Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00382903065141), received between Apr 2, 2018 and Jun 12, 2018 — classified as 3 injuries, 1 malfunction.
The most frequently reported specific device problem is “Microbial Contamination of Device” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Microbial Contamination of Device2
- Nonstandard Device2
- Crack1
- Fluid/Blood Leak1
Patient problems
- Bacterial Infection3
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
InjuryJun 12, 2018Nonstandard Device
IT WAS REPORTED THAT A (B)(6) MALE PATIENT WITH A DIAGNOSIS OF NEUROBLASTOMA AND A TUNNELED CENTRAL LINE HAD A POSITIVE BLOOD CULTURE FOR SERRATIA MARCESCENS ON (B)(6) 2018. THE PATIENT WAS TREATED AND DISCHARGED. THE PFGE PATTERNS FOR THIS INCIDENT ARE INDISTINGUISHABLE FROM THE ORIGINAL CO OUTBREAK PATTERN. THE DEVICE IMPLICATED WITH THIS INCIDENT IS A 3ML BD POSIFLUSH¿ HEPARIN LOCK FLUSH SYRINGE, IN 10ML SYRINGE, 100 USP UNITS/ML. OUT OF AN ABUNDANCE OF CAUTION AND IN THE INTEREST OF PUBLIC HEALTH, BD VOLUNTARILY RECALLED CERTAIN LOTS OF BD POSIFLUSH¿ HEPARIN LOCK FLUSH AND BD¿ PRE-FILLED NORMAL SALINE FLUSH SYRINGES DUE TO A POTENTIAL FOR CONTAMINATION WITH SERRATIA MARCESCENS BACTERIUM. BD WAS NOTIFIED BY THE U.S. FOOD AND DRUG ADMINISTRATION (FDA) AND CENTERS FOR DISEASE CONTROL AND PREVENTION (CDC) ABOUT A POTENTIAL EPIDEMIOLOGICAL LINK BETWEEN CATHETER RELATED BLOOD STREAM INFECTIONS AND THE S. MARCESCENS BACTERIUM. SPECIFICALLY, THE FDA AND CDC IDENTIFIED A POTENTIAL CONNECTION BETWEEN REPORTS OF INFECTION IN A SMALL NUMBER OF PATIENTS CAUSED BY S. MARCESCENS ACROSS MULTIPLE STATES. CDC¿S INITIAL INVESTIGATION FOUND THAT AFFECTED PATIENTS HAD RECEIVED TREATMENT USING CERTA
Patient problems: Bacterial Infection
MAUDE report 7594791 ↗InjuryJun 11, 2018Microbial Contamination of Device
IT WAS REPORTED THAT A (B)(6) FEMALE PATIENT WITH A DIAGNOSIS OF RESPIRATORY FAILURE AND SEPSIS AND A PICC LINE IN PLACE HAD A POSITIVE BLOOD CULTURE FOR SERRATIA MARCESCENS ON (B)(6) 2018. THE PATIENT WAS TREATED AND DISCHARGED. THE PFGE PATTERNS FOR THIS INCIDENT ARE INDISTINGUISHABLE FROM THE ORIGINAL CO OUTBREAK PATTERN. THE DEVICE IMPLICATED WITH THIS INCIDENT IS A 3ML BD POSIFLUSH¿ HEPARIN LOCK FLUSH SYRINGE, IN 10ML SYRINGE, 100 USP UNITS/ML. OUT OF AN ABUNDANCE OF CAUTION AND IN THE INTEREST OF PUBLIC HEALTH, BD VOLUNTARILY RECALLED CERTAIN LOTS OF BD POSIFLUSH¿ HEPARIN LOCK FLUSH AND BD¿ PRE-FILLED NORMAL SALINE FLUSH SYRINGES DUE TO A POTENTIAL FOR CONTAMINATION WITH SERRATIA MARCESCENS BACTERIUM. BD WAS NOTIFIED BY THE U.S. FOOD AND DRUG ADMINISTRATION (FDA) AND CENTERS FOR DISEASE CONTROL AND PREVENTION (CDC) ABOUT A POTENTIAL EPIDEMIOLOGICAL LINK BETWEEN CATHETER RELATED BLOOD STREAM INFECTIONS AND THE S. MARCESCENS BACTERIUM. SPECIFICALLY, THE FDA AND CDC IDENTIFIED A POTENTIAL CONNECTION BETWEEN REPORTS OF INFECTION IN A SMALL NUMBER OF PATIENTS CAUSED BY S. MARCESCENS ACROSS MULTIPLE STATES. CDC¿S INITIAL INVESTIGATION FOUND THAT AFFECTED PATIENTS HAD RECEIVED TREAT
Patient problems: Bacterial Infection
MAUDE report 7589890 ↗InjuryJun 6, 2018Nonstandard Device; Microbial Contamination of Device
IT WAS REPORTED THAT A (B)(6) YEAR OLD, MALE PATIENT WITH A DIAGNOSIS OF MEDULLOBLASTOMA AND A PORT HAD A POSITIVE BLOOD CULTURE FOR SERRATIA MARCESCENS ON (B)(6) 2018. THE PATIENT WAS TREATED AND DISCHARGED. THE PFGE PATTERNS FOR THIS INCIDENT ARE INDISTINGUISHABLE FROM THE ORIGINAL CO OUTBREAK PATTERN. THE DEVICE IMPLICATED WITH THIS INCIDENT IS A 3 ML BD POSIFLUSH¿ HEPARIN LOCK FLUSH SYRINGE, IN 10 ML SYRINGE, 100 USP UNITS/ML. OUT OF AN ABUNDANCE OF CAUTION AND IN THE INTEREST OF PUBLIC HEALTH, BD VOLUNTARILY RECALLED CERTAIN LOTS OF BD POSIFLUSH¿ HEPARIN LOCK FLUSH AND BD¿ PRE-FILLED NORMAL SALINE FLUSH SYRINGES DUE TO A POTENTIAL FOR CONTAMINATION WITH SERRATIA MARCESCENS BACTERIUM. BD WAS NOTIFIED BY THE (B)(4) ABOUT A POTENTIAL EPIDEMIOLOGICAL LINK BETWEEN CATHETER RELATED BLOOD STREAM INFECTIONS AND THE S. MARCESCENS BACTERIUM. SPECIFICALLY, THE (B)(4) IDENTIFIED A POTENTIAL CONNECTION BETWEEN REPORTS OF INFECTION IN A SMALL NUMBER OF PATIENTS CAUSED BY S. MARCESCENS ACROSS MULTIPLE STATES. (B)(4)¿S INITIAL INVESTIGATION FOUND THAT AFFECTED PATIENTS HAD RECEIVED TREATMENT USING CERTAIN BD FLUSH PRODUCTS. TO DATE, THERE IS NO EVIDENCE OF BD FLUSH PRODUCT TESTING POSITIVE FO
Patient problems: Bacterial Infection
MAUDE report 7576207 ↗MalfunctionApr 2, 2018Crack; Fluid/Blood Leak
IT WAS REPORTED A BD POSIFLUSH¿ HEPARIN LOCK FLUSH SYRINGE WAS FOUND CRACKED AND EMPTY OF THE MEDICATION. THERE WAS NO REPORT OF EXPOSURE, SERIOUS INJURY OR MEDICAL INTERVENTION.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7389741 ↗
Other Catheter,intravascular,therapeutic,short-term less than 30 days devices
- BD Insyte Autoguard 382523Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days660 reports
- BD Insyte Autoguard 382533Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days2 recalls527 reports
- BD PosiFlush 306546Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days1 death608 reports
- BD Insyte Autoguard Shielded IV Catheter 381823Becton, Dickinson And Company · Catheter, intravascular, therapeutic, short-term less than 30 days582 reports
- BD Insyte Autoguard 381423Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days1 recall528 reports
- BD Insyte Autoguard 382544Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days1 recall437 reports
- BD Nexiva, BD Q-Syte 383536Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days2 recalls374 reports
- BD PosiFlush 306547Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days2 recalls1 death331 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page