Skip to content

BD Insyte Autoguard 382644Recalls & reported problems

Becton, Dickinson And Company · INSYTE AUTOG BC WING GRN 18GA X 1.16IN · Class 2

All models of BD Insyte Autoguard (33)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This BD Insyte Autoguard winged IV catheter is a short-term intravenous line used for vascular access. Its 62 reports mostly describe needles that would not retract after insertion, sometimes with the button breaking off, plus leaking, blood leaking at the site, and a few cases where the tip sheared. No deaths or injuries were reported. There are no recalls naming this device, and reports in the last 12 months (10) are fewer than the 21 in the prior 12 months.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
62
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 62 adverse event reports that list this exact device (DI 00382903826445), received between Jul 31, 2018 and Mar 4, 2026 — classified as 62 malfunctions.

The most frequently reported specific device problem is “Retraction Problem” (34 reports).

11 other recalls cover devices cleared under the same FDA submission (K201075, K251654); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 2 reportsFeb 2025: 2 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 12 reportsJun 2025: 1 reportsJul 2025: 1 reportsAug 2025: 3 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 4 reportsFeb 2026: 3 reportsMar 2026: 2 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports20242025202612
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Retraction Problem34
  • Leak/Splash16
  • Fluid/Blood Leak13
  • Device Contamination with Chemical or Other Material6
  • Break5
  • Defective Device2
  • Material Puncture/Hole2
  • Material Split, Cut or Torn2

Patient problems

  • No Clinical Signs, Symptoms or Conditions26
  • No Known Impact Or Consequence To Patient16
  • Exposure to Body Fluids12
  • Insufficient Information12
  • No Consequences Or Impact To Patient5
  • Hemorrhage/Blood Loss/Bleeding1
  • Needle Stick/Puncture1
  • Pain1

Recent adverse event reports

The 25 most recent of 62 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 4, 2026Retraction Problem; Component Missing

    NO ADDITIONAL INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24512608 ↗
  • MalfunctionMar 4, 2026Retraction Problem

    NEEDLE WOULD NOT RETRACT AFTER IV PLACED AND BUTTON BROKE. ANY ADVERSE EVENT OR SERIOUS INJURY REPORTED TO PATIENT OR HEALTHCARE PROFESSIONAL? IF YES, PLEASE PROVIDE THE DETAILS. NO.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24506940 ↗
  • MalfunctionFeb 13, 2026Leak/Splash; Retraction Problem

    NEEDLE NOT RETRACTING. THE INCIDENT HAPPENED ON (B)(6) 2026 AT 1030. THE FEATURE TO STOP THE BLEEDING BACK DID NOT WORK, CAUSING THERE TO BE A MESS FOR THE NURSE AND PATIENT, BUT DID NOT CAUSE ANY HARM. THE RN WAS ABLE TO SAFELY DISCARD OF THE NON-RETRACTED NEEDLE WITHOUT INCIDENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24336559 ↗
  • MalfunctionFeb 9, 2026Material Split, Cut or Torn

    IT WAS REPORTED BY THE CUSTOMER THAT IV CATHETER SHEARED AT THE TIP. ADDITIONAL INFORMATION RECEIVED ON 04FEB2026. WHEN WAS THIS DEFECT OBSERVED? DURING OR BEFORE USE ON PATIENT? DURING USE ON PATIENT. ANY ADVERSE EVENT OR SERIOUS INJURY REPORTED TO PATIENT? IF YES, PLEASE PROVIDE THE DETAILS. NO.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24290705 ↗
  • MalfunctionFeb 3, 2026Retraction Problem

    NEEDLE WOULD NOT RETRACT AFTER IV PLACED AND BUTTON BROKE ANY ADVERSE EVENT OR SERIOUS INJURY REPORTED TO PATIENT OR HEALTHCARE PROFESSIONAL? IF YES, PLEASE PROVIDE THE DETAILS. NO.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24247761 ↗
  • MalfunctionJan 22, 2026Retraction Problem

    NO NEW INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24153437 ↗
  • MalfunctionJan 12, 2026Material Fragmentation; Retraction Problem

    NO NEW INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24042172 ↗
  • MalfunctionJan 12, 2026Retraction Problem

    NO NEW INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24041794 ↗
  • MalfunctionJan 12, 2026Retraction Problem

    NO NEW INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24041792 ↗
  • MalfunctionDec 28, 2025Leak/Splash; Material Puncture/Hole; Retraction Problem; Leak/Splash; Material Puncture/Hole; Retraction Problem

    NO NEW INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23903519 ↗
  • MalfunctionAug 18, 2025Retraction Problem; Retraction Problem

    NO NEW INFORMATION.

    Patient problems: Needle Stick/Puncture

    MAUDE report 22834173 ↗
  • MalfunctionAug 12, 2025Break; Break

    NO NEW INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22790041 ↗
  • MalfunctionAug 12, 2025Break; Break

    IT WAS REPORTED THAT BD INSYTE AUTOG BC WING NEEDLE BROKE (2 OF 2) THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: DURING INSERTION, AN INSYTE AUTOGUARD 18G IV BENT AND SUBSEQUENTLY BROKE IN HALF. IT WAS IMMEDIATELY REMOVED FROM THE PATIENT. THIS WAS THE SECOND CATHETER FROM THE SAME LOT TO EXHIBIT THE SAME ISSUE DURING THE SAME SHIFT. BOTH DEVICES WERE REMOVED PROMPTLY, THE CHARGE RN WAS NOTIFIED, AND ALL REMAINING STOCK FROM THIS LOT WAS PULLED FROM THE ED UNTIL THE MATTER CAN BE ADDRESSED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22790040 ↗
  • MalfunctionJul 31, 2025Retraction Problem; Retraction Problem

    NO NEW INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22680258 ↗
  • MalfunctionJun 26, 2025Retraction Problem; Retraction Problem

    IT WAS REPORTED THAT BD INSYTE AUTOG BC WING NEEDLE DID NOT RETRACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: CUSTOMER EXPERIENCING PROBLEMS VERY SIMILAR TO THE ISSUES DESCRIPED IN INSYTE AUTOGUARD RECALL. LOT IS NOT INCLUDED WITH THE RECALLED LOT NUMBERS BUT THINKING IT MAY BE THE SAME ISSUE. CUSTOMER IS SCARED TO USE THE PRODUCT WITH THE AFFECTED LOT NUMBER, AND NEEDS A FAST SOLUTION AND PRODUCT SENT OUT. ADDITIONAL INFO: COULD YOU PLEASE PROVIDE A FURTHER DESCRIPTION OF THE ISSUE? WHEN HITTING BUTTON TO RETRACT NEEDLE IT IS GETTIGN CAUGHT IN THE IV CATHETER CAUSING IT TO PULL THE CATHETER BACK ANY ADVERSE EVENT OR SERIOUS INJURY REPORTED TO PATIENT OR HEALTHCARE PROFESSIONAL? IF YES, PLEASE PROVIDE THE DETAILS. NO TOTAL NUMBER OF OCCURRENCES? 5 TO 10 IN LOT PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22340621 ↗
  • MalfunctionMay 19, 2025Fluid/Blood Leak; Retraction Problem; Fluid/Blood Leak; Retraction Problem

    IT WAS REPORTED THAT BD INSYTE AUTOG BC WING NEEDLE WAS SLOW TO RETRACT LEADING TO BLOOD LEAK. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: NEEDLE SLUGGISH WHEN RETRACTING DURING SAFETY ALLOWING BLOOD TO LEAK FROM HUB.

    Patient problems: Exposure to Body Fluids; Insufficient Information

    MAUDE report 22047772 ↗
  • MalfunctionMay 19, 2025Fluid/Blood Leak; Retraction Problem; Fluid/Blood Leak; Retraction Problem

    IT WAS REPORTED THAT BD INSYTE AUTOG BC WING NEEDLE WAS SLOW TO RETRACT LEADING TO BLOOD LEAK. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: NEEDLE SLUGGISH WHEN RETRACTING DURING SAFETY ALLOWING BLOOD TO LEAK FROM HUB.

    Patient problems: Exposure to Body Fluids; Insufficient Information

    MAUDE report 22047479 ↗
  • MalfunctionMay 19, 2025Fluid/Blood Leak; Retraction Problem; Fluid/Blood Leak; Retraction Problem

    IT WAS REPORTED THAT BD INSYTE AUTOG BC WING NEEDLE WAS SLOW TO RETRACT LEADING TO BLOOD LEAK. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: NEEDLE SLOW TO RETRACT ALLOWING BLOOD TO LEAK FROM HUB.

    Patient problems: Exposure to Body Fluids; Insufficient Information

    MAUDE report 22047477 ↗
  • MalfunctionMay 19, 2025Fluid/Blood Leak; Retraction Problem; Fluid/Blood Leak; Retraction Problem

    IT WAS REPORTED THAT BD INSYTE AUTOG BC WING NEEDLE WAS SLOW TO RETRACT LEADING TO BLOOD LEAK. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: NEEDLE SLOW TO RETRACT WHEN SAFETY PRESSED ALLOWING BLOOD TO LEAK FROM HUB.

    Patient problems: Exposure to Body Fluids; Insufficient Information

    MAUDE report 22047474 ↗
  • MalfunctionMay 19, 2025Fluid/Blood Leak; Retraction Problem; Fluid/Blood Leak; Retraction Problem

    IT WAS REPORTED THAT BD INSYTE AUTOG BC WING NEEDLE WAS SLOW TO RETRACT LEADING TO BLOOD LEAK. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: NEEDLE SLOWLY RETRACTING WHEN SAFETY PRESSED, ALLOWING BLOOD TO LEAK FROM THE HUB.

    Patient problems: Exposure to Body Fluids; Insufficient Information

    MAUDE report 22047466 ↗
  • MalfunctionMay 19, 2025Fluid/Blood Leak; Retraction Problem; Fluid/Blood Leak; Retraction Problem

    IT WAS REPORTED THAT BD INSYTE AUTOG BC WING NEEDLE WAS SLOW TO RETRACT LEADING TO BLOOD LEAK. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: NEEDLE SLOW TO SAFETY WHEN BUTTON PRESSED, ALLOWING BLOOD TO LEAK BACK.

    Patient problems: Exposure to Body Fluids; Insufficient Information

    MAUDE report 22047465 ↗
  • MalfunctionMay 19, 2025Fluid/Blood Leak; Retraction Problem; Fluid/Blood Leak; Retraction Problem

    IT WAS REPORTED THAT BD INSYTE AUTOG BC WING NEEDLE WAS SLOW TO RETRACT LEADING TO BLOOD LEAK. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: NEEDLE SLUGGISH WHEN RETRACTING DURING SAFETY ALLOWING BLOOD TO LEAK FROM HUB.

    Patient problems: Exposure to Body Fluids; Insufficient Information

    MAUDE report 22047464 ↗
  • MalfunctionMay 19, 2025Fluid/Blood Leak; Retraction Problem; Fluid/Blood Leak; Retraction Problem

    IT WAS REPORTED THAT BD INSYTE AUTOG BC WING NEEDLE WAS SLOW TO RETRACT LEADING TO BLOOD LEAK. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: NEEDLE SLOWLY RETRACTING OR GRINDING FEELING WITH SAFETY PRESSED ALLOWING BLOOD TO LEAK FROM HUB.

    Patient problems: Exposure to Body Fluids; Insufficient Information

    MAUDE report 22046960 ↗
  • MalfunctionMay 19, 2025Fluid/Blood Leak; Retraction Problem; Fluid/Blood Leak; Retraction Problem

    IT WAS REPORTED THAT BD INSYTE AUTOG BC WING NEEDLE WAS SLOW TO RETRACT LEADING TO BLOOD LEAK. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: NEEDLE TRACKING VERY SLOWING WITH GRINDING FEELING ALLOWING BLOOD TO LEAK BACK FROM HUB.

    Patient problems: Exposure to Body Fluids; Insufficient Information

    MAUDE report 22046958 ↗
  • MalfunctionMay 19, 2025Fluid/Blood Leak; Retraction Problem; Fluid/Blood Leak; Retraction Problem

    IT WAS REPORTED THAT BD INSYTE AUTOG BC WING NEEDLE WAS SLOW TO RETRACT LEADING TO BLOOD LEAK. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: BLOOD LEAKING FROM HUB WHEN SAFETY PRESSED, NEEDLE SLOW TO RETRACT.

    Patient problems: Exposure to Body Fluids; Insufficient Information

    MAUDE report 22046957 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K201075, K251654. One submission often covers many models, so check the product and lot details in each record.

  • Z-2007-2025 · May 21, 2025 · Open, Classified
    REF: 381434 BD Insyte Autoguard Shielded IV Catheter, 20 GA x 1.16 in (1.1 x 30 mm) 60 mL/min, STERILEEO, RxOnly
  • Z-2005-2025 · May 21, 2025 · Open, Classified
    REF: 382544 BD Insyte Autoguard Shielded IV Catheter, 18 GA x 1.16 in (1.3 x 30 mm) 95 mL/min, STERILEEO, RxOnly
  • Z-2000-2025 · May 21, 2025 · Open, Classified
    REF: 382633, BD Insyte Autoguard BC Winged, Shielded IV Catheter with Blood Control Technology, 20GA x 1.00 in (1.1x25 mm) 63 mL/min, STERILEEO, RxOnly
  • Z-2004-2025 · May 21, 2025 · Open, Classified
    REF: 381811, BD Insyte Autoguard BC Pro Winged Shielded IV Catheter, 18 GA x 1.16 in (1.3 x 30 mm) 95 mL/min, STERILEEO, RxOnly
  • Z-2006-2025 · May 21, 2025 · Open, Classified
    REF: 381923 BD Insyte Autoguard Winged Shielded IV Catheter, 22 GA x 1.00 in (0.9 x 25 mm) 35 mL/min, STERILEEO, RxOnly
  • Z-2002-2025 · May 21, 2025 · Open, Classified
    REF: 381433, BD Insyte Autoguard Shielded IV Catheter, 20GA x 1.00 in (1.1x25 mm) 65 mL/min, STERILEEO, RxOnly
  • Z-1999-2025 · May 21, 2025 · Open, Classified
    REF: 381423, BD Insyte Autoguard, Shielded IV Catheter, 22GA X 1.00 IN (0.9 X 25 MM) 35mL/min, STERILEEO, RxONLY
  • Z-2001-2025 · May 21, 2025 · Open, Classified
    REF: 381533, BD Insyte Autoguard Winged, Shielded IV Catheter, 20 GA x 1.00 in (1.1 x 25 mm) 65mL/min, STERILEEO,RxOnly
  • Z-2003-2025 · May 21, 2025 · Open, Classified
    REF: 382944, BD Insyte Autoguard BC Pro Winged Shielded IV Catheter, 18 GA x 1.16 in (1.3 x 30 mm) 95 mL/min, STERILEEO, RxOnly
  • Z-2458-2024 · Jun 26, 2024 · Open, Classified
    BD Insyte Autoguard BC, Shielded IV Catheter with Blood Control Technology, REF: 382533, 20 GA x 1.00 in (1.1x25 mm) 63 mL/min, Rx Only. BD Insyte Autoguard BC Shielded IV Catheter
  • Z-0400-2024 · Oct 4, 2023 · Open, Classified
    BD Insyte Autoguard BC Shielded IV Catheter with Blood Control Technology 20GA X 1.00in, REF: 382533

Other Catheter,intravascular,therapeutic,short-term less than 30 days devices

All Catheter,intravascular,therapeutic,short-term less than 30 days devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page