Skip to content

Insyte Autoguard 381912Recalls & reported problems

Becton, Dickinson And Company · BD Insyte™ Autoguard™ Winged Shielded Catheter IV 24 GA x 0.75 IN · Class 2

All models of Insyte Autoguard (20)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This BD Insyte Autoguard winged IV catheter is a short-term intravenous line used for infusions. Most of the 36 reports describe the needle being slow to retract or the plastic fraying and shearing during insertion, along with leaking and fluid leaking into surrounding tissue. There are 6 injuries and no deaths. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
35
Deaths reported
0
Injuries reported
6
Recalls naming it
0

Summary

The FDA's MAUDE database contains 35 adverse event reports that list this exact device (DI 00382903819126), received between Apr 13, 2024 and Jun 8, 2026 — classified as 6 injuries, 29 malfunctions.

The most frequently reported specific device problem is “Retraction Problem” (16 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 2 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 1 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 2 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 2 reportsMar 2026: 8 reportsApr 2026: 7 reportsMay 2026: 3 reportsJun 2026: 3 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520268
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Retraction Problem16
  • Adverse Event Without Identified Device or Use Problem6
  • Break2
  • Defective Component2
  • Device Contamination with Chemical or Other Material2
  • Fluid/Blood Leak2
  • Connection Problem1
  • Leak/Splash1

Patient problems

  • No Clinical Signs, Symptoms or Conditions28
  • Extravasation6
  • Necrosis6
  • Swelling/ Edema6
  • Foreign Body In Patient1

Recent adverse event reports

The 25 most recent of 35 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 8, 2026

    RN ATTEMPTED PIV [PERIPHERAL IV] INSERTION; CATHETER FRAYED DURING INSERTION. THE CATHETER WAS REMOVED AND REPLACED WITH A NEW DEVICE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25425925 ↗
  • MalfunctionJun 8, 2026

    PRODUCT "SHEARED" IN PATIENT'S SKIN DURING THE IV INSERTION. CATHETER WAS REMOVED FROM PATIENT FULLY INTACT.

    Patient problems: Foreign Body In Patient

    MAUDE report 25425878 ↗
  • MalfunctionJun 3, 2026Retraction Problem

    IT WAS REPORTED THAT THE SAFETY DEVICE DID NOT WORK DURING USE. NEEDLE STICK MECHANISM NOT AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25382735 ↗
  • MalfunctionMay 27, 2026Retraction Problem

    IT WAS REPORTED THAT THE NEEDLE WAS SLOW TO RETRACT. "DESCRIBE CUSTOMER CONCERN, DETAILS OF EVENT INCLUDING HOW THE ISSUE WAS RESOLVED THE HEAD NURSE FROM THE ONCOLOGY CENTER AT NKTI EXPERIENCED A DELAY IN NEEDLE RETRACTION WHILE USING THE INSYTE AUTOGUARD G24 FOR A PATIENT UNDERGOING CHEMOTHERAPY INFUSION. INITIALLY, SHE BELIEVED THE ISSUE WAS ISOLATED TO A SINGLE DEVICE. HOWEVER, UPON REPEATED USE OF DEVICES FROM THE SAME BATCH AND LOT NUMBER, SHE CONSISTENTLY OBSERVED THE SAME DELAY. AFTER INSERTION, THE NEEDLE TOOK A FEW SECONDS TO FULLY RETRACT, UNLIKE OTHER BATCHES WHERE THE NEEDLE RETRACTS IMMEDIATELY. DESCRIBE IMPACT TO PATIENT, HCP, USER OR OTHER, INCLUDING RECOVERY, IF APPLICABLE? UPON CHECKING 3¿5 PIECES FROM THE SAME BATCH AND LOT NUMBER, THE SUPERVISOR DECIDED TO ESCALATE AND FORMALLY REPORT THE INCIDENT. ACCORDING TO THE END USERS, THE DELAYED NEEDLE RETRACTION CAUSED STRESS FOR BOTH NURSES AND PATIENTS. THEY ALSO EXPRESSED CONCERN REGARDING THE AFFECTED BATCH AND REQUESTED ITS REPLACEMENT".

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25301839 ↗
  • MalfunctionMay 20, 2026Retraction Problem

    IT WAS REPORTED THAT THE NEEDLE WAS SLOW TO RETRACT. "DESCRIBE CUSTOMER CONCERN, DETAILS OF EVENT INCLUDING HOW THE ISSUE WAS RESOLVED THE HEAD NURSE FROM THE ONCOLOGY CENTER AT (B)(6) EXPERIENCED A DELAY IN NEEDLE RETRACTION WHILE USING THE INSYTE AUTOGUARD G24 FOR A PATIENT UNDERGOING CHEMOTHERAPY INFUSION. INITIALLY, SHE BELIEVED THE ISSUE WAS ISOLATED TO A SINGLE DEVICE. HOWEVER, UPON REPEATED USE OF DEVICES FROM THE SAME BATCH AND LOT NUMBER, SHE CONSISTENTLY OBSERVED THE SAME DELAY. AFTER INSERTION, THE NEEDLE TOOK A FEW SECONDS TO FULLY RETRACT, UNLIKE OTHER BATCHES WHERE THE NEEDLE RETRACTS IMMEDIATELY. DESCRIBE IMPACT TO PATIENT, HCP, USER OR OTHER, INCLUDING RECOVERY, IF APPLICABLE? UPON CHECKING 3¿5 PIECES FROM THE SAME BATCH AND LOT NUMBER, THE SUPERVISOR DECIDED TO ESCALATE AND FORMALLY REPORT THE INCIDENT. ACCORDING TO THE END USERS, THE DELAYED NEEDLE RETRACTION CAUSED STRESS FOR BOTH NURSES AND PATIENTS. THEY ALSO EXPRESSED CONCERN REGARDING THE AFFECTED BATCH AND REQUESTED ITS REPLACEMENT".

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25233900 ↗
  • MalfunctionMay 15, 2026Material Frayed; Leak/Splash

    IT WAS REPORTED THAT INSYTE AUTOGUARD WINGED LEAKED. THE NURSE ATTEMPTED A PERIPHERALLY INSERTED IV PLACEMENT IN THE PATIENT'S LEFT HAND. THE IV SITE WAS LEAKING, WITH THE NEEDLE STILL IN THE CATHETER. AS THE NURSE PULLED THE IV OUT, IT WAS THAT THE TIP OF THE CATHETER WAS FRAYED. NO INFILTRATION OR EXTRAVASATION WAS NOTED. ADDITIONAL INFORMATION 5/7/2026: ANY ADVERSE EVENT OR SERIOUS INJURY REPORTED TO PATIENT OR HEALTHCARE PROFESSIONAL? IF YES, PLEASE PROVIDE THE DETAILS. NO ADVERSE OR SERIOUS INJURY TO THE PATIENT. CAN YOU PLEASE PROVIDE AN EXACT DATE OF EVENT? (B)(6) 2026.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25188544 ↗
  • MalfunctionApr 29, 2026

    THE NURSE ATTEMPTED A PERIPHERALLY INSERTED IV PLACEMENT IN THE PATIENT'S LEFT HAND. THE IV SITE WAS LEAKING, WITH THE NEEDLE STILL IN THE CATHETER. AS THE NURSE PULLED THE IV OUT, IT WAS THAT THE TIP OF THE CATHETER WAS FRAYED. NO INFILTRATION OR EXTRAVASATION WAS NOTED. THE IV WAS PLACED IN A BIOHAZARD BAG TO BE SENT TO SUPPLY CHAIN QA [QUALITY ASSURANCE] NURSE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25028071 ↗
  • InjuryApr 13, 2026Adverse Event Without Identified Device or Use Problem

    NO ADDITIONAL INFORMATION.

    Patient problems: Extravasation; Necrosis; Swelling/ Edema

    MAUDE report 24866363 ↗
  • InjuryApr 13, 2026Adverse Event Without Identified Device or Use Problem

    NO ADDITIONAL INFORMATION.

    Patient problems: Extravasation; Necrosis; Swelling/ Edema

    MAUDE report 24866362 ↗
  • InjuryApr 13, 2026Adverse Event Without Identified Device or Use Problem

    NO ADDITIONAL INFORMATION.

    Patient problems: Extravasation; Necrosis; Swelling/ Edema

    MAUDE report 24866345 ↗
  • InjuryApr 13, 2026Adverse Event Without Identified Device or Use Problem

    NO ADDITIONAL INFORMATION.

    Patient problems: Extravasation; Necrosis; Swelling/ Edema

    MAUDE report 24866341 ↗
  • InjuryApr 13, 2026Adverse Event Without Identified Device or Use Problem

    NO ADDITIONAL INFORMATION.

    Patient problems: Extravasation; Necrosis; Swelling/ Edema

    MAUDE report 24866337 ↗
  • InjuryApr 13, 2026Adverse Event Without Identified Device or Use Problem

    NO ADDITIONAL INFORMATION.

    Patient problems: Extravasation; Necrosis; Swelling/ Edema

    MAUDE report 24866335 ↗
  • MalfunctionMar 25, 2026Retraction Problem

    NO ADDITIONAL INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24689002 ↗
  • MalfunctionMar 25, 2026Retraction Problem

    NO ADDITIONAL INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24688976 ↗
  • MalfunctionMar 25, 2026Retraction Problem

    NO ADDITIONAL INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24688974 ↗
  • MalfunctionMar 25, 2026Retraction Problem

    NO ADDITIONAL INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24688956 ↗
  • MalfunctionMar 25, 2026Retraction Problem

    NO ADDITIONAL INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24688954 ↗
  • MalfunctionMar 25, 2026Retraction Problem

    NO ADDITIONAL INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24688573 ↗
  • MalfunctionMar 25, 2026Retraction Problem

    NO ADDITIONAL INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24688565 ↗
  • MalfunctionMar 2, 2026Retraction Problem

    IT WAS REPORTED THAT THE NEEDLE WAS SLOW TO RETRACT. INSYTE AUTOGUARD INSIDE THE IV INSERTION KIT ISSUED IN HAS BEEN REPORTED HAVING A DEFECT WHILE THEY PUSH THE BUTTON FOR NEEDLE ENCAPSULATION OF THE CATHETER. THEY HAVE MENTIONED THAT THEY HAVE A HARD TIME USING IT AS IT DELAYS NEEDLE ENCAPSULATION WHICH MAKES THE BUTTON HARDER TO PRESS DURING INSERTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24482428 ↗
  • MalfunctionFeb 26, 2026Material Protrusion/Extrusion

    NO NEW INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24462016 ↗
  • MalfunctionFeb 18, 2026Retraction Problem

    NO NEW INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24380828 ↗
  • MalfunctionOct 31, 2025Retraction Problem; Retraction Problem

    DATE OF OCCURRENCE: ON (B)(6) 2025. THE PRODUCT EASILY PROTRUDES DURING PERIPHERAL VENOUS PUNCTURE; TOUCHING THE DEVICE ACTIVATES IT, CAUSING THE NEEDLE TO PROTRUDE AND RESULTING IN LOSS OF VENOUS ACCESS. WAS THE CAUSE OF THE OCCURRENCE DETECTED? IF SO, DESCRIBE THE CAUSE. PRODUCT EASILY PROTRUDE DURING PERIPHERAL VENOUS PUNCTURE. AFTER THE CAUSE OF THE OCCURRENCE WAS DETECTED, WERE MEASURES TAKEN? IF SO, DESCRIBE THE MEASURES. YES. ALWAYS PERFORM VENOUS PUNCTURE WITH TWO TECHNICIANS TO ASSIST IN MOBILIZING THE LIMB TO BE PUNCTURED, TO AVOID OR MINIMIZE CATHETER PROTRUSION DURING PERIPHERAL VENOUS PUNCTURE. DID THIS OCCURRENCE CAUSE OR COULD IT CAUSE HEALTH PROBLEMS? IF YES, LOSS OF VENOUS ACCESS, FAMILY DISSATISFACTION, PATIENT EXPOSED TO A NEW PROCEDURE. DID THE EVENT LEAD TO HOSPITALIZATION? IF YES, PLACE OF HOSPITALIZATION: NA. CASE EVOLUTION: TEMPORARY INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23441566 ↗
  • MalfunctionAug 28, 2025Retraction Problem; Retraction Problem

    THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: INSYTE AUTOGUARD INSIDE THE IV INSERTION KIT ISSUED IN AHMC HAS BEEN REPORTED HAVING A DEFECT WHILE THEY PUSH THE BUTTON FOR NEEDLE ENCAPSULATION OF THE CATHETER. THEY HAVE MENTIONED THAT THEY HAVE A HARD TIME USING IT AS IT DELAYS NEEDLE ENCAPSULATION WHICH MAKES THE BUTTON HARDER TO PRESS DURING INSERTION. NO HARM/SERIOUS INJURY TO PATIENT/HCP.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22912001 ↗

Other Catheter, intravascular, therapeutic, short-term less than 30 days devices

All Catheter, intravascular, therapeutic, short-term less than 30 days devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page