BD Insyte Autoguard 381457Recalls & reported problems
Becton, Dickinson And Company · INSYTE AUTOGUARD GRAY 16GA X 1.77IN · Class 2
All models of BD Insyte Autoguard (33)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00382903814572), received between Jun 8, 2021 and Mar 19, 2025 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Defective Component” (1 report).
11 other recalls cover devices cleared under the same FDA submission (K201075, K251654); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Defective Component1
- Device Contamination with Chemical or Other Material1
- Leak/Splash1
- Retraction Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 19, 2025Retraction Problem; Retraction Problem
IT WAS REPORTED THAT BD INSYTE AUTOGUARD NEEDLE DID NOT RETRACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: WHEN ACCESSING PATIENT'S SINGLE LUMEN POWERFLOW PORT WITH A 16 G X 1.77 INCH INSYTE AUTOGUARD NEEDLE, THE IV NEEDLE DID NOT RETRACT WHEN PRESSING THE BUTTON LOT# 4082066.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21646272 ↗MalfunctionOct 8, 2024Defective Component; Defective Component
NO ADDITIONAL INFORMATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20403652 ↗MalfunctionJun 8, 2021Leak/Splash; Device Contamination with Chemical or Other Material; Leak/Splash; Device Contamination with Chemical or Other Material
IT WAS REPORTED THAT INSYTE AUTOGUARD GRAY 16GA X 1.77IN CATHETER SEPARATED FROM THE HUB. THIS OCCURRED ON 2 OCCASIONS. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: IT WAS REPORTED BY THE HEALTH PROFESSIONAL THAT THE CATHETER DETACHED AT THE HUB TWICE. AN APHERESIS PATIENT WAS "STUCK" 3 TIMES INSTEAD OF ONCE, THIS IS THE HARM TO THE PATIENT AS THE RESULT OF THE PRODUCT FAILURE. THE PATIENT WAS NOT CAUSED FURTHER HARM BECAUSE OF ACTIVE RECOVERY BY THE RN'S EXPERTISE. THE NURSE WAS NOT HARMED. ON (B)(6) 2021 AN APHERESIS NURSE ENTERED THE ABOVE SB SAFE(INTERNAL TRACKING SYSTEM) FOR AN ANGIOCATH (BD INSYTE AUTO GUARD LOT# 9224642 EXP 7/31/2022) THAT DETACHED AT THE HUB TWICE (TWO DIFFERENT ANGIOCATHS) DURING A POWER FLOW PORT ANGIOCATH INSERTION REQUIRED FOR AN APHERESIS TREATMENT. WE SWITCHED TO A DIFFERENT LOT# AND DID NOT HAVE ANY FURTHER ISSUES. WE HAVE THE BOX OF LOT NUMBERS IN QUESTION QUARANTINED AND ARE USING THE NEW LOT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11963546 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K201075, K251654. One submission often covers many models, so check the product and lot details in each record.
- Z-1999-2025 · May 21, 2025 · Open, ClassifiedREF: 381423, BD Insyte Autoguard, Shielded IV Catheter, 22GA X 1.00 IN (0.9 X 25 MM) 35mL/min, STERILEEO, RxONLY
- Z-2007-2025 · May 21, 2025 · Open, ClassifiedREF: 381434 BD Insyte Autoguard Shielded IV Catheter, 20 GA x 1.16 in (1.1 x 30 mm) 60 mL/min, STERILEEO, RxOnly
- Z-2006-2025 · May 21, 2025 · Open, ClassifiedREF: 381923 BD Insyte Autoguard Winged Shielded IV Catheter, 22 GA x 1.00 in (0.9 x 25 mm) 35 mL/min, STERILEEO, RxOnly
- Z-2003-2025 · May 21, 2025 · Open, ClassifiedREF: 382944, BD Insyte Autoguard BC Pro Winged Shielded IV Catheter, 18 GA x 1.16 in (1.3 x 30 mm) 95 mL/min, STERILEEO, RxOnly
- Z-2002-2025 · May 21, 2025 · Open, ClassifiedREF: 381433, BD Insyte Autoguard Shielded IV Catheter, 20GA x 1.00 in (1.1x25 mm) 65 mL/min, STERILEEO, RxOnly
- Z-2004-2025 · May 21, 2025 · Open, ClassifiedREF: 381811, BD Insyte Autoguard BC Pro Winged Shielded IV Catheter, 18 GA x 1.16 in (1.3 x 30 mm) 95 mL/min, STERILEEO, RxOnly
- Z-2005-2025 · May 21, 2025 · Open, ClassifiedREF: 382544 BD Insyte Autoguard Shielded IV Catheter, 18 GA x 1.16 in (1.3 x 30 mm) 95 mL/min, STERILEEO, RxOnly
- Z-2000-2025 · May 21, 2025 · Open, ClassifiedREF: 382633, BD Insyte Autoguard BC Winged, Shielded IV Catheter with Blood Control Technology, 20GA x 1.00 in (1.1x25 mm) 63 mL/min, STERILEEO, RxOnly
- Z-2001-2025 · May 21, 2025 · Open, ClassifiedREF: 381533, BD Insyte Autoguard Winged, Shielded IV Catheter, 20 GA x 1.00 in (1.1 x 25 mm) 65mL/min, STERILEEO,RxOnly
- Z-2458-2024 · Jun 26, 2024 · Open, ClassifiedBD Insyte Autoguard BC, Shielded IV Catheter with Blood Control Technology, REF: 382533, 20 GA x 1.00 in (1.1x25 mm) 63 mL/min, Rx Only. BD Insyte Autoguard BC Shielded IV Catheter
- Z-0400-2024 · Oct 4, 2023 · Open, ClassifiedBD Insyte Autoguard BC Shielded IV Catheter with Blood Control Technology 20GA X 1.00in, REF: 382533
Other Catheter,intravascular,therapeutic,short-term less than 30 days devices
- BD Insyte Autoguard 382523Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days660 reports
- BD Insyte Autoguard 382533Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days2 recalls527 reports
- BD PosiFlush 306546Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days1 death608 reports
- BD Insyte Autoguard Shielded IV Catheter 381823Becton, Dickinson And Company · Catheter, intravascular, therapeutic, short-term less than 30 days582 reports
- BD Insyte Autoguard 381423Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days1 recall528 reports
- BD Insyte Autoguard 382544Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days1 recall437 reports
- BD Nexiva, BD Q-Syte 383536Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days2 recalls374 reports
- BD PosiFlush 306547Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days2 recalls1 death331 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page