BD Insyte Autoguard 382547Recalls & reported problems

Becton, Dickinson And Company · INSYTE AUTOG BC GRN 18GA X 1.88IN · Class 2

All models of BD Insyte Autoguard (33)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
10
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 00382903825479), received between Dec 19, 2017 and Feb 17, 2026 — classified as 10 malfunctions.

The most frequently reported specific device problem is “Retraction Problem” (4 reports).

11 other recalls cover devices cleared under the same FDA submission (K201075, K251654); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 2 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 2 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 2 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Retraction Problem4
  • Activation Problem1
  • Contamination1
  • Contamination /Decontamination Problem1
  • Leak/Splash1
  • Premature Separation1
  • Unsealed Device Packaging1

Patient problems

  • No Clinical Signs, Symptoms or Conditions7
  • No Known Impact Or Consequence To Patient2
  • Discomfort1
  • Insufficient Information1

Recent adverse event reports

The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 17, 2026Activation Problem

    THE CUSTOMER IS REPORTING A RETRACTION FAILURE THREADED THE CATHETER OFF THE NEEDLE AND PRESSED THE BUTTON ON THE DEVICE TO RETRACT THE NEEDLE, BUT THE NEEDLE DID NOT RETRACT, AND THE BUTTON REMAINED STUCK IN THE DEPRESSED STATE. ADDITIONAL INFORMATION 2/11/2026: IN THE MEDSUN FORM MENTIONED THAT THE DATE OF THE EVENT AS ¿(B)(6).¿. COULD YOU PLEASE SPECIFY THE EXACT DATE WHEN THE REPORTED ISSUE HAPPENED? I DO NOT HAVE THE EXACT DATE. ANY ADVERSE EVENT OR SERIOUS INJURY REPORTED TO PATIENT OR HEALTHCARE PROFESSIONAL? IF YES, PLEASE PROVIDE THE DETAILS. NO INJURY TO THE PATIENT OR TEAM MEMBER

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24370748 ↗
  • MalfunctionDec 30, 2025Contamination /Decontamination Problem; Contamination /Decontamination Problem

    NO ADDITIONAL INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23923703 ↗
  • MalfunctionDec 8, 2025Contamination; Contamination

    NO NEW INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23744737 ↗
  • MalfunctionJan 17, 2025Retraction Problem; Retraction Problem

    IT WAS REPORTED THAT BD INSYTE AUTOG BC NEEDLE DID NOT RETRACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: OUR PRE-OP MANAGER JUST CALLED, AND THEY ARE HAVING MORE ISSUES WITH THE INSYTE CATHETERS. PATIENTS ARE HAVING MORE INFILTRATED IV, PAIN AT THE SITE AND BRUISING. THESE ISSUES ARE OCCURRING AFTER THE IV HAS BEEN PLACED, WE DO NOT HAVE ANY PRODUCT SAVED, OR LOT# IDENTIFIED. WE HAVE ALSO HAD REPORTS FROM OUR IMAGING AND IV THERAPY DEPARTMENTS, SIMILAR TO WHAT WE HAVE BEEN SEEING WITH NEEDLES NOT COMPLETELY RETRACTING, DISENGAGED WITHOUT BUTTON BEING PUSHED, ETC. WOULD LIKE TO SEE A MEETING ON THE SCHEDULE SOONER RATHER THAN LATER. BECAUSE OF THE PATIENT ISSUES OF INFILTRATING, PAIN AND BRUISING, WE ARE NEEDING TO BRING IN AN ALTERNATE PRODUCT RIGHT AWAY.

    Patient problems: Discomfort; No Clinical Signs, Symptoms or Conditions

    MAUDE report 21179503 ↗
  • MalfunctionNov 7, 2024Retraction Problem; Retraction Problem

    IT WAS REPORTED THAT BD INSYTE AUTOG BC PARTIAL NEEDLE RETRACTION. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: IT WAS REPORTED BY CUSTOMER THAT THE NEEDLE NOT FULLY RETRACTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20630825 ↗
  • MalfunctionNov 7, 2024Retraction Problem; Retraction Problem

    NO ADDITIONAL INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20630701 ↗
  • MalfunctionJul 8, 2024Retraction Problem; Retraction Problem

    IT WAS REPORTED THAT BD INSYTE AUTOG BC NEEDLE DID NOT RETRACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: CUSTOMER REPORTS ISSUE WITH INSYTE CATHETER "WE HAVE RECEIVED FEEDBACK THAT OUR IV CATHETERS FOR OUR IV US IN THE ED ARE MALFUNCTIONING; THE SAFETY MECHANISM TO RETRACT THE NEEDLE IS NOT WORKING CORRECTLY RESULTING IN THE NEEDLE BEING UNSHEATHED AND BLOOD RUNNING OUT THE PLASTIC CHAMBER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19692338 ↗
  • MalfunctionJul 3, 2024Premature Separation; Premature Separation

    NO ADDITIONAL INFORMATION.

    Patient problems: Insufficient Information

    MAUDE report 19665660 ↗
  • MalfunctionNov 14, 2019Leak/Splash; Leak/Splash

    IT WAS REPORTED THAT AN UNSPECIFIED NUMBER OF 18G X 1.88IN (1.3 X 48 MM) INSYTE AUTOGUARD BC EXPERIENCED CATHETER ADAPTER/CONNECTOR/HUB CRACKED/BROKEN WHICH WAS NOTED PRIOR TO USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: MATERIAL NO: 382547 BATCH NO: UNKNOWN. HOLE IN THE GREEN PART OF THE HUB.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9324568 ↗
  • MalfunctionDec 19, 2017Unsealed Device Packaging

    IT WAS REPORTED THAT THE PACKAGES OF A BD INSYTE¿ AUTOGUARD¿ BC WERE OPENED BEFORE USE. NO INJURY OR MEDICAL INTERVENTION.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7132036 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K201075, K251654. One submission often covers many models, so check the product and lot details in each record.

  • Z-2007-2025 · May 21, 2025 · Open, Classified
    REF: 381434 BD Insyte Autoguard Shielded IV Catheter, 20 GA x 1.16 in (1.1 x 30 mm) 60 mL/min, STERILEEO, RxOnly
  • Z-2006-2025 · May 21, 2025 · Open, Classified
    REF: 381923 BD Insyte Autoguard Winged Shielded IV Catheter, 22 GA x 1.00 in (0.9 x 25 mm) 35 mL/min, STERILEEO, RxOnly
  • Z-2004-2025 · May 21, 2025 · Open, Classified
    REF: 381811, BD Insyte Autoguard BC Pro Winged Shielded IV Catheter, 18 GA x 1.16 in (1.3 x 30 mm) 95 mL/min, STERILEEO, RxOnly
  • Z-2002-2025 · May 21, 2025 · Open, Classified
    REF: 381433, BD Insyte Autoguard Shielded IV Catheter, 20GA x 1.00 in (1.1x25 mm) 65 mL/min, STERILEEO, RxOnly
  • Z-2003-2025 · May 21, 2025 · Open, Classified
    REF: 382944, BD Insyte Autoguard BC Pro Winged Shielded IV Catheter, 18 GA x 1.16 in (1.3 x 30 mm) 95 mL/min, STERILEEO, RxOnly
  • Z-2001-2025 · May 21, 2025 · Open, Classified
    REF: 381533, BD Insyte Autoguard Winged, Shielded IV Catheter, 20 GA x 1.00 in (1.1 x 25 mm) 65mL/min, STERILEEO,RxOnly
  • Z-2005-2025 · May 21, 2025 · Open, Classified
    REF: 382544 BD Insyte Autoguard Shielded IV Catheter, 18 GA x 1.16 in (1.3 x 30 mm) 95 mL/min, STERILEEO, RxOnly
  • Z-2000-2025 · May 21, 2025 · Open, Classified
    REF: 382633, BD Insyte Autoguard BC Winged, Shielded IV Catheter with Blood Control Technology, 20GA x 1.00 in (1.1x25 mm) 63 mL/min, STERILEEO, RxOnly
  • Z-1999-2025 · May 21, 2025 · Open, Classified
    REF: 381423, BD Insyte Autoguard, Shielded IV Catheter, 22GA X 1.00 IN (0.9 X 25 MM) 35mL/min, STERILEEO, RxONLY
  • Z-2458-2024 · Jun 26, 2024 · Open, Classified
    BD Insyte Autoguard BC, Shielded IV Catheter with Blood Control Technology, REF: 382533, 20 GA x 1.00 in (1.1x25 mm) 63 mL/min, Rx Only. BD Insyte Autoguard BC Shielded IV Catheter
  • Z-0400-2024 · Oct 4, 2023 · Open, Classified
    BD Insyte Autoguard BC Shielded IV Catheter with Blood Control Technology 20GA X 1.00in, REF: 382533

Other Catheter,intravascular,therapeutic,short-term less than 30 days devices

All Catheter,intravascular,therapeutic,short-term less than 30 days devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page