Skip to content

BD Kiestra™ ReadA 446948Recalls & reported problems

BD Kiestra B.V. · BD Kiestra™ ReadA · Class 1

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. Feb 6, 2026Class IIOpen, Classified

    In certain situations, following a system reboot, one or more modules may have become unreachable via the remote connectivity interface. When this occurred, there was potential for delays in plate retrieval from the incubator while connectivity was restored. This issue was intermittent and did not affect test results, performance of the assay, or sample integrity.

    Z-1658-2026 · Root cause: Software Manufacturing/Software Deployment · Recall details · FDA record ↗

Other BATH, INCUBATORS/WATER, ALL devices

All BATH, INCUBATORS/WATER, ALL devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page