Skip to content

BD Luer-Lok 309653Recalls & reported problems

Becton, Dickinson And Company · Syringe 50ml LL Tip · Class 2

All models of BD Luer-Lok (48)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Quick summary

This BD Luer-Lok syringe is a piston syringe. The 942 reports mostly describe contamination on the syringe, leaks and splashes, and labels or volume readings that are off, and some note discoloration found when opened. The reports include no deaths and 4 injuries, and no recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
928
Deaths reported
0
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 928 adverse event reports that list this exact device (DI 00382903096534), received between Jan 19, 2017 and Sep 28, 2026 — classified as 4 injuries, 924 malfunctions.

The most frequently reported specific device problem is “Device Contamination with Chemical or Other Material” (295 reports).

10 other recalls cover devices cleared under the same FDA submission (K110771); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 7 reportsJan 2024: 6 reportsFeb 2024: 12 reportsMar 2024: 3 reportsApr 2024: 4 reportsMay 2024: 10 reportsJun 2024: 6 reportsJul 2024: 7 reportsAug 2024: 7 reportsSep 2024: 4 reportsOct 2024: 8 reportsNov 2024: 6 reportsDec 2024: 9 reportsJan 2025: 11 reportsFeb 2025: 4 reportsMar 2025: 6 reportsApr 2025: 6 reportsMay 2025: 11 reportsJun 2025: 5 reportsJul 2025: 6 reportsAug 2025: 12 reportsSep 2025: 8 reportsOct 2025: 10 reportsNov 2025: 11 reportsDec 2025: 5 reportsJan 2026: 16 reportsFeb 2026: 2 reportsMar 2026: 7 reportsApr 2026: 5 reportsMay 2026: 8 reportsJun 2026: 6 reportsJul 2026: 0 reportsAug 2026: 3 reportsSep 2026: 2 reportsOct 2026: 0 reportsNov 2026: 0 reports20242025202616
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Contamination with Chemical or Other Material295
  • Leak/Splash258
  • Contamination78
  • Device Markings/Labelling Problem67
  • Contamination /Decontamination Problem48
  • Volume Accuracy Problem47
  • Break44
  • Fluid/Blood Leak21

Patient problems

  • No Clinical Signs, Symptoms or Conditions462
  • No Known Impact Or Consequence To Patient317
  • No Patient Involvement62
  • No Consequences Or Impact To Patient61
  • Insufficient Information18
  • No Information4
  • Emotional Changes2
  • Bacterial Infection1

Recent adverse event reports

The 25 most recent of 928 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 28, 2026Contamination

    It was reported, Syringe 50 ML LL, contamination.\\The following was provided by the initial reporter: I have been notified of 2 more of the 50ml BD syringes that are contaminated.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50915174 ↗
  • MalfunctionSep 8, 2026Contamination

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50676787 ↗
  • InjuryAug 4, 2026Fluid/Blood Leak; Separation Problem

    It was reported, Syringe 50ml LL Tip 1ml, The plunger became detached during the preparation of a hazardous medication, causing leak and staff exposure.The following was provided by the initial reporter: This letter serves to formally report a complaint regarding the BD Luer Lok 50 mL syringe, lot numbers 6134989 and 6099628 (lots in circulation at the time of the incident). The plunger became detached during the preparation of a hazardous medication.The preparation is performed in a closed system, and the medication is drawn from a 2 mL vial. This is the first time such an incident has occurred. While pulling back on the plunger to withdraw BCG, the plunger came completely out of the syringe barrel. The plunger's removal caused a spill of hazardous medication, exposing personnel to significant risks. Fortunately, the preparer was wearing appropriate protective equipment, and the spill was quickly contained.For this preparation, the 60 mL syringe used (graduated up to 50 mL) is prefilled with 50 mL of saline. The closed system transfer device is then attached to the syringe tip and to the BCG vial. Next, 2 mL of saline is injected into the vial to dilute the powdered medication. Th

    Patient problems: Emotional Changes

    MAUDE report 50132945 ↗
  • InjuryAug 4, 2026Fluid/Blood Leak; Separation Problem

    It was reported, Syringe 50ml LL Tip 1ml, The plunger became detached during the preparation of a hazardous medication, causing leak and staff exposure.The following was provided by the initial reporter: This letter serves to formally report a complaint regarding the BD Luer Lok 50 mL syringe, lot numbers 6134989 and 6099628 (lots in circulation at the time of the incident). The plunger became detached during the preparation of a hazardous medication.The preparation is performed in a closed system, and the medication is drawn from a 2 mL vial. This is the first time such an incident has occurred. While pulling back on the plunger to withdraw BCG, the plunger came completely out of the syringe barrel. The plunger's removal caused a spill of hazardous medication, exposing personnel to significant risks. Fortunately, the preparer was wearing appropriate protective equipment, and the spill was quickly contained.For this preparation, the 60 mL syringe used (graduated up to 50 mL) is prefilled with 50 mL of saline. The closed system transfer device is then attached to the syringe tip and to the BCG vial. Next, 2 mL of saline is injected into the vial to dilute the powdered medication. Th

    Patient problems: Emotional Changes

    MAUDE report 50132918 ↗
  • MalfunctionAug 4, 2026Contamination

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50129493 ↗
  • MalfunctionJun 15, 2026Contamination

    IT WAS REPORTED, SYRINGE 50 ML, FOREIGN MATTER. THE FOLLOWING WAS PROVIDED BYT HE INITIAL REPORTER: ON 6/1/2026, ONE X BD 50 ML SYRINGE LUER-LOK TIP (REF 309653) (LOT 6099625 EXP 2031-03-31) WAS OPENED AND THE SYRINGE BARREL (CLOSEST TO SYRINGE TIP) AND INSIDE OF PACKAGING WAS DISCOLORED AND STICKY. NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25503880 ↗
  • MalfunctionJun 7, 2026Pumping Problem

    IT IS REPORTED, SYRINGE 50ML LL TIP 1ML, TCI PUMP, IT PRESENTS CONNECTION ERROR (2 UNITS). THE EVENT OCCURRED DURING THE USE OF THE PRODUCT AND DID NOT CAUSE HARM TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25422428 ↗
  • MalfunctionJun 3, 2026Leak/Splash

    IT WAS REPORTED, WHILE DRAWING UP A MEDICATION INTO A SYRINGE, THE MEDICATION STARTED SEEPING PAST THE PLUNGER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25373516 ↗
  • MalfunctionJun 1, 2026Contamination

    IT WAS REPORTED, EARWAX-FOREIGN MATTER. CUSTOMER VERBATIM: EXTRINSIC PARTICULATE (CERUMEN - EARWAX) OBSERVED IN FINAL PRODUCT. WHEN OBSERVED: THE EVENT WAS DETECTED ON DAY 10 OF THE FORM-FILL PROCESS. MATERIALS IN CONTACT: POINT-OF-CONTACT ITEMS INCLUDED CONICAL TUBES, SYRINGES, PIPETTES, AND PATIENT MATERIAL BAGS. NUMBER OF OCCURRENCES: THIS IS LIMITED TO BAG 2 OF 2 FOR LOT 26BC4062. WAS AFFECTED PRODUCT USED ON A PATIENT? NO ADDITIONAL INFORMATION: COULD YOU PLEASE CLARIFY WHERE EXACTLY ON THE PRODUCT (SYRINGE) WAS THE FOREIGN MATTER FOUND? -THE SYRINGE WAS USED DURING DAY 10 FORMULATION AND FILL PROCESSING WHICH OCCURS IN GRADE A ENVIRONMENT WHERE THE PERSONS ARE FULLY GOWNED AND EVEN NEW STERILE GLOVES ARE DONNED. THE SYRINGE PACKAGE ARE INTACT UNTIL FINALLY OPEN INTO BSC. THE AIM OF THE INVESTIGATION IS TO PROVIDED THE EVALUATION OF MANUFACTURING ENVIRONMENT. WHAT ARE THE AREAS WERE HUMAN /PRODUCT INTERVENTION OCCURS DURING INITIAL RAW MATERIAL TO FINAL PACKAGING AND WHAT CONTROLS DOES THE SITE HAVE TO EXCLUDE ANY EXTRINSIC PARTICULATES IN PLACE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25348057 ↗
  • MalfunctionJun 1, 2026Contamination

    IT WAS REPORTED, EARWAX-FOREIGN MATTER. CUSTOMER VERBATIM: EXTRINSIC PARTICULATE (CERUMEN - EARWAX) OBSERVED IN FINAL PRODUCT. WHEN OBSERVED: THE EVENT WAS DETECTED ON DAY 10 OF THE FORM-FILL PROCESS. MATERIALS IN CONTACT: POINT-OF-CONTACT ITEMS INCLUDED CONICAL TUBES, SYRINGES, PIPETTES, AND PATIENT MATERIAL BAGS. NUMBER OF OCCURRENCES: THIS IS LIMITED TO BAG 2 OF 2 FOR LOT 26BC4062. WAS AFFECTED PRODUCT USED ON A PATIENT? NO ADDITIONAL INFORMATION: COULD YOU PLEASE CLARIFY WHERE EXACTLY ON THE PRODUCT (SYRINGE) WAS THE FOREIGN MATTER FOUND? -THE SYRINGE WAS USED DURING DAY 10 FORMULATION AND FILL PROCESSING WHICH OCCURS IN GRADE A ENVIRONMENT WHERE THE PERSONS ARE FULLY GOWNED AND EVEN NEW STERILE GLOVES ARE DONNED. THE SYRINGE PACKAGE ARE INTACT UNTIL FINALLY OPEN INTO BSC. THE AIM OF THE INVESTIGATION IS TO PROVIDED THE EVALUATION OF MANUFACTURING ENVIRONMENT. WHAT ARE THE AREAS WERE HUMAN /PRODUCT INTERVENTION OCCURS DURING INITIAL RAW MATERIAL TO FINAL PACKAGING AND WHAT CONTROLS DOES THE SITE HAVE TO EXCLUDE ANY EXTRINSIC PARTICULATES IN PLACE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25348056 ↗
  • MalfunctionJun 1, 2026Contamination

    IT IS REPORTED, SYRINGE 50 ML TIP 1ML, FINE, ROUND, MAGENTA-COLORED PARTICULATE WAS OVERSERVED IN 1 UNIT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25348055 ↗
  • MalfunctionMay 26, 2026Contamination /Decontamination Problem

    IT WAS REPORTED BY THE CUSTOMER THAT APPEARS TO HAVE A PIECE OF THE PLASTIC IN THE BARREL WHEN THEY ARE DRAWING SOLUTION. NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25278720 ↗
  • MalfunctionMay 13, 2026Contamination

    IT IS REPORTED, SYRINGE 50ML LL TIP 1M, FOREIGN MATTER- DARK ROUND, CLEAR LONG THIN PARTICULATE INSIDE SYRINGE NEAR TIP (2 OCCURRENCES). NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25161097 ↗
  • MalfunctionMay 13, 2026Contamination

    IT IS REPORTED, SYRINGE 50ML LL TIP 1M, FOREIGN MATTER- WHITE FIBER, TOWARDS LUER (3 OCCURRENCES). NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25161089 ↗
  • MalfunctionMay 13, 2026Contamination

    IT IS REPORTED, FOREIGN MATTER- WHITE PARTICULATE, LOOSE, TOWARDS LUER (3). NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25161079 ↗
  • MalfunctionMay 11, 2026Contamination

    IT WAS REPORTED BY CUSTOMER THAT THEY NOTICED THAT THE 50ML SYRINGE LUER LOK TIP FROM BD (REF (B)(4) HAS A RESIDUE ON THE INSIDE OF THE PACKAGING. ADDITIONAL INFORMATION QUANTITY AFFECTED: 1 EA. WAS THEIR INJURY? ERROR OCCURRED BUT DID NOT REACH THE INDIVIDUAL THE EVENT DATE IS (B)(6) 2026.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25137106 ↗
  • MalfunctionMay 6, 2026Device Contamination with Chemical or Other Material

    IT IS REPORTED, SYRINGE 50ML LL TIP 1ML, BLACK STUFF INSIDE SYRINGE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25093774 ↗
  • MalfunctionMay 6, 2026Contamination

    IT IS REPORTED, SYRINGE 50ML LL TIP 1ML, HAD A STICK RESIDUE ON THE BLACK STOPPER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25093770 ↗
  • MalfunctionMay 6, 2026Contamination

    IT IS REPORTED, SYRINGE 50ML LL TIP 1ML, TIP HAS A DEFECT/PIECE OF PLASTIC ON THE END PREVENTING PROPER THREADING TO IV PORT CONNECTOR NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25093057 ↗
  • MalfunctionApr 15, 2026Contamination

    IT WAS REPORTED , PARTICLE IN 50 ML SYRINGE. LASER INDUCED BREAKDOWN SPECTROSCOPY WAS PERFORMED, AND WAS FOUND TO BE ''LIKELY POLY (ACRYLONITRILE BUTADIENE)''.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24885364 ↗
  • MalfunctionApr 15, 2026Contamination /Decontamination Problem

    NO ADDITIONAL INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24885357 ↗
  • MalfunctionApr 8, 2026Device Contamination with Chemical or Other Material

    IT WAS REPORTED BY THE CUSTOMER THAT SMALL BLACK SPECK FOUND INSIDE A STOCK SYRINGE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24817153 ↗
  • MalfunctionApr 6, 2026Contamination

    IT IS REPORTED LARGE BLACK MARKING ON TIP. NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24789719 ↗
  • MalfunctionApr 1, 2026Contamination

    NONE OF THESE SYRINGES WERE USED IN PATIENT CARE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24753724 ↗
  • MalfunctionMar 31, 2026Material Integrity Problem

    IT IS REPORTED, PLUNGER HAD A RUBBER GASKET DISCREPANCY ON PLUNGER, SHEDDING PARTICLES INTO PLUNGER, CONTAMINATING INFUSION AND CAUSING WASTE OF CANCER MEDICATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24743390 ↗

FDA clearance and classification

Cleared through Special 510(k) K110771 on May 13, 2011 (decision: Substantially Equivalent). Submitted by Becton, Dickinson and Company as “Bd Single Use, Hypodermic Syringe”. The FDA review took 53 days. FDA record ↗

The FDA regulates product code FMF (syringe, piston) under 21 CFR 880.5860 as a Class II device. 21 CFR 880.5860 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Recalls under the same FDA submission

These recalls involve devices cleared under K110771. One submission often covers many models, so check the product and lot details in each record.

  • Z-2614-2026 · Apr 28, 2026 · Open, Classified
    SYRINGE LUER LOK 10CC-200/BOX BD Product Code: 302995 Mentor Product Code: B9604 Syringe, Piston
  • Z-2749-2017 · Sep 16, 2016 · Terminated
    BD 3ml Syringe Luer-Lok" Tip with BD PrecisionGlide" Needle 23G x 1 (0.6mm x 25mm) Catalog #309571 The BD Single Use, Hypodermic Syringe is intended for use by health care professi
  • Z-2751-2017 · Sep 16, 2016 · Terminated
    BD 3ml Syringe Luer-Lok" Tip Catalog 309657 The BD Single Use, Hypodermic Syringe is intended for use by health care professionals for general purpose fluid aspiration/injection
  • Z-2747-2017 · Sep 16, 2016 · Terminated
    BD 5ml Syringe Luer-Lok" Tip with Blunt Fill Needle 18G x 1 1/2 (1.2mm x 40mm) Catalog #305062
  • Z-2752-2017 · Sep 16, 2016 · Terminated
    BD 10mL Syringe Luer-Lok Tip Catalog number 301997 The BD Single Use, Hypodermic Syringe is intended for use by health care professionals for general purpose fluid aspiration/injec
  • Z-2753-2017 · Sep 16, 2016 · Terminated
    BD 3ml Syringer Luer-Lok Tip Catalog #309604 The BD Single Use, Hypodermic Syringe is intended for use by health care professionals for general purpose fluid aspiration/injection
  • Z-2750-2017 · Sep 16, 2016 · Terminated
    BD 3ml Syringe Luer-Lok" Tip with BD PrecisionGlide" Needle 18G x 1 1/2 (1.2mm x 40mm) Catalog #309580 The BD Single Use, Hypodermic Syringe is intended for use by health care prof
  • Z-1937-2016 · Mar 25, 2016 · Terminated
    Henry Schein Single Use Syringe Luer-Lock, 10 mL, Intended for use by health care professionals for general purpose fluid aspiration/injection
  • Z-1936-2016 · Mar 25, 2016 · Terminated
    Henry Schein Single Use Syringe Luer-Lock, 10 mL, Intended for use by health care professionals for general purpose fluid aspiration/injection
  • Z-1055-2015 · Dec 19, 2014 · Terminated
    BD 10 mL Syringe BD Luer-Lok Tip with Blunt Plastic Cannula, Not for Skin Injection ; Catalog #303348, BD 10 mL Syringe Luer-Lok Tip with Blunt Fill Needle 18G x 1 1/2 (1.2mm x 4

Common questions about the BD Luer-Lok 309653

Has the BD Luer-Lok 309653 been recalled?

No FDA recall record in our data names this exact device identifier (DI 00382903096534), based on the FDA recall file dated Nov 27, 2026. However, 10 recalls involve devices cleared under the same FDA submission. They may or may not include this model, so check the product, model and lot details in those records.

What problems have been reported with the BD Luer-Lok 309653?

The FDA's MAUDE database lists 928 adverse event reports for this device, received between Jan 19, 2017 and Sep 28, 2026. In those reports, the most common specific device problems are “Device Contamination with Chemical or Other Material” (295), “Leak/Splash” (258) and “Contamination” (78); the most common patient problems are “Emotional Changes” (2) and “Bacterial Infection” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the BD Luer-Lok 309653?

None of the reports are classified as deaths, 4 as injuries and 924 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the BD Luer-Lok 309653 FDA approved?

It was cleared through Special 510(k) K110771 on May 13, 2011. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the BD Luer-Lok 309653, and is it still sold?

The FDA's device database lists Becton, Dickinson And Company as the company responsible for it (brand BD Luer-Lok, model 309653). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Syringe, piston devices

All Syringe, piston devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page