Skip to content

STATSTRIP 1.75 (REF 53398)Recalls & reported problems

Nova Biomedical Corporation · Statstrip Glucose Meter with Battery and Quick Reference Guide · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
12
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 12 adverse event reports that list this exact device (DI 00385480533980), received between Apr 21, 2017 and May 2, 2025 — classified as 12 malfunctions.

The most frequently reported specific device problem is “Patient Data Problem” (7 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Patient Data Problem7
  • Failure to Transmit Record1
  • High Test Results1
  • Imprecise or Erratic Results1
  • Insufficient Device Problem Information1
  • Melted1
  • Overheating of Device1
  • Smoking1

Patient problems

  • No Clinical Signs, Symptoms or Conditions11
  • Insufficient Information3
  • No Code Available1

Recent adverse event reports

The 12 most recent of 12 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 2, 2025High Test Results; High Test Results

    THE COMPLAINANT REPORTED "CONFLICTING RESULTS BETWEEN 2 GLUCOMETERS AND PATIENT'S DEXCOM. NURSE ALSO SAID THE PATIENT'S DEXCOM WAS "OUT OF WACK" AS WHEN JUICE WAS GIVEN THE DEXCOM READING DID NOT CHANGE."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21953028 ↗
  • MalfunctionFeb 10, 2022Patient Data Problem; Patient Data Problem

    THE CUSTOMER REPORTS THAT A STATSTRIP METER SCANNED A PATIENT ARMBAND INCORRECTLY. WHEN ONE PATIENT WAS SCANNED, A DIFFERENT PATIENT'S DATA APPEARED. AFTER A RESCAN, THE CORRECT PATIENT'S INFORMATION APPEARED SO THE RESPONSIBLE METER'S SERIAL NUMBER WAS NOT TAKEN NOTE OF. THERE WAS NO PATIENT HARM OR INTERVENTION REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13505963 ↗
  • MalfunctionJan 26, 2022Patient Data Problem; Patient Data Problem

    THE CUSTOMER REPORTS THAT A STATSTRIP METER SCANNED A PATIENT ARMBAND INCORRECTLY. WHEN ONE PATIENT WAS SCANNED A DIFFERENT PATIENT'S DATA APPEARED. AFTER A RESCAN THE CORRECT PATIENT'S INFORMATION APPEARED SO THE METER'S SERIAL NUMBER WAS NOT TAKEN NOTE OF. THERE WAS NO PATIENT HARM OR INTERVENTION REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13359388 ↗
  • MalfunctionJul 8, 2021Failure to Transmit Record; Failure to Transmit Record

    THE CUSTOMER REPORTED A STATSTRIP GLUCOSE HOSPITAL METER MALFUNCTIONED WHICH REQUIRED HAVING TO USE AN ADDITIONAL METER TO OBTAIN GLUCOSE RESULTS FOR A PATIENT. THE EVENT OCCURRED IN THE PULMONARY CLINIC WHILE ATTEMPTING TO OBTAIN GLUCOSE MEASUREMENTS FOR A RAPID RESPONSE INCIDENT. THERE WAS NO INDICATION OF PATIENT HARM AS A SECOND STATSTRIP GLUCOSE HOSPITAL METER WAS AVAILABLE FOR USE AFTER RUNNING THE REQUIRED QUALITY CONTROL MATERIAL. THE REPORTED STATSTRIP GLUCOSE HOSPITAL METER MALFUNCTIONED BY REQUIRING QUALITY CONTROL MATERIAL TO BE RUN AGAIN, DESPITE HAVING BEEN RUN IN THE LAST 24 HOURS AS REQUIRED, AND BY NOT RECOGNIZING THREE CERTIFIED OPERATORS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12135760 ↗
  • MalfunctionMay 11, 2021Imprecise or Erratic Results; Imprecise or Erratic Results

    THE CUSTOMER REPORTED DISCREPANT PATIENT RESULTS. A TEST WAS RUN ON (B)(6) ON THE STATSTRIP 1.75 METER SN: (B)(4) AND A FLOW ERROR OCCURRED. IT WAS RERUN AND WAS LESS THAN 10, A DRAW WAS DONE AND IT WAS 532. THE PATIENT WAS MOVED TO ANOTHER FLOOR AND TESTED WITH ANOTHER METER AND RECEIVED GOOD RESULTS. THERE WAS NO PATIENT HARM OR MEDICAL INTERVENTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11809037 ↗
  • MalfunctionMar 17, 2021Patient Data Problem; Patient Data Problem

    A CUSTOMER REPORTED THAT GLUC CONN METERS (1.75) MIS-SCANNING 1D BAR SHOWING INCORRECT PATIENT ID'S. THE ACCOUNT HAS 1 & 2 D BAR CODES ON PATIENTS WRISTS. THERE HAS BEEN NO REPORT OF PATIENT HARM OR ADVERSE EVENT.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 11507600 ↗
  • MalfunctionMar 17, 2021Patient Data Problem; Patient Data Problem

    A CUSTOMER REPORTED THAT GLUC CONN METERS (1.75) MIS-SCANNING 1D BAR SHOWING INCORRECT PATIENT ID'S. THE ACCOUNT HAS 1 & 2 D BAR CODES ON PATIENTS WRISTS. THERE HAS BEEN NO REPORT OF PATIENT HARM OR ADVERSE EVENT.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 11507527 ↗
  • MalfunctionMar 17, 2021Patient Data Problem; Patient Data Problem

    A CUSTOMER REPORTED THAT GLUC CONN METERS (1.75) MIS-SCANNING 1D BAR SHOWING INCORRECT PATIENT ID'S. THE ACCOUNT HAS 1 & 2 D BAR CODES ON PATIENTS WRISTS. THERE HAS BEEN NO REPORT OF PATIENT HARM OR ADVERSE EVENT.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 11507503 ↗
  • MalfunctionMar 17, 2021Patient Data Problem; Patient Data Problem

    A CUSTOMER REPORTED THAT GLUC CONN METERS (1.75) MIS-SCANNING 1D BAR SHOWING INCORRECT PATIENT ID'S. THE ACCOUNT HAS 1 & 2 D BAR CODES ON PATIENTS WRISTS. THERE HAS BEEN NO REPORT OF PATIENT HARM OR ADVERSE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11507427 ↗
  • MalfunctionMar 17, 2021Patient Data Problem; Patient Data Problem

    A CUSTOMER REPORTED THAT GLUC CONN METERS (1.75) MIS-SCANNING 1D BAR SHOWING INCORRECT PATIENT ID'S. THE ACCOUNT HAS 1 & 2 D BAR CODES ON PATIENTS WRISTS. THERE HAS BEEN NO REPORT OF PATIENT HARM OR ADVERSE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11507402 ↗
  • MalfunctionDec 22, 2020Melted; Overheating of Device; Smoking; Overheating of Device

    THE CUSTOMER REPORTED THAT THE STATSTRIP WIRELESS METER (1.75) S/N: (B)(4) HAD BEEN SITTING IN THE DOCKING STATION S/N:(B)(4) FOR APPROXIMATELY ONE WEEK. WHEN THE METER WAS REMOVED FROM THE DOCKING STATION, THERE WAS A SMALL PUFF OF BLACK SMOKE, THE PLASTIC CASE WAS SLIGHTLY MELTED AND THE CONNECTION PINS APPEARED RED. THE BATTERY WAS REMOVED AND THE DOCKING STATION UNPLUGGED. THERE WAS NO PATIENT HARM OR MEDICAL INTERVENTION REPORTED. THIS IS REPORT 2 OF 2.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11059406 ↗
  • MalfunctionApr 21, 2017Insufficient Device Problem Information

    FACILITY CALLED TO REPORT CONCERN REGARDING A PATIENT (BABY) GLUCOSE RESULT. DUE TO THE RESULT THE BABY WAS MOVED TO THE NICU.

    Patient problems: No Code Available

    MAUDE report 6513482 ↗

Other Prescription use blood glucose meter for near-patient testing devices

All Prescription use blood glucose meter for near-patient testing devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page