Skip to content

BD Cathena Safety IV Catheter 386810Recalls & reported problems

Becton, Dickinson And Company · CATHENA 20GX1.00IN WINGED BC · Class 2

All models of BD Cathena Safety IV Catheter (33)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
7
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00382903868100), received between Apr 13, 2021 and May 6, 2025 — classified as 1 injury, 6 malfunctions.

The most frequently reported specific device problem is “Leak/Splash” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Leak/Splash2
  • Retraction Problem2
  • Activation Problem1
  • Break1
  • Contamination1
  • Material Perforation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5
  • Needle Stick/Puncture2

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 6, 2025Retraction Problem; Activation Problem; Retraction Problem; Activation Problem

    IT WAS REPORTED THAT BD CATHENA 20GX1.00IN WINGED BC NEEDLE SHIELDING FAILED. DATE OF OCCURRENCE: (B)(6) 2025. REFERENCE: 386810. BATCH NUMBER: 4180596. DESCRIPTION OF DEFECT: DURING THE INSERTION OF A PERIPHERAL VENOUS LINE WITH THE SECURE WINGED CATHETER, A NURSE PRICKED HERSELF DUE TO A SECURING RELEASE FAULT. IN ADDITION, THE PATIENT HAD SEPSIS CAUSED BY HIGHLY RESISTANT BACTERIA. THE NURSE WAS TREATED AS AN EMERGENCY ACCORDING TO THE BLOOD EXPOSURE ACCIDENT (BEA) PROTOCOL. THE DEFECTIVE MEDICAL DEVICE WAS RETAINED. WOULD YOU LIKE THE CASE TO BE RETURNED FOR EXPERT APPRAISAL?

    Patient problems: Needle Stick/Puncture

    MAUDE report 21969394 ↗
  • MalfunctionMar 27, 2023Leak/Splash; Leak/Splash

    IT WAS REPORTED THAT THE BD CATHENA¿ SAFETY IV CATHETER WITH BD MULTIGUARD¿ TECHNOLOGY LEAKED AT THE BASE DURING THE INSERTION. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM FRENCH: "LEAK AT THE BASE OF THE CATHETER DURING INSERTION, AFTER SUCCESSFUL PUNCTURE OF THE VEIN AND CONNECTION TO THE VEIN GUARD."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16620070 ↗
  • InjuryMar 7, 2023Retraction Problem; Retraction Problem

    IT WAS REPORTED THAT THE BD CATHENA¿ IV CATHETER SAFETY MECHANISM WILL NOT ACTIVATE AND CAUSE A NEEDLE STICK INJURY. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: DURING THE INSERTION OF THE 20G CATHETER, THE SAFETY DEVICE DID NOT WORK. THE NURSE THEREFORE PRICKED HERSELF.

    Patient problems: Needle Stick/Puncture

    MAUDE report 16496448 ↗
  • MalfunctionDec 23, 2022Contamination; Contamination

    IT WAS REPORTED THAT THE BD CATHENA¿ SAFETY IV CATHETER WITH BD MULTIGUARD¿ TECHNOLOGY EXPERIENCED FOREIGN MATTER ON THE DEVICE CANNULA. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: RN ABOUT TO INSERT IV IN PT'S ARM AND NOTICED A SMALL PIECE OF PLASTIC AT THE TIP OF THE CANNULA. IV WAS NOT USED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16043366 ↗
  • MalfunctionDec 20, 2021Material Perforation; Material Perforation

    IT WAS REPORTED THAT AFTER REMOVING THE BD CATHENA¿ SAFETY IV CATHETER WITH BD MULTIGUARD¿ TECHNOLOGY FROM THE PACKAGING AND MOVING IT AROUND, THE NURSE FOUND THAT THE NEEDLE HAD PIERCED THROUGH THE TIP OF THE CATHETER. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "RN WAS ABOUT TO START AN IV ON A PATIENT. WHEN THE RN OPENED THE PACKAGING, THE RN SLID THE CATHETER BACK AND FORTH, AND NOTICED A PIECE OF PLASTIC AT THE TIP OF THE CANNULA."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13048338 ↗
  • MalfunctionDec 8, 2021Break; Break

    IT WAS REPORTED THAT THE BD CATHENA¿ SAFETY IV BLOOD COLLECTION WAS DEFECTIVE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: IT WAS REPORTED DEFECTIVE CATHETER .

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12964970 ↗
  • MalfunctionApr 13, 2021Leak/Splash; Leak/Splash

    IT WAS REPORTED THAT THE BLOOD CONTROL FEATURE ON THE BD CATHENA¿ SAFETY IV CATHETER WITH WINGS BD MULTIGUARD¿ TECHNOLOGY FAILED TO WORK DURING USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM FRENCH TO ENGLISH: "THE MECHANISM THAT ENSURES THE NON-RETURN OF BLOOD DID NOT WORK DURING THE INSTALLATION OF A KT".

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11657181 ↗

Other Catheter, intravascular, therapeutic, short-term less than 30 days devices

All Catheter, intravascular, therapeutic, short-term less than 30 days devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page