Adverse event reports
7
Deaths reported
1
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00385480536578), received between Sep 29, 2020 and Oct 22, 2024 — classified as 1 death, 1 injury, 5 malfunctions.

The most frequently reported specific device problem is “High Test Results” (3 reports).

1 other recall cover devices cleared under the same FDA submission (K141907); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • High Test Results3
  • Low Test Results3
  • Inability to Irrigate1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5
  • Aneurysm1
  • Exposure to Body Fluids1

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • DeathOct 22, 2024Low Test Results; Low Test Results

    ON (B)(6) 2024 PATIENT WAS ADMITTED WITH AN ANEURYSM AND UNDERWENT CARDIAC SURGERY TO PLACE VALVES (UNSPECIFIED). LOW PCO2 RESULT INCOMPATIBLE WITH THE PATIENT'S CLINIC. PATIENT EXPIRED THURSDAY EVENING (B)(6) 2024.

    Patient problems: Aneurysm

    MAUDE report 20511410 ↗
  • InjuryDec 9, 2022Inability to Irrigate; Inability to Irrigate

    THE OPERATOR OF THE DEVICE SUSTAINED EYE EXPOSURE TO FLUSH FLUID WHEN ATTEMPTING TO CLEAR A VISIBLE CLOT IN THE PROBE OF THE ANALYZER. THE OPERATOR WAS NOT WEARING EYE PROTECTION.

    Patient problems: Exposure to Body Fluids

    MAUDE report 15952812 ↗
  • MalfunctionApr 5, 2022Low Test Results; Low Test Results

    THE CUSTOMER IS REPORTING VARIABLE RESULTS FOR PCO2 ON THEIR PRIME ANALYZER. NO PATIENT HARM OR INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14014720 ↗
  • MalfunctionApr 5, 2022Low Test Results; Low Test Results

    THE CUSTOMER IS REPORTING VARIABLE RESULTS FOR PCO2 ON THEIR PRIME ANALYZER. NO PATIENT HARM OR INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14014669 ↗
  • MalfunctionApr 5, 2022High Test Results; High Test Results

    THE CUSTOMER IS REPORTING VARIABLE RESULTS FOR PCO2 ON THEIR PRIME ANALYZER. NO PATIENT HARM OR INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14014485 ↗
  • MalfunctionApr 5, 2022High Test Results; High Test Results

    THE CUSTOMER IS REPORTING VARIABLE RESULTS FOR PCO2 ON THEIR PRIME ANALYZER. NO PATIENT HARM OR INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14014402 ↗
  • MalfunctionSep 29, 2020High Test Results; High Test Results

    CUSTOMER REPORTED ON (B)(6) 2020 THAT PCO2 PATIENT RESULTS OBTAINED ON A NOVA PRIME CCS COMP W/SCANNER ANALYZER SN (B)(4) WERE HIGH COMPARED TO THEIR LAB ANALYZER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10601007 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K141907. One submission often covers many models, so check the product and lot details in each record.

  • Z-2732-2017 · Jun 19, 2017 · Terminated
    Nova Biomedical Stat Profile Prime Analyzer as follows: PN 52855 Stat Profile Prime ABG Analyzer PN 53655 Stat Profile Prime ABG Analyzer with Scanner PN 52856 Stat Profile Prime C

Other CONTROLS FOR BLOOD-GASES, (ASSAYED AND UNASSAYED) devices

All CONTROLS FOR BLOOD-GASES, (ASSAYED AND UNASSAYED) devices →