Skip to content

Insyte Autoguard 381034Recalls & reported problems

Becton, Dickinson And Company · BD Insyte™ Autoguard™ BC Pro Shielded IV Catheter Blood Control Technology 20 GA x 1.16 IN · Class 2

All models of Insyte Autoguard (20)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This BD Insyte Autoguard BC Pro Shielded IV catheter is a short-term intravenous line with a needle that retracts into the catheter after use. The 46 reports mostly describe the needle failing to retract, plastic pieces or loose fragments breaking off, and blood or fluid leaking or splashing; some mention small pieces of catheter left in the patient. Reports include 11 injuries and no deaths. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
46
Deaths reported
0
Injuries reported
11
Recalls naming it
0

Summary

The FDA's MAUDE database contains 46 adverse event reports that list this exact device (DI 00382903810345), received between Aug 24, 2018 and Jul 2, 2026 — classified as 11 injuries, 35 malfunctions.

The most frequently reported specific device problem is “Retraction Problem” (11 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 1 reportsJan 2024: 1 reportsFeb 2024: 2 reportsMar 2024: 1 reportsApr 2024: 2 reportsMay 2024: 2 reportsJun 2024: 4 reportsJul 2024: 3 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 2 reportsNov 2024: 0 reportsDec 2024: 3 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 6 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 4 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 1 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520266
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Retraction Problem11
  • Break9
  • Contamination6
  • Leak/Splash6
  • Defective Component4
  • Adverse Event Without Identified Device or Use Problem3
  • Fluid/Blood Leak3
  • Component Missing2

Patient problems

  • No Clinical Signs, Symptoms or Conditions32
  • Foreign Body In Patient6
  • Insufficient Information3
  • Foreign Body Embolism2
  • Thrombosis/Thrombus2
  • Needle Stick/Puncture1
  • Perforation of Vessels1
  • Skin Inflammation/ Irritation1

Recent adverse event reports

The 25 most recent of 46 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 2, 2026Material Puncture/Hole

    IT WAS DETERMINED THAT BD INSYTE AUTOG BC GLOBAL NON FOREIGN MATTER WAS FOUND DURING INVESTIGATION. ORIGINAL REPORT: "IT WAS REPORTED THAT THE NEEDLE HAD POPPED OUT OF THE CATHETER BEFORE USE." "SALES REP MENTIONED THAT: WHEN THEY OPENED THE PACKAGE, THE NEEDLE WAS STICKING OUT OF THE CATHETER. THEY HAVE REPORTED THIS SAME ISSUE NUMEROUS TIMES IN THE PAST. NATURALLY, THEY DID NOT PERFORM THE PUNCTURE."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25689283 ↗
  • MalfunctionMar 24, 2026Contamination

    IT WAS REPORTED THAT WHEN CUSTOMER OPENED THE BOX TO CHECK, THERE WAS A PORTION OF THE LABEL ON THE INDIVIDUAL PACKAGES THAT HAD BLACK TAPE ON IT BEFORE USE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24674613 ↗
  • MalfunctionDec 17, 2025Contamination; Contamination

    RN NOTED A ROUGH EDGE TO THE END OF THE CATHETER, APPEARS TO BE A LOOSE PIECE OF THE CATHETER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23834852 ↗
  • MalfunctionDec 17, 2025Contamination; Material Integrity Problem; Contamination; Material Integrity Problem

    THE ROUGH PLASTIC EDGE WAS LOOSE ON THIS ONE, CREATING A SMALL FB VISIBLE ON THE BEVEL. THIS TOOK PLACE BETWEEN NOV 10 AND NOV 28, THERE WERE A FEW INCIDENTS OF THE SAME ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23834843 ↗
  • MalfunctionDec 17, 2025Contamination; Contamination

    IT WAS REPORTED THAT, ¿MULTIPLE BLACK MARKS ON THE CATHETER. THE ONLY DEVICE FOUND WITH THESE MARKS. STORED WITH THE OTHER CATHETERS UNDER THE SAME CONDITIONS.¿ NO PATIENT INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23832430 ↗
  • MalfunctionDec 16, 2025Complete Blockage; Material Separation; Complete Blockage; Material Separation

    ADDITIONAL INFORMATION: CAN YOU PLEASE RECONFIRM AN EXACT DATE OF EVENT IN FORMAT DD-MMM-YYYY? 04-11-2025.

    Patient problems: Insufficient Information

    MAUDE report 23816297 ↗
  • MalfunctionJun 23, 2025Retraction Problem; Activation Problem; Retraction Problem; Activation Problem

    IT WAS REPORTED THAT BD IAG BC PRO GLOBAL NEEDLE DID NOT RETRACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: DURING CATHETER INSERTION, THE MANDREL CAN NO LONGER RETRACT. THIS COULD LEAD TO A BLOOD EXPOSURE ACCIDENT. SATURDAY (B)(6). COULD YOU CONFIRM THE IMPACT ON PATIENTS OF ALL THE EVENTS THAT HAVE OCCURRED? RESTING A CATHETER. WHEN YOU PRESSED THE BUTTON, DID THE NEEDLE FULLY RETRACT? NO. DID THE NEEDLE RETRACT SLOWER THAN USUAL? HAS NOT RETRACTED. DID THE NEEDLE START TO RETRACT AND THEN STOP, LEAVING THE NEEDLE EXPOSED? NO. DID THE NEEDLE NOT MOVE AT ALL AFTER PRESSING THE BUTTON? YES. DID THE BUTTON GO DOWN? NO.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22310221 ↗
  • MalfunctionJun 17, 2025Retraction Problem; Retraction Problem

    NO ADDITIONAL INFORMATION

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22229303 ↗
  • MalfunctionJun 16, 2025Retraction Problem; Retraction Problem

    IT WAS REPORTED THAT BDIAG BC PRO GLOBAL NEEDLE DID NOT RETRACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: FOR SEVERAL WEEKS OR EVEN MONTHS, DURING CATHETER INSERTION, THE DURING CATHETER INSERTION, THE MANDREL CAN NO LONGER RETRACT. THIS COULD LEAD TO A BLOOD EXPOSURE ACCIDENT. 6/3/25 ¿ YOU MENTIONED SEVERAL EVENTS. COULD YOU CONFIRM THE NUMBER OF EVENTS THAT HAVE OCCURRED? PLEASE ALSO LET US KNOW THE DATE OF THE EVENTS. TUESDAY 6 MAY, MONDAY 26 MAY, SATURDAY MAY 31 ¿ COULD YOU CONFIRM THE IMPACT ON PATIENTS OF ALL THE EVENTS THAT HAVE OCCURRED? CATHETER REMOVAL ¿ WHEN YOU PRESSED THE BUTTON, DID THE NEEDLE FULLY RETRACT? NO ¿ DID THE NEEDLE RETRACT SLOWER THAN USUAL? HAS NOT RETRACTED ¿ DID THE NEEDLE START TO RETRACT AND THEN STOP, LEAVING THE NEEDLE EXPOSED? NO ¿ DID THE NEEDLE NOT MOVE AT ALL AFTER PRESSING THE BUTTON? YES ¿ DID THE BUTTON GO DOWN? NO

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22215518 ↗
  • MalfunctionJun 16, 2025Retraction Problem; Retraction Problem

    IT WAS REPORTED THAT BDIAG BC PRO GLOBAL NEEDLE DID NOT RETRACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: FOR SEVERAL WEEKS OR EVEN MONTHS, DURING CATHETER INSERTION, THE DURING CATHETER INSERTION, THE MANDREL CAN NO LONGER RETRACT. THIS COULD LEAD TO A BLOOD EXPOSURE ACCIDENT. ON (B)(6) 2025: YOU MENTIONED SEVERAL EVENTS. COULD YOU CONFIRM THE NUMBER OF EVENTS THAT HAVE OCCURRED? PLEASE ALSO LET US KNOW THE DATE OF THE EVENTS. TUESDAY (B)(6) 2025, MONDAY (B)(6) 2025, SATURDAY (B)(6) 2025. COULD YOU CONFIRM THE IMPACT ON PATIENTS OF ALL THE EVENTS THAT HAVE OCCURRED? CATHETER REMOVAL. WHEN YOU PRESSED THE BUTTON, DID THE NEEDLE FULLY RETRACT? NO. DID THE NEEDLE RETRACT SLOWER THAN USUAL? HAS NOT RETRACTED. DID THE NEEDLE START TO RETRACT AND THEN STOP, LEAVING THE NEEDLE EXPOSED? NO. DID THE NEEDLE NOT MOVE AT ALL AFTER PRESSING THE BUTTON? YES. DID THE BUTTON GO DOWN? NO.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22215515 ↗
  • MalfunctionJun 4, 2025Crack; Leak/Splash; Crack; Leak/Splash

    IT WAS REPORTED THAT BD INSYTE AUTOG BC GLOBAL HUB IS CRACKED AND LEAKS. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: IT WAS REPORTED THAT THE CATHETER HUB WAS CRACKED, AND BLOOD WAS LEAKING FROM THE CATHETER HUB.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22140771 ↗
  • MalfunctionJun 2, 2025Component Missing; Component Missing

    IT WAS REPORTED THAT BD INSYTE AUTOG BC GLOBAL NEEDLE PIERCING POUCH. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: COMPLAIN/COMMENT: NEEDLE ON CANNULA WITHIN PACKAGING DOES NOT HAVE A PLASTIC CAP. CANNULAR NEEDLE WAS FOUND STICKING OUT OF THE END OF THE PACKAGING DUE TO NO PLASTIC PROTECTIVE CAP BEING ON TOP OF THE NEEDLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22127004 ↗
  • MalfunctionMay 19, 2025Retraction Problem; Retraction Problem

    IT WAS REPORTED THAT BD IAG BC PRO GLOBAL NEEDLE RETRACTED PREMATURELY. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: NEEDLE RETACTED BEFORE IV THROUGH PATIENTS SKIN, WHITE RETRACTOR BUTTON DEFINITELY NOT PRESSED. NO ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22045625 ↗
  • MalfunctionMar 31, 2025Retraction Problem; Retraction Problem

    NO ADDITIONAL INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21733926 ↗
  • MalfunctionFeb 18, 2025Retraction Problem; Retraction Problem

    IT WAS REPORTED THAT BD INSYTE AUTOG BC NEEDLE DID NOT RETRACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: NURSE STATING CANNULA WAS NOT RETRACTING WHEN BUTTON PRESSED, TRIED MULTIPLE CANNULA'S WITH THE. LOT: 4239978 FOR SAME ISSUES. TRIED NEW LOT: 4199766 SAME ISSUES WOULD NOT RETRACT. TRIED LOT: 3362802. WOULD NOT RETRACT AS WELL. TRIED LOT 4030760, WOULD ALSO NOT RETRACT RIGHT AWAY. IT APPEARS AS IF BUTTON IS GETTING STUCK. MULTIPLE STAFF TRIED WAS NOT USER ERROR. ALL CANNULA'S REMOVED FROM FLOOR AND SENT TO MATERIALS MANAGEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21403343 ↗
  • InjuryDec 29, 2024Break; Break

    IT WAS REPORTED THAT BD IAG BC PRO GLOBAL CATHETER TIP NOT PRESENT WHEN CATHETER WAS REMOVED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM ITALIAN TO ENGLISH: THE GARRISON WAS REMOVED DUE TO LIQUID SPILLAGE OF INFUSION. DURING REMOVAL IT WAS EVIDENT THAT THE DISTAL PORTION OF THE CANNULA WAS NOT PRESENT CONSEQUENCE HOSPITALIZATION (B)(6). THE DEVICE WAS IMPLANTED IN THE PATIENT; THE PATIENT SUFFERED DAMAGE AS HE UNDERWENT A CHEST CT SCAN AND CHEST X-RAY TO IDENTIFY THE RESIDUAL PIECE BELTED; IT WAS THE NURSING STAFF WHO HIGHLIGHTED THE PROBLEM DURING THE REMOVAL OF THE DEVICE; THERE WAS NO LEAKAGE OF LIQUID; MEDICAL INTERVENTION WAS NECESSARY; UNFORTUNATELY, THE DEVICE HAS BEEN DELETED AND I CAN'T SUPPLY THE LOT.

    Patient problems: Foreign Body In Patient

    MAUDE report 21032736 ↗
  • MalfunctionDec 10, 2024Contamination; Contamination

    IT WAS REPORTED THAT BD INSYTE AUTOG BC GLOBAL FOREIGN MATTER THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM JAPANESE TO ENGLISH: THIS IS A REPORT ABOUT FM IN THE NEEDLE. ALTHOUGH IT HAS NOT YET BEEN OPENED, THERE ARE BLACK SPOTS ON THE INDWELLING NEEDLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20896173 ↗
  • MalfunctionDec 5, 2024Defective Component; Defective Component

    IT WAS REPORTED THAT BD IAG BC PRO GLOBAL CATHETER BROKE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER TRANSLATED FROM FRENCH TO ENGLISH: WHEN THE CATHETER WAS INSERTED, THE OPERATOR INJECTED AND THEN FITTED THE CATHETER. AT THIS POINT, A PIECE OF PLASTIC CAME DETACHED. UNFORTUNATELY, THE DEVICE WAS THROWN AWAY. (B)(6); HAS THERE BEEN ANY HARM TO PATIENTS/HEALTHCARE PROFESSIONALS? THERE IS NO DAMAGE TO THE PATIENT, THERE IS A RUPTURE OF THE CATHETER FOR WHICH IT IS NECESSARY TO MAKE A CHANGE AND PERFORM DOUBLE PROCEDURE TO THE PATIENT. COULD YOU SHARE THE PHOTOS OF THE INCIDENT? PHOTOGRAPHIC EVIDENCE IS ADDED. WAS THERE A NEED FOR MEDICAL AND/OR SURGICAL INTERVENTION DUE TO THE INCIDENCE (IMAGING TESTS, SURGERY, DRUG ADMINISTRATION, ETC.)? (DETAIL) IT WAS NOT NECESSARY. WAS THERE EXPOSURE OF BLOOD/CHEMOTHERAPY TO MUCOUS MEMBRANES (EYES, NOSE AND MOUTH) OR SKIN? IF YES, INDICATE WHETHER THE EXPOSURE WAS FROM THE PATIENT OR THE PROFESSIONAL AND WHAT MEASURES WERE TAKEN. (DETAIL) THERE WAS NO EXPOSURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20855150 ↗
  • MalfunctionOct 11, 2024Fluid/Blood Leak; Fluid/Blood Leak

    NO ADDITIONAL INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20432997 ↗
  • MalfunctionOct 3, 2024Material Perforation; Material Perforation

    NO ADDITIONAL INFORMATION.

    Patient problems: Perforation of Vessels

    MAUDE report 20360431 ↗
  • InjuryJul 26, 2024Break; Break

    IT WAS REPORTED THAT BD IAG BC PRO GLOBAL CATHETER BROKE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: TIP OF CANNULA HAS BROKEN OFF AS MO WITHDRAWING WHOLE UNIT. MO ATTEMPTING TO WITHDRAW BLOOD SAMPLE WHEN NOTICED BALLOONING AND PT COMPLAINED OF SORENESS IN LEFT CUBITAL FOSSA. WITHDREW CANNULA AND TIP SNAPPED. TORNIQUET APPLIED AND URGENT USS REQUIRED. TIP HAS BEEN SURGICALLY REMOVED - LODGED IN SUBCUTANEOUS TISSUE.

    Patient problems: Foreign Body In Patient

    MAUDE report 19840381 ↗
  • MalfunctionJul 25, 2024Material Puncture/Hole; Material Puncture/Hole

    IT WAS REPORTED THAT BD IAG BC PRO GLOBAL NEEDLE PIERCED THE CATHETER. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM FRENCH TO ENGLISH: THE DEPARTMENT REPORTS THAT THE CANULA CAME OUT OF ITS SHEATH DURING THE PROCEDURE. THIS RESULTED IN BLEEDING AND THEN A HEMATOMA. 9JUL. HAVE HEALTH CARE WORKERS BEEN EXPOSED TO BLOOD OR BODY FLUIDS? NO, PUT ON A PAIR OF GLOVES WERE THERE ANY PROBLEMS WITH THE SECURITY SYSTEM? (RETRACTION FAILURE, PROTECTION FAILURE, SAFETY DEVICE FAILED TO COVER THE NEEDLE PROPERLY. INDICATE IF THE DEVICE MALFUNCTIONED BEFORE OR DURING USE). ACCORDING TO THE DOCTOR: "WHEN I SAW THAT THE NEEDLE HAD PIERCED THE PLASTIC SHEATH I TOLD MYSELF THAT RETRACTING IT COULD PRODUCE BLOOD SPLASHES, I PREFERRED TO LEAVE IT AS IT WAS. THE DEVICE MALFUNCTIONED DURING USE, IT WAS IN THE PATIENT'S VEIN". HAVE OTHER MEASURES BEEN TAKEN? REMOVAL OF THE COMPRESSION DEVICE WITH A COMPRESS SET UP OF AN ALCOHOLIC THOUGHT. DISCARDED DEVICE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19826841 ↗
  • MalfunctionJul 17, 2024Leak/Splash; Leak/Splash

    NO ADDITIONAL INFORMATION

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19766185 ↗
  • MalfunctionJun 24, 2024Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT BD INSYTE AUTOGUARD BC PRO LEAKED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: UPON INSERTION, THE AUTOGUARD IMMEDIATELY BEGAN BLEEDING IN THE CATHETER, TIME < 1 SEC, NOTHING ELSE ATTACHED TO CATHETER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19593311 ↗
  • MalfunctionJun 17, 2024Defective Component; Defective Component

    IT WAS REPORTED THAT BD INSYTE AUTOG BC BLOOD LEAKED BEYOND THE BLOOD CONTROL. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: THE CUSTOMER COMPLAINED THAT THE BLOOD CONTROL VALVE DO NOT FUNCTION PROPERLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions