BD PosiFlush 306414Recalls & reported problems
Becton, Dickinson And Company · SYRINGE 5ML HEPARIN 10 UNIT · Class 2
All models of BD PosiFlush (19)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 11
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 00382903064144), received between Apr 1, 2019 and Nov 20, 2025 — classified as 1 injury, 10 malfunctions.
The most frequently reported specific device problem is “Infusion or Flow Problem” (5 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Infusion or Flow Problem5
- Device Markings/Labelling Problem1
- Filling Problem1
- Leak/Splash1
- Material Separation1
- Missing Information1
- Volume Accuracy Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions9
- Chemical Exposure1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 20, 2025Filling Problem; Filling Problem
IT WAS REPORTED THAT THE BD SYRINGE 5ML HEPARIN 10 UNIT WAS EMPTY. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: HEPARIN SYRINGE EMPTY BEFORE USE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23610684 ↗MalfunctionMay 9, 2025Infusion or Flow Problem; Infusion or Flow Problem
MATERIAL #306414, BATCH # 434180N. IT WAS REPORTED BY THE CUSTOMER THAT THE SYRINGE LACKED HEPARIN/FLUID. VERBATIM: RCC RECEIVED A COMPLAINT VIA EMAIL. EMAIL(S) ATTACHED. SHARING TWO ADDITIONAL EVENTS OF AIR IN SYRINGE AND LACK OF HEPARIN/FLUID. THESE WERE NOT SAVED FOR RETURN. ASKING FOR ADDITIONAL INVESTIGATION INTO THE MANUFACTURING DEFECT. THE CLOSURE LETTERS I RECEIVED FOR THE PRIOR EVENTS STATE NO ISSUES WITH THE PRODUCTION PROCESS BUT THIS IS AN INCREASING SAFETY TREND FOR THIS PRODUCT. EVENT DATE: 4/18/2025. REF 306414. LOT 434180N. EXP 2026-06-30. EVENT DATE: 4/19/2025. REF 306414. LOT 43410N. EXP 2026-06-30.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21994747 ↗MalfunctionMay 6, 2025Infusion or Flow Problem; Infusion or Flow Problem
MATERIAL # 306414 BATCH # 429953N. IT WAS REPORTED BY THE CUSTOMER THAT THE SYRINGE IS EMPTY OF ANY FLUID, BUT THE SYRINGE IS PULLED BACK, CAPPED AND PACKAGED AS THOUGH THERE IS FLUID IN THE SYRINGE. VERBATIM: REPORTING AN ADDITIONAL EVENT THAT OCCURRED TODAY. NOTED BY USER AT TIME OF PREPARATION FOR ADMINISTRATION TO THE PATIENT. LOT #4 29953N EXP: 2026-04-30 PLEASE PROVIDE ME WITH A CONTACT FOR A MANAGER OR DIRECTOR FOR BD¿S QUALITY ASSURANCE TEAM. THIS IS OUR 6TH REPORT OF FINDING A SYRINGE THAT IS EMPTY OF ANY FLUID, BUT THE SYRINGE IS PULLED BACK, CAPPED AND PACKAGED AS THOUGH THERE IS FLUID IN THE SYRINGE. THE FLUID IS CLEAR, SO THE RISK IF THIS IS NOT NOTICED IS THAT AIR IS INJECTED INTO THE PATIENT INSTEAD OF FLUID. IF THIS IS A SMALL INFANT OR A CHILD WITH A RIGHT TO LEFT SHUNT, THE CONSEQUENCES COULD BE CATASTROPHIC.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21968510 ↗MalfunctionMay 5, 2025Infusion or Flow Problem; Infusion or Flow Problem
MATERIAL #306414, LOT #429953N. VERBATIM: THIS IS OUR 6TH REPORT OF FINDING A SYRINGE THAT IS EMPTY OF ANY FLUID, BUT THE SYRINGE IS PULLED BACK, CAPPED AND PACKAGED AS THOUGH THERE IS FLUID IN THE SYRINGE. THE FLUID IS CLEAR, SO THE RISK IF THIS IS NOT NOTICED IS THAT AIR IS INJECTED INTO THE PATIENT INSTEAD OF FLUID. IF THIS IS A SMALL INFANT OR A CHILD WITH A RIGHT TO LEFT SHUNT, THE CONSEQUENCES COULD BE CATASTROPHIC.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21962301 ↗MalfunctionApr 29, 2025Infusion or Flow Problem; Infusion or Flow Problem
MATERIAL #306414 BATCH # 434180N IT WAS REPORTED BY THE CUSTOMER THAT THE SYRINGE LACKED HEPARIN/FLUID. VERBATIM: RCC RECEIVED A COMPLAINT VIA EMAIL. EMAIL(S) ATTACHED. SHARING TWO ADDITIONAL EVENTS OF AIR IN SYRINGE AND LACK OF HEPARIN/FLUID. THESE WERE NOT SAVED FOR RETURN. ASKING FOR ADDITIONAL INVESTIGATION INTO THE MANUFACTURING DEFECT. THE CLOSURE LETTERS I RECEIVED FOR THE PRIOR EVENTS STATE NO ISSUES WITH THE PRODUCTION PROCESS BUT THIS IS AN INCREASING SAFETY TREND FOR THIS PRODUCT. EVENT DATE: (B)(6) 2025. REF (B)(4). LOT 434180N. EXP 2026-06-30. EVENT DATE: (B)(6) 2025. REF (B)(4). LOT 43410N. EXP 2026-06-30.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21926439 ↗MalfunctionApr 3, 2025Infusion or Flow Problem; Infusion or Flow Problem
MATERIAL #306414. BATCH #429953N. IT WAS REPORTED BY THE CUSTOMER THAT A SYRINGE WAS NOTED TO HAVE AIR IN THE SYRINGE AND NO FLUID/HEPARIN. THE USER NOTED THE ABSENCE OF FLUID UPON OPENING AND PRIMING FOR INSTALLATION INTO AN ARTERIAL LINE IN THE NICU. VERBATIM: RCC RECEIVED A COMPLAINT VIA EMAIL. EMAIL(S) ATTACHED. REPORTING AN ADDITIONAL SAFETY EVENT WITH THE BD POSIFLUSH PREFILLED HEPARIN LOCK FLUSH SYRINGE, 5 ML SYRINGE 5 ML HEPARIN FILL, 10 USP UNITS/ML, SKU 306414. A SYRINGE WAS NOTED TO HAVE AIR IN THE SYRINGE AND NO FLUID/HEPARIN. THE USER NOTED THE ABSENCE OF FLUID UPON OPENING AND PRIMING FOR INSTALLATION INTO AN ARTERIAL LINE IN THE NICU. ITEM SAVED AND AVAILABLE FOR RETURN. LOT NUMBER 429953N, EXPIRATION DATE OF 04/30/2026. IT IS THE SAME LOT NUMBER AS THE PREVIOUS REPORT FROM 1/13/25, BD REFERENCE NUMBER (B)(4). ADDITIONAL INFO: 1. ANY ADVERSE EVENT OR SERIOUS INJURY REPORTED TO PATIENT OR HEALTHCARE PROFESSIONAL? IF YES, PLEASE PROVIDE THE DETAILS. NO, CAUGHT BY USER JUST PRIOR TO ADMINISTRATION TO PATIENT. 2. CAN YOU PLEASE PROVIDE AN EXACT DATE OF EVENT IN FORMAT DD-MMM-YYYY? (B)(6) 2025. 3. TOTAL NUMBER OF OCCURRENCES? ONE EVENT ON THAT DATE, THREE EVENTS TOTAL. 4.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21761572 ↗MalfunctionDec 2, 2024Material Separation; Material Separation
NO ADDITIONAL INFORMATION RECEIVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20827874 ↗MalfunctionSep 29, 2023Volume Accuracy Problem; Volume Accuracy Problem
IT WAS REPORTED WHILE USING BD POSIFLUSH¿ PREFILLED HEPARIN LOCK FLUSH SYRINGE THE LABEL WAS MISSING. THERE WAS NO REPORT OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: ONE OF OUR INTERN PHARMACISTS WAS DELIVERING THE PRODUCT INTO OUR PYXIS MACHINE ON THE INPATIENT FLOORS. PRODUCT IS NOT LABELED CORRECTLY/NO LABEL IS PLACED ON THE SYRINGES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17848256 ↗MalfunctionMay 10, 2023Missing Information; Missing Information
IT WAS REPORTED THAT THE BD POSIFLUSH¿ PRE-FILLED HEPARIN LOCK FLUSH SYRINGE WAS MISSING THE LABEL. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: HEPARIN SYRINGE ONE DID NOT HAVE ANY LABEL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16908380 ↗InjuryOct 15, 2019Leak/Splash; Leak/Splash
IT WAS REPORTED THAT SYRINGE 5ML HEPARIN 10 UNIT LEAKED FLUID WHICH SPLASHED INTO THE NURSES EYE DURING USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: MATERIAL NO: 306414 BATCH NO: UNKNOWN IT WAS REPORTED THAT WHILE INJECTING HEPARIN FROM PREFILLED SYRINGE INTO THE PATIENT'S LINE DURING A BLOOD DRAWING PROCEDURE FLUID FROM THE SYRINGE SPLASHED BACK INTO THE NURSES EYE. VERBATIM: WHILE INJECTING HEPARIN FROM PREFILLED SYRINGE INTO THE PATIENT'S LINE DURING A BLOOD DRAWING PROCEDURE FLUID FROM THE SYRINGE SPLASHED BACK INTO THE NURSES EYE. PATIENT IS ON ONCOLOGY UNIT, BUT NOT OTHER DRUGS OF THERAPIES COULD HAVE CONTRIBUTED AT THE TIME OF THE EVENT. INJECTION OF HEPARIN INTO LINE DURING BLOOD DRAW TO PREVENT CLOTTING.
Patient problems: Chemical Exposure
MAUDE report 9191928 ↗MalfunctionApr 1, 2019Device Markings/Labelling Problem
IT WAS REPORTED THAT A UNSPECIFIED BD SYRINGE HAD LABELING ERROR. THE FOLLOWING WAS REPORTED, "MATERIAL NO. UNKNOWN, BATCH NO. UNKNOWN. IT WAS REPORTED THE HEPARIN LOCKS NO LONGER HAVE BARCODES ON THEM. "GOOD AFTERNOON, WE HAVE BEEN NOTICING OUR HEPARIN AND NS FLUSHES NO LONGER HAVE BARCODES ON THEM? ARE WE ORDERING THE WRONG PRODUCTS OR IS THIS A PERMANENT CHANGE? THANKS."
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8471652 ↗
Other Heparin, vascular access flush devices
- Excelsior 60051Medline Industries, Inc. · Heparin, vascular access flush1 recall0 reports
- PENDING EMZH60051Medline Industries, Inc. · Heparin, vascular access flush1 recall0 reports
- BD PosiFlush 306424Becton, Dickinson And Company · Heparin, vascular access flush23 reports
- BD PosiFlush 306413Becton, Dickinson And Company · Heparin, vascular access flush7 reports
- BD PosiFlush 306423Becton, Dickinson And Company · Heparin, vascular access flush4 reports
- Heparin Lock Flush Syringe 10 Units/mL 3mL in 12mL syringeMedefil, Inc. · Heparin, Vascular Access Flush1 report
- Heparin Lock Flush Syringe 100 Unit/mL 3mL in 6mL syringeMedefil, Inc. · Heparin, Vascular Access Flush1 report
- Heparin Lock Flush Syringe 1 unit/mL 2 mL in 6mL SyringeMedefil, Inc. · Heparin, Vascular Access Flush1 report
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page