BD CTGCTV2 443979Recalls & reported problems

Becton, Dickinson And Company · BD CTGCTV2 · Class 1

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00382904439798), received between Dec 11, 2024 and Aug 21, 2026 — classified as 6 malfunctions.

The most frequently reported specific device problem is “False Positive Result” (3 reports).

1 other recall cover devices cleared under the same FDA submission (K210585); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 1 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • False Positive Result3
  • False Negative Result2
  • Incorrect, Inadequate or Imprecise Result or Readings1

Patient problems

  • No Clinical Signs, Symptoms or Conditions6

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 21, 2026False Positive Result

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50379882 ↗
  • MalfunctionMay 12, 2026False Negative Result

    IT WAS REPORTED THAT DURING USE OF BD MAX¿ CTGCTV2, A FALSE NEGATIVE CHLAMYDIA TRACHOMATIS (CT) PATIENT RESULT WAS OBTAINED. URINE SAMPLE WAS CT POSITIVE, HOWEVER, SWAB SAMPLE WAS CT NEGATIVE. UPON REPEAT TESTING ONCE WITH URINE AND TWICE WITH SWAB, ALL WERE CT POSITIVE. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25148827 ↗
  • MalfunctionApr 14, 2026False Negative Result

    IT WAS REPORTED THAT DURING USE OF BD MAX¿ CTGCTV2, A FALSE NEGATIVE CHLAMYDIA TRACHOMATIS (CT) PATIENT RESULT WAS OBTAINED. URINE SAMPLE WAS CT POSITIVE; HOWEVER, SWAB SAMPLE WAS CT NEGATIVE. UPON REPEAT TESTING ONCE WITH URINE AND TWICE WITH SWAB, ALL WERE CT POSITIVE. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24872505 ↗
  • MalfunctionMar 30, 2026Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED THAT DURING USE OF BD MAX¿ CTGCTV2, A DISCREPANT CHLAMYDIA TRACHOMATIS (CT) PATIENT RESULT WAS OBTAINED. SAMPLE INITIALLY TESTED CT POSITIVE. SAMPLE WAS THEN REPEATED AND WAS CT NEGATIVE. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24716087 ↗
  • MalfunctionJul 17, 2025False Positive Result; False Positive Result

    IT WAS REPORTED THAT DURING USE OF BD CTGCTV2 FOR BD MAX¿ SYSTEM, A TRICHOMONAS VAGINALIS FALSE POSITIVE PATIENT RESULT WAS OBTAINED. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22531884 ↗
  • MalfunctionDec 11, 2024False Positive Result; False Positive Result

    IT WAS REPORTED THAT DURING USE OF BD CTGCTV2 ON BD MAX INSTRUMENT, 8 SAMPLES WERE FALSE POSITIVE FOR BOTH CHLAMYDIA TRACHOMATIS AND NEISSERIA GONORRHOEAE. AFTER INSPECTING THE REAGENTS AND INSTRUMENT, IT WAS VISIBLE THAT A "BLOODY SLIME STRING" FROM A SAMPLE WAS DRAGGED OVER THE PROCESS AREA AND CONTAMINATED THE SAMPLES. CLEANING AND ENVIRONMENTAL TESTING WERE PERFORMED. SAMPLES WERE RERUN AND WERE NEGATIVE. THERE WAS NO REPORT OF PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20902537 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K210585. One submission often covers many models, so check the product and lot details in each record.

  • Z-0588-2025 · Oct 31, 2024 · Open, Classified
    BD COR MX Instrument Version or Model: 443989 Catalog Number: 443989

Other DNA PROBE, NUCLEIC ACID AMPLIFICATION, CHLAMYDIA devices

All DNA PROBE, NUCLEIC ACID AMPLIFICATION, CHLAMYDIA devices →