StatSensor 1.0 (REF 44052)Recalls & reported problems
Nova Biomedical Corporation · StatSensor Creatinine Meter with Battery and Quick Reference Guide · Class 2
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00385480440523), received between May 11, 2021 and Dec 30, 2021 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Low Test Results” (3 reports).
5 other recalls cover devices cleared under the same FDA submission (K070068); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Low Test Results3
- Incorrect Measurement2
- Low Readings2
Patient problems
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 30, 2021Low Test Results; Low Test Results
THE CUSTOMER REPORTED DISCREPANT CREATININE RESULTS ON A STATSENSOR METER SN (B)(4)WHILE USING STRIPS WITH LOT NUMBER 4921209129. THE RESULTS WERE QUESTIONED AND THE SAMPLES WERE RUN ON A LAB CHEMISTRY ANALYZER WHICH CONFIRMED THE DISCREPANT VALUES. THE PATIENT WAS NOT TREATED BASED OFF OF THE RESULTS FROM THE METER AND THERE WERE NO ADVERSE INTERACTIONS. THERE WAS NO PATIENT HARM OR INTERVENTION REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13117443 ↗MalfunctionMay 12, 2021Incorrect Measurement; Low Test Results; Low Readings; Incorrect Measurement; Low Test Results; Low Readings
NOVA BIOMEDICAL (NOVA) WAS MADE AWARE OF A POTENTIAL ISSUE REGARDING DISCREPANT CREATININE RESULTS USING A STATSENSOR HOSPITAL METER, SERIAL NUMBER (B)(4) AND CREATININE TEST STRIPS LOT 4920177139. IT WAS REPORTED THE PATIENT DISAGREED WITH THE INITIAL RESULTS. WHEN SAMPLES WERE RUN ON A LAB ANALYZER, THE RESULTS DID NOT MATCH THOSE OF THE METER. NO INAPPROPRIATE TREATMENT WAS STARTED AND THE PATIENT WAS NOT ADVERSELY AFFECTED BY THE RESULT. THIS IS REPORT TWO OF TWO.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11812726 ↗MalfunctionMay 11, 2021Incorrect Measurement; Low Test Results; Low Readings; Incorrect Measurement; Low Test Results; Low Readings
NOVA BIOMEDICAL (NOVA) WAS MADE AWARE OF A POTENTIAL ISSUE REGARDING DISCREPANT CREATININE RESULTS USING A STATSENSOR HOSPITAL METER, SERIAL NUMBER (B)(4) AND CREATININE TEST STRIPS LOT 4920177139. IT WAS REPORTED THE PATIENT DISAGREED WITH THE INITIAL RESULTS. TESTING WITH ANOTHER METER SHOWED SIMILAR RESULTS. WHEN SAMPLES WERE RUN ON A LAB ANALYZER, THE RESULTS DID NOT MATCH EITHER OF THE METERS. NO INAPPROPRIATE TREATMENT WAS STARTED AND THE PATIENT WAS NOT ADVERSELY AFFECTED BY THE RESULT. THIS IS REPORT ONE OF TWO.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11807585 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K070068. One submission often covers many models, so check the product and lot details in each record.
- Z-0650-2014 · Oct 10, 2013 · TerminatedNova StatStrip/StatSensor Hospital Blood Glucose Meter; Nova StatStrip Lactate Hospital Meter System; Nova StatSensor Creatinine Hospital Meter, Rechargeable Lithium Batteries Sin
- Z-1472-2008 · Feb 21, 2008 · TerminatedVarta Easypack XL Lithium Polymer Batteries (P/N 41772 and 42215) Used with: Nova Biomedical StatStrip Glucose Hospital Meter (P/N 42224) and Nova Biomedical StatSensor Creatinine
- Z-0725-2008 · Nov 2, 2007 · TerminatedNova - EZ CHEM Control Solution P/N 9612, Nova Biomedical, Waltham, MA 02454
- Z-0726-2008 · Nov 2, 2007 · TerminatedNova - EZ CHEM Linearity Solution P/N 9618, Nova Biomedical, Waltham, MA 02454
- Z-0724-2008 · Nov 2, 2007 · TerminatedNova - EZ CHEM Creatinine Meter: SRS Versions 5.1 and 5.2 P/N 9600, Nova Biomedical, Waltham, MA 02454
Other ELECTRODE, ION BASED, ENZYMATIC, CREATININE devices
- i-STAT 03P84-25Abbott Point Of Care Inc. · ELECTRODE, ION BASED, ENZYMATIC, CREATININE1 recall34 reports
- STATSENSOR 1.0Nova Biomedical Corporation · ELECTRODE, ION BASED, ENZYMATIC, CREATININE5 reports
- StatSensor 1.0Nova Biomedical Corporation · ELECTRODE, ION BASED, ENZYMATIC, CREATININE2 reports
- StatSensor/StatStrip 1.75Nova Biomedical Corporation · ELECTRODE, ION BASED, ENZYMATIC, CREATININE1 report
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.