BD Nexiva™ Diffusics™ Closed IV Catheter System 383596Recalls & reported problems

Becton, Dickinson And Company · NEXIVA DIFFUSICS w/MaxZero 22G X 1.00 IN · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This BD Nexiva Diffusics closed IV catheter is a short-term intravenous line placed for patients needing IV therapy or fluids. Reports mostly describe the catheter leaking, with26 mentioning leaking or splashing, plus a missing vent plug, a hole in the material, and a safety needle shield that would not activate so the needle could not be removed. There are 27 reports, with no deaths and no injuries, and no recalls name this device. Reports jumped from 1 in the prior 12 months to 25 in the last 12 months.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
27
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 27 adverse event reports that list this exact device (DI 00382903835966), received between Aug 20, 2024 and Apr 24, 2026 — classified as 27 malfunctions.

The most frequently reported specific device problem is “Leak/Splash” (26 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 23 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 2 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports20242025202623
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Leak/Splash26
  • Activation Problem1
  • Manufacturing, Packaging or Shipping Problem1
  • Material Puncture/Hole1

Patient problems

  • No Clinical Signs, Symptoms or Conditions27

Recent adverse event reports

The 25 most recent of 27 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 24, 2026Leak/Splash

    IT WAS REPORTED BY CUSTOMER THAT CATHETER HAS LEAK. VERBATIM: COMPLAINT VIA EMAIL. EMAIL(S) ATTACHED CATHETER HAS LEAK.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24984231 ↗
  • MalfunctionApr 3, 2026Leak/Splash; Manufacturing, Packaging or Shipping Problem

    IT WAS REPORTED BY A CUSTOMER THAT THE CLEAR PLASTIC CAP (VENT PLUG) IS EITHER DISCONNECTED OR MISSING FROM THE PACKAGING AND RESULTED IN A LOT OF BLOOD LEAKAGE AND EXPOSURE DURING USE. THE CLEAR PLASTIC CAP (VENT PLUG) IS EITHER DISCONNECTED OR MISSING FROM THE PACKAGING. THE AFFECTED LOT NUMBER IS 5338239. I¿VE ATTACHED A FEW PHOTOS FOR YOUR REFERENCE. THIS ISSUE HAS, IN SOME CASES, RESULTED IN A LOT OF BLOOD LEAKAGE AND EXPOSURE DURING USE, WHICH IS OBVIOUSLY A CONCERN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24781044 ↗
  • MalfunctionJan 22, 2026Activation Problem

    IT WAS REPORTED THAT BD NEXIVA DIFFUSICS W/MAXZERO 22G X 1.00 IN SAFETY SHIELD ACTIVATION FAILED IT WAS REPORTED BY THE CUSTOMER THAT WAS TRYING TO SAFETY THE NEEDLE AFTER PLACING THE IV. THE CUSTOMER HEARD THE CLICK OF THE SAFETY FEATURES ACTIVATING AND THE NEEDLE LOCKING. WHEN TRYING TO REMOVE THE NEEDLE FROM THE CATHETER BY PULLING ON THE WHITE HE COULD NOT REMOVE THE NEEDLE DESPITE THE NEEDLE BEING LOCKED AND HE WAS NOT HOLDING THE GREY IN PLACE WITH THE CATHETER. THE CUSTOMER IS VERY FAMILIAR WITH THESE CATHETERS, AND IT HAPPENED 3 TIMES WITH THE 22G DFFX. CUSTOMER WAS TRYING TO SAFETY THE NEEDLE AFTER PLACING THE IV. THE CUSTOMER HEARD THE CLICK OF THE SAFETY FEATURES ACTIVATING AND THE NEEDLE LOCKING. WHEN TRYING TO REMOVE THE NEEDLE FROM THE CATHETER BY PULLING ON THE WHITE HE COULD NOT REMOVE THE NEEDLE DESPITE THE NEEDLE BEING LOCKED AND HE WAS NOT HOLDING THE GREY IN PLACE WITH THE CATHETER. THE CUSTOMER IS VERY FAMILIAR WITH THESE CATHETERS, AND IT HAPPENED 3 TIMES WITH THE 22G DFFX. 383596, 5189322.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24146447 ↗
  • MalfunctionJan 16, 2026Leak/Splash

    NO ADDITIONAL INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24094607 ↗
  • MalfunctionJan 16, 2026Leak/Splash

    NO ADDITIONAL INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24093808 ↗
  • MalfunctionJan 16, 2026Leak/Splash

    NO ADDITIONAL INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24093805 ↗
  • MalfunctionJan 16, 2026Leak/Splash

    NO ADDITIONAL INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24093803 ↗
  • MalfunctionJan 16, 2026Leak/Splash

    NO ADDITIONAL INFORMATION

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24093802 ↗
  • MalfunctionJan 16, 2026Leak/Splash

    NO ADDITIONAL INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24093801 ↗
  • MalfunctionJan 16, 2026Leak/Splash

    LEAKING HAPPENING WHILE FLUSHING AT THE CONNECTION OF NEXIVA HUB AND MAXZERO. A FEW DROPLETS OF SALINE LEAK OUT. POTENTIAL EXPOSURE TO BLOOD ANY ADVERSE EVENT OR SERIOUS INJURY REPORTED TO PATIENT OR HEALTHCARE PROFESSIONAL? NO ADVERSE EVENT OR SERIOUS INJURY TO PATIENT OR HEALTHCARE WORKER. CAN YOU PLEASE PROVIDE AN EXACT DATE OF EVENT IN FORMAT 29-12-2025?

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24093796 ↗
  • MalfunctionJan 16, 2026Leak/Splash

    NO ADDITIONAL INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24093528 ↗
  • MalfunctionJan 16, 2026Leak/Splash

    NO ADDITIONAL INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24093527 ↗
  • MalfunctionJan 12, 2026Leak/Splash

    IT WAS REPORTED THAT BD NEXIVA DIFFUSICS W/MAXZERO 22G X 1.00 IN LEAKED IT WAS REPORTED BY CUSTOMER THAT THE LEAKING OBSERVED AT CONNECTION OF MAXZERO AND DIFFUSICS. VERBATIM: LEAKING OBSERVED AT CONNECTION OF MAXZERO AND DIFFUSICS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24047010 ↗
  • MalfunctionJan 12, 2026Leak/Splash

    NO ADDITIONAL INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24047008 ↗
  • MalfunctionJan 12, 2026Leak/Splash

    IT WAS REPORTED THAT BD NEXIVA DIFFUSICS W/MAXZERO 22G X 1.00 IN LEAKED. IT WAS REPORTED BY CUSTOMER THAT THE LEAKING OBSERVED AT CONNECTION OF MAXZERO AND DIFFUSICS. LEAKING OBSERVED AT CONNECTION OF MAXZERO AND DIFFUSICS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24047005 ↗
  • MalfunctionJan 12, 2026Leak/Splash

    IT WAS REPORTED THAT BD NEXIVA DIFFUSICS W/MAXZERO 22G X 1.00 IN LEAKED IT WAS REPORTED BY CUSTOMER THAT THE LEAKING OBSERVED AT CONNECTION OF MAXZERO AND DIFFUSICS VERBATIM: LEAKING OBSERVED AT CONNECTION OF MAXZERO AND DIFFUSICS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24047003 ↗
  • MalfunctionJan 12, 2026Leak/Splash

    NO ADDITIONAL INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24047001 ↗
  • MalfunctionJan 9, 2026Leak/Splash

    IT WAS REPORTED THAT BD NEXIVA DIFFUSICS W/MAXZERO 22G X 1.00 IN LEAKED IT WAS REPORTED BY CUSTOMER THAT "LEAKING OBSERVED AT CONNECTION OF MAXZERO AND DIFFUSICS" VERBATIM: LEAKING OBSERVED AT CONNECTION OF MAXZERO AND DIFFUSICS

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24022358 ↗
  • MalfunctionJan 8, 2026Leak/Splash

    IT WAS REPORTED THAT BD NEXIVA DIFFUSICS W/MAXZERO 22G X 1.00 IN LEAKED IT WAS REPORTED BY CUSTOMER THAT "LEAKING OBSERVED AT CONNECTION OF MAXZERO AND DIFFUSICS." LEAKING OBSERVED AT CONNECTION OF MAXZERO AND DIFFUSICS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24005754 ↗
  • MalfunctionJan 8, 2026Leak/Splash

    NO ADDITIONAL INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24005265 ↗
  • MalfunctionJan 8, 2026Leak/Splash

    NO ADDITIONAL INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24004867 ↗
  • MalfunctionJan 2, 2026Leak/Splash

    IT WAS REPORTED THAT BD NEXIVA DIFFUSICS W/MAXZERO 22G X 1.00 IN LEAKED IT WAS REPORTED BY CUSTOMER THAT LEAKING AT CONNECTION POINT BETWEEN MAX ZERO CONNECTOR AND DIFFUSICS HUB. LEAKING AT CONNECTION POINT BETWEEN MAX ZERO CONNECTOR AND DIFFUSICS HUB.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23945105 ↗
  • MalfunctionJan 2, 2026Leak/Splash

    IT WAS REPORTED THAT BD NEXIVA DIFFUSICS W/MAXZERO 22G X 1.00 IN LEAKED IT WAS REPORTED BY CUSTOMER THAT LEAKING AT CONNECTION OF MAXZERO TO HUB OF DIFFUSICS. SMALL LEAKAGE/MOVEMENT OF FLUIDS OUT OF THE THREADS AND ONTO PATIENT OR HCP. MAINLY ON FLUSHING. LEAKING AT CONNECTION OF MAXZERO TO HUB OF DIFFUSICS. SMALL LEAKAGE/MOVEMENT OF FLUIDS OUT OF THE THREADS AND ONTO PATIENT OR HCP. MAINLY ON FLUSHING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23944616 ↗
  • MalfunctionJan 2, 2026Leak/Splash

    IT WAS REPORTED THAT BD NEXIVA DIFFUSICS W/MAXZERO 22G X 1.00 IN LEAKED IT WAS REPORTED BY CUSTOMER THAT LEAKING OBSERVED AT CONNECTION BETWEEN MAXZERO HUB AND DIFFUSICS PORT. LEAKING OBSERVED AT CONNECTION BETWEEN MAXZERO HUB AND DIFFUSICS PORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23944613 ↗
  • MalfunctionJan 2, 2026Leak/Splash

    NO ADDITIONAL INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23944228 ↗

Other Catheter, intravascular, therapeutic, short-term less than 30 days devices

All Catheter, intravascular, therapeutic, short-term less than 30 days devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page