Quick summary

This BD Insyte Autoguard shielded IV catheter is a short-term catheter used to start IV lines in a vein. Many of its 398 adverse event reports describe the safety needle not retracting as intended, along with leaks, contamination and catheters breaking or fraying. The reports list 14 injuries and no deaths, including a case where a catheter fragment had to be removed surgically. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
388
Deaths reported
0
Injuries reported
12
Recalls naming it
0

Summary

The FDA's MAUDE database contains 388 adverse event reports that list this exact device (DI 00382903818341), received between Oct 20, 2017 and Sep 28, 2026 — classified as 12 injuries, 376 malfunctions.

The most frequently reported specific device problem is “Retraction Problem” (134 reports).

11 other recalls cover devices cleared under the same FDA submission (K201075, K251654); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 6 reportsDec 2023: 1 reportsJan 2024: 4 reportsFeb 2024: 8 reportsMar 2024: 3 reportsApr 2024: 4 reportsMay 2024: 4 reportsJun 2024: 5 reportsJul 2024: 3 reportsAug 2024: 6 reportsSep 2024: 4 reportsOct 2024: 10 reportsNov 2024: 3 reportsDec 2024: 8 reportsJan 2025: 1 reportsFeb 2025: 6 reportsMar 2025: 6 reportsApr 2025: 0 reportsMay 2025: 3 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 6 reportsSep 2025: 3 reportsOct 2025: 4 reportsNov 2025: 3 reportsDec 2025: 4 reportsJan 2026: 1 reportsFeb 2026: 3 reportsMar 2026: 1 reportsApr 2026: 2 reportsMay 2026: 2 reportsJun 2026: 2 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 2 reportsOct 2026: 0 reports20242025202610
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Retraction Problem134
  • Leak/Splash57
  • Device Contamination with Chemical or Other Material43
  • Break39
  • Material Protrusion/Extrusion15
  • Material Puncture/Hole15
  • Contamination12
  • Device Markings/Labelling Problem11

Patient problems

  • No Clinical Signs, Symptoms or Conditions231
  • No Known Impact Or Consequence To Patient77
  • No Consequences Or Impact To Patient23
  • No Patient Involvement15
  • Insufficient Information13
  • Pain10
  • Extravasation7
  • No Information7

Recent adverse event reports

The 25 most recent of 388 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 28, 2026Retraction Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50922147 ↗
  • MalfunctionSep 15, 2026Contamination

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50775096 ↗
  • MalfunctionJun 22, 2026Material Puncture/Hole

    IT WAS REPORTED THAT THE CATHETER IS UNCOVERED, THE PATIENT IS PUNCTURED, IT IS DIFFICULT TO ENTER THE CATHETER INTO THE SKIN, SINCE ONLY THE NEEDLE ENTERS AND THE PLASTIC PART RETRACTS, SO THE CATHETER IS REMOVED AND THE FLOWERED PLASTIC TIP IS OBSERVED. QUANTITY: 01. TYPE OF DEFECT FOUND: THE PLASTIC PART OF THE CATHETER IS IN BAD CONDITION. ADDITIONAL INFORMATION RECEIVED ON 04 JUN 2026. I COULD SAY YES, SINCE A SECOND PUNCTURE HAD TO BE PERFORMED ON THE PATIENT. DATE OF OCCURRENCE (B)(6) 2026.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25575983 ↗
  • MalfunctionJun 17, 2026Retraction Problem

    IT WAS REPORTED THAT ¿UPON COMPLETION OF THE VVP INSERTION, THE CANULA DID NOT RETRACT FULLY (APPROXIMATELY 1 CM REMAINED PROTRUDING). I TRIED PRESSING THE BUTTON AGAIN TO RETRACT THE CANULA, BUT IT WOULD NOT RETRACT ANY FURTHER.¿ THE PATIENT DID NOT SUFFER ANY HARM. THE USER DID NOT SUFFER ANY HARM; HOWEVER, SHE NEARLY PRICKED HERSELF WITH THE CANULA USED AND COULD HAVE HAD A BLOOD EXPOSURE INCIDENT. INDEED, THE CANULA IS NORMALLY RETRACTED AFTER APPLYING THE KT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25533373 ↗
  • MalfunctionMay 11, 2026Use of Device Problem; Material Perforation; Material Twisted/Bent

    THE INVESTIGATION FINDINGS INDICATE THAT THE NEEDLE PIERCED THE CATHETER. THE CUSTOMER REPORTS THAT THE CATHETER WAS BENT. IT WAS REPORTED THAT WE HAVE RECEIVED SEVERAL COMPLAINTS FROM THE HEALTHCARE TEAM REGARDING THE PERFORMANCE OF THE DEVICE DURING VENIPUNCTURE. ACCORDING TO THE PROFESSIONALS, THE ANGIOCATH HAS PRESENTED DIFFICULTY IN PROGRESSION, WITH FREQUENT REPORTS THAT THE CATHETER ¿BENDS¿ DURING THE PUNCTURE ATTEMPT. THIS SITUATION ENDS UP MAKING VENOUS ACCESS IMPOSSIBLE ON THE FIRST ATTEMPT, RESULTING IN MULTIPLE PUNCTURES IN THE SAME PATIENT. DATE OF IDENTIFICATION OF THE OCCURRENCE; IT OCCURRED ON SEVERAL DAYS. WHEN WAS THE OCCURRENCE WITH THE PRODUCT IDENTIFIED (BEFORE STARTING THE PROCEDURE, DURING HANDLING OF THE PRODUCT BY THE PROFESSIONAL/USER, DURING USE ON THE PATIENT, OR AFTER USE ON THE PATIENT)? DURING USE ON THE PATIENT WAS THERE ANY HARM TO THE PATIENT, HEALTHCARE PROFESSIONAL, USER, OR OTHERS? (PROVIDE DETAILS); IN ADDITION TO DISCOMFORT AND ADDITIONAL RISK TO THE PATIENT, THE PROBLEM HAS GENERATED SIGNIFICANT WASTE OF MATERIALS, SINCE ANGIOCATHS THAT PRESENT THIS COMPLICATION BECOME UNUSABLE, DIRECTLY IMPACTING THE INSTITUTION'S COSTS. WAS MEDICAL AND/OR S

    Patient problems: Discomfort

    MAUDE report 25138234 ↗
  • MalfunctionMay 11, 2026Retraction Problem

    IT WAS REPORTED THAT THE NEEDLE PREMATURELY RETRACTED IN THE UNIT PACKAGE. I RECEIVED THE FOLLOWING REPORT FROM THE EMERGENCY ROOM TEAM: "DEVICE TRIGGERED SAFETY LOCK BEFORE USE, INCLUDING DISCONNECTION OF THE SILICONE, I EMPHASIZE THAT STILL IN THE PACKAGE AND WITHOUT HANDLING". ADDITIONAL INFORMATION RECEIVED ON 04-MAY-2026 IT HAPPENED ON (B)(6) 2026.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25137722 ↗
  • MalfunctionApr 23, 2026Leak/Splash; Material Deformation

    IT WAS REPORTED THAT THE HUB WAS DAMAGED AND UNABLE TO CONNECT/LEAKED. "THE TUBING DOES NOT FIT THE CATHETER (LEAK). UPON REMOVAL OF THE CATHETER, DEFORMATION IS OBSERVED AT THE SCREW-ON SECTION" UNFORTUNATELY, THE CATHETER HAS NOT BEEN RETAINED" RESPONSE RECEIVED ON:4/17/2026 I CAN CONFIRM THAT THERE HAS BEEN NO HARM TO EITHER THE PATIENT OR THE USER. RESPONSE RECEIVED ON:4/21/2026 I CAN CONFIRM THAT THE EVENT TOOK PLACE ON 16 APRIL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24965769 ↗
  • MalfunctionApr 22, 2026Contamination; Delivered as Unsterile Product

    IT WAS REPORTED THAT A HAIR WAS FOUND IN THE CRIMP OF THE MEDICAL DEVICE PACKAGING. RESPONSE RECEIVED ON:04/10/2026: THE EVENT OCCURRED ON 02/04/2026 AND WAS REPORTED TO US ON 6/04.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24954279 ↗
  • MalfunctionMar 30, 2026Detachment of Device or Device Component

    THIS MDR IS A RESULT OF AN INVESTIGATION FINDING. IT WAS REPORTED THAT WHEN CUSTOMER TRIED TO REMOVE THE PROTECTIVE CAP FROM THE INSTRUMENT, THE OUTER TUBE WAS STUCK AND ONLY THE INNER TUBE FELL OUT BEFORE USE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24714126 ↗
  • MalfunctionFeb 23, 2026Material Protrusion/Extrusion

    IT WAS REPORTED THAT THE NEEDLE PIERCED THE CATHETER. WAS THERE ANY DAMAGE TO THE PATIENT'S HEALTH? IF YES, EXPLAIN IN DETAIL. THERE WAS NO DAMAGE. DATE OF EVENT IS NOT CLEARLY DEFINED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24415121 ↗
  • MalfunctionFeb 19, 2026Retraction Problem

    NO ADDITIONAL INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24388492 ↗
  • MalfunctionFeb 12, 2026Retraction Problem

    NO ADDITIONAL INFORMATION FROM CUSTOMER.

    Patient problems: Insufficient Information

    MAUDE report 24324844 ↗
  • MalfunctionJan 26, 2026Material Split, Cut or Torn

    WHEN THE PATIENT'S IV LINE IS INSERTED, THE CATHETER OPENS. PATIENT IMPACT: NONE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24183422 ↗
  • MalfunctionDec 29, 2025Retraction Problem; Retraction Problem

    THE CATHETERS THAT WERE NOT ACTIVATING, WE CHECKED WITH THE NURSES IN CHARGE AND FOUND THAT APPROXIMATELY 20 CATHETERS HAD BEEN USED (WITH PROBLEMS) AND WHEN THE SAFETY DEVICE WAS ACTIVATED, IT DID NOT RETRACT THE NEEDLE. THERE IS NO HARM TO PATIENT HEALTH, BUT THERE IS A HIGH RISK OF STAFF INJURY DUE TO THE DEMAND FOR VENIPUNCTURE. I CANNOT GIVE YOU THE EXACT NUMBER, BUT WE HAVE FOUR SHIFTS AND IT WAS HAPPENING DAILY ON ALL FOUR SHIFTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23918150 ↗
  • MalfunctionDec 16, 2025Break; Leak/Splash; Break; Leak/Splash

    NO ADDITIONAL INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23821302 ↗
  • MalfunctionDec 16, 2025Retraction Problem; Retraction Problem

    SOMETIMES THE NEEDLE IS NOT EVEN RETRACTING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23815962 ↗
  • MalfunctionDec 9, 2025Retraction Problem; Retraction Problem

    THE NEEDLE CANNOT RETRACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23761972 ↗
  • MalfunctionNov 24, 2025Material Frayed; Nonstandard Device; Material Frayed; Nonstandard Device

    NO NEW INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23640094 ↗
  • MalfunctionNov 4, 2025Retraction Problem; Retraction Problem

    NO NEW INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23469669 ↗
  • MalfunctionNov 3, 2025Leak/Splash; Leak/Splash

    HUB IS LEAKING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23454879 ↗
  • MalfunctionOct 28, 2025Retraction Problem; Retraction Problem

    WHEN THE SAFETY BUTTON IS ACTIVATED, THE NEEDLE DOES NOT RETRACT INTO THE RETRACTION CHAMBER. PLEASE CONFIRM IF THERE HAS BEEN ANY IMPACT TO THE PATIENT. IF YES, EXPLAIN IN DETAIL. A: THERE WAS NO HARM TO THE PATIENT OR THE PROFESSIONAL, ONLY THE RISK OF EXPOSURE TO THE PATIENT'S PERFORATION AND RESIDUE. IT OCCURRED ON (B)(6) 2025, AND I BECAME AWARE OF IT ON THE SAME DATE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23410009 ↗
  • MalfunctionOct 16, 2025Retraction Problem; Retraction Problem

    IT WAS REPORTED THAT THE N20 INTRAVENOUS CATHETER DID NOT ACTIVATE AND ENDED UP LOSING ACCESS TO THE PATIENT. FAILURE OF THE SAFETY DEVICE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23319722 ↗
  • MalfunctionOct 16, 2025Insufficient Device Problem Information; Insufficient Device Problem Information

    HARD NEEDLES CANNOT BE RECYCLED. 20 AUG. THE CUSTOMER DOESN¿T RESPOND AND THERE IS NO ADDITIONAL INFORMATION IS AVAILABLE FROM THE CUSTOMER, THANK YOU!

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23316929 ↗
  • MalfunctionOct 15, 2025Contamination; Contamination

    NO NEW INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23299919 ↗
  • MalfunctionSep 17, 2025Retraction Problem; Retraction Problem

    NO ADDITIONAL INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23077459 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K201075, K251654. One submission often covers many models, so check the product and lot details in each record.

  • Z-2003-2025 · May 21, 2025 · Open, Classified
    REF: 382944, BD Insyte Autoguard BC Pro Winged Shielded IV Catheter, 18 GA x 1.16 in (1.3 x 30 mm) 95 mL/min, STERILEEO, RxOnly
  • Z-2006-2025 · May 21, 2025 · Open, Classified
    REF: 381923 BD Insyte Autoguard Winged Shielded IV Catheter, 22 GA x 1.00 in (0.9 x 25 mm) 35 mL/min, STERILEEO, RxOnly
  • Z-2002-2025 · May 21, 2025 · Open, Classified
    REF: 381433, BD Insyte Autoguard Shielded IV Catheter, 20GA x 1.00 in (1.1x25 mm) 65 mL/min, STERILEEO, RxOnly
  • Z-2001-2025 · May 21, 2025 · Open, Classified
    REF: 381533, BD Insyte Autoguard Winged, Shielded IV Catheter, 20 GA x 1.00 in (1.1 x 25 mm) 65mL/min, STERILEEO,RxOnly
  • Z-2007-2025 · May 21, 2025 · Open, Classified
    REF: 381434 BD Insyte Autoguard Shielded IV Catheter, 20 GA x 1.16 in (1.1 x 30 mm) 60 mL/min, STERILEEO, RxOnly
  • Z-2005-2025 · May 21, 2025 · Open, Classified
    REF: 382544 BD Insyte Autoguard Shielded IV Catheter, 18 GA x 1.16 in (1.3 x 30 mm) 95 mL/min, STERILEEO, RxOnly
  • Z-2004-2025 · May 21, 2025 · Open, Classified
    REF: 381811, BD Insyte Autoguard BC Pro Winged Shielded IV Catheter, 18 GA x 1.16 in (1.3 x 30 mm) 95 mL/min, STERILEEO, RxOnly
  • Z-2000-2025 · May 21, 2025 · Open, Classified
    REF: 382633, BD Insyte Autoguard BC Winged, Shielded IV Catheter with Blood Control Technology, 20GA x 1.00 in (1.1x25 mm) 63 mL/min, STERILEEO, RxOnly
  • Z-1999-2025 · May 21, 2025 · Open, Classified
    REF: 381423, BD Insyte Autoguard, Shielded IV Catheter, 22GA X 1.00 IN (0.9 X 25 MM) 35mL/min, STERILEEO, RxONLY
  • Z-2458-2024 · Jun 26, 2024 · Open, Classified
    BD Insyte Autoguard BC, Shielded IV Catheter with Blood Control Technology, REF: 382533, 20 GA x 1.00 in (1.1x25 mm) 63 mL/min, Rx Only. BD Insyte Autoguard BC Shielded IV Catheter
  • Z-0400-2024 · Oct 4, 2023 · Open, Classified
    BD Insyte Autoguard BC Shielded IV Catheter with Blood Control Technology 20GA X 1.00in, REF: 382533

Other Catheter, intravascular, therapeutic, short-term less than 30 days devices

All Catheter, intravascular, therapeutic, short-term less than 30 days devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page