Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00382903822775), received between Jan 9, 2023 and Jan 9, 2023 — classified as 1 malfunction.

The most frequently reported specific device problem is “Material Separation” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Material Separation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 9, 2023Material Separation; Material Separation

    IT WAS REPORTED WHILE USING BD ANGIOCATH¿ IV CATHETER THE CATHETER SEPARATED FROM THE HUB. THERE WAS NO REPORT OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: IT WAS STATED BY THE CUSTOMER THAT IT WAS DISCOVERED THAT THE ELONGATED SHEATH CAN BE DISCONNECTED FROM THE HUB COMPONENT OF THE SHEATH. THE HUB COMPONENT WAS UNKNOWINGLY LEFT INSITU WHILST CANNULATION OCCURRED. THIS RESULTED IN A PARTIAL CANNULATION TO ENABLE ATTACHMENT OF A BUNG AT THE CATHETER END. A FULLY ADVANCED CATHETER WOULD COVER THE HUB AND THE CANNULA COULDN¿T BE ACCESSED TO BE CAPPED OFF, RESULTING IN BLEED BACK.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16127938 ↗

Other Catheter,intravascular,therapeutic,short-term less than 30 days devices

All Catheter,intravascular,therapeutic,short-term less than 30 days devices →