Skip to content

BD Integra 305273Recalls & reported problems

Becton, Dickinson And Company · SYRINGE INTEGRA 3ML W/NDL 21X1 RB TW · Class 2

All models of BD Integra (12)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00382903052738), received between Aug 17, 2018 and May 24, 2023 — classified as 2 injuries, 3 malfunctions.

The most frequently reported specific device problem is “Difficult or Delayed Activation” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Most reported problems

Device problems

  • Difficult or Delayed Activation4
  • Retraction Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Exposure to Body Fluids1
  • Foreign Body In Patient1
  • Needle Stick/Puncture1
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 24, 2023Difficult or Delayed Activation; Difficult or Delayed Activation

    IT WAS REPORTED WHILE USING BD INTEGRA¿ RETRACTING SAFETY SYRINGE WITH 21 G X 1 IN., 3 ML THE NEEDLE RETRACTED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: SYRINGE WAS DEFECTIVE. CUSTOMER DREW UP THE MEDICATION, WHEN CUSTOMER'S WIFE WENT TO ADMINISTER MEDICATION IN LEFT BUTTOX THEY HEARD A POP. NEEDLE HAD BROKE OFF IN CUSTOMER. THERE WAS A LOT OF BLOOD. CUSTOMER WENT TO ER, ULTRASOUND DID NOT LOCATE NEEDLE. AT ER THEY DID SMALL INCISION AND ATTEMPTED TO LOCATE NEEDLE WITH TWEEZERS WITH NO LUCK. WENT TO OR NEXT DAY AND DID PRE-OPERATION X-RAY AND COULD NOT LOCATE NEEDLE. SURGEON SEWED UP INCISION AND TOOK APART THE DEFECTIVE SYRINGE. UPON TAKING APART THE SYRINGE IT WAS FOUND THAT THE NEEDLE HAD RETRACTED INTO THE SYRINGE. CUSTOMER WAS GIVEN A STITCH FOR THE INCISION AND PRESCRIBED ANTIBIOTICS AS A RESULT.

    Patient problems: Foreign Body In Patient

    MAUDE report 17000026 ↗
  • MalfunctionMar 31, 2021Difficult or Delayed Activation; Difficult or Delayed Activation

    IT WAS REPORTED THAT SYRINGE INTEGRA 3ML W/NDL 21X1 RB TW WAS NOT RETRACTING. THIS OCCURRED ON 5 OCCASIONS. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: SYRINGE NOT RETRACTING. REPORTED ISSUE - SYRINGE NOT RETRACTING. SOME VAGUE AND CONFLICTING INFORMATION HAS BEEN PROVIDED. WILL PROVIDE MORE DETAILS WHEN MORE INFORMATION PROVIDED BY THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11599557 ↗
  • MalfunctionMar 11, 2021Difficult or Delayed Activation; Difficult or Delayed Activation

    IT WAS REPORTED THAT SYRINGE INTEGRA 3ML W/NDL 21X1 RB TW NEEDLE RETRACTION PREMATURELY. THIS OCCURRED ON 2 OCCASIONS. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: NEEDLE RETRACTED PREMATURELY. NEEDLE RETRACTED SPONTANEOUSLY WHILE ADMINISTERING MEDICATION MEANING THAT NOT ALL OF THE MEDICATION COULD BE ADMINISTERED. HAS HAPPENED ON SEVERAL OCCASIONS OVER THE LAST COUPLE OF WEEKS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11468241 ↗
  • InjuryJul 6, 2020Difficult or Delayed Activation; Difficult or Delayed Activation

    IT WAS REPORTED THAT THE SYRINGE INTEGRA 3ML W/NDL 21X1 RB TW EXPERIENCED NEEDLE RETRACTION FAILURE AND WAS INVOLVED WITH A SIRTY NEEDLE STICK INJURY. IT HAS BEEN MENTIONED THAT THE NURSE WHO RECEIVED THE DIRTY NEEDLE STICK INJURY RECEIVED MEDICAL INTERVENTION AS A RESULT, THOUGH THE SPECIFIC INTERVENTION OR ANY ASSOCIATED TESTING RESULTS HAVE NOT BEEN SPECIFIED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: MATERIAL NO: 305273, BATCH NO: 8095530. CUSTOMER CALLED TO REPORT THAT NEEDLE DID NOT RETRACT CAUSING THE NURSE TO GET STUCK WITH THE NEEDLE AFTER INJECTION. SHE STATED THAT THE NURSE WAS SENT OUT PER PROTOCOL TO MEDICAL FACILITY, IT HAPPENED LAST NIGHT AND EMPLOYEE HEALTH WAS NOT OPEN AT THAT TIME.

    Patient problems: Exposure to Body Fluids; Needle Stick/Puncture

    MAUDE report 10237811 ↗
  • MalfunctionAug 17, 2018Retraction Problem

    IT WAS REPORTED WITH THE USE OF THE BD INTEGRA¿ SYRINGE WITH DETACHABLE NEEDLE THERE WAS AN ISSUE WITH SPRING FAILURE. WHEN THE NEEDLE RETRACTED, THE SPRING DID NOT RETRACT INTO THE PLUNGER. INSTEAD, IT CURLED UP IN THE SYRINGE, BETWEEN THE HUB AND PLUNGER. COULD NOT SEE THE NEEDLE IN THE SYRINGE, OR ANYWHERE SO THE PATIENT WAS GIVEN A X-RAY TO DETERMINE IF IT WAS LODGED IN THE PATIENT. IT WAS NEGATIVE AND THE NEEDLE WAS FOUND IN THE PLUNGER OF THE SYRINGE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7794931 ↗

FDA clearance and classification

Cleared through 510(k) K011103 on Jun 13, 2001 (decision: Substantially Equivalent). Submitted by Bd as “Bd Spring Based Syringe”. The FDA review took 63 days. FDA record ↗

The FDA regulates product code MEG (syringe, antistick) under 21 CFR 880.5860 as a Class II device. 21 CFR 880.5860 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the BD Integra 305273

Has the BD Integra 305273 been recalled?

No FDA recall record in our data names this exact device identifier (DI 00382903052738), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the BD Integra 305273?

The FDA's MAUDE database lists 5 adverse event reports for this device, received between Aug 17, 2018 and May 24, 2023. In those reports, the most common specific device problems are “Difficult or Delayed Activation” (4) and “Retraction Problem” (1); the most common patient problems are “Exposure to Body Fluids” (1), “Foreign Body In Patient” (1) and “Needle Stick/Puncture” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the BD Integra 305273?

None of the reports are classified as deaths, 2 as injuries and 3 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the BD Integra 305273 FDA approved?

It was cleared through 510(k) K011103 on Jun 13, 2001. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the BD Integra 305273, and is it still sold?

The FDA's device database lists Becton, Dickinson And Company as the company responsible for it (brand BD Integra, model 305273). Its record shows it as in commercial distribution.