Stat Profile Prime Plus (REF 57400)Recalls & reported problems
Nova Biomedical Corporation · Stat Profile Prime Plus® Analyzer · Class 2
Data updated Oct 13, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
The Stat Profile Prime Plus is a Nova Biomedical blood gas analyzer used in hospital labs. Its 50 FDA reports mostly describe results that seemed wrong or unstable, including low or high readings and analyzers going out of calibration or stopping entirely; 1 report describes a death and none describe injuries. One open recall, started in September 2023, covers the chance of incorrect creatinine results in the first 2-4 hours after a new BUN/Creatinine sensor cartridge is installed.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 50
- Deaths reported
- 1
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 50 adverse event reports that list this exact device (DI 00385480574006), received between Aug 28, 2019 and Mar 11, 2026 — classified as 1 death, 49 malfunctions.
The most frequently reported specific device problem is “Incorrect Measurement” (16 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- Sep 18, 2023Open, Classified
Potential risk of reporting erroneous creatinine patient results associated with the initial start-up of a new BUN/Creatinine Sensor Cartridge. The potential risk is within the first 2-4 hours after cartridge installation that may not be identified by traditional Quality Control practices.
Z-0207-2024 · Root cause: Process design · FDA record ↗
Most reported problems
Device problems
- Incorrect Measurement16
- Imprecise or Erratic Results11
- Incorrect, Inadequate or Imprecise Result or Readings9
- Low Test Results7
- Smoking3
- High Test Results2
- Misassembly During Maintenance/Repair2
- Adverse Event Without Identified Device or Use Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions19
- No Consequences Or Impact To Patient15
- No Known Impact Or Consequence To Patient7
- Insufficient Information4
- Cardiac Arrest1
- Heart Failure/Congestive Heart Failure1
- Hyperkalemia1
- Loss of consciousness1
Recent adverse event reports
The 25 most recent of 50 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 11, 2026Failure to Prime
PER CUSTOMER: "I AM WRITING TO REPORT THAT OUR NOVA STAT PROFILE PRIME PLUS WAS COMPLETELY NONFUNCTIONAL TODAY, WHICH SIGNIFICANTLY IMPACTED PATIENT TESTING AND WORKFLOW IN OUR LABORATORY. THIS DOWNTIME DIRECTLY AFFECTED PATIENT CARE. WE HAD A STAT PATIENT WITH SUSPECTED HYPERKALEMIA, AND WE WERE UNABLE TO PROVIDE CRITICAL POTASSIUM RESULTS TO THE PROVIDER DUE TO THE INSTRUMENT FAILURE. THIS SITUATION POSED A SERIOUS RISK TO TIMELY CLINICAL DECISION-MAKING." A FOLLOW UP WITH THE CUSTOMER INDICATED THE ANALYZER HAD BEEN DOWN ALL DAY AND AFTER CONSUMABLES WERE REPLACED THE ANALYZER HAS BEEN BUSY PRIMING/CALIBRATING AND MOST OF THE TEST ARE LOCKED OUT. THE AUDIT LOG INDICATED SLOPE ERRORS ON SEVERAL PARAMETERS.
Patient problems: Hyperkalemia
MAUDE report 24570256 ↗MalfunctionFeb 12, 2026Failure to Calibrate; Calibration Problem
CONSUMER REPORTED THAT THEY HAVE BEEN HAVING ISSUES WITH THE ANALYZER STAYING CALIBRATED FOR PO2 AND HAVE CONSISTENTLY NEEDED TO REPLACE THE CONSUMABLES.THE CONSUMER REQUESTED SERVICE DUE TO INCONSISTENCY WITH CALIBRATION.
Patient problems: Insufficient Information; Ventilator Dependent
MAUDE report 24321484 ↗MalfunctionDec 17, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
ON (B)(6) 2025 THE CUSTOMER RAN PATIENT SAMPLES FOR PH AND HCO3 (BICARBONATE). THEY BELIEVED THE RESULTS THEY RECEIVED WERE INACCURATE AND RETESTED ANOTHER SAMPLE ON A DIFFERENT ANALYZER (ISTAT). THERE WAS SIGNIFICANT DISCREPANCY BETWEEN THE 2 ANALYZERS. IT IS UNCLEAR HOW MUCH TIME HAD PASSED BETWEEN THE RETESTING OF THE SAMPLE. ALTHOUGH THERE WAS NO REPORTED TREATMENT TO THE PATIENT BASED ON THE ALLEGED INACCURATE RESULTS, THE CUSTOMER CLAIMED THAT THERE COULD HAVE BEEN AN ADVERSE EVENT HAD THEY TREATED BASED ON THOSE RESULTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23830129 ↗MalfunctionMay 2, 2025Low Test Results; Low Test Results
SODIUM RESULT INCOMPATIBLE WITH THE PATIENT'S CLINICAL DATA.
Patient problems: Heart Failure/Congestive Heart Failure; Renal Failure
MAUDE report 21952903 ↗MalfunctionNov 15, 2022Low Readings; Low Readings
CUSTOMER REPORTS DISCREPANCIES IN HB RESULTS FOR ONE PATIENT.
Patient problems: Pulmonary Dysfunction
MAUDE report 15799839 ↗MalfunctionJul 15, 2022No Apparent Adverse Event; No Apparent Adverse Event
COMPLAINANT REPORTED SUSPECT ERROR PROBLEMS OVER THE WEEKEND OF (B)(6) 2022 WHICH RESULTED IN A DELAY OF CARE FOR A PATIENT IN THE NEONATAL UNIT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15033007 ↗MalfunctionJul 7, 2022Incorrect Measurement; Incorrect Measurement
NOVA BIOMEDICAL (NOVA) WAS MADE AWARE OF A POTENTIAL ISSUE REGARDING ANALYSIS OF BICARBONATE (HCO3) AND BASE EXCESS OF BLOOD (BE-B) RESULTS DURING THE TREATMENT OF A PATIENT ON A STATPROFILE PRIME PLUS ANALYZER WITH SERIAL NUMBER (B)(4). THE RESULTS WERE REPORTED BUT THE CLINICIAN QUESTIONED THE ACCURACY OF THESE WHEN COMPARED TO THE CLINICAL PRESENTATION OF THE PATIENT. AN ADDITIONAL SAMPLE FROM THE PATIENT WAS RUN USING THE REFERENCE EPOC ANALYZER WHICH CONFIRMED THE INITIAL RESULTS WERE INACCURATE. NO PATIENT HARM OR ADVERSE TREATMENT DUE TO THE INACCURATE RESULTS WAS REPORTED.
Patient problems: Insufficient Information
MAUDE report 14944310 ↗MalfunctionJul 7, 2022Incorrect Measurement; Incorrect Measurement
NOVA BIOMEDICAL (NOVA) WAS MADE AWARE OF A POTENTIAL ISSUE REGARDING ANALYSIS OF SODIUM, POTASSIUM, CHLORIDE, AND BLOOD UREA NITROGEN (BUN) RESULTS DURING THE TREATMENT OF A PATIENT ON A STATPROFILE PRIME PLUS ANALYZER WITH SERIAL NUMBER (B)(4). THE RESULTS WERE REPORTED BUT THE CLINICIAN QUESTIONED THE ACCURACY OF THESE WHEN COMPARED TO THE CLINICAL PRESENTATION OF THE PATIENT. AN ADDITIONAL SAMPLE FROM THE PATIENT WAS RUN USING THE REFERENCE ANALYZER BECKMAN AU WHICH CONFIRMED THE INITIAL RESULTS WERE INACCURATE. NO PATIENT HARM OR ADVERSE TREATMENT DUE TO THE INACCURATE RESULTS WAS REPORTED.
Patient problems: Insufficient Information
MAUDE report 14944290 ↗MalfunctionMar 31, 2022Smoking; Smoking
CUSTOMER REPORTED THAT WHEN THE PRIME PLUS ANALYZER WAS PLUGGED IN THERE WAS A BURNING SMELL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13965970 ↗MalfunctionMar 31, 2022Smoking; Smoking
CUSTOMER REPORTED THAT WHEN THE PRIME PLUS ANALYZER WAS PLUGGED IN THERE WAS A BURNING SMELL AND THE SCREEN REMAINED BLACK.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13965873 ↗MalfunctionMar 2, 2022Imprecise or Erratic Results; Imprecise or Erratic Results
PRIME PLUS REPORTED HIGHER HEMATOCRIT RESULTS AND DOES NOT CORRELATE WELL WHEN COMPARED TO THE HEMOGLOBIN RESULTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13656430 ↗MalfunctionMar 2, 2022Imprecise or Erratic Results; Imprecise or Erratic Results
PRIME PLUS REPORTED HIGHER HEMATOCRIT RESULTS AND DOES NOT CORRELATE WELL WHEN COMPARED TO THE HEMOGLOBIN RESULTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13656294 ↗MalfunctionMar 2, 2022Imprecise or Erratic Results; Imprecise or Erratic Results
PRIME PLUS REPORTED HIGHER HEMATOCRIT RESULTS AND DOES NOT CORRELATE WELL WHEN COMPARED TO THE HEMOGLOBIN RESULTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13656279 ↗MalfunctionMar 2, 2022Imprecise or Erratic Results; Imprecise or Erratic Results
PRIME PLUS REPORTED HIGHER HEMATOCRIT RESULTS AND DOES NOT CORRELATE WELL WHEN COMPARED TO THE HEMOGLOBIN RESULTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13656247 ↗MalfunctionMar 2, 2022Imprecise or Erratic Results; Imprecise or Erratic Results
PRIME PLUS REPORTED HIGHER HEMATOCRIT RESULTS AND DOES NOT CORRELATE WELL WHEN COMPARED TO THE HEMOGLOBIN RESULTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13656186 ↗MalfunctionFeb 23, 2022Low Test Results; Low Test Results
THE CUSTOMER REPORTS THAT THE NOVA PRIME PLUS ANALYZER SN (B)(4) RETURNED A LOWER CREATININE RESULT THAN A REFERENCE ANALYZER. THE PATIENT WAS GIVEN CONTRAST BASED OFF OF THIS RESULT FOR A CT SCAN.
Patient problems: Insufficient Information
MAUDE report 13587092 ↗MalfunctionJan 19, 2022Imprecise or Erratic Results; Imprecise or Erratic Results
THE CUSTOMER INDICATED THAT THEIR PRIME PLUS ANALYZER SERIAL NUMBER PP1120040C REPORTED PATIENT RESULTS FOR SODIUM AND POTASSIUM WITH A SIGNIFICANT DIFFERENCE TO LABORATORY VALUES FROM SAME SAMPLING TIME. CUSTOMER CONFIRMED THAT THE PATIENT RESULTS WERE NOT USED FOR MEDICAL DECISIONS AND THERE WAS NO PATIENT INJURY, HARM OR ADDITIONAL TREATMENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13289443 ↗DeathNov 19, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
NOVA BIOMEDICAL (NOVA) WAS MADE AWARE OF A POTENTIAL ISSUE REGARDING MISSING CO-OXIMETRY RESULTS DURING THE TREATMENT OF A PATIENT ON A STATPROFILE PRIME PLUS ANALYZER WITH SERIAL NUMBER (B)(4). IT WAS REPORTED THE PATIENT ARRIVED INTUBATED AND WAS PRESUMED TO BE INTOXICATED LEADING TO SEVERE METHEMOGLOBINEMIA. THE PRIME PLUS ANALYZER DID NOT PRODUCE CO-OXIMETRY RESULTS WHEN RUNNING A SAMPLE AND INSTEAD DISPLAYED A "RESULTS SUSPECT" MESSAGE. THIS SAMPLE WAS RUN ON AN ADDITIONAL ANALYZER WHICH CONFIRMED THE "RESULTS SUSPECT". AFTER THREE ATTEMPTS AT RUNNING A SAMPLE, A METHEMOGLOBIN VALUE OF 55% WAS RECEIVED. WITHOUT CONFIRMATION OF THE DIAGNOSIS, THE PATIENT WAS TREATED BASED ON CLINICAL PRESENTATION. THE PATIENT LATER WENT INTO CARDIAC ARREST AND DID NOT RECOVER AFTER CARDIOPULMONARY RESUSCITATION.
Patient problems: Cardiac Arrest; Loss of consciousness
MAUDE report 12843935 ↗MalfunctionNov 7, 2021Low Test Results; Low Test Results
NOVA BIOMEDICAL (NOVA) WAS MADE AWARE OF A SITUATION WHERE AN INACCURATE RESULT FOR SODIUM (NA) WAS RECEIVED WHILE USING A STATPROFILE PRIME PLUS ANALYZER, (B)(4). THE NURSE QUESTIONED THE RESULTS AS THEY DID NOT FIT THE PATIENT'S CLINICAL PRESENTATION. ADDITIONAL SAMPLES WERE RUN ACROSS AN ADDITIONAL ANALYZER AND CONFIRMED THE INITIAL RESULTS WERE INCORRECT. NO PATIENT HARM OR ADVERSE TREATMENT DUE TO THE INACCURATE RESULTS WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12759394 ↗MalfunctionNov 7, 2021High Test Results; High Test Results
NOVA BIOMEDICAL (NOVA) WAS MADE AWARE OF A SITUATION WHERE AN INACCURATE RESULT FOR SODIUM (NA) WAS RECEIVED WHILE USING A STATPROFILE PRIME PLUS ANALYZER, (B)(4). THE NURSE QUESTIONED THE RESULTS AS THEY DID NOT FIT THE PATIENT'S CLINICAL PRESENTATION. ADDITIONAL SAMPLES WERE RUN ACROSS OTHER ANALYZERS AND CONFIRMED THE INITIAL RESULTS WERE INCORRECT. NO PATIENT HARM OR ADVERSE TREATMENT DUE TO THE INACCURATE RESULTS WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12759051 ↗MalfunctionSep 28, 2021Smoking; Insufficient Device Problem Information; Smoking; Insufficient Device Problem Information
NOVA BIOMEDICAL CORPORATE WAS MADE AWARE OF A POTENTIAL PRODUCT MALFUNCTION USING A STATPROFILE PRIME PLUS BENCHTOP ANALYZER, SERIAL NUMBER (B)(4). IT WAS REPORTED THE ANALYZER WAS PRODUCING A BURING SMELL AND SMOKE WAS COMING FROM THE MACHINE. THE ANALYZER WAS IMMEDIATELY TAKEN OUT OF SERVICE AND REMOVED FROM THE FACILITY. THERE WAS NO IMPACT ON PATIENT OR USER SAFETY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12541450 ↗MalfunctionAug 24, 2021Imprecise or Erratic Results; Imprecise or Erratic Results
THE CUSTOMER INDICATED THAT THEIR PRIME PLUS ANALYZER SERIAL NUMBER: (B)(4) REPORTED DISCREPANT PCO2 MEASUREMENT, HCO3 AND CL ARE BOTH HIGH.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12362110 ↗MalfunctionAug 24, 2021Imprecise or Erratic Results; Imprecise or Erratic Results
THE CUSTOMER INDICATED THAT THEIR PRIME PLUS ANALYZER SERIAL NUMBER: (B)(4) REPORTED DISCREPANT PCO2 MEASUREMENT, HCO3 AND CL ARE BOTH HIGH.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12361052 ↗MalfunctionAug 24, 2021Imprecise or Erratic Results; Imprecise or Erratic Results
THE CUSTOMER INDICATED THAT THEIR PRIME PLUS ANALYZER SERIAL NUMBER: (B)(4) REPORTED DISCREPANT PCO2 MEASUREMENT, HCO3 AND CL ARE BOTH HIGH.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12360107 ↗MalfunctionJul 8, 2021Imprecise or Erratic Results; Imprecise or Erratic Results
THE CUSTOMER REPORTED DISCREPANT PATIENT RESULTS. A TEST WAS RUN ON JUNE 11TH ON THE PRIME PLUS SN: (B)(4) AND REPORTED HIGH CREATININE RESULTS UTILIZING BUN/ CREAT CARD LOT: 21130038. THE PATIENTS DID NOT HAVE THE SCHEDULED PROCEDURES BASED ON THE RESULTS SINCE THE VALUES WERE QUESTIONED BY THE PHYSICIAN. THERE WAS NO PATIENT HARM OR MEDICAL INTERVENTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12136057 ↗
Other Electrode measurement, blood-gases (pco2, po2) and blood ph devices
- GEM Premier 5000 00024019255Instrumentation Laboratory Company · Electrode Measurement, Blood-Gases (Pco2, Po2) And Blood Ph5 recalls5 deaths7 reports
- ABL800 FLEX 393-800Radiometer Medical ApS · ELECTRODE MEASUREMENT, BLOOD-GASES (PCO2, PO2) AND BLOOD PH2 recalls81 reports
- ABL90 FLEX Analyzer 393-090Radiometer Medical ApS · ELECTRODE MEASUREMENT, BLOOD-GASES (PCO2, PO2) AND BLOOD PH1 recall112 reports
- i-STAT 03P88-25Abbott Point Of Care Inc. · ELECTRODE MEASUREMENT, BLOOD-GASES (PCO2, PO2) AND BLOOD PH105 reports
- GEM Premier 5000 PAK BG/ISE/GL/COOX 300 TEST 00055430010Instrumentation Laboratory Company · ELECTRODE MEASUREMENT, BLOOD-GASES (PCO2, PO2) AND BLOOD PH2 recalls3 deaths3 reports
- GEM Premier 5000 PAK BG/ISE/GL/COOX 450 TEST 00055445010Instrumentation Laboratory Company · ELECTRODE MEASUREMENT, BLOOD-GASES (PCO2, PO2) AND BLOOD PH2 recalls2 reports
- GEM Premier 5000 PAK BG/COOX 75 TEST 00055407504Instrumentation Laboratory Company · ELECTRODE MEASUREMENT, BLOOD-GASES (PCO2, PO2) AND BLOOD PH2 recalls0 reports
- GEM Premier 5000 PAK BG/ISE/GL/COOX 600 TEST 00055360010Instrumentation Laboratory Company · ELECTRODE MEASUREMENT, BLOOD-GASES (PCO2, PO2) AND BLOOD PH2 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page