NIPRO SAFETOUCH™ II SAFETY FISTULA NEEDLE 15GX1" HC-30WRecalls & reported problems
Nipro Medical Corporation · 15Gx1" Fixed Hub 11" Tubing · Class 2
All models of NIPRO SAFETOUCH II SAFETY FISTULA NEEDLE (4)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 14
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 14 adverse event reports that list this exact device (DI 00383790000062), received between Nov 9, 2020 and May 27, 2021 — classified as 14 malfunctions.
The most frequently reported specific device problem is “Retraction Problem” (12 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Retraction Problem12
- Material Fragmentation2
Patient problems
- No Clinical Signs, Symptoms or Conditions14
Recent adverse event reports
The 14 most recent of 14 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 27, 2021Material Fragmentation; Material Fragmentation
DURING DIALYSIS TREATMENT THE SMALL CLEAR PLASTIC RING BELOW THE SAFETY, HAD CRACKED AND FELL OFF. TREATMENT WAS COMPLETED AND THE NURSE WENT TO REMOVE THE AVF FISTULA NEEDLE AND NOTED IT WAS SITTING BY THE PATIENTS HAND. NO ISSUE REMOVING THE AVF NEEDLE. NO INJURIES OR MEDICAL INTERVENTIONS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11897958 ↗MalfunctionMay 27, 2021Material Fragmentation; Material Fragmentation
DURING DIALYSIS TREATMENT THE SMALL CLEAR PLASTIC RING BELOW THE SAFETY, HAD CRACKED AND FELL OFF. TREATMENT WAS COMPLETED AND THE NURSE WENT TO REMOVE THE AVF FISTULA NEEDLE AND NOTED IT WAS SITTING BY THE PATIENTS HAND. NO ISSUE REMOVING THE AVF NEEDLE. NO INJURIES OR MEDICAL INTERVENTIONS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11897875 ↗MalfunctionFeb 18, 2021Retraction Problem; Retraction Problem
NURSE MANAGER, REPORTED THAT A SAFETOUCH II NEEDLE WOULD NOT RETRACT INTO THE SAFETY UPON REMOVAL. THIS WAS NOTICED UPON REMOVING THE NEEDLE POST TREATMENT. NO BLOOD WAS LOST, NO ADVERSE EVENTS OCCURRED, NO NEEDLE STICK RESULTED. NO FURTHER INFORMATION HAS BEEN PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11346874 ↗MalfunctionFeb 18, 2021Retraction Problem; Retraction Problem
NURSE MANAGER, REPORTED THAT A SAFETOUCH II NEEDLE WOULD NOT RETRACT INTO THE SAFETY UPON REMOVAL. THIS WAS NOTICED UPON REMOVING THE NEEDLE POST TREATMENT. NO BLOOD WAS LOST, NO ADVERSE EVENTS OCCURRED, NO NEEDLE STICK RESULTED. NO FURTHER INFORMATION HAS BEEN PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11346870 ↗MalfunctionFeb 18, 2021Retraction Problem; Retraction Problem
NURSE MANAGER, REPORTED THAT A SAFETOUCH II NEEDLE WOULD NOT RETRACT INTO THE SAFETY UPON REMOVAL. THIS WAS NOTICED UPON REMOVING THE NEEDLE POST TREATMENT. NO BLOOD WAS LOST, NO ADVERSE EVENTS OCCURRED, NO NEEDLE STICK RESULTED. NO FURTHER INFORMATION HAS BEEN PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11346854 ↗MalfunctionFeb 18, 2021Retraction Problem; Retraction Problem
NURSE MANAGER, REPORTED THAT A SAFETOUCH II NEEDLE WOULD NOT RETRACT INTO THE SAFETY UPON REMOVAL. THIS WAS NOTICED UPON REMOVING THE NEEDLE POST TREATMENT. NO BLOOD WAS LOST, NO ADVERSE EVENTS OCCURRED, NO NEEDLE STICK RESULTED. NO FURTHER INFORMATION HAS BEEN PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11346840 ↗MalfunctionFeb 18, 2021Retraction Problem; Retraction Problem
NURSE MANAGER, REPORTED THAT A SAFETOUCH II NEEDLE WOULD NOT RETRACT INTO THE SAFETY UPON REMOVAL. THIS WAS NOTICED UPON REMOVING THE NEEDLE POST TREATMENT. NO BLOOD WAS LOST, NO ADVERSE EVENTS OCCURRED, NO NEEDLE STICK RESULTED. NO FURTHER INFORMATION HAS BEEN PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11346830 ↗MalfunctionFeb 18, 2021Retraction Problem; Retraction Problem
NURSE MANAGER, REPORTED THAT A SAFETOUCH II NEEDLE WOULD NOT RETRACT INTO THE SAFETY UPON REMOVAL. THIS WAS NOTICED UPON REMOVING THE NEEDLE POST TREATMENT. NO BLOOD WAS LOST, NO ADVERSE EVENTS OCCURRED, NO NEEDLE STICK RESULTED. NO FURTHER INFORMATION HAS BEEN PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11346819 ↗MalfunctionDec 23, 2020Retraction Problem; Retraction Problem
SAFETOUCH II NEEDLE WOULD NOT RETRACT INTO THE SAFETY UPON REMOVAL. THIS WAS NOTICED UPON REMOVING THE NEEDLE POST TREATMENT. NO BLOOD WAS LOST, NO ADVERSE EVENTS OCCURRED, NO NEEDLE STICK RESULTED. (3 TOTAL INCIDENTS) THIS FACILITY HAS USED SAFETOUCH II SINCE (B)(6) 2019 T HE CLINICAL SPECIALIST PROVIDED TRAINING ON WEDNESDAY, (B)(6) 2019 WEDNESDAY, (B)(6) 2019, AND WEDNESDAY, (B)(6) 2020 TO PROVIDE SUPPLEMENTAL TRAINING AND SUPPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11070510 ↗MalfunctionDec 23, 2020Retraction Problem; Retraction Problem
SAFETOUCH II NEEDLE WOULD NOT RETRACT INTO THE SAFETY UPON REMOVAL. THIS WAS NOTICED UPON REMOVING THE NEEDLE POST TREATMENT. NO BLOOD WAS LOST, NO ADVERSE EVENTS OCCURRED, NO NEEDLE STICK RESULTED. (2 TOTAL INCIDENTS) THIS FACILITY HAS USED SAFETOUCH II SINCE (B)(6) 2019. THE CLINICAL SPECIALIST PROVIDED TRAINING ON WEDNESDAY, (B)(6) 2019, WEDNESDAY, (B)(6) 2019, AND WEDNESDAY, (B)(6) 2020 TO PROVIDE SUPPLEMENTAL TRAINING AND SUPPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11070481 ↗MalfunctionDec 18, 2020Retraction Problem; Retraction Problem
SAFETOUCH II NEEDLE WOULD NOT RETRACT INTO THE SAFETY UPON REMOVAL. THIS WAS NOTICED UPON REMOVING THE NEEDLE POST TREATMENT. NO BLOOD WAS LOST, NO ADVERSE EVENTS OCCURRED, NO NEEDLE STICK RESULTED. THIS FACILITY HAS USED SAFETOUCH II SINCE (B)(6) 2019. THE CLINICAL SPECIALIST PROVIDED TRAINING ON WEDNESDAY, (B)(6) 2019, WEDNESDAY, (B)(6) 2019, AND WEDNESDAY, (B)(6) 2020 TO PROVIDE SUPPLEMENTAL TRAINING AND SUPPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11041473 ↗MalfunctionDec 18, 2020Retraction Problem; Retraction Problem
SAFETOUCH II NEEDLE WOULD NOT RETRACT INTO THE SAFETY UPON REMOVAL. THIS WAS NOTICED UPON REMOVING THE NEEDLE POST TREATMENT. NO BLOOD WAS LOST, NO ADVERSE EVENTS OCCURRED, NO NEEDLE STICK RESULTED. THIS FACILITY HAS USED SAFETOUCH II SINCE (B)(6) 2019. THE CLINICAL SPECIALIST PROVIDED TRAINING ON WEDNESDAY, (B)(6) 2019, WEDNESDAY, (B)(6) 2019, AND WEDNESDAY, (B)(6) 2020 TO PROVIDE SUPPLEMENTAL TRAINING AND SUPPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11036535 ↗MalfunctionNov 9, 2020Retraction Problem; Retraction Problem
NURSE MANAGER, REPORTED THAT A SAFETOUCH II NEEDLE WOULD NOT RETRACT INTO THE SAFETY UPON REMOVAL. FURTHER INFORMATION HAS BEEN REQUESTED INCLUDING WHAT THE AFFECTED LOT# IS. THIS WAS NOTICED UPON REMOVING THE NEEDLE POST TREATMENT. NO BLOOD WAS LOST, NO ADVERSE EVENTS OCCURRED, NO NEEDLE STICK RESULTED. ADDITIONAL DETAILS 10/22/2020: THE TREATMENT WAS COMPLETED, WITH NO INTERRUPTIONS. NO PATIENT HARM. AT THE END OF THE TREATMENT WHEN STAFF REMOVED THE NEEDLE FROM THE PATIENT ACCESS, THEY WERE UNABLE TO SLIDE THE SAFETY OVER THE NEEDLE: SOME WOULD NOT MOVE AT ALL, MY EXPERIENCE WAS THAT THEY WERE VERY DIFFICULT TO MOVE, CAUSING THE USE OF 2 HANDS, AND DRIPPING BLOOD AND SALINE IN THE PROCESS. EXACT DATE OF OCCURRENCE IS UNKNOWN. EVENT OCCURRED BETWEEN SEPTEMBER AND OCTOBER 2020.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10809100 ↗MalfunctionNov 9, 2020Retraction Problem; Retraction Problem
NURSE MANAGER, REPORTED THAT A SAFETOUCH II NEEDLE WOULD NOT RETRACT INTO THE SAFETY UPON REMOVAL. FURTHER INFORMATION HAS BEEN REQUESTED INCLUDING WHAT THE AFFECTED LOT# IS. THIS WAS NOTICED UPON REMOVING THE NEEDLE POST TREATMENT. NO BLOOD WAS LOST, NO ADVERSE EVENTS OCCURRED, NO NEEDLE STICK RESULTED. ADDITIONAL DETAILS 10/22/2020: THE TREATMENT WAS COMPLETED, WITH NO INTERRUPTIONS. NO PATIENT HARM. AT THE END OF THE TREATMENT WHEN STAFF REMOVED THE NEEDLE FROM THE PATIENT ACCESS, THEY WERE UNABLE TO SLIDE THE SAFETY OVER THE NEEDLE: SOME WOULD NOT MOVE AT ALL, MY EXPERIENCE WAS THAT THEY WERE VERY DIFFICULT TO MOVE, CAUSING THE USE OF 2 HANDS, AND DRIPPING BLOOD AND SALINE IN THE PROCESS. EXACT DATE OF OCCURRENCE IS UNKNOWN. EVENT OCCURRED BETWEEN SEPTEMBER AND OCTOBER 2020.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10808979 ↗
Other Catheter,intravascular,therapeutic,short-term less than 30 days devices
- BD Insyte Autoguard 382523Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days660 reports
- BD Insyte Autoguard 382533Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days2 recalls527 reports
- BD PosiFlush 306546Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days1 death608 reports
- BD Insyte Autoguard Shielded IV Catheter 381823Becton, Dickinson And Company · Catheter, intravascular, therapeutic, short-term less than 30 days582 reports
- BD Insyte Autoguard 381423Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days1 recall528 reports
- BD Insyte Autoguard 382544Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days1 recall437 reports
- BD Nexiva, BD Q-Syte 383536Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days2 recalls374 reports
- BD PosiFlush 306547Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days2 recalls1 death331 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page