Skip to content

BD Cathena™ Safety IV Catheter 386881Recalls & reported problems

Becton, Dickinson And Company · US CATHENA 22GX1.00IN WINGED BC · Class 2

All models of BD Cathena Safety IV Catheter (33)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00382903868810), received between Apr 16, 2021 and Sep 30, 2026 — classified as 5 malfunctions.

The most frequently reported specific device problem is “Activation Problem” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 2 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Activation Problem3
  • Activation, Positioning or Separation Problem1
  • Material Protrusion/Extrusion1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4
  • Needle Stick/Puncture1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 30, 2026Activation Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50948300 ↗
  • MalfunctionMay 12, 2026Activation Problem

    IT WAS REPORTED BY CUSTOMER THAT ""CATHENA SAFETY MECHANISM DID NOT ENGAGE. LEAVING THE NEEDLE EXPOSED"" VERBATIM: CATHENA SAFETY MECHANISM DID NOT ENGAGE. LEAVING THE NEEDLE EXPOSED".

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25150370 ↗
  • MalfunctionApr 30, 2026Activation Problem

    IT WAS REPORTED BY THE CUSTOMER THAT WE HAD ANOTHER CATHENA NOT SAFETY, VERBATIM: WE HAD ANOTHER CATHENA NOT SAFETY, THIS ONE WAS A WINGED VERSION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25045756 ↗
  • MalfunctionApr 28, 2026Activation, Positioning or Separation Problem

    IT WAS REPORTED BY THE CUSTOMER THAT THE DEVICE APPEARED BENT BEFORE ENTERING PATIENT. WHEN THE CATHETER WAS PULLED FROM THE NEEDLE, THE NEEDLE PROTECTOR DID NOT EXTEND, SO FULL NEEDLE WAS EXPOSED. SAFETY DEVICE WAS PULLED DOWN SEPARATELY, NEEDLE COVERED APPROPRIATELY AND DEVICE DISCARDED. CATHETER PULLED FROM PACKAGE TO USE ON PATIENT. WRITER NOTICED THE DEVICE APPEARED BENT BEFORE ENTERING PATIENT. WHEN THE CATHETER WAS PULLED FROM THE NEEDLE, THE NEEDLE PROTECTOR DID NOT EXTEND, SO FULL NEEDLE WAS EXPOSED. SAFETY DEVICE WAS PULLED DOWN SEPARATELY, NEEDLE COVERED APPROPRIATELY AND DEVICE DISCARDED."

    Patient problems: Needle Stick/Puncture

    MAUDE report 25017375 ↗
  • MalfunctionApr 16, 2021Material Protrusion/Extrusion; Material Protrusion/Extrusion

    IT WAS REPORTED THAT US CATHENA 22GX1.00IN WINGED BC HAD A SAFETY MECHANISM FAILURE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: MATERIAL NO.: 386881 BATCH NO.: UNKNOWN IT WAS REPORTED THAT THE 22G 1" CATHENA CATHETER SAFETY DID NOT ENGAGE WHEN OFF OF CATHETER. VERBATIM: CUSTOMER UNSURE IF SHE FORCED THE WHITE PART OFF OF CATHETER HUB PRIOR TO SAFETY OR IF IT FREELY CAME OFF. SHE WAS ABLE TO SAFETY AFTER DETACHED FROM CATHETER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11682137 ↗

Other Catheter, intravascular, therapeutic, short-term less than 30 days devices

All Catheter, intravascular, therapeutic, short-term less than 30 days devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page