Skip to content

OmniLab Advanced Plus OmniLab Advanced Plus, INTLRecalls & reported problems

Respironics, Inc. · OmniLab Advanced Plus, International · Class 2

All models of OmniLab Advanced Plus (5)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00606959029415), received between Mar 23, 2022 and Feb 29, 2024 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Electrical /Electronic Property Problem” (2 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 2 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Aug 26, 2022Class IOpen, Classified

    Motor assemblies manufactured with nonconforming plastic that contributes to off-gassing of VOCs.

    Z-1658-2022 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Electrical /Electronic Property Problem2
  • Degraded1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 29, 2024Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    THE MANUFACTURER RECEIVED INFORMATION ALLEGING A VENTILATOR INOPERATIVE CONDITION OCCURRED. THERE WAS NO HARM OR INJURY REPORTED. DURING THE EVALUATION OF THE DEVICE AT THE MANUFACTURER'S SERVICE CENTER, THE CUSTOMER COMPLAINT WAS CONFIRMED. THE DEVICES PCA AND BOTTOM ENCLOSURE WERE REPLACED AND THE FIRMWARE UPDATED TO V3.7. THE DEVICE PASSES TESTING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18810999 ↗
  • MalfunctionFeb 28, 2024Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    THE MANUFACTURER RECEIVED INFORMATION ALLEGING A VENTILATOR INOPERATIVE CONDITION OCCURRED. THERE WAS NO HARM OR INJURY REPORTED. THE MAIN PCA AND BLOWER ASSEMBLY WERE REPLACED TO RESOLVE THE REPORTED PROBLEM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18797604 ↗
  • MalfunctionMar 23, 2022Degraded; Degraded

    THE MANUFACTURER RECEIVED INFORMATION ALLEGING AN ISSUE RELATED TO A BIPAP DEVICE'S SOUND ABATEMENT FOAM. THE DEVICE WAS RETURNED AND NOISY BLOWER; BOTTOM ENCLOSURE, RIGHT SIDE ASSY, TOP ENCLOSURE DAMAGED BY UNKNOWN LIQUID AND DAMAGED INLET FOAM WERE OBSERVED DURING DEVICE EVALUATION. THERE WAS NO REPORT OF HARM OR INJURY. THIS ISSUE WAS REPORTED TO THE FDA PER 21 CFR 806. THE DEVICE WILL BE CORRECTED PER RES 88058.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13869774 ↗

Other VENTILATOR, CONTINUOUS, NON-LIFE-SUPPORTING devices

All VENTILATOR, CONTINUOUS, NON-LIFE-SUPPORTING devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page