BiPAP autoSV Advanced System One 60 Series BiPAP autoSV Advanced, 60 Series, USRecalls & reported problems
Respironics, Inc. · BiPAP autoSV Advanced System One 60 Series, US · Class 2
All models of BiPAP autoSV Advanced System One 60 Series (19)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Respironics BiPAP autoSV Advanced has 123 adverse event reports, with 7 deaths and 24 injuries. Most reports describe foam inside the unit breaking down and contamination found during service checks, and some mention headaches, dizziness or urinary problems. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 123
- Deaths reported
- 7
- Injuries reported
- 24
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 123 adverse event reports that list this exact device (DI 00606959012516), received between Nov 3, 2021 and Jul 2, 2026 — classified as 7 deaths, 24 injuries, 92 malfunctions.
The most frequently reported specific device problem is “Degraded” (119 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Degraded119
- Contamination9
- Adverse Event Without Identified Device or Use Problem6
- Overheating of Device2
- Break1
- Material Integrity Problem1
- Noise, Audible1
Patient problems
- No Clinical Signs, Symptoms or Conditions75
- Unspecified Respiratory Problem21
- Headache12
- Dizziness10
- Cancer7
- Eye Irritation7
- Unspecified Kidney or Urinary Problem7
- Asthma6
Recent adverse event reports
The 25 most recent of 123 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 2, 2026Degraded
A BIPAP AUTOSV DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER FOR SERVICE AS PART OF THE SOUND ABATEMENT FOAM RECALL PROCESS. DURING THE EVALUATION OF THE DEVICE, THE SERVICE TECHNICIAN OBSERVED VISIBLE FOAM PARTICLES. ADDITIONALLY, UNRELATED TO THE REPORTABLE MALFUNCTION, THE TECHNICIAN OBSERVED DESTROYED POWER CONNECTOR AND THE EVALUATION OF THE DEVICE DATA IDENTIFIED UNRELATED ERROR CODES. THESE ERROR CODES ARE CONSISTENT WITH NORMAL DEVICE OPERATION AND EXPECTED USE CONDITIONS AND ARE NOT CONSIDERED REPORTABLE UNDER MEDICAL DEVICE REPORTING REQUIREMENTS. NO EVIDENCE WAS IDENTIFIED TO SUGGEST THAT THESE ERROR CODES CONTRIBUTED TO OR CAUSED THE REPORTED EVENT. THE DEVICE WAS SCRAPPED BY THE SERVICE CENTER AFTER THE EVALUATION WAS COMPLETED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25698126 ↗MalfunctionJun 17, 2026Degraded
THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. A BIPAP AUTOSV ADV DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER WITH NO INITIAL ALLEGED COMPLAINT. DURING THE EVALUATION OF THE DEVICE AT THE THIRD-PARTY SERVICE CENTER, THE DEVICE WAS VISUALLY INSPECTED, AND EVIDENCE OF FOAM DEGRADATION/PARTICLES WAS FOUND. ADDITIONALLY, THE EVALUATION OF THE DEVICE DATA IDENTIFIED UNRELATED ERROR CODES. THESE ERROR CODES ARE CONSISTENT WITH NORMAL DEVICE OPERATION AND EXPECTED USE CONDITIONS AND ARE NOT CONSIDERED REPORTABLE UNDER MEDICAL DEVICE REPORTING REQUIREMENTS. NO EVIDENCE WAS IDENTIFIED TO SUGGEST THAT THESE ERROR CODES CONTRIBUTED TO OR CAUSED THE REPORTED EVENT. THE DEVICE WAS SCRAPPED BY THE SERVICE CENTER AFTER THE EVALUATION WAS COMPLETED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25538635 ↗MalfunctionApr 7, 2026Degraded
A DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. THE DEVICE WAS NOT IN PATIENT USE. DURING THE EVALUATION OF THE DEVICE, THE THIRD-PARTY SERVICE CENTER VISUALLY INSPECTED THE DEVICE AND FOUND EVIDENCE OF FOAM DEGRADATION AND DUST CONTAMINATION. ADDITIONALLY, THE EVALUATION OF THE DEVICE DATA IDENTIFIED UNRELATED ERROR CODES. THESE ERROR CODES ARE CONSISTENT WITH NORMAL DEVICE OPERATION AND EXPECTED USE CONDITIONS AND ARE NOT CONSIDERED REPORTABLE UNDER MEDICAL DEVICE REPORTING REQUIREMENTS. NO EVIDENCE WAS IDENTIFIED TO SUGGEST THAT THESE ERROR CODES CONTRIBUTED TO OR CAUSED THE REPORTED EVENT. THE DEVICE HAS BEEN SCRAPPED. AT THIS TIME, NO FURTHER INVESTIGATION CAN BE PERFORMED. IF ANY ADDITIONAL INFORMATION IS RECEIVED, A FOLLOW UP REPORT WILL BE FILED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24809676 ↗MalfunctionApr 1, 2026Degraded
(B)(6) 2026 13:05:08 EST (PLXS_PRD_DBM). REF. NOTIFICATION: (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24760970 ↗MalfunctionMar 26, 2026Degraded
THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. A BIPAP AUTOSV ADV DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER WITH NO INITIAL ALLEGED COMPLAINT. THERE WAS NO REPORT OF SERIOUS OR PERMANENT HARM OR INJURY. THERE WAS NO REPORT OF MEDICAL INTERVENTION. DURING EVALUATION NO FOAM PARTICLE WERE NOTED IN THE AIR PATH, YET FOAM DEGRADATION WAS NOTED. THE DEVICE SOUND ABATEMENT FOAM NEEDS REPLACED TO ADDRESS THE ISSUE. ADDITIONALLY, THE SERVICE TECHNICIAN ALSO NOTED DUST CONTAMINATION, AND THE EVALUATION OF THE DEVICE DATA IDENTIFIED UNRELATED ERROR CODES. THESE ERROR CODES ARE CONSISTENT WITH NORMAL DEVICE OPERATION AND EXPECTED USE CONDITIONS AND ARE NOT CONSIDERED REPORTABLE UNDER MEDICAL DEVICE REPORTING REQUIREMENTS. NO EVIDENCE WAS IDENTIFIED TO SUGGEST THAT THESE ERROR CODES CONTRIBUTED TO OR CAUSED THE REPORTED EVENT. THE DEVICE WAS SCRAPPED BY THE SERVICE CENTER AFTER THE EVALUATION WAS COMPLETED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24699278 ↗MalfunctionMar 11, 2026Degraded
THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. A BIPAP AUTOSV ADV DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER WITH NO INITIAL ALLEGED COMPLAINT. THERE WAS NO REPORT OF SERIOUS OR PERMANENT HARM OR INJURY. THERE WAS NO REPORT OF MEDICAL INTERVENTION.DURING EVALUATION NO FOAM PARTICLE WERE NOTED IN THE AIR PATH, YET FOAM DEGRADATION WAS NOTED. THE DEVICE SOUND ABATEMENT FOAM NEEDS REPLACED TO ADDRESS THE ISSUE. ADDITIONALLY, THE SERVICE TECHNICIAN ALSO NOTED DUST CONTAMINATION, AND THE EVALUATION OF THE DEVICE DATA IDENTIFIED UNRELATED ERROR CODES. THESE ERROR CODES ARE CONSISTENT WITH NORMAL DEVICE OPERATION AND EXPECTED USE CONDITIONS AND ARE NOT CONSIDERED REPORTABLE UNDER MEDICAL DEVICE REPORTING REQUIREMENTS. NO EVIDENCE WAS IDENTIFIED TO SUGGEST THAT THESE ERROR CODES CONTRIBUTED TO OR CAUSED THE REPORTED EVENT. THE DEVICE WAS SCRAPPED BY THE SERVICE CENTER AFTER THE EVALUATION WAS COMPLETED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24575882 ↗MalfunctionFeb 27, 2026Contamination; Degraded
THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. A BIPAP AUTOSV DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER WITH NO INITIAL ALLEGED COMPLAINT. DURING THE EVALUATION OF THE DEVICE AT THE THIRD-PARTY SERVICE CENTER, THE DEVICE WAS VISUALLY INSPECTED, AND NO FOAM PARTICLES WERE NOTED IN THE AIRPATH, YET FOAM DEGRADATION WAS NOTED. THE DEVICE'S SOUND ABATEMENT FOAM WAS REPLACED TO ADDRESS THE ISSUE. ADDITIONALLY, THE SERVICE TECHNICIAN ALSO NOTED DUST CONTAMINATION, AND THE EVALUATION OF THE DEVICE DATA IDENTIFIED UNRELATED ERROR CODES. THESE ERROR CODES ARE CONSISTENT WITH NORMAL DEVICE OPERATION AND EXPECTED USE CONDITIONS AND ARE NOT CONSIDERED REPORTABLE UNDER MEDICAL DEVICE REPORTING REQUIREMENTS. NO EVIDENCE WAS IDENTIFIED TO SUGGEST THAT THESE ERROR CODES CONTRIBUTED TO OR CAUSED THE REPORTED EVENT. THE DEVICE WAS REPAIRED BY THE SERVICE CENTER AFTER THE EVALUATION WAS COMPLETED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24475052 ↗MalfunctionJan 29, 2026Degraded
A BIPAP AUTO BI FLEX DEVICE WAS RETURNED TO A THIRD PARTY SERVICE CENTER FOR SERVICE AS PART OF THE SOUND ABATEMENT FOAM RECALL PROCESS WITH NO INITIAL ALLEGED COMPLAINT. DURING THE EVALUATION OF THE DEVICE, THE THIRD PARTY SERVICE CENTER VISUALLY INSPECTED THE DEVICE AND OBSERVED VISIBLE FOAM PARTICLES INSIDE THE BLOWER KIT AND NOTED A BLOWER FOAM DEGRADATION. ADDITIONALLY, THE SERVICE TECHNICIAN ALSO NOTED DUST CONTAMINATION, AND THE EVALUATION OF THE DEVICE DATA IDENTIFIED UNRELATED ERROR CODES. THESE ERROR CODES ARE CONSISTENT WITH NORMAL DEVICE OPERATION AND EXPECTED USE CONDITION THAT ARE NOT CONSIDERED REPORTABLE UNDER MEDICAL DEVICE REPORTING REQUIREMENTS. NO EVIDENCE WAS IDENTIFIED TO SUGGEST THAT THESE ERROR CODES CONTRIBUTED TO OR CAUSED THE REPORTED EVENT. THE DEVICE WAS SCRAPPED BY THE SERVICE CENTER AFTER THE EVALUATION WAS COMPLETED. THIS EVENT IS REPORTABLE UNDER FDA 21 CFR PART 803 AS IT MEETS THE CRITERIA FOR A SERIOUS INJURY AND OR PRODUCT MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24211542 ↗MalfunctionJan 8, 2026Degraded
A BIPAP AUTOSV ADV SYS ONE DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER FOR SERVICE AS PART OF THE SOUND ABATEMENT FOAM RECALL PROCESS WITH NO INITIAL ALLEGED COMPLAINT. DURING THE EVALUATION OF THE DEVICE, THE SERVICE TECHNICIAN OBSERVED VISIBLE FOAM PARTICLES INSIDE THE BLOWER KIT AND A BLOWER FOAM DEGRADATION. ADDITIONALLY, THE SERVICE TECHNICIAN ALSO NOTED DUST FAIL CONTAMINATION AND DISPLAYED ERROR CODE E066 (ERR_STUCK_ENCODER_B), E062 (ERR_ STUCK_RAMP_KEY), E022 (ERR_MOTOR_FLUX_MAGNITUDE), E032 (ERR_ LOW_PRESSURE_SUPPORT). THE DEVICE WAS SCRAPPED BY THE SERVICE CENTER AFTER THE EVALUATION WAS COMPLETED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24012925 ↗MalfunctionJan 8, 2026Degraded
THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE/RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. A BIPAP AUTOSV ADV SYS ONE DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER FOR EVALUATION. THERE WAS NO REPORT OF SERIOUS OR PERMANENT HARM OR INJURY. THERE WAS NO REPORT OF MEDICAL INTERVENTION. DURING THE EVALUATION OF THE DEVICE, THE THIRD-PARTY SERVICE CENTER VISUALLY INSPECTED THE DEVICE AND FOUND EVIDENCE OF VISIBLE FOAM. IN ADDITION, DEVICE DISPLAY ERROR CODE E093 (RTC_VALUE). DUST WAS FOUND AS CONTAMINATION. THE DEVICE WAS SCRAPPED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24001181 ↗MalfunctionJan 6, 2026Degraded
A BIPAP AUTOSV DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER AS PART OF THE UNO RECALL PROCESS WITH NO INITIAL ALLEGED COMPLAINT. DURING THE EVALUATION OF THE DEVICE, THE SERVICE TECHNICIAN OBSERVED VISIBLE FOAM PARTICLES INSIDE THE BLOWER KIT. ADDITIONALLY, THE SERVICE TECHNICIAN FOUND CONTAMINATION FROM DUST AND ERROR CODE E066 (ERR STUCK ENCODER B). THE DEVICE WAS SCRAPPED BY THE SERVICE CENTER AFTER THE EVALUATION WAS COMPLETED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23974822 ↗MalfunctionNov 24, 2025Contamination; Degraded; Contamination; Degraded
A DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. DURING THE EVALUATION OF THE DEVICE AT THE THIRD-PARTY SERVICE CENTRE, THE DEVICE WAS VISUALLY INSPECTED AND FOUND EVIDENCE OF FOAM DEGRADATION. DURING THE EVALUATION, SECONDARY FINDINGS DUST WAS OBSERVED. THE DEVICE POWERED ON AND AIRFLOW WAS CONFIRMED. THE DEVICE'S DOWNLOADED LOGS WERE REVIEWED BY THE THIRD-PARTY SERVICE CENTRE. THERE WAS 11 ERROR CODES FOUND. THE THIRD-PARTY SERVICE CENTER CONCLUDES THAT THERE WAS VISIBLE FOAM DEGRADATION AND UNIT GOT CORRECTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23629258 ↗MalfunctionNov 19, 2025Contamination; Degraded; Contamination; Degraded
A DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. DURING THE EVALUATION OF THE DEVICE AT THE THIRD-PARTY SERVICE CENTER, THE DEVICE WAS VISUALLY INSPECTED AND FOUND EVIDENCE OF FOAM DEGRADATION. DURING THE EVALUATION, DUST WAS OBSERVED. THE DEVICE POWERED ON AND AIRFLOW WAS CONFIRMED. THE DEVICE'S DOWNLOADED LOGS WERE REVIEWED BY THE THIRD-PARTY SERVICE CENTER. THERE WAS TEN ERROR CODE FOUND. THE THIRD-PARTY SERVICE CENTER CONCLUDES THAT THEY COULD CONFIRM THE CUSTOMER'S ALLEGATION AND THERE WERE VISIBLE FOAM DEGRADATION AND UNIT GOT CORRECTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23591925 ↗MalfunctionNov 19, 2025Contamination; Degraded; Contamination; Degraded
A DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. DURING THE EVALUATION OF THE DEVICE AT THE THIRD-PARTY SERVICE CENTRE, THE DEVICE WAS VISUALLY INSPECTED AND FOUND EVIDENCE OF FOAM DEGRADATION. DURING THE EVALUATION, DUST CONTAMINATION WAS OBSERVED. THE DEVICE POWERED ON AND AIRFLOW WAS CONFIRMED. THE DEVICE'S DOWNLOADED LOGS WERE REVIEWED BY THE THIRD-PARTY SERVICE CENTRE. THERE WAS 6 ERROR CODE FOUND. THE THIRD-PARTY SERVICE CENTER CONCLUDES THAT THEY COULD CONFIRM THE CUSTOMER'S ALLEGATION AND THERE WERE VISIBLE FOAM DEGRADATION AND UNIT GOT CORRECTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23591397 ↗MalfunctionNov 4, 2025Degraded; Overheating of Device; Degraded; Overheating of Device
A DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. DURING THE EVALUATION OF THE DEVICE AT THE THIRD-PARTY SERVICE CENTRE, THE DEVICE WAS VISUALLY INSPECTED AND FOUND NO EVIDENCE OF FOAM DEGRADATION. DURING THE EVALUATION, BURNT POWER CONNECTOR WAS OBSERVED. THE DEVICE POWERED ON AND AIRFLOW WAS CONFIRMED. THE DEVICE'S DOWNLOADED LOGS WERE REVIEWED BY THE THIRD-PARTY SERVICE CENTRE. THERE WAS 31 ERROR CODE FOUND. THE THIRD-PARTY SERVICE CENTRE CONCLUDES THAT THEY COULDN'T CONFIRM THE CUSTOMER'S ALLEGATION AND THERE WERE NO VISIBLE FOAM DEGRADATION AND UNIT GOT CORRECTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23460705 ↗MalfunctionNov 3, 2025Degraded; Overheating of Device; Degraded; Overheating of Device
A DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. DURING THE EVALUATION OF THE DEVICE AT THE THIRD-PARTY SERVICE CENTRE, THE DEVICE WAS VISUALLY INSPECTED AND FOUND NO EVIDENCE OF FOAM DEGRADATION. DURING THE EVALUATION, BURNT POWER CONNECTOR WAS OBSERVED. THE DEVICE POWERED ON AND AIRFLOW WAS CONFIRMED. THE DEVICE'S DOWNLOADED LOGS WERE REVIEWED BY THE THIRD-PARTY SERVICE CENTRE. THERE WAS 1 ERROR CODE FOUND. THE THIRD-PARTY SERVICE CENTER CONCLUDES THAT THERE WERE NO VISIBLE FOAM DEGRADATION AND UNIT GOT CORRECTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23450011 ↗MalfunctionAug 6, 2025Degraded; Degraded
THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. THE MANUFACTURER RECEIVED INFORMATION ALLEGING VISUALIZATION OF PARTICLES. THERE IS NO ALLEGATION OF SERIOUS OR PERMANENT HARM OR INJURY. NO MEDICAL INTERVENTION WAS SPECIFIED AS REQUIRED BY THE PATIENT. DURING THE EVALUATION OF THE DEVICE, AT THE MANUFACTURE'S THIRD-PARTY'S SERVICE CENTER THE DEVICE WAS VISUALLY INSPECTED AND FOUND EVIDENCE OF FOAM PARTICLES. THE DEVICE HAS BEEN SCRAPPED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22732941 ↗MalfunctionJul 22, 2025Degraded; Degraded
THIS COMPLAINT HAS BEEN REVIEWED, AND IT HAS BEEN DETERMINED THAT THE DEVICE WAS RETURNED TO THE THIRD-PARTY SERVICE CENTER. THE EVALUATION RESULT FOUND EXIST CONTAMINATION OF FOAM PARTICLES INSIDE THE BLOWER KIT. THERE WAS NO REPORTED PATIENT DEATH OR SERIOUS INJURY, HOWEVER THE REPORTED ISSUE/EVENT IS REPORTABLE DUE TO THE POTENTIAL TO CAUSE OR CONTRIBUTE TO A SERIOUS INJURY AS IDENTIFIED IN CAPA PR7211, IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE/RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22564083 ↗MalfunctionJul 16, 2025Degraded; Degraded
A DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. DURING THE EVALUATION OF THE DEVICE AT THE THIRD-PARTY SERVICE CENTER, THE DEVICE WAS VISUALLY INSPECTED/SCRAPPED AND FOUND EVIDENCE OF FOAM DEGRADATION. DURING THE EVALUATION, FOAM PARTICLES WERE VISIBLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22526798 ↗MalfunctionJun 13, 2025Degraded; Degraded
THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. THERE WAS NO ALLEGATION REPORTED BY THE PATIENT. THERE WAS NO REPORT OF SERIOUS PATIENT HARM OR INJURY. THERE WAS NO REPORT OF MEDICAL INTERVENTION. NO ADDITIONAL INFORMATION CAN BE REQUESTED AT THIS TIME. T LINE NOT REQUIRED) THE MANUFACTURER'S INVESTIGATION IS ONGOING. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER'S INVESTIGATION IS COMPLETE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22201089 ↗MalfunctionJun 12, 2025Degraded; Degraded
THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE/RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES.THERE WAS NO ALLEGATION REPORTED BY THE PATIENT. THERE WAS NO REPORT OF MEDICAL INTERVENTION. THE MANUFACTURER WAS MADE AWARE OF THIS COMPLAINT THROUGH A REPRESENTATIVE OF THE CUSTOMER. THE MANUFACTURER'S INVESTIGATION IS ONGOING. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER'S INVESTIGATION IS COMPLETE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22194039 ↗MalfunctionJun 12, 2025Degraded; Degraded
DeathMar 15, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE MANUFACTURER WAS CONTACTED IN REFERENCE TO AN INFORMATION REQUEST CONCERNING DISPOSING OF A DEVICE. THE MANUFACTURER WAS INFORMED THAT THE PATIENT WAS DECEASED. THERE IS NO ALLEGATION OF DEVICE MALFUNCTION. THERE IS NO ALLEGATION THAT THE DEVICE CONTRIBUTED TO THE PATIENT DEATH. NO OTHER INFORMATION IS AVAILABLE AT THIS TIME. THIS REPORT IS BEING SUBMITTED FOR AN ALLEGATION OF A PATIENT DEATH ONLY DUE TO INSUFFICIENT INFORMATION RELATED TO THE DEVICE. THE DEVICE HAS NOT BEEN RETURNED TO THE MANUFACTURER FOR INVESTIGATION. THE MANUFACTURER INVESTIGATION IS ONGOING. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER'S INVESTIGATION IS COMPLETE.
Patient problems: Insufficient Information
MAUDE report 21613602 ↗MalfunctionMar 6, 2025Degraded; Degraded
THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. THE MANUFACTURER RECEIVED INFORMATION ALLEGING AN ISSUE RELATED TO A BIPAP DEVICE'S SOUND ABATEMENT FOAM. THE PATIENT ALLEGES VISUALIZATION OF PARTICLES IN DEVICE. THERE IS NO ALLEGATION OF SERIOUS OR PERMANENT HARM OR INJURY. NO MEDICAL INTERVENTION WAS REQUIRED BY THE PATIENT. THE MANUFACTURER'S INVESTIGATION IS ONGOING. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER'S INVESTIGATION IS COMPLETE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21534120 ↗MalfunctionMar 4, 2025Degraded; Degraded
THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. THE MANUFACTURER RECEIVED INFORMATION ALLEGING VISUALIZATION OF PARTICLES. THERE IS NO ALLEGATION OF SERIOUS OR PERMANENT HARM OR INJURY. NO MEDICAL INTERVENTION WAS SPECIFIED BY THE PATIENT. THE DEVICE HAS NOT YET BEEN RETURNED TO THE MANUFACTURER FOR INVESTIGATION. THE MANUFACTURER'S INVESTIGATION IS ONGOING. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER'S INVESTIGATION IS COMPLETE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21515566 ↗
Other VENTILATOR, CONTINUOUS, NON-LIFE-SUPPORTING devices
- DreamStation BiPAP ST30 00Respironics, Inc. · VENTILATOR, CONTINUOUS, NON-LIFE-SUPPORTING5 deaths547 reports
- DreamStation BiPAP ST30 00Respironics, Inc. · VENTILATOR, CONTINUOUS, NON-LIFE-SUPPORTING3 deaths283 reports
- BiPAP AVAPS BiPAP AVAPS C Series, 30cm, INTLRespironics, Inc. · VENTILATOR, CONTINUOUS, NON-LIFE-SUPPORTING274 reports
- BiPAP autoSV Advanced System One 60 Series BiPAP autoSV Advanced w/H Tube Humid, 60 Series, USRespironics, Inc. · VENTILATOR, CONTINUOUS, NON-LIFE-SUPPORTING3 deaths209 reports
- BiPAP S/T C-Series BiPAP S/T C Series Core Pack, 30cm, USRespironics, Inc. · VENTILATOR, CONTINUOUS, NON-LIFE-SUPPORTING1 death200 reports
- DreamStation BiPAP AVAPS DreamStation BiPAP AVAPS30 w/Humidifier w/Cell, USRespironics, Inc. · VENTILATOR, CONTINUOUS, NON-LIFE-SUPPORTING4 deaths174 reports
- DreamStation BiPAP AVAPS DreamStation AVAPS30 w/Humid w/H Tube w/Cell, USRespironics, Inc. · VENTILATOR, CONTINUOUS, NON-LIFE-SUPPORTING2 deaths144 reports
- BiPAP A30 00Respironics, Inc. · VENTILATOR, CONTINUOUS, NON-LIFE-SUPPORTING2 recalls29 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page