Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00382903813339), received between Oct 27, 2022 and Jun 12, 2026 — classified as 5 malfunctions.
The most frequently reported specific device problem is “Device Damaged Prior to Use” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Device Damaged Prior to Use2
- Material Perforation2
- Device Contamination with Chemical or Other Material1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 12, 2026Material Perforation
IT WAS REPORTED BY CUSTOMER THAT STYLETS ARE STICKY WHEN TRYING TO ADVANCE. ALSO HAVING A LOT OF THE NEEDLES POKING THROUGH THE CATHETER. OBSERVED ISSUE PLEASE PROVIDE A DESCRIPTION OF THE OBSERVED ISSUE: STYLETS ARE STICKY WHEN TRYING TO ADVANCE. ALSO HAVING A LOT OF THE NEEDLES POKING THROUGH THE CATHETER. CUSTOMER RESPONSE ON 01-JUN-2026. 1. CAN YOU PLEASE PROVIDE AN EXACT DATE OF EVENT IN FORMAT MM-DD-YYYY? - THIS HAS BEEN AN ONGOING ISSUE WITH THIS LOT NUMBER. DATE I HAVE PHOTOS FROM IS 05-21-2026 2. DESCRIBE ANY PATIENT HARM, INJURY, COMPLICATION OR NEGATIVE OUTCOME THAT OCCURRED AS A RESULT OF THE EVENT. - COMPLICATION IS THE CATHETERS SEEM STICKY, WE CAN¿T ADVANCE THE CATHETER EASILY AND THE STYLET IS PIERCING THROUGH THE FLEXIBLE CATHETER 3. ANY SAMPLE AVAILABLE FOR INVESTIGATION? IF YES, ARE YOU ABLE TO PROVIDE THE ADDRESS OF THE FACILITY FOR US TO SHIP THE RETURN LABEL? - I HAVE 2 FULL BOXES (100) PLUS AN OPEN BOX OF CATHETERS WITH THE SAME LOT NUMBERS THAT WE¿VE BEEN STRUGGLING WITH XXX".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25488445 ↗MalfunctionJun 12, 2026Material Perforation
IT WAS REPORTED BY CUSTOMER THAT STYLETS ARE STICKY WHEN TRYING TO ADVANCE. ALSO HAVING A LOT OF THE NEEDLES POKING THROUGH THE CATHETER. OBSERVED ISSUE: PLEASE PROVIDE A DESCRIPTION OF THE OBSERVED ISSUE: STYLETS ARE STICKY WHEN TRYING TO ADVANCE. ALSO HAVING A LOT OF THE NEEDLES POKING THROUGH THE CATHETER. CUSTOMER RESPONSE ON 01-JUN-2026. 1. CAN YOU PLEASE PROVIDE AN EXACT DATE OF EVENT IN FORMAT MM-DD-YYYY? THIS HAS BEEN AN ONGOING ISSUE WITH THIS LOT NUMBER. DATE I HAVE PHOTOS FROM (B)(6) 2026. 2. DESCRIBE ANY PATIENT HARM, INJURY, COMPLICATION OR NEGATIVE OUTCOME THAT OCCURRED AS A RESULT OF THE EVENT. - COMPLICATION IS THE CATHETERS SEEM STICKY, WE CAN¿T ADVANCE THE CATHETER EASILY AND THE STYLET IS PIERCING THROUGH THE FLEXIBLE CATHETER. 3. ANY SAMPLE AVAILABLE FOR INVESTIGATION? IF YES, ARE YOU ABLE TO PROVIDE THE ADDRESS OF THE FACILITY FOR US TO SHIP THE RETURN LABEL? - I HAVE 2 FULL BOXES (100) PLUS AN OPEN BOX OF CATHETERS WITH THE SAME LOT NUMBERS THAT WE¿VE BEEN STRUGGLING WITH.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25488441 ↗MalfunctionAug 9, 2024Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material
IT WAS REPORTED THAT BD INSYTE-W WINGED PNK 20GA X 1.0IN HAD FOREIGN MATTER. IT WAS REPORTED BY CUSTOMER THAT THEY CALLED INTO THE QUEUE AND STATED THERE WAS A CLEAR JELLY-LIKE SUBSTANCE COMING FROM THE TIP OF THE CATHETER (VERY SMALL AMOUNT AND STICKY). HAPPENED TWO TIMES. DID NOT USE ON PT. VERBATIM: CUSTOMER CALLED INTO THE QUEUE AND STATED THERE WAS A CLEAR JELLY-LIKE SUBSTANCE COMING FROM THE TIP OF THE CATHETER (VERY SMALL AMOUNT AND STICKY). HAPPENED TWO TIMES. DID NOT USE ON PT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19951024 ↗MalfunctionOct 27, 2022Device Damaged Prior to Use; Device Damaged Prior to Use
IT WAS REPORTED THAT THE BD INSYTE-W¿ IV CATHETER EXPERIENCED A DEFECTIVE CATHETER. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: HAVE HAD SEVERAL CATHETERS BLOW IN A ROW MAKING US CONCERNED THAT THIS BOX MAY BE DEFECTIVE
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15678832 ↗MalfunctionOct 27, 2022Device Damaged Prior to Use; Device Damaged Prior to Use
IT WAS REPORTED THAT THE BD INSYTE-W¿ IV CATHETER EXPERIENCED A DEFECTIVE CATHETER. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: HAVE HAD SEVERAL CATHETERS BLOW IN A ROW MAKING US CONCERNED THAT THIS BOX MAY BE DEFECTIVE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15678830 ↗
Other Catheter,intravascular,therapeutic,short-term less than 30 days devices
- BD Insyte Autoguard 382523Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days660 reports
- BD Insyte Autoguard 382533Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days2 recalls527 reports
- BD PosiFlush 306546Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days1 death608 reports
- BD Insyte Autoguard Shielded IV Catheter 381823Becton, Dickinson And Company · Catheter, intravascular, therapeutic, short-term less than 30 days582 reports
- BD Insyte Autoguard 381423Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days1 recall528 reports
- BD Insyte Autoguard 382544Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days1 recall437 reports
- BD Nexiva, BD Q-Syte 383536Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days2 recalls374 reports
- BD PosiFlush 306547Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days2 recalls1 death331 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page