960 SIRE kit, used to test tuberculosis drug susceptibility, has 166 reports, with no deaths and no injuries. Most reports describe false resistant results, sometimes with no growth in the tube, and none list a patient symptom. No recalls name this device, and reports rose from 57 in the prior 12 months to 84 in the latest 12 months.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
166
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 166 adverse event reports that list this exact device (DI 00382902451235), received between Mar 15, 2021 and Aug 19, 2026 — classified as 166 malfunctions.

The most frequently reported specific device problem is “Incorrect, Inadequate or Imprecise Result or Readings” (161 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 1 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 2 reportsJul 2024: 1 reportsAug 2024: 4 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 2 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 1 reportsMay 2025: 1 reportsJun 2025: 2 reportsJul 2025: 6 reportsAug 2025: 39 reportsSep 2025: 1 reportsOct 2025: 3 reportsNov 2025: 1 reportsDec 2025: 4 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 75 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 3 reportsSep 2026: 0 reportsOct 2026: 0 reports20242025202675
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Incorrect, Inadequate or Imprecise Result or Readings161
  • Contamination2
  • Contamination of Device Ingredient or Reagent2
  • Device Ingredient or Reagent Problem1
  • Device Markings/Labelling Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions166

Recent adverse event reports

The 25 most recent of 166 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 19, 2026Incorrect, Inadequate or Imprecise Result or Readings

    It was reported while using the BD BACTEC¿ MGIT¿ 960 SIRE Kit, one (1) Ethambutol patient culture tube had positive growth and appeared visibly turbid. In addition, a reconstituted Ethambutol tube was inoculated without bacteria and visibly demonstrated turbid growth. A microscopic examination of a smear prepared from the tube revealed numerous cocci and other indistinct cellular forms. No clear acid-fast bacilli (AFB) were observed. There were no health impact or consequences reported.This is report 3 of 3.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50325246 ↗
  • MalfunctionAug 19, 2026Incorrect, Inadequate or Imprecise Result or Readings

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50325244 ↗
  • MalfunctionAug 19, 2026Incorrect, Inadequate or Imprecise Result or Readings

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50323587 ↗
  • MalfunctionMar 26, 2026Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING THE BD BACTEC¿ MGIT¿ 960 SIRE KIT, ONE (1) FALSE RESISTANT M. TUBERCULOSIS PATIENT RESULT WAS OBTAINED FROM A MGIT INSTRUMENT. NO CONFIRMATORY TESTING WAS REPORTED; HOWEVER, THE CUSTOMER NOTED THAT NO ACTUAL GROWTH WAS OBSERVED. REPEAT TESTING OCCURRED AND NO ADVERSE HEALTH IMPACT WAS REPORTED. THIS IS REPORT 19 OF 42.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24700781 ↗
  • MalfunctionMar 26, 2026Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING THE BD BACTEC¿ MGIT¿ 960 SIRE KIT, ONE (1) FALSE RESISTANT M. TUBERCULOSIS PATIENT RESULT WAS OBTAINED FROM A MGIT INSTRUMENT. NO CONFIRMATORY TESTING WAS REPORTED; HOWEVER, THE CUSTOMER NOTED THAT NO ACTUAL GROWTH WAS OBSERVED. REPEAT TESTING OCCURRED AND NO ADVERSE HEALTH IMPACT WAS REPORTED. THIS IS REPORT 39 OF 42.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24700642 ↗
  • MalfunctionMar 26, 2026Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING THE BD BACTEC¿ MGIT¿ 960 SIRE KIT, ONE (1) FALSE RESISTANT M. TUBERCULOSIS PATIENT RESULT WAS OBTAINED FROM A MGIT INSTRUMENT. NO CONFIRMATORY TESTING WAS REPORTED; HOWEVER, THE CUSTOMER NOTED THAT NO ACTUAL GROWTH WAS OBSERVED. REPEAT TESTING OCCURRED AND NO ADVERSE HEALTH IMPACT WAS REPORTED. THIS IS REPORT 42 OF 42.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24700638 ↗
  • MalfunctionMar 26, 2026Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING THE BD BACTEC¿ MGIT¿ 960 SIRE KIT, ONE (1) FALSE RESISTANT M. TUBERCULOSIS PATIENT RESULT WAS OBTAINED FROM A MGIT INSTRUMENT. NO CONFIRMATORY TESTING WAS REPORTED; HOWEVER, THE CUSTOMER NOTED THAT NO ACTUAL GROWTH WAS OBSERVED. REPEAT TESTING OCCURRED AND NO ADVERSE HEALTH IMPACT WAS REPORTED. THIS IS REPORT 38 OF 42.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24700636 ↗
  • MalfunctionMar 26, 2026Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING THE BD BACTEC¿ MGIT¿ 960 SIRE KIT, ONE (1) FALSE RESISTANT M. TUBERCULOSIS PATIENT RESULT WAS OBTAINED FROM A MGIT INSTRUMENT. NO CONFIRMATORY TESTING WAS REPORTED; HOWEVER, THE CUSTOMER NOTED THAT NO ACTUAL GROWTH WAS OBSERVED. REPEAT TESTING OCCURRED AND NO ADVERSE HEALTH IMPACT WAS REPORTED. THIS IS REPORT 32 OF 42.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24700635 ↗
  • MalfunctionMar 26, 2026Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING THE BD BACTEC¿ MGIT¿ 960 SIRE KIT, ONE (1) FALSE RESISTANT M. TUBERCULOSIS PATIENT RESULT WAS OBTAINED FROM A MGIT INSTRUMENT. NO CONFIRMATORY TESTING WAS REPORTED; HOWEVER, THE CUSTOMER NOTED THAT NO ACTUAL GROWTH WAS OBSERVED. REPEAT TESTING OCCURRED AND NO ADVERSE HEALTH IMPACT WAS REPORTED. THIS IS REPORT 41 OF 42.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24700633 ↗
  • MalfunctionMar 26, 2026Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING THE BD BACTEC¿ MGIT¿ 960 SIRE KIT, ONE (1) FALSE RESISTANT M. TUBERCULOSIS PATIENT RESULT WAS OBTAINED FROM A MGIT INSTRUMENT. NO CONFIRMATORY TESTING WAS REPORTED; HOWEVER, THE CUSTOMER NOTED THAT NO ACTUAL GROWTH WAS OBSERVED. REPEAT TESTING OCCURRED AND NO ADVERSE HEALTH IMPACT WAS REPORTED. THIS IS REPORT 40 OF 42.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24700632 ↗
  • MalfunctionMar 26, 2026Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING THE BD BACTEC¿ MGIT¿ 960 SIRE KIT, ONE (1) FALSE RESISTANT M. TUBERCULOSIS PATIENT RESULT WAS OBTAINED FROM A MGIT INSTRUMENT. NO CONFIRMATORY TESTING WAS REPORTED; HOWEVER, THE CUSTOMER NOTED THAT NO ACTUAL GROWTH WAS OBSERVED. REPEAT TESTING OCCURRED AND NO ADVERSE HEALTH IMPACT WAS REPORTED. THIS IS REPORT 37 OF 42.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24700630 ↗
  • MalfunctionMar 26, 2026Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING THE BD BACTEC¿ MGIT¿ 960 SIRE KIT, ONE (1) FALSE RESISTANT M. TUBERCULOSIS PATIENT RESULT WAS OBTAINED FROM A MGIT INSTRUMENT. NO CONFIRMATORY TESTING WAS REPORTED; HOWEVER, THE CUSTOMER NOTED THAT NO ACTUAL GROWTH WAS OBSERVED. REPEAT TESTING OCCURRED AND NO ADVERSE HEALTH IMPACT WAS REPORTED. THIS IS REPORT 23 OF 42.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24700629 ↗
  • MalfunctionMar 26, 2026Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING THE BD BACTEC¿ MGIT¿ 960 SIRE KIT, ONE (1) FALSE RESISTANT M. TUBERCULOSIS PATIENT RESULT WAS OBTAINED FROM A MGIT INSTRUMENT. NO CONFIRMATORY TESTING WAS REPORTED; HOWEVER, THE CUSTOMER NOTED THAT NO ACTUAL GROWTH WAS OBSERVED. REPEAT TESTING OCCURRED AND NO ADVERSE HEALTH IMPACT WAS REPORTED. THIS IS REPORT 33 OF 42.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24700623 ↗
  • MalfunctionMar 26, 2026Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING THE BD BACTEC¿ MGIT¿ 960 SIRE KIT, ONE (1) FALSE RESISTANT M. TUBERCULOSIS PATIENT RESULT WAS OBTAINED FROM A MGIT INSTRUMENT. NO CONFIRMATORY TESTING WAS REPORTED; HOWEVER, THE CUSTOMER NOTED THAT NO ACTUAL GROWTH WAS OBSERVED. REPEAT TESTING OCCURRED AND NO ADVERSE HEALTH IMPACT WAS REPORTED. THIS IS REPORT 31 OF 42.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24700618 ↗
  • MalfunctionMar 26, 2026Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING THE BD BACTEC¿ MGIT¿ 960 SIRE KIT, ONE (1) FALSE RESISTANT M. TUBERCULOSIS PATIENT RESULT WAS OBTAINED FROM A MGIT INSTRUMENT. NO CONFIRMATORY TESTING WAS REPORTED; HOWEVER, THE CUSTOMER NOTED THAT NO ACTUAL GROWTH WAS OBSERVED. REPEAT TESTING OCCURRED AND NO ADVERSE HEALTH IMPACT WAS REPORTED. THIS IS REPORT 29 OF 42.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24700617 ↗
  • MalfunctionMar 26, 2026Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING THE BD BACTEC¿ MGIT¿ 960 SIRE KIT, ONE (1) FALSE RESISTANT M. TUBERCULOSIS PATIENT RESULT WAS OBTAINED FROM A MGIT INSTRUMENT. NO CONFIRMATORY TESTING WAS REPORTED; HOWEVER, THE CUSTOMER NOTED THAT NO ACTUAL GROWTH WAS OBSERVED. REPEAT TESTING OCCURRED AND NO ADVERSE HEALTH IMPACT WAS REPORTED. THIS IS REPORT 28 OF 42.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24700616 ↗
  • MalfunctionMar 26, 2026Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING THE BD BACTEC¿ MGIT¿ 960 SIRE KIT, ONE (1) FALSE RESISTANT M. TUBERCULOSIS PATIENT RESULT WAS OBTAINED FROM A MGIT INSTRUMENT. NO CONFIRMATORY TESTING WAS REPORTED; HOWEVER, THE CUSTOMER NOTED THAT NO ACTUAL GROWTH WAS OBSERVED. REPEAT TESTING OCCURRED AND NO ADVERSE HEALTH IMPACT WAS REPORTED. THIS IS REPORT 30 OF 42.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24700614 ↗
  • MalfunctionMar 26, 2026Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING THE BD BACTEC¿ MGIT¿ 960 SIRE KIT, ONE (1) FALSE RESISTANT M. TUBERCULOSIS PATIENT RESULT WAS OBTAINED FROM A MGIT INSTRUMENT. NO CONFIRMATORY TESTING WAS REPORTED; HOWEVER, THE CUSTOMER NOTED THAT NO ACTUAL GROWTH WAS OBSERVED. REPEAT TESTING OCCURRED AND NO ADVERSE HEALTH IMPACT WAS REPORTED. THIS IS REPORT 18 OF 42.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24700606 ↗
  • MalfunctionMar 26, 2026Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING THE BD BACTEC¿ MGIT¿ 960 SIRE KIT, ONE (1) FALSE RESISTANT M. TUBERCULOSIS PATIENT RESULT WAS OBTAINED FROM A MGIT INSTRUMENT. NO CONFIRMATORY TESTING WAS REPORTED; HOWEVER, THE CUSTOMER NOTED THAT NO ACTUAL GROWTH WAS OBSERVED. REPEAT TESTING OCCURRED AND NO ADVERSE HEALTH IMPACT WAS REPORTED. THIS IS REPORT 25 OF 42.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24700605 ↗
  • MalfunctionMar 26, 2026Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING THE BD BACTEC¿ MGIT¿ 960 SIRE KIT, ONE (1) FALSE RESISTANT M. TUBERCULOSIS PATIENT RESULT WAS OBTAINED FROM A MGIT INSTRUMENT. NO CONFIRMATORY TESTING WAS REPORTED; HOWEVER, THE CUSTOMER NOTED THAT NO ACTUAL GROWTH WAS OBSERVED. REPEAT TESTING OCCURRED AND NO ADVERSE HEALTH IMPACT WAS REPORTED. THIS IS REPORT 20 OF 42.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24700601 ↗
  • MalfunctionMar 26, 2026Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING THE BD BACTEC¿ MGIT¿ 960 SIRE KIT, ONE (1) FALSE RESISTANT M. TUBERCULOSIS PATIENT RESULT WAS OBTAINED FROM A MGIT INSTRUMENT. NO CONFIRMATORY TESTING WAS REPORTED; HOWEVER, THE CUSTOMER NOTED THAT NO ACTUAL GROWTH WAS OBSERVED. REPEAT TESTING OCCURRED AND NO ADVERSE HEALTH IMPACT WAS REPORTED. THIS IS REPORT 7 OF 42.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24700595 ↗
  • MalfunctionMar 26, 2026Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING THE BD BACTEC¿ MGIT¿ 960 SIRE KIT, ONE (1) FALSE RESISTANT M. TUBERCULOSIS PATIENT RESULT WAS OBTAINED FROM A MGIT INSTRUMENT. NO CONFIRMATORY TESTING WAS REPORTED; HOWEVER, THE CUSTOMER NOTED THAT NO ACTUAL GROWTH WAS OBSERVED. REPEAT TESTING OCCURRED AND NO ADVERSE HEALTH IMPACT WAS REPORTED. THIS IS REPORT 26 OF 42.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24700590 ↗
  • MalfunctionMar 26, 2026Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING THE BD BACTEC¿ MGIT¿ 960 SIRE KIT, ONE (1) FALSE RESISTANT M. TUBERCULOSIS PATIENT RESULT WAS OBTAINED FROM A MGIT INSTRUMENT. NO CONFIRMATORY TESTING WAS REPORTED; HOWEVER, THE CUSTOMER NOTED THAT NO ACTUAL GROWTH WAS OBSERVED. REPEAT TESTING OCCURRED AND NO ADVERSE HEALTH IMPACT WAS REPORTED. THIS IS REPORT 10 OF 42.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24700589 ↗
  • MalfunctionMar 26, 2026Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING THE BD BACTEC¿ MGIT¿ 960 SIRE KIT, ONE (1) FALSE RESISTANT M. TUBERCULOSIS PATIENT RESULT WAS OBTAINED FROM A MGIT INSTRUMENT. NO CONFIRMATORY TESTING WAS REPORTED; HOWEVER, THE CUSTOMER NOTED THAT NO ACTUAL GROWTH WAS OBSERVED. REPEAT TESTING OCCURRED AND NO ADVERSE HEALTH IMPACT WAS REPORTED. THIS IS REPORT 24 OF 42.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24700588 ↗
  • MalfunctionMar 26, 2026Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING THE BD BACTEC¿ MGIT¿ 960 SIRE KIT, ONE (1) FALSE RESISTANT M. TUBERCULOSIS PATIENT RESULT WAS OBTAINED FROM A MGIT INSTRUMENT. NO CONFIRMATORY TESTING WAS REPORTED; HOWEVER, THE CUSTOMER NOTED THAT NO ACTUAL GROWTH WAS OBSERVED. REPEAT TESTING OCCURRED AND NO ADVERSE HEALTH IMPACT WAS REPORTED. THIS IS REPORT 27 OF 42.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24700587 ↗

Other SUSCEPTIBILITY TEST POWDERS, ANTIMYCOBACTERIAL devices

All SUSCEPTIBILITY TEST POWDERS, ANTIMYCOBACTERIAL devices →