Skip to content

BD Nexiva 383576Recalls & reported problems

Becton, Dickinson And Company · NEXIVA 20GA X 1.00 IN DP WITH MAXZERO · Class 2

All models of BD Nexiva (25)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
11
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 00382903835768), received between Feb 8, 2022 and Sep 16, 2024 — classified as 11 malfunctions.

The most frequently reported specific device problem is “Material Protrusion/Extrusion” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 1 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 2 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Protrusion/Extrusion2
  • Deformation Due to Compressive Stress1
  • Detachment of Device or Device Component1
  • Device Dislodged or Dislocated1
  • Difficult or Delayed Activation1
  • Leak/Splash1
  • Material Deformation1
  • Obstruction of Flow1

Patient problems

  • No Clinical Signs, Symptoms or Conditions8
  • Insufficient Information2
  • Failure of Implant1

Recent adverse event reports

The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 16, 2024Deformation Due to Compressive Stress; Deformation Due to Compressive Stress

    NO ADDITIONAL INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20227223 ↗
  • MalfunctionSep 6, 2024Leak/Splash; Leak/Splash

    IT WAS REPORTED THAT BD NEXIVA 20GA X 1.00 IN DP WITH MAXZERO TUBING DEFECTIVE / DAMAGED. IT WAS REPORTED BY CUSTOMER THAT NEXIVA CATHETER EXTENSION SET "BLEW UP" DURING CT SCAN WHEN USING ISOVUE300 AND ISOVUE370. VERBATIM: NEXIVA CATHETER EXTENSION SET "BLEW UP" DURING CT SCAN WHEN USING ISOVUE300 AND ISOVUE370. ADDITIONAL INFORMATION RECEIVED ON 16AUG. HELLO, BECAUSE THE PRODUCT WAS USED ON THE PATIENT IN OUR EMERGENCY ROOM, WE DO NOT HAVE ANY PACKAGING WITH A SPECIFIC LOT NUMBER. THE PRODUCT NUMBER IS 383576. I ALSO HAVE THE ITEM IN MY POSSESSION THAT LOOKS ALMOST MELTED. THE CLINICAL STAFF WILL NEED TO ANSWER IF ANYTHING HAPPENED TO THE PATIENT. THANK YOU. ADDITIONAL INFORMATION RECEIVED ON 16AUG. NO HARM TO PATIENT. ALL HAVE BEEN INSTRUCTED TO RETAIN ANY FUTURE DEVICES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20163582 ↗
  • MalfunctionJan 11, 2024

    ISSUE WITH THIS DEVICE BENDING INSIDE THE PATIENT'S VEIN DURING INSERTION, ALL ITEMS REMOVED FROM THE SHELF.

    Patient problems: Insufficient Information

    MAUDE report 18491107 ↗
  • MalfunctionDec 13, 2023

    NEXIVA 20 GA IV, CATHETER TIPS ARE KINKING AND FOLDING INSIDE OF PATIENTS. MULTIPLE RN'S HAVE EXPERIENCED THE SAME ISSUE.

    Patient problems: Failure of Implant

    MAUDE report 18317811 ↗
  • MalfunctionNov 21, 2023Difficult or Delayed Activation; Difficult or Delayed Activation

    NO ADDITIONAL INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18186428 ↗
  • MalfunctionAug 14, 2023Material Protrusion/Extrusion; Material Protrusion/Extrusion

    IT WAS REPORTED THE BD NEXIVA¿ CLOSED IV CATHETER SYSTEM HAD A BULGE NEAR THE EXTENSION WINGS. THE FOLLOWING WAS RECEIVED BY THE INITIAL REPORTER: PROBLEM INFORMATION. CITE CUSTOMER REF#(B)(4) 20G CATHALON INITIATED TO L ACF. WHEN RETURNED FROM CT THE NEXIVA 20G HAD A BULGE IN THE EXTENSION BY THE WINGS. THE BULGE WAS RIGHT BESIDE THE WING AS IT COMES OUT TO THE EXTENSION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17546995 ↗
  • MalfunctionJun 30, 2023Material Deformation; Material Deformation

    IT WAS REPORTED WHILE USING BD NEXIVA¿ CLOSED IV CATHETER SYSTEM THE CATHETER KINKED DURING INFUSION. THERE WAS NO REPORT OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: RN STARTED PERIPHERAL IV, FLASH NOTED, AND CATHETER THREADED WITHOUT DIFFICULTY. FLASH STOPPED AT THE END OF THE TUBING, RN ATTEMPTED TO FLUSH CATHETER AND IMMEDIATELY ENCOUNTERED RESISTANCE. PATIENT REPORTED NO PAIN AT SIDE. RN REMOVED PERIPHERAL IV CATHETER; END OF CATHETER TIP BENT SIGNIFICANTLY BUT INTACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17240471 ↗
  • MalfunctionMay 11, 2023Obstruction of Flow; Obstruction of Flow

    IT WAS REPORTED WHILE USING BD NEXIVA¿ CLOSED IV CATHETER SYSTEM ¿ DUAL PORT MAXZERO NEEDLELESS CONNECTOR THE TUBING EXPANDED. THERE WAS NO REPORT OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: DURING CT SCAN AND CONTRAST ADMINISTRATION, IV TUBING ATTACHED TO NEXIVA HERNIATED OUT, BUT DID NOT RUPTURE. IV WAS REMOVED AND RESTARTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16915984 ↗
  • MalfunctionApr 18, 2023Device Dislodged or Dislocated

    DURING CT SCAN AND CONTRAST ADMINISTRATION, IV TUBING ATTACHED TO NEXIVA HERNIATED OUT, BUT DID NOT RUPTURE. IV WAS REMOVED AND RESTARTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16761080 ↗
  • MalfunctionMar 1, 2022Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED WHILE USING NEXIVA¿ CLOSED IV CATHETER SYSTEM ¿ DUAL PORT MAXZERO NEEDLELESS CONNECTOR A THE NEEDLE DETACHED FROM THE SAFETY MECHANISM. THERE WAS NO REPORT OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: MATERIAL NO.: 383576 BATCH NO.: 0300545 IT WAS REPORTED THAT NEEDLE WAS EXPOSED AND NOT PROTECTED, AFTER IV WAS STARTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13638756 ↗
  • MalfunctionFeb 8, 2022Material Protrusion/Extrusion; Protective Measures Problem

    IV WAS STARTED USING 20G BD NEXIVA CLOSED IV CATHETER. AFTER IV WAS STARTED AND NEEDLE WAS DISCONNECTED FROM CATHETER, NEEDLE CAME OUT EXPOSED AND NOT PROTECTED

    Patient problems: Insufficient Information

    MAUDE report 13479893 ↗

Other Catheter, intravascular, therapeutic, short-term less than 30 days devices

All Catheter, intravascular, therapeutic, short-term less than 30 days devices →