Skip to content

BD QUINCKE Spinal Needles 405180Recalls & reported problems

Becton, Dickinson And Company · NEEDLE SPINAL S/SU 25GA 3-1/2IN QUINCKE · Class 2

All models of BD QUINCKE Spinal Needles (9)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This BD Quincke spinal needle is a thin needle used to deliver anesthesia or other medication near the spine. The 23 reports that name its identifier mostly describe needles that were blocked or hard to inject through, and hubs that leaked during use, with no deaths and 2 injuries mentioned. No recalls name this device, with no reports in the past year after 9 the year before.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
23
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 23 adverse event reports that list this exact device (DI 00382904051808), received between Mar 27, 2019 and Aug 11, 2025 — classified as 2 injuries, 21 malfunctions.

The most frequently reported specific device problem is “Complete Blockage” (7 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 2 reportsJan 2025: 2 reportsFeb 2025: 2 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 2 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Complete Blockage7
  • Leak/Splash6
  • Adverse Event Without Identified Device or Use Problem2
  • Contamination1
  • Device Contamination with Chemical or Other Material1
  • Device Markings/Labelling Problem1
  • Failure to Deliver1
  • Inaccurate Information1

Patient problems

  • No Clinical Signs, Symptoms or Conditions16
  • Insufficient Information3
  • No Known Impact Or Consequence To Patient2
  • Exposure to Body Fluids1
  • Headache1
  • Needle Stick/Puncture1
  • No Patient Involvement1

Recent adverse event reports

The 23 most recent of 23 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 11, 2025Leak/Splash; Leak/Splash

    NO ADDITIONAL INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22776560 ↗
  • MalfunctionAug 5, 2025Leak/Splash; Leak/Splash

    IT WAS REPORTED THAT BD UNKNOWN PRODUCT LEAKED. IT WAS REPORTED BY CUSTOMER THAT THEY HAVE TWO MORE INCIDENTS ON A LIST OF LEAKING. RCC RECEIVED A COMPLAINT VIA EMAIL. I DO HAVE 2 MORE INCIDENTS ON A LIST OF LEAKING DATES/LOT #. HOWEVER, I DIDN'T KNOW OF THE TIME LIMIT TO KEEP THE COMPLAINT OPEN. I WILL CONTACT YOU AGAIN IF THE PROVIDER AGREES TO REOPENING THE CASE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22717097 ↗
  • MalfunctionJun 9, 2025Leak/Splash; Leak/Splash

    MATERIAL #405180 LOT #UNKNOWN THE CLINIC BELIEVES THERE IS A DIFFERENCE WITH THE HUBS BECAUSE THE TAN ONE LEAKS WHEN IN USE. VERBATIM: I HAVE A CLINIC WHO IS TRYING TO ORDER #405180 THROUGH XXXX. THE PICTURE ON XXXX SHOWS A BLUE HUB, BUT WHEN THEY RECEIVE THE PRODUCT, IT IS A TAN HUB. THE CLINIC BELIEVES THERE IS A DIFFERENCE WITH THE HUBS BECAUSE THE TAN ONE LEAKS WHEN IN USE. WE WILL NOT BE ABLE TO GIVE YOU THE USED/DIRTY NEEDLES OR PROVIDE THE PACKAGING INFORMATION AS THIS HAS BEEN A PROBLEM SINCE LATE 2024. HOWEVER, WE CAN MAKE A LIST STARTING TODAY AND GOING FORWARD. WE DO MANY PROCEDURES WEEKLY THAT USE THESE NEEDLES. CAN YOU PLEASE PROVIDE AN EXACT DATE OF EVENT IN THIS FORMAT MM-DD-YYYY? CAN YOU PLEASE PROVIDE THE LOT NUMBER ASSOCIATED WITH REPORTED ISSUE? ANSWERS: I DO NOT HAVE SPECIFIC DATES, AS THERE ARE MULTIPLE OCCASIONS BEFORE DR WANTED THE LEAKING REPORTED. I HAVE A CHART ON THE COMPUTER THAT I WILL BE IMMEDIATELY ABLE TO WRITE DOWN DATES/LOT # ETC., GOING FORWARD. THIS HAS BEEN ONGOING SINCE NOVEMBER 2024. 4. DESCRIBE ANY PATIENT HARM, INJURY, COMPLICATION OR NEGATIVE OUTCOME THAT OCCURRED AS A RESULT OF THE EVENT. ANSWER: THE DR STATED THE NEEDLES LEAK AFTER THE ST

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22168807 ↗
  • MalfunctionFeb 25, 2025Complete Blockage; Complete Blockage

    ADDITIONAL INFORMATION ON EVENT FROM CUSTOMER: I HAD A CHANCE TO SPEAK WITH THE RADIOLOGIST WHO HAS EXPERIENCED THE ISSUE. MY UNDERSTANDING IS THEY ARE INJECTING PAIN MEDS INTO THE FACET JOINTS ALONG THE SPINE. THIS PROCEDURE IS EXECUTED UNDER FLUOROSCOPY. PROCEDURE DONE IN RADIOLOGY VERDUN AND NOTRE DAME IN MONTREAL HE MENTIONED THE STYLET IS IN WHEN ADVANCING THE NEEDLE OFTEN STRIKE BONE TO VALIDATE POSITIONING ¿ MOVE THE NEEDLE BACK A TOUCH, PULL THE STYLET AND INJECT THEY NEVER REUSE OR ADVANCE THE NEEDLE ONCE IT IS IN PLACE PROBLEM WITH BLOCKAGE STARTED ONCE THE HUB COLOR CHANGED. SOUNDS LIKE A COINCIDENCE. THEY HAVE A LONG HISTORY USING THIS NEEDLE FOR THIS APPLICATION. I ASKED ONCE HE PULLS THE NEEDLE OUT HAS, HE CHECKED FOR PATENCY ¿ HE SAID THEY HAVE AND THEY ARE BLOCKED. PROBLEM STILL ONGOING AS OF YESTERDAY. I RECOMMENDED OUR 23G QUINCKE TO SEE IF THAT HELPS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21462449 ↗
  • MalfunctionFeb 21, 2025Complete Blockage; Complete Blockage

    NO ADDITIONAL INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21434586 ↗
  • MalfunctionJan 27, 2025Complete Blockage; Complete Blockage

    NO ADDITIONAL INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21242178 ↗
  • MalfunctionJan 5, 2025Complete Blockage; Complete Blockage

    MATERIAL #: 405180, BATCH#: UNKNOWN. VERBATIM: RCC RECEIVED A COMPLAINT VIA EMAIL. EMAIL(S) ATTACHED. REQUESTING IF THERE IS CHANGE IN THE QUINCKE SPINAL NEEDLE (REF 405180). STATES THE NEEDLE HUB WAS BLUE IN THE PAST AND NOW IT IS ORANGE. INQUIRING WHEN THAT CHANGE WAS MADE. ALSO, STATES THE DOCTORS ARE HAVING ISSUES USING THE NEEDLES. THEY HAVE TRIED SEVERAL DIFFERENT LOTS AND THEY STATE THERE IS A FLUID RESISTANCE WHEN THEY INJECT MEDICATIONS IN THE NEEDLE.

    Patient problems: Insufficient Information

    MAUDE report 21076441 ↗
  • MalfunctionDec 26, 2024Complete Blockage; Complete Blockage

    MATERIAL#: 405180, BATCH NUMBER#: 4211030. VERBATIM#: A DOCTOR WAS USING BD NEEDLES, REF# 405180, AND FOUND THE NEEDLES TO BE CLOGGED. SHE TRIED SEVERAL AND THEY DID NOT WORK AND CHECKED THE SYRINGES WHICH WERE OK. THE FAULTY ONES WERE IN LOT # 4211030. WE HAVE 21 UNOPENED NEEDLES FROM ONE OPENED BOX. WE HAVE 6 MORE BOXES OF 25 UNOPENED OF THAT SAME LOT NUMBER. NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21019274 ↗
  • MalfunctionDec 11, 2024Leak/Splash; Leak/Splash

    MATERIAL#: 405180, BATCH#: UNKNOWN. SINCE THE ANESTHESIA NEEDLES HAVE CHANGED TO ORANGE, THERE IS NOW LEAKING AT THE HUB WHEN INJECTING LIPIODOL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20901864 ↗
  • MalfunctionAug 6, 2024Inaccurate Information; Inaccurate Information

    MATERIAL # 405180 LOT # 4134383, 4015086 VERBATIM: RCC RECEIVED A COMPLAINT VIA PHONE. PIR ATTACHED. CALLER STATES THEY HAVE PRODUCT CODE 405180 WITH LOT#4134383 AND 4015086. CALLER STATES THAT THEY ARE LABELED DIFFERENTLY BY 1.1MM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19923518 ↗
  • MalfunctionJul 7, 2024Device Markings/Labelling Problem; Device Markings/Labelling Problem

    MATERIAL #: 405180 BATCH#: UNKNOWN. IT WAS REPORTED BY CUSTOMER THAT PLEASE ADVISE IF THE ITEMS SHOWN IN THE PICTURE IS SAME. ONE IS BLUE AND THE OTHER IS RED AND THE ITEM NO. IS SAME FOR BOTH. QUINCKE SPINAL NEEDLE 405180. PLEASE ADVISE IF THE ITEMS SHOWN IN THE PICTURE IS SAME. ONE IS BLUE AND THE OTHER IS RED AND THE ITEM NO. IS SAME FOR BOTH. QUINCKE SPINAL NEEDLE 405180.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 19686667 ↗
  • MalfunctionAug 9, 2023Contamination; Contamination

    IT WAS REPORTED WHILE USING BD® SPINAL NEEDLE 25 G X 3 1/2 IN. WITH QUINCKE BEVEL FOREIGN MATTER WAS FOUND. THERE WAS NO REPORT OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: QUANTITY AFFECTED: 1 BX. SERIAL/LOT NUMBER: (B)(6). ARE ANY SAMPLES AVAILABLE FOR RETURN? YES. REPORTED ISSUE: PERXXX: "WE WERE INFORMED THAT 3 SPINAL NEEDLES FROM LOT# 3062941 HAD A METALLIC SUBSTANCE ON THE INSIDE OF THE STERILE PACKAGING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17494140 ↗
  • MalfunctionMay 8, 2023Packaging Problem; Packaging Problem

    IT WAS REPORTED THAT THE BD¿ QUINCKE SPINAL NEEDLE EXPERIENCED MIXED PRODUCT TYPE IN PACKAGING. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: THE CUSTOMER REPORTED THAT THEY RECEIVED A BOX OF BD SPINAL NEEDLES #405180 LOT # 2334994 WITHOUT A DIFFERENT SIZE TIP THAN THE 25G THEY ORDERED AND LISTED ON THE PACKAGING. CUSTOMER OPENED SPINAL NEEDLE 405180 AND THE NEEDLE HAD A DIFFERENT TIP OTHER THAN THE 25GX3.5 AS THE PACKAGING INDICATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16892307 ↗
  • InjuryApr 13, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT 30 BD¿ QUINCKE SPINAL NEEDLES CAUSED SPINAL HEADACHES TO OCCUR FOLLOWING USE. THERE WAS NO LONG-TERM ADVERSE IMPACT REPORTED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "5G QUINCKIE SPINAL NEEDLES CAUSE A HIGHER INCIDENCE OF SPINAL HEADACHE FOLLOWING USE AND OUR ANESTHESIOLOGISTS HAVE STATED THAT THIS NEEDLE IS NOT APPROPRIATE FOR USE IN THIS SETTING, AND THEY WILL NOT USE THEM ON OUR PREGNANT POPULATION. PATIENT INJURY? (Y/N) NO LONG-TERM ADVERSE IMPACT, BUT SPINAL NEEDLES CAUSE A HIGHER INCIDENCE OF SPINAL HEADACHE.

    Patient problems: Headache

    MAUDE report 16740492 ↗
  • MalfunctionFeb 15, 2023Complete Blockage; Complete Blockage

    IT WAS REPORTED THAT 2 OF THE BD¿ SPINAL NEEDLE WITH QUINCKE BEVEL TIPS WERE BLOCKED AND UNABLE TO INJECT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: NEEDLES TIPS WERE BLOCKED AND WAS UNABLE TO INJECT

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16382982 ↗
  • MalfunctionFeb 14, 2023Complete Blockage; Complete Blockage

    IT WAS REPORTED THAT 1BD¿ QUINCKE SPINAL NEEDLE EACH FROM LOTS 2073452 AND 2159499 WERE CLOGGED DURING USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "THE TIPS ARE NOT POTENT AND CANNOT INJECT/USE THEM."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16375785 ↗
  • MalfunctionFeb 14, 2023Failure to Deliver; Failure to Deliver

    IT WAS REPORTED WHILE USING BD¿ SPINAL NEEDLE WITH QUINCKE BEVEL, STERILE, SINGLE USE, THE NEEDLE WAS CLOGGED. THERE WAS NO REPORT OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: NEEDLES TIPS WERE BLOCKED AND WAS UNABLE TO INJECT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16375670 ↗
  • MalfunctionAug 17, 2022Microbial Contamination of Device; Microbial Contamination of Device

    IT WAS REPORTED THAT THE BD¿ SPINAL NEEDLE WITH QUINCKE BEVEL EXPERIENCED FOREIGN MATTER IN CANNULA. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: NEEDLE WITH FOREIGN MATTER (HAIR).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15245717 ↗
  • MalfunctionAug 8, 2022Mechanical Problem

    LEFT HIP INJECTION BEING PERFORMED USING 22GX3.5IN BD WHITACRE PENCIL POINT SPINAL NEEDLE. EXTREME DIFFICULTY EXPERIENCED WHEN TRYING TO PENETRATE NEEDLE TIP THROUGH SKIN. ONCE THROUGH, NEEDLE WAS VERY FLIMSY AND HARD TO MANEUVER.

    Patient problems: Insufficient Information

    MAUDE report 15182518 ↗
  • MalfunctionOct 22, 2020Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material

    IT WAS REPORTED THAT 63 NEEDLES SPINAL S/SU 25GA 3-1/2IN QUINCKE WERE FOUND DISCOLORED BEFORE USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "REPORTED ISSUE: DISCOLORATION".

    Patient problems: No Patient Involvement

    MAUDE report 10723559 ↗
  • InjuryFeb 6, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT WHILE USING THE BD NEEDLE SPINAL S/SU 25GA 3-1/2IN QUINCKE THE HEALTHCARE PROVIDER RECEIVED A CONTAMINATED NEEDLE STICK. VERBATIM: IT WAS REPORTED THAT A NEEDLE STICK INJURY OCCURRED ON THE CLINICIAN AFTER HAVING USED THE NEEDLE ON THE PATIENT. IN THIS OCCURRENCE THE CLINICIAN RECAPPED THE NEEDLE AFTER USING IT, THEN WENT TO USE IT AGAIN AND THE NEEDLE RECOILED AND STUCK THE TECH. IN THIS NEEDLE STICK OCCURRENCE, CLINICIAN USED THE NEEDLE ON THE PATIENT, RECAPPED IT AND WERE GOING TO USE IT AGAIN WHEN THE NEEDLE 'RECOILED' AND STUCK THE TECH.

    Patient problems: Exposure to Body Fluids; Needle Stick/Puncture

    MAUDE report 9676627 ↗
  • MalfunctionOct 25, 2019Leak/Splash; Leak/Splash

    IT HAS BEEN REPORTED THAT THE NEEDLE SPINAL S/SU 25GA 3-1/2IN QUINCKE HAS BEEN FOUND EXPERIENCING LEAKAGE DURING USE. THE FOLLOWING HAS BEEN PROVIDED BY THE INITIAL REPORTER: WHEN INTRODUCING THE DEVICE IN THE PATIENTS IN ORDER TO PERFORM THE THERAPY ON ONCOLOGIC PATIENTS, IT IS OBSERVED THE NEEDLE BENDS, OCCURRING LEAKAGE OF THE CYTOTOXIC MEDICATION.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9242221 ↗
  • MalfunctionMar 27, 2019Leak/Splash

    IT WAS REPORTED THAT A NEEDLE SPINAL S/SU 25GA 3-1/2IN QUINCKE HAD LEAKAGE. THE FOLLOWING WAS REPORTED, "AT THE TIME OF ADMINISTERING INTRATHECAL ONCOLOGICAL MEDICATION, THE SYRINGE DOES NOT TIE WITH THE SPINAL NEEDLE HUB SO THE MEDICATION LEAKS, ALSO IN THE PUNCTURE IS VERY SOFT."

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8457994 ↗

Other NEEDLE, CONDUCTION, ANESTHETIC (W/WO INTRODUCER) devices

All NEEDLE, CONDUCTION, ANESTHETIC (W/WO INTRODUCER) devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page