Skip to content

BD MAX™ ExK™ DNA-3 442821Recalls & reported problems

Becton, Dickinson And Company · BD MAX™ ExK™ DNA-3 · Class 1

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
7
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00382904428211), received between Jun 17, 2021 and May 14, 2025 — classified as 7 malfunctions.

The most frequently reported specific device problem is “Contamination” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Contamination2
  • Contamination of Device Ingredient or Reagent2
  • False Positive Result2
  • Incorrect, Inadequate or Imprecise Result or Readings1

Patient problems

  • No Clinical Signs, Symptoms or Conditions7

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 14, 2025False Positive Result; False Positive Result

    IT WAS REPORTED WHEN USING KIT BD MAX EXK DNA-3 USA DISCREPANT RESULTS WERE OBTAINED FOR C. AURIS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22021913 ↗
  • MalfunctionJul 13, 2022Contamination of Device Ingredient or Reagent; Contamination of Device Ingredient or Reagent

    IT WAS REPORTED WHILE TESTING WITH THE BD MAX¿ EXK¿ DNA-3 INSTRUMENT PROTEUS SP. CONTAMINATION WAS PRESENT WITH LOT#: 1306435. THERE WAS NO REPORT OF CONFIRMATORY TESTING. THERE WAS NO REPORT OF PATIENT IMPACT. THE FOLLOWING HAS BEEN PROVIDED BY THE INITIAL REPORTER: IT WAS REPORTED BY THE CUSTOMER THAT THEY ARE HAVING ISSUES WITH THE LOT#: 1306435. ISSUE IS PROTEUS SP. CONTAMINATION OF THE KITS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15009061 ↗
  • MalfunctionMay 10, 2022Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED THAT WHILE USING BD MAX¿ EXK¿ DNA-3 THE CUSTOMER OBSERVED BIOLOGICAL CONTAMINATION OF SEVERAL KITS. THIS EVENT OCCURRED 2 TIME(S). THERE IS NO INDICATION THAT FALSE RESULTS WERE REPORTED TO HEALTH CARE PRACTITIONERS. THERE IS NO INDICATION OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "CUSTOMER REPORTS WIDESPREAD CONTAMINATION ACROSS SEVERAL KITS OF 442821 - LOT #2025615.", "I CAN CONFIRM THAT THE NEW LOT (2025615) WAS POSITIVE FOR ENTEROCOCCUS FAECALIS IN TWO REPLICATE WATER BLANK SAMPLES."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14350237 ↗
  • MalfunctionNov 16, 2021False Positive Result; False Positive Result

    IT WAS REPORTED THAT WHILE TESTING WITH BD MAX¿ EXK¿ DNA-3 FALSE POSITIVE RESULTS WERE OBTAINED. MULTIPLE ATTEMPTS HAVE BEEN MADE TO OBTAIN ADDITIONAL INFORMATION, THE CUSTOMER HAS NOT RESPONDED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: CONSISTENT POSITIVE RESULTS WITH EACH TEST ON THIS LOT#.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12822087 ↗
  • MalfunctionOct 15, 2021Contamination; Contamination

    IT WAS REPORTED THAT WHILE USING 10 BD MAX¿ EXK¿ DNA-3CLINICAL SAMPLE CONCENTRATOR CONTAMINATION WAS OBSERVED BY THE LABORATORY PERSONNEL. THERE WAS NO INDICATION THAT RESULTS WERE REPORTED OUT AND THERE WAS NO REPORT OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: " ¿ CUSTOMER PROBLEM: RECEIVED 10 KITS FROM LOT NUMBER WITH A KNOWN E. FEACALIS CONTAMINATION ISSUE. "

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12639961 ↗
  • MalfunctionJun 23, 2021Contamination; Contamination

    IT WAS REPORTED THAT WHILE USING 25 BD MAX¿ EXK¿ DNA-3 CONTAMINATION WAS OBSERVED BY THE LABORATORY PERSONNEL. THERE WAS NO INDICATION THAT RESULTS WERE REPORTED OUT AND THERE WAS NO REPORT OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: " CUSTOMER ALLEGES THE KITS BEING A SOURCE OF E. FAECALIS CONTAMINATION OF THEIR RUNS"

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12052005 ↗
  • MalfunctionJun 17, 2021Contamination of Device Ingredient or Reagent; Contamination of Device Ingredient or Reagent

    IT WAS REPORTED THAT WHILE USING BD MAX¿ EXK¿ DNA-3 FALSE POSITIVE RESULTS CONTAMINATION WAS OBSERVED BY THE LABORATORY PERSONNEL. THERE WAS NO INDICATION THAT RESULTS WERE REPORTED OUT AND THERE WAS NO REPORT OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: " CUSTOMER REPORTS THAT REPLACEMENT RECEIVED UNDER CASE (B)(4). IS ALSO POSITIVE FOR E. FAECALIS "

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12019110 ↗

Other CLINICAL SAMPLE CONCENTRATOR devices

All CLINICAL SAMPLE CONCENTRATOR devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page