BiPAP S/T C-Series BiPAP S/T C Series, 30cm, USRecalls & reported problems
Respironics, Inc. · BiPAP S/T C Series, 30cm, US · Class 2
All models of BiPAP S/T C-Series (7)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Respironics BiPAP S/T ventilator delivers breathing support. Most of the 52 reports describe foam inside the device breaking down, with visible particles and dust found in the blower during service; others mention odor, headaches, dizziness and shortness of breath. There were no deaths and 10 injuries, with 42 malfunction reports. No recalls name this device. In the past 12 months there were 16 reports, up from 4 in the year before.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 51
- Deaths reported
- 0
- Injuries reported
- 10
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 51 adverse event reports that list this exact device (DI 00606959012547), received between Jan 18, 2022 and Aug 11, 2026 — classified as 10 injuries, 41 malfunctions.
The most frequently reported specific device problem is “Degraded” (50 reports).
1 other recall cover devices cleared under the same FDA submission (K102465); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Degraded50
- Contamination5
- Device Emits Odor1
Patient problems
- No Clinical Signs, Symptoms or Conditions36
- Headache6
- Unspecified Respiratory Problem5
- Asthma4
- Dizziness4
- Cancer3
- Dyspnea3
- Respiratory Tract Infection3
Recent adverse event reports
The 25 most recent of 51 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 11, 2026Degraded
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50202437 ↗MalfunctionJun 17, 2026Degraded
A BIPAP S/T C SERIES WAS RETURNED TO A THIRD-PARTY SERVICE CENTER FOR SERVICE AS PART OF THE SOUND ABATEMENT FOAM RECALL PROCESS. DURING THE EVALUATION OF THE DEVICE, THE SERVICE TECHNICIAN OBSERVED VISIBLE FOAM PARTICLES. ADDITIONALLY, UNRELATED TO THE REPORTABLE MALFUNCTION, THE TECHNICIAN OBSERVED MISSING KNOB AND THE EVALUATION OF THE DEVICE DATA IDENTIFIED UNRELATED ERROR CODES. THESE ERROR CODES ARE CONSISTENT WITH NORMAL DEVICE OPERATION AND EXPECTED USE CONDITIONS AND ARE NOT CONSIDERED REPORTABLE UNDER MEDICAL DEVICE REPORTING REQUIREMENTS. NO EVIDENCE WAS IDENTIFIED TO SUGGEST THAT THESE ERROR CODES CONTRIBUTED TO OR CAUSED THE REPORTED EVENT. THE DEVICE WAS SCRAPPED BY THE SERVICE CENTER AFTER THE EVALUATION WAS COMPLETED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25534901 ↗MalfunctionMay 26, 2026Degraded
A BIPAP S/T C SERIES DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER FOR SERVICE AS PART OF THE SOUND ABATEMENT FOAM RECALL PROCESS. DURING THE EVALUATION OF THE DEVICE AT THE THIRD-PARTY SERVICE CENTER, THE DEVICE WAS VISUALLY INSPECTED, AND EVIDENCE OF FOAM PARTICLES. ADDITIONALLY, THE SERVICE TECHNICIAN IDENTIFIED UNRELATED ERROR CODES. THESE ERROR CODES ARE CONSISTENT WITH NORMAL DEVICE OPERATION AND EXPECTED USE CONDITIONS AND ARE NOT CONSIDERED REPORTABLE UNDER MEDICAL DEVICE REPORTING REQUIREMENTS. NO EVIDENCE WAS IDENTIFIED TO SUGGEST THAT THESE ERROR CODES CONTRIBUTED TO OR CAUSED THE REPORTED EVENT. THE DEVICE WAS SCRAPPED BY THE SERVICE CENTER AFTER THE EVALUATION WAS COMPLETED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25282695 ↗MalfunctionMay 25, 2026Degraded
THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. A DREAMSTATION AUTO CPAP DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER WITH NO INITIAL ALLEGED COMPLAINT. DURING THE EVALUATION OF THE DEVICE AT THE THIRD-PARTY SERVICE CENTER THE DEVICE WAS VISUALLY INSPECTED AND EVIDENCE OF FOAM DEGRADATION WAS FOUND. IN ADDITION, THE DEVICE WAS SCRAPPED BY THE SERVICE CENTER AFTER THE EVALUATION WAS COMPLETED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25272339 ↗MalfunctionMay 4, 2026Degraded
A BIPAP S/T C SERIES DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER FOR SERVICE AS PART OF THE SOUND ABATEMENT FOAM RECALL PROCESS WITH NO INITIAL ALLEGED COMPLAINT. DURING THE EVALUATION OF THE DEVICE AT THE THIRD-PARTY SERVICE CENTER, THE DEVICE WAS VISUALLY INSPECTED, THE SERVICE TECHNICIAN OBSERVED VISIBLE FOAM PARTICLES, AND FOAM DEGRADATION WAS NOTED. ADDITIONALLY, THE SERVICE TECHNICIAN ALSO NOTED NO CONTAMINATION, DESTROYED POWER CONNECTOR, AND THE EVALUATION OF THE DEVICE DATA IDENTIFIED UNRELATED ERROR CODES. THESE ERROR CODES ARE CONSISTENT WITH NORMAL DEVICE OPERATION AND EXPECTED USE CONDITIONS AND ARE NOT CONSIDERED REPORTABLE UNDER MEDICAL DEVICE REPORTING REQUIREMENTS. NO EVIDENCE WAS IDENTIFIED TO SUGGEST THAT THESE ERROR CODES CONTRIBUTED TO OR CAUSED THE REPORTED EVENT. THE DEVICE WAS SCRAPPED BY THE SERVICE CENTER AFTER THE EVALUATION WAS COMPLETED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25071235 ↗MalfunctionMar 16, 2026Degraded
A BIPAP S/T DEVICE WAS RETURNED TO THE THIRD-PARTY SERVICE CENTER IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. DURING THE EVALUATION OF THE DEVICE AT THE SERVICE CENTER, THE DEVICE WAS VISUALLY INSPECTED AND FOUND EVIDENCE OF FOAM DEGRADATION. DURING THE EVALUATION, FOAM PARTICLES WERE OBSERVED, FLUIDS CONTAMINATION AND DUST CONTAMINATION WAS FOUND IN THE DEVICE. ADDITIONALLY, THE EVALUATION OF THE DEVICE DATA IDENTIFIED UNRELATED ERROR CODES. THESE ERROR CODES ARE CONSISTENT WITH NORMAL DEVICE OPERATION AND EXPECTED USE CONDITIONS AND ARE NOT CONSIDERED REPORTABLE UNDER MEDICAL DEVICE REPORTING REQUIREMENTS. NO EVIDENCE WAS IDENTIFIED TO SUGGEST THAT THESE ERROR CODES CONTRIBUTED TO OR CAUSED THE REPORTED EVENT. THE DEVICE WAS SCRAPPED. A FINAL REPORT IS BEING SUBMITTED. IF ANY ADDITIONAL INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE FILED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24611821 ↗MalfunctionMar 12, 2026Degraded
A BIPAP S/T C SERIES DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER WITH NO INITIAL ALLEGED COMPLAINT. THERE WAS NO REPORT OF SERIOUS OR PERMANENT HARM OR INJURY. THERE WAS NO REPORT OF MEDICAL INTERVENTION. DURING THE EVALUATION OF THE DEVICE AT THE THIRD-PARTY SERVICE CENTER, THE DEVICE WAS VISUALLY INSPECTED, AND EVIDENCE OF FOAM DEGRADATION/PARTICLES WAS FOUND INSIDE THE BLOWER KIT. ADDITIONALLY, THE SERVICE TECHNICIAN ALSO NOTED DUST CONTAMINATION, AND THE EVALUATION OF THE DEVICE DATA IDENTIFIED UNRELATED ERROR CODES. THESE ERROR CODES ARE CONSISTENT WITH NORMAL DEVICE OPERATION AND EXPECTED USE CONDITIONS AND ARE NOT CONSIDERED REPORTABLE UNDER MEDICAL DEVICE REPORTING REQUIREMENTS. NO EVIDENCE WAS IDENTIFIED TO SUGGEST THAT THESE ERROR CODES CONTRIBUTED TO OR CAUSED THE REPORTED EVENT. THE DEVICE WAS SCRAPPED BY THE SERVICE CENTER AFTER THE EVALUATION WAS COMPLETED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24584168 ↗MalfunctionFeb 10, 2026Degraded
A BIPAP S/T DEVICE WAS RETURNED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. DURING THE EVALUATION OF THE DEVICE, THE SERVICE TECHNICIAN OBSERVED VISIBLE FOAM PARTICLES. ADDITIONALLY, THE SERVICE TECHNICIAN ALSO NOTED CONTAMINATION. THE DEVICE WAS SCRAPPED BY THE SERVICE CENTER AFTER THE EVALUATION WAS COMPLETED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24305264 ↗MalfunctionFeb 4, 2026Degraded
A BIPAP S/T SERIES DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER FOR SERVICE AS PART OF THE SOUND ABATEMENT FOAM RECALL PROCESS WITH NO INITIAL ALLEGED COMPLAINT. DURING THE EVALUATION OF THE DEVICE, THE SERVICE TECHNICIAN OBSERVED VISIBLE FOAM PARTICLES INSIDE THE BLOWER KIT AND A BLOWER FOAM DEGRADATION. ADDITIONALLY, THE SERVICE TECHNICIAN ALSO NOTED DUST FAIL CONTAMINATION, AND THE EVALUATION OF THE DEVICE DATA IDENTIFIED UNRELATED ERROR CODES. THESE ERROR CODES ARE CONSISTENT WITH NORMAL DEVICE OPERATION AND EXPECTED USE CONDITIONS AND ARE NOT CONSIDERED REPORTABLE UNDER MEDICAL DEVICE REPORTING REQUIREMENTS. NO EVIDENCE WAS IDENTIFIED TO SUGGEST THAT THESE ERROR CODES CONTRIBUTED TO OR CAUSED THE REPORTED EVENT. THE DEVICE WAS SCRAPPED BY THE SERVICE CENTER AFTER THE EVALUATION WAS COMPLETED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24258879 ↗MalfunctionFeb 3, 2026Degraded
A BIPAP DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER FOR SERVICE AS PART OF THE SOUND ABATEMENT FOAM RECALL PROCESS WITH NO INITIAL ALLEGED COMPLAINT. DURING THE EVALUATION OF THE DEVICE, THE THIRD-PARTY SERVICE CENTER VISUALLY INSPECTED THE DEVICE AND OBSERVED VISIBLE FOAM PARTICLES INSIDE THE BLOWER KIT AND NOTED A BLOWER FOAM DEGRADATION. ADDITIONALLY, THE SERVICE TECHNICIAN ALSO NOTED DUST CONTAMINATION, AND THE EVALUATION OF THE DEVICE DATA IDENTIFIED UNRELATED ERROR CODES. THESE ERROR CODES ARE CONSISTENT WITH NORMAL DEVICE OPERATION AND EXPECTED USE CONDITION THAT ARE NOT CONSIDERED REPORTABLE UNDER MEDICAL DEVICE REPORTING REQUIREMENTS. NO EVIDENCE WAS IDENTIFIED TO SUGGEST THAT THESE ERROR CODES CONTRIBUTED TO OR CAUSED THE REPORTED EVENT. THE DEVICE WAS SCRAPPED BY THE SERVICE CENTER AFTER THE EVALUATION WAS COMPLETED. THIS EVENT IS REPORTABLE UNDER FDA 21 CFR PART 803 AS IT MEETS THE CRITERIA FOR A SERIOUS INJURY AND OR PRODUCT MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24248580 ↗MalfunctionFeb 3, 2026Degraded
THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. A BIPAP S/T C DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER WITH NO INITIAL ALLEGED COMPLAINT. DURING THE EVALUATION OF THE DEVICE AT THE THIRD-PARTY SERVICE CENTER, THE DEVICE WAS VISUALLY INSPECTED, AND EVIDENCE OF FOAM DEGRADATION/PARTICLES WAS FOUND INSIDE THE BLOWER KIT. ADDITIONALLY, THE SERVICE TECHNICIAN ALSO NOTED DUST CONTAMINATION, AND THE EVALUATION OF THE DEVICE DATA IDENTIFIED UNRELATED ERROR CODES. THESE ERROR CODES ARE CONSISTENT WITH NORMAL DEVICE OPERATION AND EXPECTED USE CONDITIONS AND ARE NOT CONSIDERED REPORTABLE UNDER MEDICAL DEVICE REPORTING REQUIREMENTS. NO EVIDENCE WAS IDENTIFIED TO SUGGEST THAT THESE ERROR CODES CONTRIBUTED TO OR CAUSED THE REPORTED EVENT. THE DEVICE WAS SCRAPPED BY THE SERVICE CENTER AFTER THE EVALUATION WAS COMPLETED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24247845 ↗MalfunctionFeb 2, 2026Degraded
THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE/RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. A BIPAP S/T C SERIES DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER WITH NO INITIAL ALLEGED COMPLAINT. DURING THE EVALUATION OF THE DEVICE, THE SERVICE TECHNICIAN OBSERVED VISIBLE FOAM PARTICLES INSIDE THE BLOWER KIT AND A BLOWER FOAM DEGRADATION. ADDITIONALLY, THE SERVICE TECHNICIAN ALSO NOTED DUST FAIL CONTAMINATION, AND THE EVALUATION OF THE DEVICE DATA IDENTIFIED UNRELATED ERROR CODES. THESE ERROR CODES ARE CONSISTENT WITH NORMAL DEVICE OPERATION AND EXPECTED USE CONDITIONS AND ARE NOT CONSIDERED REPORTABLE UNDER MEDICAL DEVICE REPORTING REQUIREMENTS. NO EVIDENCE WAS IDENTIFIED TO SUGGEST THAT THESE ERROR CODES CONTRIBUTED TO OR CAUSED THE REPORTED EVENT. THE DEVICE UNIT IS FUNCTIONAL. FOR MORE INFORMATION, SEE THE ATTACHMENT. THE DEVICE WAS SCRAPPED BY THE SERVICE CENTER AFTER THE EVALUATION WAS COMPLETED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24234134 ↗MalfunctionJan 30, 2026Degraded
THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. A BIPAP S/T DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER WITH NO INITIAL ALLEGED COMPLAINT. DURING THE EVALUATION OF THE DEVICE AT THE THIRD-PARTY SERVICE CENTER THE DEVICE WAS VISUALLY INSPECTED AND EVIDENCE OF FOAM DEGRADATION WAS FOUND INSIDE THE BLOWER KIT. MOREOVER, THE DEVICE HAD DUST CONTAMINATION. ADDITIONALLY, THE EVALUATION OF THE DEVICE DATA IDENTIFIED UNRELATED ERROR CODES. THESE ERROR CODES ARE CONSISTENT WITH NORMAL DEVICE OPERATION AND EXPECTED USE CONDITIONS AND ARE NOT CONSIDERED REPORTABLE UNDER MEDICAL DEVICE REPORTING REQUIREMENTS. NO EVIDENCE WAS IDENTIFIED TO SUGGEST THAT THESE ERROR CODES CONTRIBUTED TO OR CAUSED THE REPORTED EVENT. IN ADDITION, THE DEVICE WAS SCRAPPED BY THE SERVICE CENTER AFTER THE EVALUATION WAS COMPLETED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24223738 ↗MalfunctionNov 25, 2025Contamination; Degraded; Contamination; Degraded
A DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. DURING THE EVALUATION OF THE DEVICE AT THE THIRD-PARTY SERVICE CENTER, THE DEVICE WAS VISUALLY INSPECTED AND FOUND EVIDENCE OF FOAM DEGRADATION. DURING THE EVALUATION SECONDARY FINDINGS ARE OBSERVED. DUST WAS OBSERVED. THE DEVICE POWERED ON AND AIRFLOW WAS CONFIRMED. THE DEVICE'S DOWNLOADED LOGS WERE REVIEWED BY THE THIRD-PARTY SERVICE CENTER. THERE WAS NO ERROR CODES FOUND. THE THIRD-PARTY SERVICE CENTRE CONCLUDES THAT THEY COULD CONFIRM THE CUSTOMER'S ALLEGATION AND THERE WAS VISIBLE FOAM DEGRADATION AND UNIT GOT CORRECTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23642348 ↗MalfunctionNov 18, 2025Contamination; Degraded; Contamination; Degraded
A DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. DURING THE EVALUATION OF THE DEVICE AT THE THIRD-PARTY SERVICE CENTRE, THE DEVICE WAS VISUALLY INSPECTED AND FOUND EVIDENCE OF FOAM DEGRADATION. DURING THE EVALUATION, SECONDARY FINDINGS AND DUST CONTAMINATION WAS OBSERVED. THE DEVICE POWERED ON AND AIRFLOW WAS CONFIRMED. THE DEVICE'S DOWNLOADED LOGS WERE REVIEWED BY THE THIRD-PARTY SERVICE CENTRE. THERE WAS ZERO ERROR CODE FOUND. THE THIRD-PARTY SERVICE CENTER CONCLUDES THAT THERE WAS VISIBLE FOAM DEGRADATION AND UNIT GOT CORRECTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23579892 ↗MalfunctionJul 17, 2025Degraded; Degraded
A DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. DURING THE EVALUATION OF THE DEVICE AT THE THIRD-PARTY SERVICE CENTER, THE DEVICE WAS VISUALLY INSPECTED/SCRAPPED AND FOUND EVIDENCE OF FOAM DEGRADATION. DURING THE EVALUATION, FOAM PARTICLES WERE VISIBLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22529193 ↗MalfunctionJun 17, 2025Degraded; Degraded
THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. THE MANUFACTURER PREVIOUSLY RECEIVED INFORMATION ALLEGING VISUALIZATION OF PARTICLES RELATED TO A BIPAP DEVICE'S SOUND ABATEMENT FOAM. THERE WAS NO REPORT OF PATIENT HARM OR INJURY. THERE WAS NO REPORT OF MEDICAL INTERVENTION. THE MANUFACTURER'S INVESTIGATION IS ONGOING. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER'S INVESTIGATION IS COMPLETE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22217249 ↗MalfunctionJun 13, 2025Degraded; Degraded
THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. THERE WAS NO ALLEGATION REPORTED BY THE PATIENT. THERE WAS NO REPORT OF SERIOUS PATIENT HARM OR INJURY. THERE WAS NO REPORT OF MEDICAL INTERVENTION. THE MANUFACTURER'S INVESTIGATION IS ONGOING. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER'S INVESTIGATION IS COMPLETE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22201353 ↗MalfunctionNov 17, 2024Degraded; Degraded
THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. THE MANUFACTURER RECEIVED INFORMATION ALLEGING VISUALIZATION OF PARTICLES IN THE AIR PATH. THERE IS NO ALLEGATION OF SERIOUS OR PERMANENT HARM OR INJURY. NO MEDICAL INTERVENTION WAS REQUIRED BY THE PATIENT. THE DEVICE HAS NOT YET BEEN RETURNED TO THE MANUFACTURER FOR EVALUATION. THREE ATTEMPTS TO HAVE THE DEVICE RETURNED FOR EVALUATION AND INVESTIGATION WERE UNSUCCESSFUL. AT THIS TIME, NO FURTHER INVESTIGATION CAN BE PERFORMED. IF ANY ADDITIONAL INFORMATION IS RECEIVED, A FOLLOW UP REPORT WILL BE FILED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20701986 ↗MalfunctionAug 15, 2024Degraded; Degraded
THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. THE MANUFACTURER RECEIVED INFORMATION ALLEGING VISUALIZATION OF PARTICLES IN THE AIR PATH. THERE IS NO ALLEGATION OF SERIOUS OR PERMANENT HARM OR INJURY. NO MEDICAL INTERVENTION WAS REQUIRED BY THE PATIENT. THE MANUFACTURER'S INVESTIGATION IS ONGOING. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER'S INVESTIGATION IS COMPLETE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19992497 ↗MalfunctionAug 14, 2024Degraded; Degraded
THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. THE MANUFACTURER RECEIVED INFORMATION ALLEGING VISUALIZATION OF PARTICLES. THERE WAS NO REPORT OF HARM OR INJURY. NO MEDICAL INTERVENTION WAS REQUIRED BY THE PATIENT. THE DEVICE HAS NOT BEEN RETURNED TO THE MANUFACTURER. AT THIS TIME, WE ARE UNABLE TO CONFIRM THE ALLEGED MALFUNCTION. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER'S INVESTIGATION IS COMPLETE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19985186 ↗MalfunctionAug 14, 2024Degraded; Degraded
THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. THE MANUFACTURER RECEIVED INFORMATION ALLEGING VISUALIZATION OF PARTICLES. THERE WAS NO REPORT OF HARM OR INJURY. NO MEDICAL INTERVENTION WAS REQUIRED BY THE PATIENT. THE DEVICE HAS NOT BEEN RETURNED TO THE MANUFACTURER. AT THIS TIME, WE ARE UNABLE TO CONFIRM THE ALLEGED MALFUNCTION. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER'S INVESTIGATION IS COMPLETE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19982521 ↗MalfunctionAug 12, 2024Degraded; Degraded
THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE/RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. THE MANUFACTURER RECEIVED INFORMATION ALLEGING AN ISSUE RELATED TO A BIPAP S/T DEVICE'S SOUND ABATEMENT FOAM. THE PATIENT HAS ALLEGED VISUALIZATION OF PARTICLES. THERE WAS NO REPORT OF SERIOUS OR PERMANENT HARM OR INJURY. NO MEDICAL INTERVENTION WAS REQUIRED BY THE PATIENT. THE MANUFACTURER'S INVESTIGATION IS ONGOING. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER'S INVESTIGATION IS COMPLETE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19967698 ↗MalfunctionAug 12, 2024Degraded; Degraded
THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE/RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. THE MANUFACTURER RECEIVED INFORMATION ALLEGING AN ISSUE RELATED TO A BIPAP S/T DEVICE'S SOUND ABATEMENT FOAM. THE PATIENT HAS ALLEGED VISUALIZATION OF PARTICLES. THERE WAS NO REPORT OF SERIOUS OR PERMANENT HARM OR INJURY. NO MEDICAL INTERVENTION WAS REQUIRED BY THE PATIENT. THE MANUFACTURER'S INVESTIGATION IS ONGOING. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER'S INVESTIGATION IS COMPLETE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19967620 ↗MalfunctionAug 12, 2024Degraded; Degraded
THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. THE MANUFACTURER RECEIVED INFORMATION ALLEGING AN ISSUE RELATED TO A BIPAP S/T DEVICE'S SOUND ABATEMENT FOAM. THE MANUFACTURER RECEIVED INFORMATION ALLEGING VISUALIZATION OF PARTICLES. THERE WAS NO REPORT OF HARM OR INJURY. NO MEDICAL INTERVENTION WAS REQUIRED BY THE PATIENT. THE DEVICE HAS NOT BEEN RETURNED TO THE MANUFACTURER. AT THIS TIME, WE ARE UNABLE TO CONFIRM THE ALLEGED MALFUNCTION. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER'S INVESTIGATION IS COMPLETE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19960148 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K102465. One submission often covers many models, so check the product and lot details in each record.
- Z-1973-2021 · Jun 14, 2021 · Open, ClassifiedDreamStation ASV, DreamStation ST, AVAPS, SystemOne ASV4, C-Series ASV, C-Series S/T and AVAPS, OmniLab Advanced+ Continuous Ventilator, Non-life Supporting The Philips Respironic
Other VENTILATOR, CONTINUOUS, NON-LIFE-SUPPORTING devices
- DreamStation BiPAP ST30 00Respironics, Inc. · VENTILATOR, CONTINUOUS, NON-LIFE-SUPPORTING5 deaths547 reports
- DreamStation BiPAP ST30 00Respironics, Inc. · VENTILATOR, CONTINUOUS, NON-LIFE-SUPPORTING3 deaths283 reports
- BiPAP AVAPS BiPAP AVAPS C Series, 30cm, INTLRespironics, Inc. · VENTILATOR, CONTINUOUS, NON-LIFE-SUPPORTING274 reports
- BiPAP autoSV Advanced System One 60 Series BiPAP autoSV Advanced w/H Tube Humid, 60 Series, USRespironics, Inc. · VENTILATOR, CONTINUOUS, NON-LIFE-SUPPORTING3 deaths209 reports
- BiPAP S/T C-Series BiPAP S/T C Series Core Pack, 30cm, USRespironics, Inc. · VENTILATOR, CONTINUOUS, NON-LIFE-SUPPORTING1 death200 reports
- DreamStation BiPAP AVAPS DreamStation BiPAP AVAPS30 w/Humidifier w/Cell, USRespironics, Inc. · VENTILATOR, CONTINUOUS, NON-LIFE-SUPPORTING4 deaths174 reports
- DreamStation BiPAP AVAPS DreamStation AVAPS30 w/Humid w/H Tube w/Cell, USRespironics, Inc. · VENTILATOR, CONTINUOUS, NON-LIFE-SUPPORTING2 deaths144 reports
- BiPAP A30 00Respironics, Inc. · VENTILATOR, CONTINUOUS, NON-LIFE-SUPPORTING2 recalls29 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page