Nova Biomedical Corporation · Stat Profile pHOx Ultra Co-Oximeter Analyzer · Class 1

All models of STAT PROFILE (13)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
14
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 14 adverse event reports that list this exact device (DI 00385480420136), received between Jun 7, 2019 and May 23, 2022 — classified as 14 malfunctions.

The most frequently reported specific device problem is “Misassembly During Maintenance/Repair” (12 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Misassembly During Maintenance/Repair12
  • Low Test Results10
  • High Test Results6
  • Incorrect, Inadequate or Imprecise Result or Readings3

Patient problems

  • No Known Impact Or Consequence To Patient12
  • No Clinical Signs, Symptoms or Conditions1
  • No Information1

Recent adverse event reports

The 14 most recent of 14 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 23, 2022Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    THE CUSTOMER REPORTED THAT THEIR PHOX ULTRA ANALYZER, SERIAL NUMBER (B)(4) GAVE A CRITICAL LOW ICA RESULT FOR A PATIENT. THE TEST WAS RUN ON (B)(6) 2022 USING ICA SENSOR LOT NUMBER 22033023. WHEN SAME SAMPLE WAS REPEATED ON A DIFFERENT PHOX ULTRA, THE RESULT WAS NORMAL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14476543 ↗
  • MalfunctionOct 9, 2019Incorrect, Inadequate or Imprecise Result or Readings

    DISCREPANT PATIENT PH RESULT ON NOVA PHOX ULTRA COOX ANALYZER.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9171679 ↗
  • MalfunctionJun 7, 2019Misassembly During Maintenance/Repair; High Test Results; Low Test Results; Misassembly During Maintenance/Repair

    ON (B)(6) 2019: CUSTOMER REPORTED LOW NA AND HIGH CL RESULTS FOR A PATIENT IN EMERGENCY DEPARTMENT USING A NOVA PHOX ULTRA COOX ANALYZER WHEN COMPARED TO THE LABORATORY REFERENCE ANALYZER.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8680652 ↗
  • MalfunctionJun 7, 2019Misassembly During Maintenance/Repair; Low Test Results; Misassembly During Maintenance/Repair

    ON (B)(6) 2019: CUSTOMER REPORTED LOW NA RESULT FOR A PATIENT IN EMERGENCY DEPARTMENT USING A NOVA PHOX ULTRA COOX ANALYZER WHEN COMPRED TO THE LABORATORY REFERENCE ANALYZER.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8680650 ↗
  • MalfunctionJun 7, 2019Misassembly During Maintenance/Repair; High Test Results; Low Test Results; Misassembly During Maintenance/Repair

    ON (B)(6) 2019: CUSTOMER REPORTED LOW NA AND HIGH CL RESULTS FOR A PATIENT IN EMERGENCY DEPARTMENT USING A NOVA PHOX ULTRA COOX ANALYZER WHEN COMPARED TO THE LABORATORY REFERENCE ANALYZER

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8680648 ↗
  • MalfunctionJun 7, 2019Misassembly During Maintenance/Repair; High Test Results; Low Test Results; Misassembly During Maintenance/Repair

    ON (B)(6) 2019: CUSTOMER REPORTED LOW NA AND HIGH CL RESULTS FOR A PATIENT IN EMERGENCY DEPARTMENT USING A NOVA PHOX ULTRA COOX ANALYZER WHEN COMPARED TO THE LABORATORY REFERENCE ANALYZER.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8680647 ↗
  • MalfunctionJun 7, 2019Misassembly During Maintenance/Repair; High Test Results; Misassembly During Maintenance/Repair

    ON (B)(6) 2019: CUSTOMER REPORTED HIGH CL RESULT FOR A PATIENT IN EMERGENCY DEPARTMENT USING A NOVA PHOX ULTRA COOX ANALYZER WHEN COMPARED TO THE LABORATORY REFERENCE ANALYZER.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8680646 ↗
  • MalfunctionJun 7, 2019Misassembly During Maintenance/Repair; Low Test Results; Misassembly During Maintenance/Repair

    ON (B)(6) 2019: CUSTOMER REPORTED LOW NA RESULT FOR A PATIENT IN EMERGENCY DEPARTMENT USING A NOVA PHOX ULTRA COOX ANALYZER WHEN COMPARED TO THE LABORATORY REFERENCE ANALYZER.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8680645 ↗
  • MalfunctionJun 7, 2019Misassembly During Maintenance/Repair; High Test Results; Low Test Results; Misassembly During Maintenance/Repair

    ON (B)(6) 2019: CUSTOMER REPORTED LOW NA AND HIGH CL RESULTS FOR A PATIENT IN EMERGENCY DEPARTMENT USING A NOVA PHOX ULTRA COOX ANALYZER WHEN COMPARED TO THE LABORATORY REFERENCE ANALYZER.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8680644 ↗
  • MalfunctionJun 7, 2019Misassembly During Maintenance/Repair; Low Test Results; Misassembly During Maintenance/Repair

    ON (B)(6) 2019: CUSTOMER REPORTED LOW NA RESULTS FOR A PATIENT IN EMERGENCY DEPARTMENT USING A NOVA PHOX ULTRA COOX ANALYZER WHEN COMPARED TO THE LABORATORY REFERENCE ANALYZER.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8680643 ↗
  • MalfunctionJun 7, 2019Misassembly During Maintenance/Repair; Low Test Results; Misassembly During Maintenance/Repair

    ON (B)(6) 2019: CUSTOMER REPORTED LOW NA RESULT FOR A PATIENT IN EMERGENCY DEPARTMENT USING A NOVA PHOX ULTRA COOX ANALYZER WHEN COMPARED TO THE LABORATORY REFERENCE ANALYZER.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8680642 ↗
  • MalfunctionJun 7, 2019Misassembly During Maintenance/Repair; Low Test Results; Misassembly During Maintenance/Repair

    ON (B)(6) 2019: CUSTOMER REPORTED LOW NA RESULT FOR A PATIENT IN EMERGENCY DEPARTMENT USING A NOVA PHOX ULTRA COOX ANALYZER WHEN COMPARED TO THE LABORATORY REFERENCE ANALYZER.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8680641 ↗
  • MalfunctionJun 7, 2019Misassembly During Maintenance/Repair; High Test Results; Low Test Results; Misassembly During Maintenance/Repair

    ON (B)(6) 2019: CUSTOMER REPORTED LOW NA AND HIGH CL RESULTS FOR A PATIENT IN EMERGENCY DEPARTMENT USING A NOVA PHOX ULTRA COOX ANALYZER WHEN COMPARED TO THE LABORATORY REFERENCE ANALYZER.

    Patient problems: No Information

    MAUDE report 8680634 ↗
  • MalfunctionJun 7, 2019Misassembly During Maintenance/Repair; Incorrect, Inadequate or Imprecise Result or Readings; Misassembly During Maintenance/Repair

    ON (B)(6) 2019: CUSTOMER REPORTED LOW NA AND HIGH CL RESULTS FOR A PATIENT IN EMERGENCY DEPARTMENT USING A NOVA PHOX ULTRA COOX ANALYZER WHEN COMPARED TO THE LABORATORY REFERENCE ANALYZER.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8680632 ↗