BD Cathena™ Safety IV Catheter 386860Recalls & reported problems
Becton, Dickinson And Company · US CATHENA 24GX0.75IN STRAIGHT BC · Class 2
All models of BD Cathena Safety IV Catheter (33)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 14
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 14 adverse event reports that list this exact device (DI 00382903868605), received between Jul 7, 2022 and Feb 5, 2026 — classified as 14 malfunctions.
The most frequently reported specific device problem is “Crack” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Crack4
- Material Protrusion/Extrusion3
- Leak/Splash2
- Retraction Problem2
- Activation Problem1
- Activation, Positioning or Separation Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions12
- Exposure to Body Fluids1
- Insufficient Information1
Recent adverse event reports
The 14 most recent of 14 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 5, 2026Leak/Splash
IT WAS REPORTED BY CUSTOMER THAT 24G PIVS LEAKING DURING APPOINTMENTS IN OUTPATIENT INFUSION. VERBATIM: RCC RECEIVED A COMPLAINT VIA EMAIL. MULTIPLE 24G PIVS [PERIPHERAL INTRAVENOUS LINES] LEAKING DURING APPOINTMENTS IN OUTPATIENT INFUSION AT SITE WHERE PIV CONNECTS TO CATHETER EXTENSION SET. PIV DOES NOT LEAK WHEN CHECKING BLOOD RETURN BUT WHEN FLUSHING, PIV LEAKS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24269187 ↗MalfunctionJan 21, 2026
MULTIPLE 24G PIVS [PERIPHERAL INTRAVENOUS LINES] LEAKING DURING APPOINTMENTS IN OUTPATIENT INFUSION AT SITE WHERE PIV CONNECTS TO CATHETER EXTENSION SET. PIV DOES NOT LEAK WHEN CHECKING BLOOD RETURN BUT WHEN FLUSHING, PIV LEAKS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24130209 ↗MalfunctionNov 25, 2025Activation, Positioning or Separation Problem; Activation, Positioning or Separation Problem
IT WAS REPORTED THAT BD US CATHENA 24GX0.75IN STRAIGHT BC SAFETY SHIELD ACTIVATION FAILED IT WAS REPORTED BY CUSTOMER THAT CATHENA DID NOT SAFETY APPROPRIATELY, EXPOSING BEVEL TO PROVIDER AFTER INSERTING INTO PATIENT. VERBATIM: CATHENA DID NOT SAFETY APPROPRIATELY, EXPOSING BEVEL TO PROVIDER AFTER INSERTING INTO PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23646719 ↗MalfunctionAug 22, 2025Crack; Crack
IT WAS REPORTED THAT BD CATHENA 24GX0.75IN STRAIGHT BC - CRACK. TODAY, THREE PATIENTS REPORTED LEAKAGE OF MEDICATION FROM THEIR IV SITES. UPON ASSESSMENT, THE IV SITES WERE FOUND TO BE DAMP. THE IV CATHETERS WERE REMOVED, AND INSPECTION REVEALED THAT THE CONNECTOR COMPONENTS OF ALL THREE CATHETERS WERE SLIGHTLY CRACKED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22874188 ↗MalfunctionAug 22, 2025Crack; Crack
IT WAS REPORTED THAT BD CATHENA 24GX0.75IN STRAIGHT BC - CRACK. TODAY, THREE PATIENTS REPORTED LEAKAGE OF MEDICATION FROM THEIR IV SITES. UPON ASSESSMENT, THE IV SITES WERE FOUND TO BE DAMP. THE IV CATHETERS WERE REMOVED, AND INSPECTION REVEALED THAT THE CONNECTOR COMPONENTS OF ALL THREE CATHETERS WERE SLIGHTLY CRACKED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22874176 ↗MalfunctionAug 22, 2025Crack; Crack
IT WAS REPORTED THAT BD CATHENA 24GX0.75IN STRAIGHT BC - CRACK. TODAY, THREE PATIENTS REPORTED LEAKAGE OF MEDICATION FROM THEIR IV SITES. UPON ASSESSMENT, THE IV SITES WERE FOUND TO BE DAMP. THE IV CATHETERS WERE REMOVED, AND INSPECTION REVEALED THAT THE CONNECTOR COMPONENTS OF ALL THREE CATHETERS WERE SLIGHTLY CRACKED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22874164 ↗MalfunctionJul 3, 2025Activation Problem; Activation Problem
IT WAS REPORTED THAT BD US CATHENA 24GX0.75IN STRAIGHT BC NEEDLE RETRACTION FAILURE IT WAS REPORTED BY CUSTOMER THAT THE NEEDLE NOT RETRACTING FROM HUB. VERBATIM: RCC RECEIVED A COMPLAINT VIA EMAIL. EMAIL(S) ATTACHED. CAN YOU SEND ME THE PR FOR: AWARENESS DATE: 6/11/2025. NEEDLE NOT RETRACTING FROM HUB.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22395473 ↗MalfunctionJun 30, 2025Retraction Problem; Retraction Problem
IT WAS REPORTED THAT BD US CATHENA 24GX0.75IN STRAIGHT BC NEEDLE RETRACTION FAILED IT WAS REPORTED BY CUSTOMER THAT NEEDLE NOT RETRACTING FROM HUB RCC RECEIVED A COMPLAINT VIA EMAIL. I HAVE CC¿D XXX AND XXX FROM XXXXXX TO REPORT THE COMPLAINTS BELOW: AWARENESS DATE: 6/16/2025, WE HAD ANOTHER XXXXX INCIDENT WHERE THE NEEDLE CAUGHT INSIDE THE CATHETER AND THE RN WAS NOT ABLE TO REMOVE IT, THEREFORE CAUSING THE PATIENT TO ENDURE ANOTHER PERIPHERAL STICK. #24 GAUGE BD CATHENA LOT#: 2309256, REF#: 386859 AWARENESS DATE: 6/11/2025, NEEDLE NOT RETRACTING FROM HUB.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22363266 ↗MalfunctionFeb 6, 2025Material Protrusion/Extrusion; Material Protrusion/Extrusion
IT WAS REPORTED UNKNOWN BD CATHETER SAFETY MECHANISM WILL NOT DISENGAGE FROM CATHETER / STUCK AT HUB. IT WAS REPORTED BY CUSTOMER THAT IV WAS STARTED ON PATIENT AND WHEN THE NEEDLE WAS RETRACTED OUT OF THE PATIENT, THE WHITE DEVICE THAT BLOCKS THE NEEDLE FROM GOING BACK IN CLAMPS THE BASE OF THE YELLOW 24G CATHETER AND DOES NOT ALLOW IT TO DETACH FROM THE NEEDLE DEVICE. IV SET WAS REMOVED FROM PATIENT AND ANOTHER IV WAS STARTED WITH SUCCESSFUL DETACHMENT. VERBATIM#: IV WAS STARTED ON PATIENT AND WHEN THE NEEDLE WAS RETRACTED OUT OF THE PATIENT, THE WHITE DEVICE THAT BLOCKS THE NEEDLE FROM GOING BACK IN CLAMPS THE BASE OF THE YELLOW 24G CATHETER AND DOES NOT ALLOW IT TO DETACH FROM THE NEEDLE DEVICE. IV SET WAS REMOVED FROM PATIENT AND ANOTHER IV WAS STARTED WITH SUCCESSFUL DETACHMENT. OF NOTE, WHEN MANAGER INQUIRED IF THIS WAS THE FIRST TIME THAT THIS HAD OCCURRED, THE STAFF ADVISED THAT IT HAS HAPPENED ON AND OFF IN THE PAST RANDOMLY. THE INCIDENT WAS NOT REPORTED AT THE TIME BY THE STAFF. I FOUND OUT IT HAD OCCURRED PREVIOUSLY AT THE TIME OF THE INCIDENT. I INQUIRED ABOUT THE BOXES WHEN IT OCCURRED AND I WAS TOLD MONTHS AGO AND WE NO LONGER HAD LOT INFO AVAILABLE SO I AM UNSURE O
Patient problems: Insufficient Information
MAUDE report 21325627 ↗MalfunctionFeb 6, 2025Material Protrusion/Extrusion; Material Protrusion/Extrusion
IT WAS REPORTED THAT BD US CATHENA 24GX0.75IN STRAIGHT BC SAFETY MECHANISM WILL NOT DISENGAGE FROM CATHETER / STUCK AT HUB. IT WAS REPORTED BY CUSTOMER THAT IV WAS STARTED ON PATIENT AND WHEN THE NEEDLE WAS RETRACTED OUT OF THE PATIENT, THE WHITE DEVICE THAT BLOCKS THE NEEDLE FROM GOING BACK IN CLAMPS THE BASE OF THE YELLOW 24G CATHETER AND DOES NOT ALLOW IT TO DETACH FROM THE NEEDLE DEVICE. IV SET WAS REMOVED FROM PATIENT AND ANOTHER IV WAS STARTED WITH SUCCESSFUL DETACHMENT. IV WAS STARTED ON PATIENT AND WHEN THE NEEDLE WAS RETRACTED OUT OF THE PATIENT, THE WHITE DEVICE THAT BLOCKS THE NEEDLE FROM GOING BACK IN CLAMPS THE BASE OF THE YELLOW 24G CATHETER AND DOES NOT ALLOW IT TO DETACH FROM THE NEEDLE DEVICE. IV SET WAS REMOVED FROM PATIENT AND ANOTHER IV WAS STARTED WITH SUCCESSFUL DETACHMENT. OF NOTE, WHEN MANAGER INQUIRED IF THIS WAS THE FIRST TIME THAT THIS HAD OCCURRED, THE STAFF ADVISED THAT IT HAS HAPPENED ON AND OFF IN THE PAST RANDOMLY. ITEM: BD CATHENA SAFETY IV CATHETER, LOT: 4204903, REF: 386860.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21325622 ↗MalfunctionJun 17, 2024Material Protrusion/Extrusion; Material Protrusion/Extrusion
IT WAS REPORTED THAT BD US CATHENA 24GX0.75IN STRAIGHT BC SAFETY MECHANISM WILL NOT ACTIVATE. IT WAS REPORTED BY CUSTOMER THAT THE SAFETY DID NOT CLICK, STATING THE ¿SAFETY COVER WAS NOT THERE¿, AND THAT THE IV CATHETER IS SUPPOSED TO CLICK AUTOMATICALLY TO ENGAGE THE SAFETY. THE SAFETY MECHANISM IS A PASSIVE SYSTEM, MEANING THAT WHEN THE NURSE PULLS BACK ON THE FROSTED PART AFTER INSERTING THE IV INTO A PATIENT, THE NEEDLE RETRACTS BACK INTO THAT FROSTED PART. WHEN FULLY OUT, THE WHOLE PART DISENGAGES FROM THE SPRING IN THE HUB, CREATING A LOCKED SAFETY ¿COVERING¿ FOR THE SHARP END OF THE NEEDLE. VERBATIM: RCC RECEIVED A COMPLAINT VIA EMAIL. THAT UPON REMOVING THE IV CATHETER, THE SAFETY DID NOT CLICK, STATING THE ¿SAFETY COVER WAS NOT THERE¿, AND THAT THE IV CATHETER IS SUPPOSED TO CLICK AUTOMATICALLY TO ENGAGE THE SAFETY. THE SAFETY MECHANISM IS A PASSIVE SYSTEM, MEANING THAT WHEN THE NURSE PULLS BACK ON THE FROSTED PART AFTER INSERTING THE IV INTO A PATIENT, THE NEEDLE RETRACTS BACK INTO THAT FROSTED PART. WHEN FULLY OUT, THE WHOLE PART DISENGAGES FROM THE SPRING IN THE HUB, CREATING A LOCKED SAFETY ¿COVERING¿ FOR THE SHARP END OF THE NEEDLE. MANUFACTURER RESPONSE FOR PERIPHERA
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19548240 ↗MalfunctionJun 6, 2024Leak/Splash; Leak/Splash
MATERIAL #: 386860; BATCH#: 2274376; IT WAS REPORTED BY CUSTOMER THAT VALVE VOT OCCLUDING AS EXPECTED UPON INSERTION OF 24G CATHENA PIV COMPLAINT RECEIVED.
Patient problems: Exposure to Body Fluids
MAUDE report 19476826 ↗MalfunctionAug 29, 2023Crack; Crack
IT WAS REPORTED THAT THE BD CATHENA¿ SAFETY IV CATHETER EXPERIENCED CATHETER CRACKED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: CUSTOMER USED A 24G NEEDLE IN HER L AC. WHEN RETRACTING THE NEEDLE, CUSTOMER NOTICED THE CATHETER BENT. USER PULLED BACK AND GOT BLOOD RETURN HOWEVER, THEY THEN NOTICED THERE WAS BLOOD COMING FROM THE IV SITE. USER ONCE AGAIN, PULLED THE CATHETER BACK AND THATS WHEN USER NOTICED IT WAS THE ACTUAL CATHETER THAT WAS CRACKED. WHEN FLUSHING THE LINE THE SALINE WAS GOING RIGHT THROUGH THE TUBING. IV WAS REMOVED AND A 22G NEEDLE WAS INSERTED INSTEAD. PATIENT DID NOTICE WHAT WAS HAPPENING, HOWEVER SHE STATES SHE FEELS FINE AND OFFERS NO COMPLAINTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17646893 ↗MalfunctionJul 7, 2022Retraction Problem; Retraction Problem
IT WAS REPORTED THAT THE BD CATHENA¿ SAFETY IV CATHETER EXPERIENCED A SAFETY MECHANISM THAT WOULD NOT ACTIVATE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "CATHENA WAS NOT ABLE TO SAFETY."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14951487 ↗
Other Catheter, intravascular, therapeutic, short-term less than 30 days devices
- BD Insyte Autoguard 382523Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days660 reports
- BD Insyte Autoguard 382533Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days2 recalls527 reports
- BD PosiFlush 306546Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days1 death608 reports
- BD Insyte Autoguard Shielded IV Catheter 381823Becton, Dickinson And Company · Catheter, intravascular, therapeutic, short-term less than 30 days582 reports
- BD Insyte Autoguard 381423Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days1 recall528 reports
- BD Insyte Autoguard 382544Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days1 recall437 reports
- BD Nexiva, BD Q-Syte 383536Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days2 recalls374 reports
- BD PosiFlush 306547Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days2 recalls1 death331 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page