Insyte Autoguard 381833Recalls & reported problems
Becton, Dickinson And Company · BD Insyte™ Autoguard™ Shielded IV Catheter 20 GA x 1.00 IN · Class 2
All models of Insyte Autoguard (20)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 13
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 13 adverse event reports that list this exact device (DI 00382903818334), received between Oct 12, 2022 and Apr 30, 2026 — classified as 2 injuries, 11 malfunctions.
The most frequently reported specific device problem is “Retraction Problem” (6 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Retraction Problem6
- Device Contamination with Chemical or Other Material3
- Adverse Event Without Identified Device or Use Problem1
- Component Missing1
- Contamination /Decontamination Problem1
- Fluid/Blood Leak1
- Material Integrity Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions10
- Perforation of Vessels2
- Pain1
- Skin Inflammation/ Irritation1
Recent adverse event reports
The 13 most recent of 13 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 30, 2026Material Integrity Problem
IT WAS REPORTED THAT THE CATHETER TIP HAD A BULGE. OUR ED STAFF IS HAVING TROUBLE WITH ALL IV CATHETERS WITH THIS LOT #. THEY ARE HAVING TROUBLE ADVANCING THE CATHETER AFTER THE INITIAL PUNCTURE OF THE VEIN, SOME NOT ABLE TO ADVANCE AND BLOWING THE VEIN. AFTER INSPECTION, SEVERAL STAFF BELIEVE THERE IS A SLIGHT BULGE NEAR THE TIP OF THE VIALON CATHETER THAT IS ABNORMAL AND MAY CONTRIBUTE TO THE DIFFICULTY ADVANCING.
Patient problems: Perforation of Vessels
MAUDE report 25045754 ↗MalfunctionDec 14, 2025Retraction Problem; Retraction Problem
MalfunctionOct 24, 2025Retraction Problem; Retraction Problem
NO ADDITIONAL INFORMATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23376310 ↗InjuryAug 7, 2025Contamination /Decontamination Problem; Contamination /Decontamination Problem
EVENT DETAILS: AN EMERGENCY ROOM NURSE ALERTED ME TODAY TO THE PRESENCE OF A FOREIGN BODY, PRESUMABLY A SMALL PIECE OF HARD PLASTIC ON THE OUTSIDE OF THE CATHETER ADAPTER OF A SHORT AUTOGUARD 20G 25MM KT REF 381833, LOT 4310283, WHICH SHE SAW WHILE PUSHING THE KT INTO A PATIENT'S VEIN. SHE IMMEDIATELY REMOVED THE KT. THIS SMALL PIECE (WHICH SHE CALLS A ¿SPANGLE¿ IN HER STATEMENT) FELL TO THE FLOOR AS THE KT WAS WITHDRAWN. WITH OTHER COLLEAGUES, SHE OPENED A DOZEN OTHER KTS FROM THE SAME BATCH AND FOUND A SIMILAR SMALL PIECE ON A 2ND KT THAT I RECOVERED AND VISUALIZED. AS NO OTHER BATCH OF THIS SIZE WAS AVAILABLE IN THE PHARMACY, I WAS UNABLE TO QUARANTINE THIS BATCH.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22745441 ↗InjuryMar 27, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT BD INSYTE AUTOGUARD POSSIBLE IV SITE INFECTION IN THE PRESENCE OF OTHER INFECTIOUS PROCESSES. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM FRENCH TO ENGLISH: ¿DATE OF EVENT: (B)(6) 2025, EQUIPMENT CONCERNED: BD INSYTE AUTOGUARD MODEL: 381833 LOT :4310283 PER: (B)(6) 2027 PROBLEM ENCOUNTERED: AFTER REMOVAL OF CATHETER, WOUND AND PAIN WITH INFLAMMATORY INDURATION OF A PATIENT. ACTION TAKEN: ALCOHOL AND WATER COMPRESS ON THE WOUND. ATYPICAL TREATMENT, ANTIBIOTIC. PATIENT IMPACT: INFLAMMATORY PAIN FEVER THE DEVICE IS NOT AVAILABLE PHOTO AND DOCTOR'S DESCRIPTION ATTACHED¿ ATTACHED IS AN INFECTION FOLLOWING A PERIPHERAL CATHETER INSERTED IN SMR ON (B)(6) 2025 AT 12 NOON BEFORE A COLONOSCOPY SCHEDULED FOR THE AFTERNOON. THE PATIENT WAS VOMITING AND IS DIABETIC. THE CATHETER WAS REMOVED THE SAME EVENING AFTER THE EXAMINATION. IN THE WEEKEND (B)(6) THAT FOLLOWED THE APPEARANCE OF PAIN WITH INFLAMMATORY INDURATION, COMPRESS ALCOHOL AND WATER APPLIED. ON MONDAY (B)(6) NOTICED THIS, PATIENT HAD COVID + FEVER. TREATED BY AUGMENTIN. NO MAJOR BIOLOGICAL INFLAMMATORY SYNDROME: 36 ON (B)(6) AND 46 ON (B)(6). APYRETIC SINCE (B)(6).
Patient problems: Pain; Skin Inflammation/ Irritation
MAUDE report 21711035 ↗MalfunctionOct 9, 2024Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material
NO ADDITIONAL INFORMATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20413234 ↗MalfunctionOct 4, 2024Fluid/Blood Leak; Fluid/Blood Leak
NO ADDITIONAL INFORMATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20371209 ↗MalfunctionOct 4, 2024Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material
IT WAS REPORTED THAT BD INSYTE AUTOGUARD FOREIGN MATTER FOUND ON NEEDLE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM FRENCH TO ENGLISH DATE OF OCCURRENCE: ON (B)(6) 2024. INCIDENT DESCRIPTION: WHILE HANDLING THE MANDREL, A SUSPICIOUS BLACKENING WAS SPOTTED ON THE CANULA AS WELL AS ON THE CATHETER ADAPTER. CURRENT PATIENT CONDITION: NO DIRECT CONSEQUENCES FOR THE PATIENT, PROBLEM IDENTIFIED BEFORE USE. RISK OF IATROGENIA IF CONTAMINATION OF THE BATCH IS CONFIRMED. ACTIONS TAKEN IN THE CARE FACILITY TO MANAGE THE PATIENT: USE OF ANOTHER DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20367440 ↗MalfunctionJan 16, 2024Retraction Problem; Retraction Problem
IT WAS REPORTED THAT THE BD INSYTE¿ AUTOGUARD¿ SHIELDED IV CATHETERS' NEEDLE WAS DID NOT RETRACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "ONCE I FINISHED INSERTING THE IV, I WENT TO PUSH THE BUTTON FOR THE NEEDLE RETRACTION AND THE BUTTON DIDN'T DEPRESS AND THE NEEDLE DID NOT RETRACT. I OBSERVED THERE WAS NO OBVIOUS DEFECT OF THE NEEDLE OR THE REST OF THE DEVICE. THE DEVICE WAS THE INSYTE AUTOGUARD SHIELDED IV CATHETER 20G. THE NEEDLE NOT RETRACTING PUT ME AT A POTENTIAL RISK THAT DID NOT HAPPEN, ONCE I REALIZED THE DEVICE WAS DEFECTIVE, I PLACED THE NEEDLE ON THE FLOOR AWAY FROM ME AND THE PATIENT UNTIL I COMPLETED THE IV INSERTION AND PLACED THE NEEDLE INTO THE SHARPS CONTAINER PROMPTLY FOLLOWING THE TASK. EXPOSED NEEDLE. NO HARM TO THE PATIENT OCCURRED."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18521251 ↗MalfunctionDec 21, 2023Retraction Problem; Retraction Problem
NO ADDITIONAL INFORMATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18371196 ↗MalfunctionDec 12, 2023Retraction Problem; Component Missing; Retraction Problem; Component Missing
IT WAS REPORTED THAT BD INSYTE AUTOGUARD HAD A NEEDLE RETRACTION PROBLEM WITH REPORT OF A MISSING SPRING. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM FRENCH TO ENGLISH: A 3AP NURSE WAS INJURED ON FRIDAY, (B)(6) 2023, WHILE ACTIVATING THE SAFETY OF A PERIPHERAL VENOUS CATHETER. SHE HELD THE CATHETER BETWEEN HER THUMB AND FOREFINGER. WHEN SHE ACTIVATED THE SECURITY, BY PRESSING THE WHITE BUTTON WITH HER THUMB, SHE FELT A STING ? A PINCH ? AT THE INDEX LEVEL. WHEN SHE LOOKED AT THE DEVICE AND COMPARED IT WITH ANOTHER DEVICE SHE HAD JUST OPENED, SHE NOTICED THAT A PIECE OF THE PLASTIC PART PROTECTING THE SPRING WAS MISSING. THE NURSE DID NOT KEEP THE OFFENDING CATHETER. IT WAS A BD INSYTE AUTOGUARD 20 GA REF 381833. CURRENTLY, DEVICES FROM THE LOT 3055420 EXPIRY DATE 2026-02-28 ARE USED IN THE SERVICE. CONSEQUENCES : NO WOUND A PRIORI BUT POTENTIAL RISK OF EXPOSURE TO THE PATIENT'S BLOOD LOCATION OF THE DEVICE : NOT PRESERVED, NOT AVAILABLE FOR EXPERTISE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18313295 ↗MalfunctionDec 6, 2023Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material
IT WAS REPORTED THAT BD INSYTE AUTOGUARD HAS FOREIGN MATTER INSIDE THE CAP. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: 1. PLEASE DESCRIBE THE SPECIFIC ISSUE FOUND ON DEFECTIVE PRODUCT? BLACK SPECS NOTED INSIDE THE CAP OF THE NEEDLE UPON OPENING A NEW ANGIOCATH 2. HOW WAS THE PATIENT OUTCOME? ARE THERE ANY CLINICAL SIGNS, HEALTH CONSEQUENCES OR IMPACT? NO IMPACT TO PATIENT, THE PRODUCT WAS SWITCHED TO A NEW ANGIOCATH AFTER CAREFUL INSPECTION. 3. ANY ADVERSE EVENT OR SERIOUS INJURY REPORTED TO PATIENT OR HEALTH CARE PROFESSIONAL? NO 4. ARE YOU ABLE TO PROVIDE THE DATES OF THE EVENTS IN THE FORMAT OF MM-DD-YYYY? IF UNKNOWN, CAN STATE UNKNOWN.(B)(6) 2023.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18279662 ↗MalfunctionOct 12, 2022Retraction Problem; Retraction Problem
IT WAS REPORTED THAT THE BD INSYTE¿ AUTOGUARD¿ IV CATHETER NEEDLE WOULD NOT RETRACT WHEN THE SAFETY DEVICE BUTTON WAS PRESSED. THIS OCCURRED WITH 5 CATHETERS. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM PORTUGUESE: "I REPORT 5 CASES OF CATHETER NUMBER 20 THAT DID NOT WORK SAFETY VALVE TO PROTECT THE NEEDLE. NURSING ACTIVATES THE WHITE BUTTON AND DOES NOT TRIGGER THE SAFETY DEVICE."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15588016 ↗
Other Catheter, intravascular, therapeutic, short-term less than 30 days devices
- BD Insyte Autoguard 382523Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days660 reports
- BD Insyte Autoguard 382533Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days2 recalls527 reports
- BD PosiFlush 306546Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days1 death608 reports
- BD Insyte Autoguard Shielded IV Catheter 381823Becton, Dickinson And Company · Catheter, intravascular, therapeutic, short-term less than 30 days582 reports
- BD Insyte Autoguard 381423Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days1 recall528 reports
- BD Insyte Autoguard 382544Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days1 recall437 reports
- BD Nexiva, BD Q-Syte 383536Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days2 recalls374 reports
- BD PosiFlush 306547Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days2 recalls1 death331 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page